Title: Mapping electrical properties heterogeneity of tumor using boundary informed electrical properties tomography (BIEPT) at 7T
Purposes

To develop and evaluate a boundary informed electrical properties tomography (BIEPT) technique for high‐resolution imaging of tumor electrical properties (EPs) heterogeneity on a rodent tumor xenograft model.

Methods

Tumor EP distributions were inferred from a reference area external to the tumor, as well as internal EP spatial variations derived from a plurality of relative transmit B1measurements at 7T. Edge sparsity constraint was enforced to enhance numerical stability. Phantom experiments were performed to determine the imaging accuracy and sensitivity for structures of various EP values, as well as geometrical sizes down to 1.5 mm. Numerical simulation of a realistic rodent model was used to quantify the algorithm performance in the presence of noise. Eleven athymic rats with human breast cancer xenograft were imaged in vivo, and representative pathological samples were acquired for comparison.

Results

Reconstructed EPs of the phantoms correspond well to the ground truth acquired from dielectric probe measurements, with the smallest structure reliably detectable being 3 mm. EPs heterogeneity inside a tumor is successfully retrieved in both simulated and experimental cases. In vivo tumor imaging results demonstrate similar local features and spatial patterns to anatomical MRI and pathological slides. The imaged conductivity of necrotic tissue is higher than that of viable tissues, which agrees with our expectation.

Conclusion

BIEPT enables robust detection of tumor EPs heterogeneity with high accuracy and sensitivity to small structures. The retrieved quantitative EPs reflect tumor pathological features (e.g., necrosis). These results provide strong rationale to further expand BIEPT studies toward pathological conditions where EPs may yield valuable, non‐invasive biomarkers.

 
more » « less
NSF-PAR ID:
10075789
Author(s) / Creator(s):
 ;  ;  ;  ;  ;  ;  
Publisher / Repository:
Wiley Blackwell (John Wiley & Sons)
Date Published:
Journal Name:
Magnetic Resonance in Medicine
Volume:
81
Issue:
1
ISSN:
0740-3194
Page Range / eLocation ID:
p. 393-409
Format(s):
Medium: X
Sponsoring Org:
National Science Foundation
More Like this
  1. Abstract Background

    Current preclinical models of metastatic prostate cancer (PCa) require sophisticated technologies and/or genetically engineered cells for the noninvasive monitoring of tumors in remote sites, such as bone. Recent developments in circulating tumor DNA (ctDNA) analysis provide an alternative method for noninvasive tumor monitoring at a low cost. Here, we sought to evaluate human Alu and LINE‐1 ctDNA for the longitudinal measurement of subcutaneous and intratibial human PCa xenograft growth and response to ionizing radiation (IR) through comparison with standard slide caliper and bioluminescence measurements.

    Material and Methods

    Intratibial and subcutaneous xenografts were established in male athymic nude mice using LNCaP cells that stably express firefly luciferase. A subset of tumors was treated with a single dose of IR (CT‐guided focal IR, 6 Gy). Tumor measurements were simultaneously taken by slide caliper (subcutaneous only), in vivo bioluminescence imaging, and quantitative real‐time PCR (qPCR) of human‐specific Alu and LINE‐1 ctDNA for several weeks.

    Results

    Levels of ctDNA and bioluminescence increased concordantly with subcutaneous and intratibial tumor growth. A statistically significant correlation (Spearman) was observed between ctDNA and subcutaneous tumor volume (LINE‐1,r = .94 and Alu,r = .95,p < .0001), ctDNA and bioluminescence (LINE‐1,r = .66 and Alu,r = .60,p < .002), and bioluminescence and tumor volume (r = .66,p = .0003). Bioluminescence and ctDNA were also significantly correlated in intratibial tumors (LINE‐1,r = .82 and Alu,r = .81,p < .0001). Following external beam IR, the tumor responses varied briefly by method of measurement, but followed a similar trend. Statistically significant correlations were maintained between ctDNA and slide caliper measurement in irradiated subcutaneous tumors (LINE‐1,r = .64 and Alu,r = .44,p < .02), and ctDNA and bioluminescence in intratibial tumors (LINE‐1,r = .55,p = .018).

    Conclusions

    Real‐time qPCR of circulating human Alu and LINE‐1 DNA provides an accurate measurement of subcutaneous and intratibial xenograft burden that is comparable with conventional bioluminescence imaging and slide caliper measurement. Transient differences in measurements were observed following tumor‐targeted IR, but overall all measurements mirrored tumor growth and response.

     
    more » « less
  2. Abstract Purpose

    One standard method, proton resonance frequency shift, for measuring temperature using magnetic resonance imaging (MRI), in MRI‐guided surgeries, fails completely below the freezing point of water. Because of this, we have developed a new methodology for monitoring temperature with MRI below freezing. The purpose of this paper is to show that a strong temperature dependence of the nuclear relaxation timeT1in soft silicone polymers can lead to temperature‐dependent changes of MRI intensity acquired withT1weighting. We propose the use of silicone filaments inserted in tissue for measuring temperature during MRI‐guided cryoablations.

    Methods

    The temperature dependence ofT1in bio‐compatible soft silicone polymers was measured using nuclear magnetic resonance spectroscopy and MRI. Phantoms, made of bulk silicone materials and put in an MRI‐compatible thermal container with dry ice, allowed temperature measurements ranging from –60°C to + 20°C.T1‐weighted gradient echo images of the phantoms were acquired at spatially uniform temperatures and with a gradient in temperature to determine the efficacy of using these materials as temperature indicators in MRI. Ex vivo experiments on silicone rods, 4 mm in diameter, inserted in animal tissue were conducted to assess the practical feasibility of the method.

    Results

    Measurements of nuclear relaxation times of protons in soft silicone polymers show a monotonic, nearly linear, change with temperature (R2 > 0.98) and have a significant correlation with temperature (Pearson'sr > 0.99,p < 0.01). Similarly, the intensity of the MR images in these materials, taken with a gradient echo sequence, are also temperature dependent. There is again a monotonic change in MRI intensity that correlates well with the measured temperature (Pearson'sr < ‐0.98 andp < 0.01). The MRI experiments show that a temperature change of 3°C can be resolved in a distance of about 2.5 mm. Based on MRI images and external sensor calibrations for a sample with a gradient in temperature, temperature maps with 3°C isotherms are created for a bulk phantom. Experiments demonstrate that these changes in MRI intensity with temperature can also be seen in 4 mm silicone rods embedded in ex vivo animal tissue.

    Conclusions

    We have developed a new method for measuring temperature in MRI that potentially could be used during MRI‐guided cryoablation operations, reducing both procedure time and cost, and making these surgeries safer.

     
    more » « less
  3. Purpose

    The purpose of this study is to develop double diffusion encoding (DDE) MRI methods for clinical use. Microscopic diffusion anisotropy measurements from DDE promise greater specificity to changes in tissue microstructure compared with conventional diffusion tensor imaging, but implementation of DDE sequences on whole‐body MRI scanners is challenging because of the limited gradient strengths and lengthy acquisition times.

    Methods

    A custom single‐refocused DDE sequence was implemented on a 3T whole‐body scanner. The DDE gradient orientation scheme and sequence parameters were optimized based on a Gaussian diffusion assumption. Using an optimized 5‐min DDE acquisition, microscopic fractional anisotropy (μFA) maps were acquired for the first time in multiple sclerosis patients.

    Results

    Based on simulations and in vivo human measurements, six parallel and six orthogonal diffusion gradient pairs were found to be the minimum number of diffusion gradient pairs necessary to produce a rotationally invariant measurement of μFA. Simulations showed that optimal precision and accuracy of μFA measurements were obtained using b‐values between 1500 and 3000 s/mm2. The μFA maps showed improved delineation of multiple sclerosis lesions compared with conventional fractional anisotropy and distinct contrast from T2‐weighted fluid attenuated inversion recovery and T1‐weighted imaging.

    Conclusion

    The μFA maps can be measured using DDE in a clinical setting and may provide new opportunities for characterizing multiple sclerosis lesions and other types of tissue degeneration. Magn Reson Med 80:507–520, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

     
    more » « less
  4. Abstract

    Imaging of cardiac tissue structure plays a critical role in the treatment and understanding of cardiovascular disease. Optical coherence tomography (OCT) offers the potential to provide valuable, high‐resolution imaging of cardiac tissue. However, there is a lack of comprehensive OCT imaging data of the human heart, which could improve identification of structural substrates underlying cardiac abnormalities. The objective of this study was to provide qualitative and quantitative analysis of OCT image features throughout the human heart. Fifty human hearts were acquired, and tissues from all chambers were imaged with OCT. Histology was obtained to verify tissue composition. Statistical differences between OCT image features corresponding to different tissue types and chambers were estimated using analysis of variance. OCT imaging provided features that were able to distinguish structures such as thickened collagen, as well as adipose tissue and fibrotic myocardium. Statistically significant differences were found between atria and ventricles in attenuation coefficient, and between adipose and all other tissue types. This study provides an overview of OCT image features throughout the human heart, which can be used for guiding future applications such as OCT‐integrated catheter‐based treatments or ex vivo investigation of structural substrates.

     
    more » « less
  5. Background

    There is growing interest in detecting cerebro‐cerebellar circuits, which requires adequate blood oxygenation level dependent contrast and signal‐to‐noise ratio (SNR) throughout the brain. Although 7T scanners offer increased SNR, coverage of commercial head coils is currently limited to the cerebrum.

    Purpose

    To improve cerebellar functional MRI (fMRI) at 7T with high permittivity material (HPM) pads extending the sensitivity of a commercial coil.

    Study Type

    Simulations were used to determine HPM pad configuration and assess radiofrequency (RF) safety. In vivo experiments were performed to evaluate RF field distributions and SNR and assess improvements of cerebellar fMRI.

    Subjects

    Eight healthy volunteers enrolled in a prospective motor fMRI study with and without HPM.

    Field Strength/Sequence

    Gradient echo (GRE) echo planar imaging for fMRI, turbo FLASH for flip angle mapping, GRE sequence for SNR maps, and T1‐weighted MPRAGE were acquired with and without HPM pads at 7T.

    Assessment

    Field maps, SNR maps, and anatomical images were evaluated for coverage. Simulation results were used to assess SAR levels of the experiment. Activation data from fMRI experiments were compared with and without HPM pads.

    Statistical Tests

    fMRI data were analyzed using FEAT FSL for each subject followed by group level analysis using paired t‐test of acquisitions with and without HPM.

    Results

    Simulations showed 52% improvement in transmit efficiency in cerebellum with HPM and SAR levels well below recommended limits. Experiments showed 27% improvement in SNR in cerebellum and improvement in coverage on T1‐weighted images. fMRI showed greater cerebellar activation in individual subjects with the HPM pad present (Z > = 4), especially in inferior slices of cerebellum, with 59% average increase in number of activated voxels in the cerebellum. Group‐level analysis showed improved functional activation (Z > = 2.3) in cerebellar regions with HPM pads without loss of measured activation elsewhere.

    Data Conclusion

    HPM pads can improve cerebellar fMRI at 7T with a commonly‐used head coil without compromising RF safety.

    Level of Evidence: 2

    Technical Efficacy: Stage 1

    J. MAGN. RESON. IMAGING 2018;48:431–440.

     
    more » « less