<?xml-model href='http://www.tei-c.org/release/xml/tei/custom/schema/relaxng/tei_all.rng' schematypens='http://relaxng.org/ns/structure/1.0'?><TEI xmlns="http://www.tei-c.org/ns/1.0">
	<teiHeader>
		<fileDesc>
			<titleStmt><title level='a'>Recent developments in the AMS system at the Nuclear Science Laboratory: Impacts on radionuclide sensitivities and current capabilities</title></titleStmt>
			<publicationStmt>
				<publisher></publisher>
				<date>02/01/2021</date>
			</publicationStmt>
			<sourceDesc>
				<bibl> 
					<idno type="par_id">10209712</idno>
					<idno type="doi">10.1016/j.nimb.2020.12.009</idno>
					<title level='j'>Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms</title>
<idno>0168-583X</idno>
<biblScope unit="volume">488</biblScope>
<biblScope unit="issue">C</biblScope>					

					<author>M. Skulski</author><author>T. Anderson</author><author>L. Callahan</author><author>A.M. Clark</author><author>A.D. Nelson</author><author>D. Robertson</author><author>E. Stech</author><author>P. Collon</author>
				</bibl>
			</sourceDesc>
		</fileDesc>
		<profileDesc>
			<abstract><ab><![CDATA[In recent years, the AMS system at the Nuclear Science Laboratory of the University of Notre Dame has introduced a series of upgrades which have dramatically increased the capabilities for the measurement of trace radionuclides. These upgrades have solved issues with isotopic selectivity in the injection system, allowed for the measurement of heavier radionuclides, and improved the quality of both ion beams and the data obtained with them. The specifics of the installed equipment, as well as preliminary results on isotopic sensitivities for previously measured and newly-developed radionuclides, are discussed. Finally, the impacts on the development of planned radionuclide measurements in the future, including measurements of the actinides, will be mentioned.]]></ab></abstract>
		</profileDesc>
	</teiHeader>
	<text><body xmlns="http://www.tei-c.org/ns/1.0" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:xlink="http://www.w3.org/1999/xlink">
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="1.">Introduction</head><p>Accelerator Mass Spectrometry (AMS) is a method for particle identification based on isotopic and isobaric discrimination, using methods from nuclear and atomic physics. AMS takes the traditional elements of mass spectrometry, which is typically performed at the keV scale, and introduces an accelerator for particle analysis on the MeV scale. Because of its selectivity, AMS is used for the measurement of trace concentrations of isotopes (typically radionuclides), capable of measuring content of one part in 10 16 <ref type="bibr">[1]</ref>. The radionuclides are typically measured with respect to one or multiple stable beam current(s) to report an isotopic ratio, or with respect to a spike isotope when a stable beam is unavailable.</p><p>The AMS program of the Nuclear Science Laboratory (NSL) of the University of Notre Dame, begun in 2003, has successfully developed the capabilities for the measurement of several radionuclides with a focus towards nuclear astrophysics. The Browne-Buechner spectrograph at the NSL, initially set up for the measurement of nuclear reactions <ref type="bibr">[2]</ref>, was altered for the use of the Gas-Filled Magnet (GFM) technique for isobar separation. This system, both the GFM technique and the NSL AMS detection system, has been described in detail in previous literature <ref type="bibr">[3]</ref>. The measurement techniques for the radionuclides 36 Cl, 44 Ti, and 60 Fe have been established with this system, despite some indication that the detection limits for each isotope were influenced by neighboring masses tailing through the injection system.</p><p>Along with the upgrade to the spectrograph, and the switch from a single-cathode National Electrostatics Corporation (NEC) Source of Negative Ions via Cesium Sputtering (SNICS) to a Multi-Cathode Source of Negative Ions via Cesium Sputtering (MC-SNICS), the 10 MV FN Tandem Van de Graaff accelerator facility has continued to successfully produce and accelerate the light ion beams for which it was designed for numerous experimental setups. These setups include a dedicated beamline for AMS measurements, two dedicated beamlines for studies involving nuclear structure, a dedicated beamline for production of short-lived radioactive nuclei for both nuclear structure and nuclear astrophysics, as well as several multi-purpose stations to accommodate different target or detector needs. As the trend towards the acceleration of heavier masses throughout the lab began for AMS analysis of heavier radionuclides and production of heavier short-lived radionuclides, there were signs that the MC-SNICS injection system (using a single dipole magnet) was no longer satisfactory. AMS analysis of the mediummass radionuclide 36 Cl showed signs that its stable isotopes 35 Cl and 37 Cl were tailing into the injection system, and were observed in the spectra from the AMS detection system. While careful tuning of the system was able to remove these chlorine isotopes, the problem only increased with higher masses. 53 Mn faced issues in the previous system during preliminary measurements, suffering from contamination from tailing of 52 Cr and 54 Cr, stable isotopes of 53 Cr, 53 Mn's isobaric contaminant.</p><p>The injection of Zr -for analysis of 93 Zr showed no clear separation of the zirconium isotopes as can be seen in Fig. <ref type="figure">1</ref>. This tailing resulted in isotopic interference in the AMS detection system that produced a detection limit of 93 Zr/Zr &#8776; 1 &#215; 10 -5 , prohibiting any useful 93 Zr measurements (such as the measurement of nuclear fuel cladding which require a detection limit of lower than 10 -7 <ref type="bibr">[4]</ref>). Funding for an  improved injection system was applied for, and awarded through, the Major Research Instrumentation (MRI) program funded through the National Science Foundation (NSF). Although 93 Zr measurement was the catalyst for this upgrade, advances towards AMS analysis of heavier radionuclides (such as 129 I and 236 U) would undoubtedly benefit from this improvement.</p><p>This article represents a report on a number of upgrades made to the FN accelerator facility not only to support the AMS program but also the increasing demand for heavier accelerated masses. Besides the injection upgrade, this included the use of the accelerator's gas stripping system and the design and installation of a high energy offset Faraday cup following the analyzing magnet to complement the low energy injection system. Additionally, a time-of-flight (TOF) system was integrated into the AMS beamline to allow the measurement of 129 I (discriminating against interfering 127 I) as well as the actinides.</p><p>Finally, a redesign of the AMS beamline, including a series of new beamline components, was completed to improve the beam optics prior to ion analysis and to eliminate scattering in the beamline observed in previous spectra (example shown in Fig. <ref type="figure">2</ref>). The current status of the AMS system at the NSL (shown in Fig. <ref type="figure">3</ref>) and the impact of the recent upgrades on measurements of radionuclides will be introduced, as will the current, preliminary sensitivity levels 1   1 Sensitivity level is synonymous with detection limit, both meaning the lowest measurable isotopic ratio for an isotope of interest.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.">Upgrades</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.1.">Injection system</head><p>Both the initial and current injection systems are shown in Fig. <ref type="figure">4</ref>. The previous MC-SNICS injection relied solely on a 60 &#8226; dipole magnet for ion selection based on magnetic rigidity, with the helium ion source (HIS) occupying the 30 &#8226; port, which has been retained for use exclusively by the HIS. The new system, purchased from NEC, includes a 45 &#8226; spherical electrostatic analyzer (&#119903; = 0.3 m, max voltage &#177;15 kV on either plate) for energy selection, which reduces the energy spread of the beam (by appropriate manipulation of downstream slits), followed by a 90 &#8226; dipole magnet (&#119903; = 0.457 m) with double focusing. The image slit positions for the dipole magnet are optimized to obtain the necessary resolution for a given mass region, maintained at &#177;3 mm horizontal and &#177;5 mm vertical for medium masses (30-110 amu) and being reduced further at higher masses as required. The new injection magnet also has a higher maximum magnetic field (1.375 T) for &#119872;&#119864;/&#119876; 2 &#8776; 19 MeV amu, which allows operation of the MC-SNICS at typical operating voltages (&#8764;60 kV) for masses up to 300 amu, whereas the previous magnet limited the full range of ion source voltages to masses less than &#8764;100 amu.</p><p>Following the injection magnet, a chamber housing two offset Faraday cups allows for the measurement of stable beam currents while an isotope of interest is injected through the accelerator system for ion detection. There is a cup on either side of the chamber for measurement of lighter and heavier mass beams for the radioisotope being measured.</p><p>The injection system includes the ability to bias the magnet vacuum box up to 15 kV. This changes the energy of beams before and after the magnet, altering their trajectories to be directed into an offset Faraday cup on the low energy side of the accelerator, or to allow the stable beams to be sent through the accelerator system for current measurement in a downstream Faraday cup for transmission measurements. This allows for near-continuous (&#8764;10 Hz) monitoring of the ion source output, improving the accuracy of reported isotopic concentrations. The magnet bias is incorporated into the fast-switching system, which controls the magnet bias and current measurement in offset Faraday cups during a measurement sequence. During the periods in a sequence when the isotope of interest is not being injected through the system, a veto signal is sent to the data acquisition system to prevent signals from interferences from being accepted and entering detector spectra.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.2.">FN accelerator system</head><p>The FN Tandem accelerator's primary method for electron stripping are carbon foils inside the terminal, which have relatively long lives and produce excellent charge state abundances for the acceleration of light ions. In cases where this stripping method is not adequate, the FN also has two other stripping methods -a gas stripping system in the terminal as a substitute for terminal foil stripping, and a second foil stripper halfway through the high energy side of the accelerator used in combination with the terminal foil stripper. In the early years of the AMS group, neither the second foil stripper nor the gas stripper were used as the dedicated stripping method for any isotope beams.</p><p>The second stripper is located in the mid-plane of the high energy column in the FN Tandem, and allows stripping to higher charge states by making use of the additional acceleration from the terminal to the mid-plane. This can allow higher beam energies (compared to single foil stripping) to be obtained at an equal terminal voltage. The second stripper has been proven critical for providing an ion beam with enough energy and sufficient intensity in the measurement of 60 Fe and 93 Zr.</p><p>The gas stripper, the traditional stripping method for heavy ions, is not subject to the rapid degradation observed in foil strippers caused by heavy ions, and energy/angular straggling of heavy ion beams through the gas stripper is reduced compared to foil stripping. In recent years, the previous string material, used in the accelerator tank to adjust the stripper's thermomechanical leak, was replaced with a kevlar braid which has been much more resilient.  </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.3.">AMS system</head><p>Following the analyzing magnet on the high energy side of the accelerator, an offset Faraday cup was located based on ion optics calculations shown in Fig. <ref type="figure">5</ref>. This cup intercepts beams injected through the accelerator as part of the fast switching system. Measurement of stable beam currents through the accelerator allows the transmission of the system to be monitored during a radionuclide measurement, necessary for achieving the required accuracy of AMS concentrations. This is particularly important when there is risk of stripper degradation, where the abundance of the selected charge state can change with changing stripper thickness. While the cup was designed specifically for the measurement of 127 I measured during 129 I AMS analysis, it is movable such that any beam with a magnetic rigidity up to &#8764;1% less than the radionuclide of interest can be measured, with the rigidity limited only by spatial constraints in the analyzing magnet and the beamline immediately downstream.</p><p>Arguably the most significant element introduced into the beamline was the time-of-flight (TOF) region, which has enabled the AMS measurement of multiple heavy isotopes. The TOF system utilizes two microchannel plate (MCP) detectors over a &#8764;2.5 m flight path and was initially developed for the measurement of 129 I, but the TOF system will be utilized for the measurement of several actinide isotopes ( 236 U, 237 Np) for research projects that are in development. The MCP detectors were obtained from the Racah Institute of Physics from the Hebrew University of Jerusalem, having been used previously for 129 I measurement <ref type="bibr">[5]</ref>.</p><p>Along with the implementation of the TOF system, several components on the AMS beamline (beamline following the switching magnet in Fig. <ref type="figure">3</ref>) were installed, shown in Fig. <ref type="figure">6</ref>. These pieces of equipment were used to replace identical components in the scattering chamber that served as the optical focus before the Browne-Buechner  spectrograph, which has been uninstalled from the beamline for the present time, but can be reinstalled if desired. The constriction into the scattering chamber had been known to cause regions of bad vacuum and ion scattering that led to energy tailing in detector spectra (see Fig. <ref type="figure">2</ref>). The pieces include two retractable apertures for focusing, a retractable Faraday cup for stable current measurement during tuning, and a retractable silicon detector at the same position as the Faraday cup for energy measurement (specifically for use along with the TOF system).</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.">Results</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.1.">Impacts on injection selectivity</head><p>Following the installation of the upgrade to the low-energy injection system, the injection of 93 Zr and several higher-mass isotope beams were studied to evaluate the improvement that was obtained. For all mass resolution studies, the bias applied to the magnet vacuum chamber is used to change the injected mass at a constant magnetic field to avoid hysteresis in the magnet.</p><p>As can be seen from the upgraded injection of zirconium elemental ions in Fig. <ref type="figure">7a</ref> and zirconium oxide ions in Fig. <ref type="figure">7b</ref>, the zirconium isotopes can be easily resolved, presenting a dramatic improvement over the previous system. This greatly expanded the capabilities for 93 Zr AMS measurements by essentially eliminating isotopic interference as well as increasing the signal intensity.</p><p>Progressing to even heavier nuclides, the injection of 127 I -(for the measurement of the radionuclide 129 I) was investigated utilizing the commonly-used sample material of silver iodide mixed with niobium powder packed into a copper sample holder. The magnet bias range corresponding to 129 I is clear of any interference, shown in Fig. <ref type="figure">7c</ref>.</p><p>Exploration to further mass regions, including lead dioxide (Fig. <ref type="figure">7d</ref>) and uranium oxide (Fig. <ref type="figure">7e</ref>), have produced initial results that demonstrate partial isotopic separation. The ultimate limitation has been in deciding the operating position of the injection slits, which requires a compromise between the reduction of ion beam intensity and increased mass resolution.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.2.">Impacts on radioisotope sensitivities</head><p>Following the installation of the upgrade components, several radionuclides that were previously measured, and several additional radioisotopes, were studied to establish new sensitivity levels for the AMS system. Table <ref type="table">1</ref> shows the list of radioisotopes measured, along with the corresponding measurement parameters and the resulting sensitivities. Each isotope is discussed further in the following paragraphs.</p><p>14 C (t 1&#8725;2 = 5.7 &#215; 10 3 yr), arguably the most well-known and widelyapplicable isotope measured using AMS (primarily for radiocarbon dating), has been in routine measurement. While the focus of AMS applications has primarily been nuclear astrophysics, the 14 C program has been a strong, undergraduate-driven project that serves as an excellent opportunity for undergraduate training in accelerator operation, data acquisition, and data analysis. 14 C solely requires energy measurement, as interfering isotopes are easily separable through magnetic rigidity and velocity selection. The isobar, 14 N, is suppressed by using a negative ion source as nitrogen does not form a negative ion. Both the measurement technique, as well as the chemistry procedure used to produce sample graphite, have been explored alongside undergraduates for the purposes of supporting departments such as Art and Biology at the University of Notre Dame. The system has been used to measure samples (at a beam energy of 10.0 MeV) with a 14 C/C isotopic ratio of 1 &#215; 10 -12 , consistent with 14 C/C values both modern to pre-nuclear. Measurement of ''blank'' material (''dead carbon'' originating from coal from a mine in Virginia) suggests a background level of 5 &#215; 10 -15 , similarly obtained for both raw coal and graphitized coal.</p><p>36 Cl (t 1&#8725;2 = 3.01 &#215; 10 5 yr) has primarily geological and astrophysical applications, though the measurements of 36 Cl have focused on the astrophysical, measuring cross sections for its production in the Early Solar System (ESS) <ref type="bibr">[6]</ref><ref type="bibr">[7]</ref><ref type="bibr">[8]</ref><ref type="bibr">[9]</ref>. The current limit of the 36 Cl/Cl isotopic ratios measurable using a beam energy of 74.7 MeV is 5 &#215; 10 -14 , partially owed to the residual 36 S interfering in the regions of interest of 36 Cl, with spectra shown in Fig. <ref type="figure">8</ref>. It is possible that sulfur reduction chemistry can be used to improve this sensitivity, and will be a subject of active exploration for future 36 Cl measurements. At greater sensitivity levels, reactions with lower cross sections, and lower energies for those reactions already studied, will become feasible.</p><p>41 Ca (t 1&#8725;2 = 9.94 &#215; 10 4 yr) has geological and astrophysical applications, similar to 36 Cl, but also has potential biomedical applications. Like 36 Cl, the measurements of 41 Ca are focused towards its astrophysical production in the ESS. While the method and system settings are still under development, preliminary measurements of 41 Ca using GFM at 87.2 MeV have been promising, with a sensitivity level of 41 Ca/Ca of 1 &#215; 10 -13 . The CaF -molecule is currently injected because of its relative production compared to CaF - 3 in the ion source, but using CaF - 3 has been reported to partially suppress the production of the isobar 41 K and will be examined <ref type="bibr">[10]</ref>. The current detection limit is well below the levels predicted to be produced for planned cross sections studies (10 -10 -10 -12 ), and thus is not expected to prohibit any measurements in the immediate future. </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Table 1</head><p>Summary of measurement parameters and sensitivity levels for radionuclides measured by AMS in the NSL.</p><p>Transmission is defined as the ratio of the particle current measured in the Faraday cup at the end of the AMS beamline to the Faraday cup following the ion source injection system. Sensitivity refers to the concentration of a radioisotope measured in a ''blank'' sample; samples above this concentration can be measured above background in the NSL's AMS system. 60 Fe (t 1&#8725;2 = 2.62 &#215; 10 6 yr) is exclusively of astrophysical importance, and was initially measured <ref type="bibr">[11]</ref> for verification of its halflife <ref type="bibr">[12]</ref>. At 113.0 MeV, obtained using the second foil stripper, the sensitivity level that is currently obtained is 60 Fe/Fe = 1 &#215; 10 -12 . This is a result of the isobaric contaminant 60 Ni ( 60 Co is not present in sufficient quantities unless artificially introduced). For measurements of 60 Fe in ferromanganese ocean crusts and ocean sediments, the most common use of measuring 60 Fe, a background of 1 &#215; 10 -16 should be obtained. 60 Fe was the first isotope to employ the second foil stripper for AMS measurements, and future measurements will explore higher energy to further separate 60 Ni. 93 Zr (t 1&#8725;2 = 1.61 &#215; 10 6 yr) is a radionuclide produced astrophysically in s-process nucleosynthesis as well as in nuclear fuel through neutron captures on fuel rod cladding. The measurement of 93 Zr is difficult due to the presence of the isobar 93 Nb, and typically requires the use of larger accelerator facilities. The measurements at the NSL have utilized the existing GFM detection system, a Bragg curve detector <ref type="bibr">[13]</ref>, and an alternate ionization chamber after <ref type="bibr">[14]</ref>, with the latter two operated outside of GFM. The lowest detection level obtained using a beam energy of 132.4 MeV with the Browne-Buechner spectrograph's detection system (operated without using the GFM) is estimated to be 93 Zr/Zr = 1 &#215; 10 -10 using a niobium-reduced zirconium oxide sample (zirconium oxide processed via ion-exchange chromatography for niobium removal), though the alternate ionization chamber may be able to improve this and will be explored in the future. The sensitivity level is estimated by extrapolating the region of interest for 93 Zr from 92 Zr and 94 Zr, shown in Fig. <ref type="figure">9</ref>. The low energy injection upgrade has succeeded in removing other zirconium isotopes from entering the spectra, so future attempts at separation will focus on experimental methods and beam settings for removal of 93 Nb, including additional niobium reduction chemistry for a potential improvement in sensitivity.</p><p>129 I (t 1&#8725;2 = 1.57 &#215; 10 7 yr) is produced naturally in the atmosphere through cosmic-ray-induced reactions with xenon isotopes. However, its largest contribution on Earth comes from neutron-induced fission of common nuclear fuels 235 U and 239 Pu, and its subsequent release from fuel reprocessing centers. Measurements of 129 I have focused on measuring the impact of these 129 I releases in the Great Lakes region. The interference for 129 I measurements comes not from the isobar 129 Xe, which is suppressed by using a negative ion source, but from the isotope 127 I. 129 I is separated from 127 I using the time-of-flight method, along with energy measurement, as shown in Fig. <ref type="figure">10</ref>. The current detection limit is 129 I/I = 6 &#215; 10 -14 at 79.0 MeV, which is only slightly higher than the 3 &#215; 10 -14 of the iodine carrier solution used in sample preparation (from Woodward Iodine Corporation, Woodward, OK) <ref type="bibr">[5,</ref><ref type="bibr">15]</ref>, which is the elemental iodine with the lowest 129 I/I that is currently known to be commercially available.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.">Future work</head><p>At present, the measurement capabilities for several new radionuclides are being developed, as are methods for improving those already being measured. A new split-anode ionization chamber is being designed following <ref type="bibr">[16]</ref> for use with GFM, specifically for the measurement of 53 Mn. 53 Mn (t 1&#8725;2 = 3.74 &#215; 10 6 yr) is produced in explosive  silicon burning, but is a valuable isotope in a geological context, where the measurement of 53 Mn produced through spallation of the Earth's surface is used in erosion studies. While the injection upgrade prohibits tailing from neighboring chromium isotopes through the AMS system, the 53 Cr isobaric interference cannot be removed this way. Separation from 53 Cr presents a significant challenge, and the use of the split-anode ionization chamber that is under development to measure position as well as energy loss, compared to the distinct detectors used to measure these beam characteristics currently, should prove invaluable for successful particle identification.</p><p>Additionally, the TOF system will be used for measurement of several of the actinides along with the implementation of the compact ionization chamber inspired by <ref type="bibr">[16]</ref><ref type="bibr">[17]</ref><ref type="bibr">[18]</ref>. The samples for the actinides originate from the MANTRA collaboration <ref type="bibr">[19]</ref>, having been irradiated by reactor neutrons at Idaho National Laboratory under several neutron filters for determining integrated neutron capture yields over specific energy ranges. Isotopically-purified samples will be measured for single, double, and possibly triple neutron capture cross-sections. This is in no way the extent of the applications of the actinides, but will serve as an initial project for the actinide program's development.</p><p>Two accelerator developments are also being installed, of benefit to all users of the FN Tandem. An upgrade to 100% SF 6 as the insulating gas will soon be complete, allowing higher maximum terminal voltages (up to 10 MV without alteration of charging components) to be explored and described in future publications. This will allow for higher energies for several AMS radioisotopes, in particular 60 Fe and 93 Zr, which have been energy-limited by the accelerator. Secondly, the gas stripping system is planned to be improved through the installation of a turbo pump for recirculating the stripping gas. Not only does this reduce the use of gas in the system, but it also helps confine the stripping region to the stripping canal instead of extending into the accelerator tubes.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="5.">Conclusion</head><p>The AMS system at the NSL has increased its capabilities for the measurement of several radionuclides thanks to an upgraded lowenergy injection system. Several other improvements including the design and implementation of a high energy offset Faraday cup and a time-of-flight system have further improved the capabilities of the AMS system not only in terms of the measurable radionuclides, but also the repeatability of the system. Several more radionuclides are planned for measurement in the near future, pending the implementation of two new detectors that will work within the existing system but greatly enhance the capabilities for the measurement of more challenging radionuclides. </p></div>