Title: Imputation of the continuous arterial line blood pressure waveform from non-invasive measurements using deep learning
Abstract

In two-thirds of intensive care unit (ICU) patients and 90% of surgical patients, arterial blood pressure (ABP) is monitored non-invasively but intermittently using a blood pressure cuff. Since even a few minutes of hypotension increases the risk of mortality and morbidity, for the remaining (high-risk) patients ABP is measured continuously using invasive devices, and derived values are extracted from the recorded waveforms. However, since invasive monitoring is associated with major complications (infection, bleeding, thrombosis), the ideal ABP monitor should be both non-invasive and continuous. With large volumes of high-fidelity physiological waveforms, it may be possible today to impute a physiological waveform from other available signals. Currently, the state-of-the-art approaches for ABP imputation only aim at intermittent systolic and diastolic blood pressure imputation, and there is no method that imputes the continuous ABP waveform. Here, we developed a novel approach to impute the continuous ABP waveform non-invasively using two continuously-monitored waveforms that are currently part of the standard-of-care, the electrocardiogram (ECG) and photo-plethysmogram (PPG), by adapting a deep learning architecture designed for image segmentation. Using over 150,000 min of data collected at two separate health systems from 463 patients, we demonstrate that our model provides a highly accurate prediction of the continuous ABP waveform (root mean square error 5.823 (95% CI 5.806–5.840) mmHg), as well as the derived systolic (mean difference 2.398 ± 5.623 mmHg) and diastolic blood pressure (mean difference − 2.497 ± 3.785 mmHg) compared to arterial line measurements. Our approach can potentially be used to measure blood pressure continuously and non-invasively for all patients in the acute care setting, without the need for any additional instrumentation beyond the current standard-of-care.

 
more » « less
NSF-PAR ID:
10283461
Author(s) / Creator(s):
; ; ; ; ; ; ;
Publisher / Repository:
Nature Publishing Group
Date Published:
Journal Name:
Scientific Reports
Volume:
11
Issue:
1
ISSN:
2045-2322
Format(s):
Medium: X
Sponsoring Org:
National Science Foundation
More Like this
  1. Abstract

    Indwelling arterial lines, the clinical gold standard for continuous blood pressure (BP) monitoring in the pediatric intensive care unit (PICU), have significant drawbacks due to their invasive nature, ischemic risk, and impediment to natural body movement. A noninvasive, wireless, and accurate alternative would greatly improve the quality of patient care. Recently introduced classes of wireless, skin‐interfaced devices offer capabilities in continuous, precise monitoring of physiologic waveforms and vital signs in pediatric and neonatal patients, but have not yet been employed for continuous tracking of systolic and diastolic BP—critical for guiding clinical decision‐making in the PICU. The results presented here focus on materials and mechanics that optimize the system‐level properties of these devices to enhance their reliable use in this context, achieving full compatibility with the range of body sizes, skin types, and sterilization schemes typically encountered in the PICU. Systematic analysis of the data from these devices on 23 pediatric patients, yields derived, noninvasive BP values that can be quantitatively validated against direct recordings from arterial lines. The results from this diverse cohort, including those under pharmacological protocols, suggest that wireless, skin‐interfaced devices can, in certain circumstances of practical utility, accurately and continuously monitor BP in the PICU patient population.

     
    more » « less
  2. Abstract

    Wearable devices for continuous monitoring of arterial pulse waves have the potential to improve the diagnosis, prognosis, and management of cardiovascular diseases. These pulse wave signals are often affected by the contact pressure between the wearable device and the skin, limiting the accuracy and reliability of hemodynamic parameter quantification. Here, a continuous hemodynamic monitoring device that enables the simultaneous recording of dual‐channel bioimpedance and quantification of pulse wave velocity (PWV) and blood pressure (BP) is reported. The investigations demonstrate the effect of contact pressure on bioimpedance and PWV. The pulsatile bioimpedance magnitude reached its maximum when the contact pressure approximated the mean arterial pressure of the subject. PWV is employed to continuously quantify BP while maintaining comfortable contact pressure for prolonged wear. The mean absolute error and standard deviation of the error compared to the reference value are determined to be 0.1 ± 3.3 mmHg for systolic BP, 1.3 ± 3.7 mmHg for diastolic BP, and −0.4 ± 3.0 mmHg for mean arterial pressure when measurements are conducted in the lying down position. This research demonstrates the potential of wearable dual‐bioimpedance sensors with contact pressure guidance for reliable and continuous hemodynamic monitoring.

     
    more » « less
  3. Abstract

    Right ventricular (RV) failure remains a significant clinical burden particularly during the perioperative period surrounding major cardiac surgeries, such as implantation of left ventricular assist devices (LVADs), bypass procedures or valvular surgeries. Device solutions designed to support the function of the RV do not keep up with the pace of development of left‐sided solutions, leaving the RV vulnerable to acute failure in the challenging hemodynamic environments of the perioperative setting. This work describes the design of a biomimetic, soft, conformable sleeve that can be prophylactically implanted on the pulmonary artery to support RV ventricular function during major cardiac surgeries, through afterload reduction and augmentation of flow. Leveraging electrohydraulic principles, a technology is proposed that is non‐blood contacting and obviates the necessity for drivelines by virtue of being electrically powered. In addition, the integration of an adjacent is demonstrate, continuous pressure sensing module to support physiologically adaptive control schemes based on a real‐time biological signal. In vitro experiments conducted in a pulsatile flow‐loop replicating physiological flow and pressure conditions show a reduction of mean pulmonary arterial pressure of 8 mmHg (25% reduction), a reduction in peak systolic arterial pressure of up to 10 mmHg (20% reduction), and a concomitant 19% increase in diastolic pulmonary flow. Computational simulations further predict substantial augmentation of cardiac output as a result of reduced RV ventricular stress and RV dilatation.

     
    more » « less
  4. Estimating central aortic blood pressure is important for cardiovascular health and risk prediction purposes. Cardiovascular system is a multi-channel dynamical system that yields multiple blood pressures at various body sites in response to central aortic blood pressure. This paper concerns the development and analysis of an observer-based approach to de-convolution of unknown input in a class of coprime multi-channel systems applicable to non-invasive estimation of central aortic blood pressure. A multi-channel system yields multiple outputs in response to a common input. Hence, the relationship between any pair of two outputs constitutes a hypothetical input-output system with unknown input embedded as a state. The central idea underlying our approach is to derive the unknown input by designing an observer for the hypothetical input-output system. In this paper, we developed an unknown input observer (UIO) for input de-convolution in coprime multi-channel systems. We provide a universal design algorithm as well as meaningful physical insights and inherent performance limitations associated with the algorithm. The validity and potential of our approach was illustrated using a case study of estimating central aortic blood pressure waveform from two non-invasively acquired peripheral arterial pulse waveforms. The UIO could reduce the root-mean-squared error associated with the central aortic blood pressure by up to 27.5% and 28.8% against conventional inverse filtering and peripheral arterial pulse scaling techniques. 
    more » « less
  5. Abstract

    Continuous monitoring of arterial blood pressure is clinically important for diagnosing and managing cardiovascular diseases. Soft electronic devices with skin‐like properties show promise in various applications, including the human‐machine interface, the Internet of Things, and health monitoring. Herein, the use of add‐on soft electronic interfaces addresses the connection challenges between soft electrodes and rigid data acquisition circuitry for bioimpedance monitoring of cardiac signals, including heart rate and cuffless blood pressure is reported. Nanocomposite films in add‐on electrodes provide robust electrical and mechanical contact with the skin and the rigid circuitry. Bioimpedance sensors composed of add‐on electrodes offer continuous blood pressure monitoring with high accuracy. Specifically, the bioimpedance collected with add‐on nanocomposite electrodes shows a signal‐to‐noise ratio of 37.0 dB, higher than the ratio of 25.9 dB obtained with standard silver/silver chloride (Ag/AgCl) gel electrodes. Although the sample set is low, the continuously measured systolic and diastolic blood pressure offer accuracy of −2.0 ± 6.3 mmHg and −4.3 ± 3.9 mmHg, respectively, confirming the grade A performance based on the IEEE standard. These results show promise in bioimpedance measurements with add‐on soft electrodes for cuffless blood pressure monitoring.

     
    more » « less