			<titleStmt><title level='a'>An inter-comparison of autonomous in situ instruments for ocean CO2 measurements under laboratory-controlled conditions</title></titleStmt>
			<publicationStmt>
				<publisher></publisher>
				<date>03/01/2022</date>
			</publicationStmt>
			<sourceDesc>
				<bibl> 
					<idno type="par_id">10324487</idno>
					<idno type="doi">10.1016/j.marchem.2022.104085</idno>
					<title level='j'>Marine Chemistry</title>
<idno>0304-4203</idno>
<biblScope unit="volume">240</biblScope>
<biblScope unit="issue">C</biblScope>					

					<author>Qipei Shangguan</author><author>Adam Prody</author><author>Taylor S. Wirth</author><author>Ellen M. Briggs</author><author>Todd R. Martz</author><author>Michael D. DeGrandpre</author>
				</bibl>
			</sourceDesc>
		</fileDesc>
		<profileDesc>
			<abstract><ab><![CDATA[Observations of the marine CO 2 system are important for understanding ocean acidification, air-sea CO 2 fluxes and the marine carbon cycle in general. A variety of autonomous in situ instruments have been developed and used towards this end. There is, however, a lack of inter-comparison studies of currently available sensor technologies. In this study, a total of 10 instruments including commercially available pH, partial pressure of CO 2 (pCO 2 ), and total alkalinity (A T ) sensors were tested and compared in a 5000 L seawater tank located at Scripps Institution of Oceanography (SIO), California, USA. The test took place over ~12 days (August 16 to 28, 2016)   where conditions in the tank were artificially varied to encompass a wide range of A T , pH and pCO 2 as well as temperature and salinity. To assess accuracy, independent measurements of pH, pCO 2 , dissolved inorganic carbon (DIC) and A T were made using benchtop instrumentation. We also evaluated internal consistency, comparing the measured parameter from the sensors with the calculated parameter, e.g. pCO 2 calculated from A T and pH compared with directly measured pCO 2 . DIC can be precisely derived (within ±5 μmol/kg) over a wide range of conditions from measured pH or pCO 2 paired with A T . Sensor accuracy established by discrete samples is sufficient for short-term and seasonal dynamical studies, but their ability in determining long-term (e.g. climate) variability could not be evaluated because of the brevity of the study. These results provide insights into sensor performance and strategies for data quality control for future studies of ocean acidification and carbon cycling.]]></ab></abstract>
		</profileDesc>
	</teiHeader>
	<text><body xmlns="http://www.tei-c.org/ns/1.0" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:xlink="http://www.w3.org/1999/xlink">
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="1.">Introduction</head><p>The oceans currently act as a sink for atmospheric CO 2 , and are absorbing 25% of modern anthropogenic emissions <ref type="bibr">(Sabine et al., 2004;</ref><ref type="bibr">Watson et al., 2020)</ref>. This uptake of CO 2 has decreased seawater pH and calcium carbonate (CaCO 3 ) mineral saturation state (&#937;), a process known as ocean acidification (OA). Sustained observations of CO 2 chemistry provide critically needed data for understanding not only OA, but spatiotemporal variability and magnitudes of air-sea CO 2 fluxes for use in regional and global carbon budgets <ref type="bibr">(Sutton et al., 2017;</ref><ref type="bibr">Takahashi et al., 2002)</ref>. The temporal resolution of ship-based measurement programs, such as the Hawaii Ocean Time-series <ref type="bibr">(Karl and Lukas, 1996)</ref> and the Bermuda Atlantic Time-series <ref type="bibr">(Michaels and Knap, 1996)</ref> are limited by high capital and labor investments. Such programs are useful for characterizing long-term trends in the open ocean, but cannot capture episodic events, such as phytoplankton blooms <ref type="bibr">(Westberry et al., 2016)</ref> and storms <ref type="bibr">(Wu et al., 2020)</ref> that influence seasonal and interannual CO 2 levels. Hourly to monthly variability on local scales cannot be resolved using surface inorganic carbon data based on sparse sampling. For example, coastal environments have rapid fluctuations in the CO 2 system through biological production and short-term physical influences such as upwelling <ref type="bibr">(Cai et al., 2020;</ref><ref type="bibr">Gac et al., 2020;</ref><ref type="bibr">Harris et al., 2013;</ref><ref type="bibr">Takeshita et al., 2015)</ref>. Ecosystems such as kelp forests and coral reefs often exhibit significant short-term dynamics <ref type="bibr">(Cyronak et al., 2020;</ref><ref type="bibr">Hofmann et al., 2011;</ref><ref type="bibr">Pezner et al., 2021)</ref>, requiring high-frequency measurements to decipher. Compared to ship-based measurements, ocean datasets collected using autonomous, in situ instruments have significant advantages for characterizing these complex systems <ref type="bibr">(Islam et al., 2017;</ref><ref type="bibr">Wang et al., 2019;</ref><ref type="bibr">Claustre et al., 2020)</ref>.</p><p>The inorganic carbon system can be fully characterized by quantifying two of the four commonly measured inorganic carbon parameters along with peripheral variables (e.g. temperature and salinity): total alkalinity (A T ), pH, the partial pressure of CO 2 (pCO 2 ), and dissolved inorganic carbon (DIC) <ref type="bibr">(Millero, 2007;</ref><ref type="bibr">Orr et al., 2018)</ref>. Ideally, we would have autonomous sensors for each of the four parameters, but some of the measurement techniques are challenging to adapt for autonomous measurements because of their complexity, power consumption, and other limitations such as reagent consumption. Instruments are commercially available for measurements of pH (e.g. <ref type="bibr">Seidel et al., 2008;</ref><ref type="bibr">Martz et al., 2010)</ref>, pCO 2 (e.g. <ref type="bibr">DeGrandpre et al., 1995;</ref><ref type="bibr">Friederich et al., 1995;</ref><ref type="bibr">Clarke et al., 2017</ref>) and A T <ref type="bibr">(Spaulding et al., 2014;</ref><ref type="bibr">Shangguan et al., 2021)</ref>. There are also new technologies that potentially further extend our in situ observational capabilities, such as DIC <ref type="bibr">(Liu et al., 2013;</ref><ref type="bibr">Fassbender et al., 2015;</ref><ref type="bibr">Wang et al., 2015)</ref> and A T sensors <ref type="bibr">(Briggs et al., 2017</ref>), but they have not been commercialized and are not yet widely available.</p><p>The assessment of biological and biogeochemical responses to small changes in the CO 2 system requires accurate, precise and stable in situ measurements <ref type="bibr">(Williams et al., 2017)</ref>. Thus, it is important to rigorously evaluate instrument performance under controlled conditions rather than on a remote mooring or other platform where access is limited. The quality of data from varying instruments can be assessed through comparison with bottle samples, over-determinations of the CO 2 system, as well as measuring the same parameter with multiple instruments. Past inorganic carbon inter-comparison studies have focused on shipboard underway pCO 2 <ref type="bibr">(Arruda et al., 2020;</ref><ref type="bibr">K&#246;rtzinger et al., 2000;</ref><ref type="bibr">Macovei et al., 2021)</ref> and underway pH measurements <ref type="bibr">(R&#233;rolle et al., 2016)</ref>, in situ pH sensors <ref type="bibr">(Okazaki et al., 2017)</ref>, and simultaneous measurements of pCO 2 , pH and/or DIC <ref type="bibr">(Wang et al., 2007;</ref><ref type="bibr">Ribas-Ribas et al., 2014;</ref><ref type="bibr">Gray et al., 2011)</ref>. The Alliance for Coastal Technologies (ACT) has also documented performance evaluations of in situ pH and pCO 2 sensors <ref type="bibr">(Tamburri et al., 2011)</ref>. With the increased use and availability of autonomous instruments since most of these past inter-comparisons were completed, there is a continued need for inter-comparison studies of the variety of autonomous inorganic carbon instruments now available.</p><p>The goal of this research is to extensively evaluate a suite of in situ CO 2 system sensors in a controlled environment. To accomplish this, a subset of commercially available autonomous A T , pH, and pCO 2 instruments were tested in a 5000 L seawater tank (Fig. <ref type="figure">1</ref>) at Scripps Institution of Oceanography (SIO). The pH and pCO 2 instruments have been widely used for autonomous in situ seawater measurements (e.g. <ref type="bibr">Gray et al., 2018;</ref><ref type="bibr">DeGrandpre et al., 2020;</ref><ref type="bibr">McLaughlin et al., 2017;</ref><ref type="bibr">Sutton et al., 2019)</ref>, while A T instruments have only recently become available <ref type="bibr">(Spaulding et al., 2014;</ref><ref type="bibr">Pezner et al., 2021)</ref>. The tank chemistry was artificially varied to encompass a wide range of conditions (temperature, salinity, and inorganic carbon species). The 12-day test provides valuable insights into the sensor performance, data quality control, and strategies for using the sensors in studies of inorganic carbon cycling.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.">Materials</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.1.">Instrumentation</head><p>As shown in Table <ref type="table">1</ref>, this study used two SAMI-pHs (Sunburst Sensors, LLC; p66 and p87) <ref type="bibr">(Seidel et al., 2008)</ref>, two SeapHOx sensors (SP020 and SP032) <ref type="bibr">(Bresnahan et al., 2014)</ref> along with an in situ sensor for pCO 2 (Sunburst Sensors, LLC; SAMI-CO 2 ) <ref type="bibr">(DeGrandpre et al., 1995)</ref>. The SAMI-pH is based on the spectrophotometric pH method using meta-cresol purple (mCP) indicator <ref type="bibr">(Clayton and Byrne, 1993)</ref>. A pump and a valve draw in seawater and mix with indicator before spectrophotometric detection. The SeapHOx uses a potentiometric pH sensor, the Honeywell Durafet <ref type="bibr">(Martz et al., 2010)</ref>. The SAMI-CO 2 uses a gaspermeable membrane where a colorimetric pH indicator equilibrates with ambient seawater pCO 2 . The equilibrated solution is pumped into a fiber-optic cell for detection. A benchtop analyzer, the SuperCO 2 (Sunburst Sensors, LLC), was used as a pCO 2 reference measurement <ref type="bibr">(Hales et al., 2004)</ref>. The SuperCO 2 is based on non-dispersive infrared (NDIR) gas detection (model LI-850, Li-COR, Inc.) after the seawater stream equilibrates with a gas stream in a showerhead equilibrator. Three identical A T instruments (SAMI-alk) were also tested (Sunburst Sensors, LLC, alk00, 01, and 03) <ref type="bibr">(Martz et al., 2006;</ref><ref type="bibr">Spaulding et al., 2014)</ref>. The SAMI-alk was developed based on Tracer Monitored Titration (TMT), where an acidified indicator solution is incrementally pumped into a mixed fiber-optic flow cell and used to simultaneously quantify pH and dilution factor (volumetric ratio between indicator solution and sample). A T can then be solved by nonlinear least-square analysis (NLLS) from a sequence of titration points. A CTD (Conductivity, Temperature, and Depth) (Sea-Bird Scientific; SBE 37-SI) recorded temperature and salinity in the tank.</p><p>The 10 instruments were placed in the SIO test tank (Fig. <ref type="figure">1</ref>). The tank is 305 cm in diameter and 84 cm in height and contains ~5000 L of seawater. The tank was filled with seawater pumped from the Scripps Pier one week prior to the experiment. Tank water is first run through  sand filters at the pier. Once in the tank, seawater is sterilized with a DEL Ozone Generator for 3 h and then continuously filtered through a 1 &#956;m bag filter <ref type="bibr">(Part Number F3AB00013,</ref><ref type="bibr">Eaton)</ref>. Seawater is internally circulated through the tank with a pumping rate of ~170 L/min (model DB7P-M227, Finish Thompson, Inc.). Although tank temperature is not under precise control, a manually-operated heat exchange unit allows rapid tank temperature changes on-demand. The tank was open to the atmosphere allowing gas exchange and evaporation. Instruments were deployed close to the edge of the tank for easy-access (Fig. <ref type="figure">1</ref>).</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.2.">Chemicals</head><p>Nanopure water (17.9 M&#937;&#8901;cm specific resistance) was obtained from a Barnstead water purification system (Thermo Scientific) and used for all solutions. Chemicals used to modify the CO 2 chemistry for the tank seawater included 0.1 N HCl (Fischer Scientific) and a solution with A T ~ 0.23 mol/kg and DIC ~ 0.17 mol/kg prepared from dried NaHCO 3 and Na 2 CO 3 powder (analytical grade, Fischer Scientific). This carbonate solution was used to increase A T for the tank seawater (see Seawater CO 2 chemistry in the tank).</p><p>Bromocresol purple sodium salt (BCP) (indicator grade, Sigma-Aldrich), NaCl powder (analytical grade, Fischer Scientific), and 0.1 N HCl were used to prepare titrant solutions for the SAMI-alks <ref type="bibr">(Spaulding et al., 2014)</ref>. The titrant solution contained approximately 6.5 &#215; 10 -5 mol/kg BCP, 8.3 &#215; 10 -4 mol/kg HCl and 0.68 mol/kg NaCl. Salty carbonate standards for SAMI-alks were made from Na 2 CO 3 (ACS primary standard grade, Alfa Aesar) and 0.1 N HCl with 0.68 mol/kg NaCl solution. Alkaline bromothymol blue (BTB) solutions for SAMI-CO 2 contained 5.6 &#215; 10 -5 mol/kg BTB and 8.5 &#215; 10 -5 mol/kg NaOH prepared from BTB powder (ACS Reagent grade, Sigma-Aldrich) and certified 0.1 N NaOH (Fischer Scientific) <ref type="bibr">(DeGrandpre et al., 1995)</ref>. The SAMI-pH used purified mCP (original powder from Sigma-Aldrich, indicator grade mCP). Indicator purification procedures are described in <ref type="bibr">DeGrandpre et al. (2014)</ref>. The mCP concentration was ~3.8 &#215; 10 -4 mol/kg with a pH of 7.8 adjusted by using 0.1 N NaOH and HCl. Indicator and carbonate standard solutions were stored in gas-impermeable bags.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.">Procedure</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.1.">Instrument calibration</head><p>Calibration of the SAMI-alks was performed as described in <ref type="bibr">Spaulding et al. (2014)</ref> using certified reference materials (CRMs) (Batch 158, A T = 2226.6 &#956;mol/kg) <ref type="bibr">(Dickson et al., 2003)</ref>. A calibration factor was obtained by comparing the A T output from SAMI-alks to the certified A T . Calibration factors vary among three SAMI-alks and depend on the combined uncertainties of the optical pathlength and molar absorptivities, as discussed below. In addition to the CRM calibration, synthetic carbonate standards were measured in situ twice per day during the deployment through a three-way selection valve in the SAMIalk.</p><p>Calibration and pre-deployment preparation of the SAMI-CO 2 , SAMI-pH, and SeapHOx were performed as described in <ref type="bibr">DeGrandpre et al. (1995)</ref>, <ref type="bibr">Seidel et al. (2008), and</ref><ref type="bibr">Bresnahan et al., 2014, respectively.</ref> All SAMI instrument calibrations (pCO 2 ) and validations (pH) were performed in the DeGrandpre lab before shipment to the SIO test facility. Specifically, the SAMI-CO 2 was calibrated by determining its response over variable CO 2 concentrations in a water-filled temperaturecontrolled tank while CO 2 was simultaneously monitored by an NDIR CO 2 analyzer (model LI-840, Li-COR, Inc.) with a membrane equilibrator <ref type="bibr">(Hales et al., 2004)</ref>. The SAMI-pH accuracy was validated to within 0.004 pH units using tris buffer (DelValls and <ref type="bibr">Dickson, 1998)</ref>. Calibrations of the SeapHOx systems were performed by the Martz lab group before testing began by applying a single calibration point with reference to spectrophotometric pH (see Seawater CO 2 chemistry in the tank). SuperCO 2 periodically measured standard CO 2 gas at three different concentrations (202.7, 800.0, and 1503.0 ppm), which were used to correct for drift following the descriptions in <ref type="bibr">Dickson et al. (2007)</ref>.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.2.">Seawater CO 2 chemistry in the tank</head><p>The experiment was designed to reduce temporal and spatial mismatch to minimize errors from non-sensor sources. All instruments were configured to start measurements on the hour. The SeapHOx, CTD, SAMI-pH, and SAMI-CO 2 instruments were set to 15 min measurement intervals. The SAMI-alks were set to 1-h intervals due to reagent and power limitations. The SuperCO 2 recorded pCO 2 every four seconds. All data are reported at Universal Time (UTC) throughout this paper.</p><p>Three replicate discrete samples (0.5 L each) were drawn from the tank on the hour for benchtop analyses ~four times per day. A total of samples for pH, 50 samples for A T and 43 samples for DIC were collected. The pH samples were analyzed by spectrophotometry immediately after collection using purified mCP. The A T and DIC samples were poisoned and stored following standard protocol <ref type="bibr">(Dickson et al., 2007)</ref>, and were analyzed within three days using CRMs for quality assurance (Batch 156). A T was determined by open-cell potentiometric titration <ref type="bibr">(Gran, 1952)</ref> and a nonlinear least squares approach <ref type="bibr">(Dickson et al., 2007)</ref>. DIC was determined by acidifying samples and quantifying the extracted CO 2 gas by infrared detection <ref type="bibr">(Goyet and Snover, 1993)</ref>. Bottle sample results were compared to measurements from the autonomous instruments.</p><p>Seven different tests were run during the 12-day experiment listed here in chronological order and labeled in the time-series data in Fig. <ref type="figure">2a:</ref> 1. Static environment (August 16, 21:00): this test was to determine the stability of the instruments keeping conditions as constant as possible. All instruments operated without any purposeful changes in conditions over a 72 h period. 2. Increase A T and DIC (August 19 at 22:00): this test was to determine instrument performance at higher A T and DIC. The A T was raised by adding a 2 L carbonate solution with A T ~ 0.23 mol/kg and DIC ~ 0.17 mol/kg. The solution was used to simulate a natural A T increase of ~90 &#956;mol/kg and ~ 70 &#956;mol/kg increase in DIC and was prepared based on a mass balance calculation. Measurements were made for 24 h after addition of the carbonate solution. 3. Decrease A T with constant DIC (August 20 at 22:00): this test continued to evaluate the instruments' performance over a range of A T , pH, and pCO 2 . The A T was lowered using 1 L of a 0.1 N HCl solution. These conditions were observed for 48 h. 4. Decrease pCO 2 with constant A T (August 22 at 22:00): this test was performed to evaluate instrument response to changes of lowered pCO 2 and elevated pH with constant A T . The pCO 2 was lowered without changing A T by bubbling compressed air through an air stone into the tank. The test was observed for 24 h. 5. Increase pCO 2 with constant A T (August 23 at 22:00): this test was used to evaluate instrument response to elevated pCO 2 and lowered pH changes by dissolving small amounts of dry ice in the tank. The test was observed for 24 h. 6. Temperature change (August 24 at 22:00): this test was performed to evaluate instruments' response with varying temperature. Temperature was decreased by ~7&#8226; C from ambient (24 &#8226; C) using a closed loop cooling tower attached to the test tank. The cooling tower maintained the lower temperature for 24 h, and was shut off thereafter allowing the tank to warm naturally for another 24 h. 7. Salinity change (dilution) (August 26 at 22:00): this period evaluated the instruments' response to changes in salinity by adding 480 L of deionized water to dilute the tank by ~10%. Due to the tank's maximum capacity, ~500 L of seawater was drained from the tank to allow addition of the deionized water. This test was observed for h.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.3.">Data analysis</head><p>To assess sensor performance, we used accuracy defined as the difference between results obtained by sensor and bottle measurements (sensor valuesbottle values). For internal consistency comparisons, the differences are expressed as mean difference (measuredcalculated) &#177; one standard deviation (SD). All pH were determined on the total hydrogen ion concentration scale. pCO 2 was directly used without converting to carbon dioxide fugacity (fCO 2 ). All CO 2 calculations were made using CO2SYS version 2.1 <ref type="bibr">(Pierrot et al., 2006)</ref> and also the more recent version in <ref type="bibr">Orr et al. (2018)</ref> that allows assessment of uncertainties in CO 2 calculations. The carbonic acid dissociation constants of <ref type="bibr">Lueker et al. (2000)</ref>, the total-boron-to-salinity ratio from <ref type="bibr">Lee et al. (2010)</ref>, and K HSO4 from <ref type="bibr">Dickson (1990)</ref> were used. The choice of <ref type="bibr">Lueker et al. (2000)</ref> was to be consistent with recent evaluations of CO 2 system calculations <ref type="bibr">(Fong and Dickson, 2019;</ref><ref type="bibr">Patsavas et al., 2015;</ref><ref type="bibr">Raimondi et al., 2019)</ref>. Phosphate and silicate concentrations were assumed to be zero. Some data sets were averaged or interpolated for internal consistency comparisons. SuperCO 2 data was averaged to 15-min intervals to be in sync with the SAMI-CO 2 and other instruments. SAMI-alk data were interpolated to 15-min intervals. Based on the rapid changes in temperature and salinity, 95% mixing time for the tank was estimated to be 1.5 h and 1.9 h from temperature and salinity data, respectively, by using a first-order reaction model. Sensor and bottle measurements immediately after the addition of the NaHCO 3 /H 2 CO 3 and HCl solutions were excluded from the accuracy assessment because the tank was not fully homogeneous.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.">Results and discussion</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.1.">Tank conditions</head><p>The tank was exposed to the air during the 12-day experiment, allowing for natural CO 2 air-water exchange, temperature fluctuations and evaporation. From August 16-24 temperature slowly decreased from 24.8 &#8226; C to 24.0 &#8226; C while salinity increased from 33.7 to 34.0 due to evaporation (Fig. <ref type="figure">2a</ref>). As indicated above and in Fig. <ref type="figure">2a</ref>, temperature and salinity tests were performed after August 24 at 22:00, artificially altering these parameters. The sawtooth pattern on August 25 is due to the unstable cooling tower temperature regulation with a resolution of ~1-2 &#8226; C. The temperature test was accompanied by a salinity drop of ~0.5 because the seawater stored in the heat exchanger and associated piping remained at a salinity of ~33.5 while the tank water had evaporated to a higher salinity before the test began. This inflow was later found to produce negligible effects on A T (Fig. <ref type="figure">2b</ref>), but a temporary increase of ~40 &#956;mol/kg in DIC (discussed in Internal Consistency section). Dilution (August 26 at 22:00) led to a further salinity decrease of ~3.2. The initial measured A T (2238 &#956;mol/kg) (Fig. <ref type="figure">2b</ref>) is typical of the range of 2200-2250 &#956;mol/kg reported by <ref type="bibr">Bockmon et al. (2013)</ref> for this area. The A T then increased by ~10 &#956;mol/kg due to evaporation during the 3-day static conditions. After the addition of carbonate/bicarbonate solution (August 19 at 22:00), A T increased from ~2250 to 2350 &#956;mol/ kg, followed by a ~ 190 &#956;mol/kg decrease due to acid addition (August 20 at 22:00). The final step of dilution (August 26 at 22:00) led to another ~190 &#956;mol/kg decrease of A T .</p><p>The pH and pCO 2 ranged from ~7.50 to 7.97 (Fig. <ref type="figure">2c</ref>) and ~ 500 to 1430 &#956;atm (Fig. <ref type="figure">2d</ref>), respectively, using the approximate range measured by each sensor. They were relatively stable during the static period with a ~ 0.01 unit decrease in pH and ~ 15 &#956;atm increase in pCO 2 caused by slow gas exchange with the high indoor air pCO 2 . Adding carbonate/bicarbonate solution (August 19 at 22:00) did not dramatically change pH and pCO 2 because they are not strongly affected by simultaneous increase of A T and DIC due to buffering. The largest variation was observed by HCl addition (August 20 at 22:00) which caused a ~ 0.47 unit decrease in pH and ~ 960 &#956;atm increase in pCO 2 .</p><p>After addition of the acid, the CO 2 chemistry evolved by CO 2 degassing into the air, evidenced by gradually increased pH and decreased pCO 2 from August 21 to 22 until more rapid changes were created by bubbling with compressed air (August 22 at 22:00). The dry ice addition (August 23 at 22:00) led to a decrease of pH of ~0.05 pH units and an increase of pCO 2 of ~110 &#956;atm. On August 24 at 22:00, as the temperature dropped by ~7&#8226; C, pH increased by ~0.12 unit and pCO 2 decreased by ~190 &#956;atm. By using CO2sys with bottle A T (2180 &#956;mol/kg) and DIC (2032 &#956;mol/kg) at this time, we determined that the temperature change from 24 &#8226; C to 17 &#8226; C would increase pH by 0.10 units and decrease pCO 2 by 188 &#956;atm, close to the observed changes indicating that the seawater inflow from the cooling system did not strongly affect pCO 2 . The effect of dilution (August 26 at 22:00) on pH and pCO 2 is small, similar to the test of increased A T .</p><p>To summarize, A T , pH, and pCO 2 encompassed a large range characteristic of coastal to open ocean surface waters and deep open ocean waters <ref type="bibr">(Cai et al., 2010;</ref><ref type="bibr">Carter et al., 2016;</ref><ref type="bibr">Jiang et al., 2008;</ref><ref type="bibr">Millero, 2007;</ref><ref type="bibr">Takahashi et al., 2002)</ref>. A detailed analysis of sensor performance is presented in the following sections.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.2.">SAMI-alk data</head><p>SAMI-alk accuracy depends on accuracy of parameters that form its theoretical basis and calculations, which are titrant acidity (C a/i ), combined optical pathlength (b) -molar absorption (&#949;) coefficients (&#949;b), and indicator and carbonate thermodynamic constants <ref type="bibr">(Martz et al., 2006)</ref>. Of these variables, &#949;b is an instrument-dependent property that requires careful experimental determinations for each SAMI-alk. The parameter was only measured on SAMI-alk00 for this experiment, however, and the other instruments' optical cells were disassembled before we realized its variability. With recent evaluations and A T modeling <ref type="bibr">(Shangguan et al., 2021)</ref>, we found that errors in &#949;b create concentration dependent A T errors that are not corrected by the single point CRM calibration. These errors were evident in the raw SAMI-alk data where values were close to the CRM (2226.6 &#956;mol/kg), but increased away from the lower A T level (~20 &#956;mol/kg on August 27). Adjustment of &#949;b was found to reduce these concentration dependent errors; consequently, we used the initial discrete sample A T and final A T (August 27) to determine the optimal &#949;b and this reduced the sensor errors for the entire data set. Optimized &#949;b varied by 1.6% for three SAMI-alks and are within the manufacturing tolerance (~5%) controlled by placement of the optical fibers in the flow cell. Each &#949;b can be precisely determined to ~0.3% based on replicate measurements of known concentrations of indicator species (n = 5 typically). These results highlight the importance of &#949;b determination for SAMI-alk calibration if a wide range of A T is encountered. After &#949;b optimization, the SAMI-alk accuracies relative to discrete samples are 0.9 &#177; 4.8, 0.0 &#177; 5.8 and -0.5 &#177; 6.0 &#956;mol/kg (n = 45) for SAMI-alk00, alk01 and alk02, respectively (Fig. <ref type="figure">3a</ref>). Using SAMI-alk01 as a reference (SAMI-alk00 or alk02 -alk01), the differences for SAMI-alk00 and SAMI-alk02 are -1.4 &#177; 4.2 &#956;mol/kg and -0.2 &#177; 6.6 &#956;mol/kg (n = 240) (Fig. <ref type="figure">3b</ref>). The difference between SAMI-alk00 and SAMI-alk02 is 1.3 &#177; 6.7 &#956;mol/kg (n = 240, not shown).</p><p>Instrument performance can also be evaluated by examining the twice daily measurements of in situ standards (22 total for each SAMI, Fig. <ref type="figure">3c</ref>). The mean values and standard deviations are 2209.7 &#177; 5.4 &#956;mol/kg for SAMI-alk00, 2216.4 &#177; 6.9 &#956;mol/kg for SAMI-alk01, and 2214.1 &#177; 4.0 &#956;mol/kg for SAMI-alk02. Upon further inspection, the first 4 standard measurements of SAMI-alk01 were affected by poor optical blanks likely due to bubbles temporarily entrained in the optical cell. Replacing measured blanks with accurate blanks from other subsequent measurements recover these four standards and improve the results to 2213.8 &#177; 4.1 &#956;mol/kg. Consequently, the in situ standards provide an independent validation of the accuracy and stability of the SAMI-alks during the experiment.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.3.">pH sensor inter-comparison</head><p>All four pH sensors (Table <ref type="table">1</ref>) exhibited excellent accuracy (mean errors &lt;0.004 with one standard deviation &lt; &#177;0.008) compared with discrete pH samples, n = 48 (Figs. <ref type="figure">2c</ref> and<ref type="figure">4a</ref>). These comparisons are also included in Tables <ref type="table">2</ref> and<ref type="table">3</ref> for clarity. The errors for all pH sensors are significantly correlated (p &lt; 0.05, not shown). Around August 21 at the time of high pCO 2 (Fig. <ref type="figure">2</ref>), consistent negative pH errors may have resulted from degassing during sampling and analysis that increased the discrete bottle pH <ref type="bibr">(Bockmon and Dickson, 2015)</ref>. Therefore, the standard deviation of accuracy was at least partially due to the quality of discrete samples (Table <ref type="table">2</ref>). We also noted positive A T errors over the same period (Fig. <ref type="figure">3a</ref>) possibly because of loss of CO 2 in the SAMI-alk titration cell for the higher temperature measurements affecting nonlinear least squares algorithm in its data processing <ref type="bibr">(Spaulding et al., 2014)</ref>. Larger errors after the dilution test (August 26 at 22:00) were found for all pH sensors possibly caused by inadequate mixing in the tank. Samples were taken through a spigot located near the bottom of the tank, not from the sensor locations. In terms of mean errors, SAMI-pH (p66) and two SeapHOx are more consistent (+0.003), but SAMI-pH (p87) has a negative mean error (-0.003), making a significant but constant offset between the two SAMI-pHs (see next).</p><p>Sensor-to-sensor pH comparisons were also performed (Table <ref type="table">3</ref>). In addition to comparison of intra-category pH sensors, data from SAMI-pH (p66) and SeapHOx (SP032) was used as an example of inter-category comparison. The mean difference between the two SAMI-pHs was 0.0059 &#177; 0.0037 pH units (n = 1115) (Fig. <ref type="figure">4b</ref> and Table <ref type="table">3</ref>). The value of &#177;0.0037 is interpreted as the quadratic addition of the noise of SAMI-pH (i.e. &#177;0.0026 = &#177;0.0037/1.4), close to the previously reported value of ~ &#177; 0.003 based on replicate sample measurements <ref type="bibr">(Lai et al., 2018)</ref>. The automated pH perturbation correction is the major source of noise <ref type="bibr">(Lai et al., 2018)</ref>, but this correction is important for accuracy of indicator-based pH measurements <ref type="bibr">(Clayton and Byrne, 1993;</ref><ref type="bibr">Li et al., 2020)</ref>. SAMI-pHs differ slightly in the commercial optical components (e.g. interference filters, LEDs) for construction. Experience has shown that SAMI-pH is accurate within ~0.004 by using the same mCP molar absorptivities for every SAMI-pH to calculate pH. Differences among SAMI-pHs (e.g., 0.0059 in this experiment) can be reduced to ~0.002 by accurately measuring their individual molar absorptivities <ref type="bibr">(Seidel et al., 2008)</ref>. For the comparison of two SeapHOx (Fig. <ref type="figure">4c</ref> and Table <ref type="table">3</ref>), the overall difference was 0.0013 &#177; 0.0053 pH units (n = 1115). Before the temperature testing (August 24 at 22:00), the two SeapHOx showed excellent consistency (0.0011 &#177; 0.0028, n = 841, Fig. <ref type="figure">4c</ref>). The sawtooth pattern in Fig. <ref type="figure">4c</ref> is due to the lagging response of the SeapHOx to the rapidly changing temperature (Fig. <ref type="figure">2a</ref>). The Durafet electrode used on a SeapHOx is mostly situated in an air-filled sensor housing and has a large internal reference gel volume that slows down thermal equilibration <ref type="bibr">(Martz et al., 2010)</ref>. This property is instrument-specific leading to the observed transient difference. The sawtooth pattern is not observed in the SAMI-pH comparison data (Fig. <ref type="figure">4b</ref>) because the indicator equilibrium response is very rapid <ref type="bibr">(Liu et al., 2011)</ref>. However, there are some short spikes from August 25 to 26 (Fig. <ref type="figure">4b</ref>) because optics and electronics inside the SAMI housing have slightly different temperature from that of surrounding waters potentially causing optical drift during rapid temperature changes. The comparison between SAMI-pH (p66) and SeapHOx (SP032) (Fig. <ref type="figure">4d</ref>) resulted in a mean difference of -0.0024 &#177; 0.0045 pH units (n = 1115). Comparisons summarized in Table <ref type="table">3</ref> indicate that there are no statistical differences in pH results from SAMI-pH and SeapHOx. Errors in the SeapHOx caused by the rapid temperature changes were mostly averaged out given the data length.   pH sensor pairs Mean difference &#177; SD SAMI-pH (p66-p87) 0.0059 &#177; 0.0037 SeapHOx (SP020-SP032) -0.0013 &#177; 0.0053 SAMI-pH (p66) -SeapHOx (SP032) -0.0024 &#177; 0.0045 SAMI-pH (p66) -SeapHOx (SP020) -0.0011 &#177; 0.0041 SAMI-pH (p87) -SeapHOx (SP032) -0.0083 &#177; 0.0045 SAMI-pH (p87) -SeapHOx (SP020) -0.0069 &#177; 0.0039 4.4. pCO 2 data assessment SAMI-CO 2 data (shown in Fig. <ref type="figure">2d</ref>) accuracy was evaluated by comparing with calculated pCO 2 from discrete pH and DIC (Fig. <ref type="figure">5a</ref>) and directly measured pCO 2 by the SuperCO 2 (Fig. <ref type="figure">5b</ref>). Around the time when compressed air bubbling was initiated (August 22 22:00, Fig. <ref type="figure">2</ref>), it appears that an adjustment to the SuperCO 2 caused a leak in the connection between the equilibrator and LI-COR (based on large deviations from the SAMI-CO 2 data). Therefore, SuperCO 2 data were only used prior to the leak. SAMI-CO 2 errors relative to the calculated pCO 2 were -9.3 &#177; 17.7 &#956;atm (measuredcalculated pCO 2 , n = 40) (Fig. <ref type="figure">5a</ref>).</p><p>The error is dependent on the pCO 2 value (R 2 = 0.40, p &lt; 0.05). Uncertainty in both the pH and DIC measurements contribute to the pCO 2 uncertainty <ref type="bibr">(Millero, 2007;</ref><ref type="bibr">Orr et al., 2018)</ref> in addition to the carbonate system equilibrium constants (Table <ref type="table">4</ref>). Five groups of bottle data were chosen to cover the pCO 2 , temperature and salinity range in this experiment. Uncertainties assigned to pH (&#177;0.01 pH unit), DIC (&#177;2 &#956;mol/kg), and equilibrium constants were taken from <ref type="bibr">Orr et al. (2018)</ref>.</p><p>Table <ref type="table">4</ref> shows that pCO 2 calculation uncertainty increases with higher pCO 2 (R 2 = 1.00, p &lt; 0.05), suggesting the comparison in Fig. <ref type="figure">5a</ref> is susceptible to high pCO 2 calculation uncertainty. Using pH and A T to calculate pCO 2 did not change the comparison statistically as differences using different pairs were 1.6 &#177; 3.3 &#956;atm (n = 38). Additionally, the high pCO 2 levels after addition of HCl on August 21 from SAMI-CO 2 have larger uncertainty because it was only calibrated from ~200 to 690 &#956;atm. It was difficult to predict the increase in pCO 2 from the dry ice addition because an unquantifiable amount of CO 2 (s) sublimation is lost to the atmosphere. Compared to SuperCO 2 (Fig. <ref type="figure">5b</ref>, SAMI-CO 2 -SuperCO 2 ), a concentration-dependent difference was observed separated by the transient spike on August 20 at 22:00: before the addition of HCl, the difference was -1.6 &#177; 2.7 &#956;atm (n = 364) for pCO 2 between 510 and 530 &#956;atm; the difference increased to 15.5 &#177; 9.3 &#956;atm (n = 168) between 1058 and 1426 &#956;atm. The short spikes on August 18 at 19:30 and August 21 at 11:30 (Fig. <ref type="figure">5b</ref>, red arrows) were caused by a blank measurement by SAMI-CO 2 which pumps deionized water (DI) stored in a reagent bag to the optical cell. This step is typically performed every 255 measurements to improve SAMI-CO 2 long-term stability by accounting for any changes in light intensity, but it can affect subsequent pCO 2 measurement due to the carryover of the blank solution <ref type="bibr">(Lai et al., 2018)</ref>.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.5.">Internal consistency</head><p>Internal consistency can be evaluated using a number of different strategies. We focus on computation of pCO 2 and DIC because these are the two most commonly desired parameters directly used for air-sea flux calculations and water column mass balances, respectively. First, pH and pCO 2 are similar in terms of CO 2 system calculations. pH combined with A T (pH-A T ) yields calculated pCO 2 , which can be compared with the data from SAMI-CO 2 . This comparison is valuable in view of the popularity of utilizing the pH sensors and salinity-derived A T to quantify CO 2 gas exchange (e.g. <ref type="bibr">Gray et al., 2018;</ref><ref type="bibr">Williams et al., 2017)</ref>. A similar comparison can be made for pH by using the pCO 2 -A T pair. Secondly, DIC time series that have been commonly used to model biological and physical processes can be calculated using different sensor pairs, specifically pCO 2 or pH combined with salinity-derived A T (e.g. <ref type="bibr">Briggs et al., 2018;</ref><ref type="bibr">Islam et al., 2017;</ref><ref type="bibr">Martz et al., 2009)</ref>. Previous studies have shown that high noise level results if pCO 2 and pH are used as input parameters <ref type="bibr">(Gray et al., 2011;</ref><ref type="bibr">Raimondi et al., 2019)</ref>. Our calculations  * Uncertainties assigned to the input CO 2 system variables and equilibrium constants were taken from <ref type="bibr">Orr et al. (2018)</ref>. Temperature and salinity were assumed to be perfectly measured (i.e., uncertainty of zero units).</p><p>reaffirm this point (shown below). SAMI-alk02 was randomly selected as input A T among the three SAMI-alks to simplify the analysis. Other SAMI-alk time-series gave similar results and thus, the comparisons are not shown. Results between directly measured values and those calculated from other paired parameters are presented in Table <ref type="table">5</ref>. In terms of pCO 2 , a concentrationdependent difference (p &lt; 0.05) (calculatedmeasured pCO 2 ) is identified for all of the pCO 2 internal consistency tests (Fig. <ref type="figure">6</ref>) that was also evident in the bottle-based comparisons (see pCO 2 data assessment). At high pCO 2 conditions (1058 and 1426 &#956;atm between acid addition and bubbling compressed air), SAMI-CO 2 data errors are very similar to the SuperCO 2 data (Fig. <ref type="figure">6</ref>). Thus, all calculations from different pH and pCO 2 sensors suggest the errors are mostly caused by thermodynamic constant errors. Larger differences (calculatedmeasured pCO 2 ) at high pCO 2 were also previously observed from ship-based studies <ref type="bibr">(McElligott et al., 1998;</ref><ref type="bibr">Patsavas et al., 2015)</ref> that use spectrophotometric pH and DIC to calculate pCO 2 . Fig. <ref type="figure">6</ref> provides a more direct comparison between measured and calculated pCO 2 because it eliminates the difficulty in handling discrete pCO 2 samples and all measurements and calculations were made at in situ temperatures. <ref type="bibr">Fong and Dickson (2019)</ref> proposed a series of systematic adjustments to carbonate thermodynamic constants to reduce differences in measured and calculated pH. These adjustments do not significantly affect calculated pCO 2 from A T and DIC <ref type="bibr">(Fong and Dickson, 2019)</ref>, thus consistency in terms of pCO 2 is not clear when pH is used as an input parameter. Further more systematic studies are needed to improve our understanding of CO 2 system internal consistency.</p><p>Fig. <ref type="figure">6</ref> results do not imply errors for studies that use pH sensors to calculate surface-water pCO 2 because the pCO 2 range (Fig. <ref type="figure">2d</ref>) in this work is above most observed surface values. <ref type="bibr">Sulpis et al. (2020)</ref> verified the carbonate thermodynamic constants in <ref type="bibr">Lueker et al. (2000)</ref> above 8 &#8226; C using combined global ocean data products, but this assessment is restricted to surface-water pCO 2 with a maximum of 446 &#956;atm. In terms of calculations for pH from pCO 2 (calculatedmeasured pH) (Table <ref type="table">5</ref>), they are not as good as the Tables <ref type="table">2</ref> and<ref type="table">3</ref>  ~2075 &#956;mol/kg was observed matching the period when temperature dropped most rapidly (August 24 at 20:00 to August 25 at 2:00). We do not attribute this to poor mixing in the tank because pH sensors and SAMI-CO 2 were located at different positions (Fig. <ref type="figure">1</ref>) and their DIC calculations all have the same pattern. It could be because the seawater from the cooling tower piping system had a different DIC. The final step of dilution (August 26 at 22:00) led to a ~ 185 &#956;mol/kg decrease of DIC.</p><p>Calculated DIC from any of the five pairs closely track the discrete DIC (Fig. <ref type="figure">7a</ref>). Errors (calculateddiscrete DIC) significantly correlate with each other (p &lt; 0.05, not shown), suggesting the uncertainties of the DIC samples, SAMI-alk02 data or carbonate thermodynamic constants dominate this comparison. Large errors are often associated with large DIC variations, for example, addition of carbonate solution on August 19 at 22:00, further highlighting the sampling difficulty. Notably, excellent agreement was obtained among the five calculated DIC time series (Table <ref type="table">6</ref> and Fig. <ref type="figure">7c</ref>) (DIC from pH sensors -DIC from SAMI-CO 2 ). The differences are within &#177;5 &#956;mol/kg, a value comparable to the minimum adjustment value of 4 &#956;mol/kg in GLODAPv.2020 that considers typical shipboard DIC precision <ref type="bibr">(Olsen et al., 2020)</ref>. However, the climate goal of &#177;2 &#956;mol/kg <ref type="bibr">(Bockmon and Dickson, 2015)</ref> has not been fulfilled. This comparison reduces uncertainty consistently applied to DIC calculations from different pH/pCO 2 sensors and A T . Lastly, DIC calculated from pH and pCO 2 is noisy due to this pair's high sensitivity to random noise and is thus not shown in Fig. <ref type="figure">7</ref>. For example, DIC time series calculated from SAMI-pH (p66) and SAMI-CO 2 shows a difference of -14 &#177; 48 &#956;mol/kg (n = 1053) compared with that from SAMI-alk02 and SAMI-CO 2 .</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="5.">Conclusion</head><p>This work compares multiple groups of sensors in a well-controlled environment over a time span of 12 days. A total of 10 instruments were deployed in a 5000 L tank (Table <ref type="table">1</ref>). The tank chemistry was modified to vary A T , pH, pCO 2 , temperature, and salinity while keeping various properties constant for each individual test, which allowed us to examine individual instrument response and the ability to calculate CO parameters over a wide range. Except for SuperCO 2 's leaking connection, all of the instruments worked well with all possible data acquired. Each instruments' performance demonstrated adequate precision and accuracy for capturing the modified conditions representative of natural variability that could be observed in coastal, open ocean and deep ocean environments. Averaged accuracy for three SAMI-alks is 0.1 &#177; 5.5 &#956;mol/ kg. pH sensors including SAMI-pH and SeapHOx have an mean error within 0.004 with one standard deviation within &#177;0.008 pH units. SAMI-CO 2 agrees well with SuperCO 2 within the calibration range (-1.6 &#177; 2.7 &#956;atm). Results in this study have also shown that, like shipboard or laboratory measurement programs, pH or pCO 2 can be combined with A T to fully characterize the CO 2 system. In high CO waters, calculated pCO 2 may be subject to the uncertainties from CO system thermodynamic constants <ref type="bibr">(McElligott et al., 1998)</ref>, but DIC time series calculated from pH-A T and pCO 2 -A T pairs agree to within &#177;5 &#956;mol/kg. Increasingly, in situations where A T sensors are not available, empirical algorithms serve to estimate A T from correlated variables such as salinity and temperature <ref type="bibr">(Lee et al., 2006;</ref><ref type="bibr">Carter et al., 2016)</ref>. This strategy has been adopted in many buoy-based studies (e.g. <ref type="bibr">Gray et al., 2011;</ref><ref type="bibr">Briggs et al., 2018;</ref><ref type="bibr">Harris et al., 2013;</ref><ref type="bibr">Islam et al., 2017)</ref> and a rapidly growing number of papers based on profiling float pH data (e.g. <ref type="bibr">Bushinsky et al., 2019;</ref><ref type="bibr">Williams et al., 2018)</ref>. By comparing calculated</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Table 5</head><p>Results of comparisons between directly measured values and those calculated from other parameters. DIC from different pairs, co-deployment of pH and pCO 2 sensors with estimated A T can be used for in situ data quality control and serve as a sensitive indicator of drift, e.g., due to sensor biofouling <ref type="bibr">(Gray et al., 2011)</ref>. Autonomous sensors that are capable of measuring two parameters are making progress <ref type="bibr">(Wang et al., 2015;</ref><ref type="bibr">Briggs et al., 2017;</ref><ref type="bibr">Briggs et al., 2020)</ref> and give rise to new opportunities for future testing and  applications. This 12-day study cannot detect problems in long-term drift and reliability, which require experiments over longer periods. Challenges such as biofouling and particles are not present in the tank, but commonly affect sensor performance <ref type="bibr">(Manov et al., 2004)</ref>. There are also other sensor-related environmental conditions that are not controlled, such as pressure effects in deep-sea oceans <ref type="bibr">(Johnson et al., 2016)</ref> and large salinity swings caused by tidal cycles <ref type="bibr">(Gonski et al., 2018)</ref>. Therefore, it is paramount that rigorous data quality control be routinely implemented in sensor deployments.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Declaration of Competing Interest</head><p>Michael D. DeGrandpre is a co-owner of Sunburst Sensors, LLC, the company that manufactures the SAMI sensor technologies.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Table 6</head><p>Results of DIC calculations. The DIC calculated from SAMI-CO 2 and SAMI-alk02 was used as a reference (n = 1053). Discrete DIC was also used to determine the accuracy of DIC calculations (n = 38). </p></div></body>