Title: Effect of pulsatility on shear‐induced extensional behavior of Von Willebrand factor
Abstract Background

Patients with continuous flow ventricular assist devices (CF‐VADs) are at high risk for non‐surgical bleeding, speculated to associate with the loss of pulsatility following CF‐VAD placement. It has been hypothesized that continuous shear stress causes elongation and increased enzymatic degradation of von Willebrand Factor (vWF), a key player in thrombus formation at sites of vascular damage. However, the role of loss of pulsatility on the unravelling behavior of vWF has not been widely explored.

Methods

vWF molecules were immobilized on the surface of microfluidic devices and subjected to various pulsatile flow profiles, including continuous flow and pulsatile flow of different magnitudes,dQ/dt(i.e., first derivative of flow rate) of pulsatility and pulse frequencies to mimic in vivo shear flow environments with and without CF‐VAD support. VWF elongation was observed using total internal reflection fluorescence (TIRF) microscopy. Besides, the vWF level is measured from the patients’ blood sample before and after CF‐VAD implantation from a clinical perspective. To our knowledge, this work is the first in providing direct, visual observation of single vWF molecule extension under controlled‐pulsatile shear flow.

Results

Unravelling of vWF (total sample sizen ~ 200 molecules) is significantly reduced under pulsatile flow (p < 0.01) compared to continuous flow. An increase in the magnitude of pulsatility further reduces unravelling lengths, while lower frequency of pulsatility (20 vs. 60 pulses per min) does not have a major effect on the maximum or minimum unravelling lengths. Evaluation of CF‐VAD patient blood samples (n = 13) demonstrates that vWF levels decreased by ~40% following CF‐VAD placement (p < 0.01), which correlates to single‐molecule observations from a clinical point of view.

Conclusions

Pulsatile flow reduces unfolding of vWF compared to continuous flow and a lower pulse frequency of 20 pulses/minute yielded comparable vWF unfolding to 60 pulses/minute. These findings could shed light on non‐surgical bleeding associated with the loss of pulsatility following CF‐VAD placement.

 
more » « less
Award ID(s):
2004475
NSF-PAR ID:
10446025
Author(s) / Creator(s):
 ;  ;  ;  ;  ;  ;  
Publisher / Repository:
Wiley-Blackwell
Date Published:
Journal Name:
Artificial Organs
Volume:
46
Issue:
5
ISSN:
0160-564X
Page Range / eLocation ID:
p. 887-898
Format(s):
Medium: X
Sponsoring Org:
National Science Foundation
More Like this
  1. Background

    Empty nose syndrome (ENS) remains highly controversial, with aggressive inferior turbinate reduction (ITR) or mucociliary dysfunction frequently implicated. However, the appropriate degree of ITR is highly debatable.

    Methods

    We applied individual computed tomography (CT)‐based computational fluid dynamics (CFD) to 5 patients receiving relatively aggressive ITR but with no ENS symptoms, and compared them to 27 symptomatic ENS patients who all had histories of aggressive ITRs, and 42 healthy controls. Patients’ surgical outcomes were confirmed with 22‐item Sino‐Nasal Outcome Test (SNOT‐22) (ITR: 6.40 ± 4.56; ENS: 58.2 ± 15.9; healthy: 13.2 ± 14.9), Nasal Obstruction Symptom Evaluation (NOSE) scores (ITR: 4.00 ± 2.24; ENS: 69.4 ± 17.1; healthy: 11.9 ± 12.9), and Empty Nose Syndrome 6‐Item Questionnaire (ENS6Q) (≥11 for ENS).

    Results

    Both aggressive ITR without ENS symptoms and symptomatic ENS patients had significantly lower nasal resistance (ITR: 0.059 ± 0.020 Pa·s/mL; ENS: 0.052 ± 0.015 Pa·s/mL; healthy: 0.070 ± 0.021 Pa·s/mL) and higher cross‐sectional areas surrounding the inferior turbinate (ITR: 0.94 ± 0.21 cm2; ENS: 1.19 ± 1.05 cm2; healthy: 0.42 ± 0.22 cm2) than healthy controls. The lack of significant differences among patient groups indicated similar degrees of surgeries between ITR with and without ENS symptom cohorts. However, symptomatic ENS patients have paradoxical significantly less airflow in the inferior meatus (ITR: 47.7% ± 23.6%; ENS: 25.8% ± 17.6%; healthy: 36.5 ± 15.9%; bothp< 0.01), but higher airflow around the middle meatus (ITR: 49.7% ± 22.6%; ENS: 66.5% ± 18.3%; healthy: 49.9% ± 15.1%,p< 0.0001) than aggressive ITR without symptoms and controls. Aggressive ITR patients have increased inferior meatus airflow as expected (p< 0.05). This imbalanced airflow produced less inferior wall‐shear‐stress distribution among symptomatic ENS patients only (ITR: 42.45% ± 11.4%; ENS: 32.2% ± 12.6%; healthy: 49.7% ± 9.9%). ENS patients (n = 12) also had impaired nasal trigeminal function, as measured by menthol lateralization detection thresholds (ITR: 15.2 ± 1.2; ENS: 10.3 ± 3.9; healthy: 13.8 ± 3.09, bothp< 0.0001). Surprisingly, aggressive ITR patients without ENS symptoms have better menthol lateralization detection thresholds (LDTs) than healthy controls.

    Conclusion

    Although turbinate tissue loss is linked with ENS, the degree of ITR that might distinguish postoperative patient satisfaction in their nasal breathing vs development of ENS symptoms is unclear. Our results suggest that a combination of distorted nasal aerodynamics and loss of mucosal sensory function may potentially lead to ENS symptomology.

     
    more » « less
  2. Abstract

    Ventricular assist devices (VADs), among which the HeartMate 3 (HM3) is the latest clinically approved representative, are often the therapy of choice for patients with end‐stage heart failure. Despite advances in the prevention of pump thrombosis, rates of stroke and bleeding remain high. These complications are attributed to the flow field within the VAD, among other factors. One of the HM3’s characteristic features is an artificial pulse that changes the rotor speed periodically by 4000 rpm, which is meant to reduce zones of recirculation and stasis. In this study, we investigated the effect of this speed modulation on the flow fields and stresses using high‐resolution computational fluid dynamics. To this end, we compared Eulerian and Lagrangian features of the flow fields during constant pump operation, during operation with the artificial pulse feature, and with the effect of the residual native cardiac cycle. We observed good washout in all investigated situations, which may explain the low incidence rates of pump thrombosis. The artificial pulse had no additional benefit on scalar washout performance, but it induced rapid variations in the flow velocity and its gradients. This may be relevant for the removal of deposits in the pump. Overall, we found that viscous stresses in the HM3 were lower than in other current VADs. However, the artificial pulse substantially increased turbulence, and thereby also total stresses, which may contribute to clinically observed issues related to hemocompatibility.

     
    more » « less
  3. Abstract

    Stress suppresses pulsatile luteinising hormone (LH) secretion in a variety of species, although the mechanism underlying this inhibition of reproductive function remains unclear. Metabolic stress, particularly hypoglycaemia, is a clinically‐relevant stress type that is modelled with bolus insulin injection (insulin‐induced hypoglycaemia). The present study utilised ovariectomised C57BL/6 mice to test the hypothesis that acute hypoglycaemia suppresses pulsatile LH secretion via central mechanisms. Pulsatile LH secretion was measured in 90‐minute sampling periods immediately prior to and following i.p. injection of saline or insulin. The secretion of LH was not altered over time in fed animals or acutely fasted (5 hours) animals following an i.p. saline injection. By contrast, insulin elicited a robust suppression of pulsatile LH secretion in fasted animals, preventing LH pulses in five of six mice. To identify the neuroendocrine site of impairment, a kisspeptin challenge was performed in saline or insulin pre‐treated animals in a cross‐over design. LH secretion in response to exogenous kisspeptin was not different between animals pre‐treated with saline or insulin, indicating normal gonadotrophin‐releasing hormone cell and pituitary responses during acute hypoglycaemia. Based on this finding, the effect of insulin‐induced hypoglycaemia on arcuate kisspeptin (Kiss1) cell function was determined using c‐Fos as a marker of neuronal activation. Insulin caused a significant suppression in the percentage of Kiss1 cells in the arcuate nucleus that contained c‐Fos compared to saline‐injected controls. Taken together, these data support the hypothesis that insulin‐induced hypoglycaemia suppresses pulsatile LH secretion in the female mouse via predominantly central mechanisms, which culminates in the suppression of the arcuate Kiss1 population.

     
    more » « less
  4. Abstract

    Microfluidics has enabled a revolution in the manipulation of small volumes of fluids. Controlling flows at larger scales and faster rates, ormacrofluidics, has broad applications but involves the unique complexities of inertial flow physics. We show how such effects are exploited in a device proposed by Nikola Tesla that acts as a diode or valve whose asymmetric internal geometry leads to direction-dependent fluidic resistance. Systematic tests for steady forcing conditions reveal that diodicity turns on abruptly at Reynolds number$${\rm{Re}}\approx 200$$Re200and is accompanied by nonlinear pressure-flux scaling and flow instabilities, suggesting a laminar-to-turbulent transition that is triggered at unusually low$${\rm{Re}}$$Re. To assess performance for unsteady forcing, we devise a circuit that functions as an AC-to-DC converter, rectifier, or pump in which diodes transform imposed oscillations into directed flow. Our results confirm Tesla’s conjecture that diodic performance is boosted for pulsatile flows. The connections between diodicity, early turbulence and pulsatility uncovered here can inform applications in fluidic mixing and pumping.

     
    more » « less
  5. Hagemeyer, Christoph E (Ed.)
    The search persists for a safe and effective agent to lyse arterial thrombi in the event of acute heart attacks or strokes due to thrombotic occlusion. The culpable thrombi are composed either primarily of platelets and von Willebrand Factor (VWF), or polymerized fibrin, depending on the mechanism of formation. Current thrombolytics were designed to target red fibrin-rich clots, but may be not be efficacious on white VWF-platelet-rich arterial thrombi. We have developed an in vitro system to study the efficacy of known and proposed thrombolytic agents on white clots formed from whole blood in a stenosis with arterial conditions. The agents and adjuncts tested were tPA, ADAMTS-13, abciximab, N-acetyl cysteine, and N,N’-Diacetyl-L-cystine (DiNAC). Most of the agents, including tPA, had little thrombolytic effect on the white clots. In contrast, perfusion of DiNAC lysed thrombi as quickly as 1.5 min, which ranged up to 30 min at lower concentrations, and resulted in an average reduction in surface area of 71 ± 20%. The clot burden was significantly reduced compared to both tPA and a saline control ( p <0.0001). We also tested the efficacy of all agents on red fibrinous clots formed in stagnant conditions. DiNAC did not lyse red clots, whereas tPA significantly lysed red clot over 48 h ( p <0.01). These results lead to a novel use for DiNAC as a possible thrombolytic agent against acute arterial occlusions that could mitigate the risk of hyper-fibrinolytic bleeding. 
    more » « less