			<titleStmt><title level='a'>Characterization of a Spatially resolved multi-element laser ablation ion source</title></titleStmt>
			<publicationStmt>
				<publisher></publisher>
				<date>02/01/2022</date>
			</publicationStmt>
			<sourceDesc>
				<bibl> 
					<idno type="par_id">10331292</idno>
					<idno type="doi">10.1016/j.ijms.2021.116763</idno>
					<title level='j'>International Journal of Mass Spectrometry</title>
<idno>1387-3806</idno>
<biblScope unit="volume">472</biblScope>
<biblScope unit="issue">C</biblScope>					

					<author>K. Murray</author><author>C. Chambers</author><author>D. Chen</author><author>Z. Feng</author><author>J. Fraser</author><author>Y. Ito</author><author>Y. Lan</author><author>S. Mendez</author><author>M. Medina Peregrina</author><author>H. Rasiwala</author><author>L. Richez</author><author>N. Roy</author><author>R. Simpson</author><author>J. Dilling</author><author>W. Fairbank</author><author>A.A. Kwiatkowski</author><author>T. Brunner</author>
				</bibl>
			</sourceDesc>
		</fileDesc>
		<profileDesc>
			<abstract><ab><![CDATA[A laser ablation ion source (LAS) is a powerful tool by which diverse species of ions can be produced for mass spectrometer calibration or surface study applications. It is necessary to frequently shift the laser position on the target to selectively ablate materials in a controlled manner, and to mitigate degradation of the target surface caused by ablation. An alternative to mounting the target onto a rotation wheel or x À y translation stage, is to shift the laser spot position with a final reflection from a motorized kinematic mirror mount. Such a system has been developed, assembled and characterized with a two axis motorized mirror and various metal targets. In the system presented here, ions are ablated from the target surface and guided by a 90 quadrupole bender to a Faraday cup where the ion current is measured. Spatially resolved scans of the target are produced by actuating the mirror motors, thus moving the laser spot across the target, and performing synchronous measurements of the ion current to construct 2D images of a target surface which can be up to 50 mm in diameter. The spatial resolution of the system has been measured by scanning the interfaces between metals such as steel and niobium, where it was demonstrated that the LAS can selectively ablate an area of diameter z50 mm. This work informs the development of subsequent LAS systems, that are intended to serve as multi-element ion sources for commercial and custom-built time-of-flight mass spectrometers, or to selectively study surface specific regions of samples.]]></ab></abstract>
		</profileDesc>
	</teiHeader>
	<text><body xmlns="http://www.tei-c.org/ns/1.0" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:xlink="http://www.w3.org/1999/xlink">
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="1.">Introduction</head><p>The use of lasers in the production of ion beams, whether by resonance ionisation or ablation, has been demonstrated in many applications, see e.g. <ref type="bibr">Refs. [1e9]</ref>. Ions may be produced in vacuum, or at higher pressures with different species of ambient gases <ref type="bibr">[10]</ref>. Since the process of ablation is inherently destructive, it is typically necessary to shift the laser spot position on a target as the target surface degrades. Moreover, it may be desirable to selectively ablate specific areas of a multi-element target for applications such as mass spectrometer calibration. In high-precision mass-spectrometry applications, the positioning of a laser beam on the target surface is typically achieved by mounting the target onto a rotating wheel or x &#192; y translation stage <ref type="bibr">[11e18]</ref>. In a recent development, the laser spot position was randomised by the movement of an external focusing lens which was mounted onto an x &#192; y translation stage <ref type="bibr">[19]</ref>. An alternative method, presented here, is to shift the laser spot position on the target with the final reflection into the vacuum chamber, using a high-precision motorized kinematic mirror mount.</p><p>The Laser Ablation Ion Source (LAS) presented in this work is being developed to provide ions with well known mass to calibrate a multi-reflection time of flight mass spectrometer <ref type="bibr">[20]</ref>. This development is part of the Ba-tagging development <ref type="bibr">[21]</ref> that is being pursued for a potential upgrade to the nEXO detector <ref type="bibr">[22]</ref>.</p><p>The LAS uses a pulsed UV laser to produce ion bunches from a custom target in an Ultra-High Vacuum (UHV) system. For ablation to occur, an incident laser spot fluence of approximately 0.1e10 J/ cm 2 is required <ref type="bibr">[23e27]</ref>. Target materials are chosen such that they span a large mass range for the purposes of mass-spectrometer calibration. The resulting ion bunches are guided by a quadrupole bender to a Faraday cup where the ion current is measured to characterize the performance of the setup. A 2D image of the target can be reproduced by rastering the ablation laser spot across the surface of the target and recording the ion current at each location. Since the ablation yield is material dependent <ref type="bibr">[28]</ref>, the measured ion current changes for different materials, thus it is possible to distinguish physical features of the target. The resolution of physical features of the target allows for the establishment of a coordinate system on the target surface, which can be used to selectively ablate different materials. This work characterizes the spatial resolution with which the LAS may selectively ablate the surface of a target.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.">Experimental setup</head><p>The LAS consists of a vacuum chamber where the ions are produced and measured, attached to an optical breadboard where the laser beam optics are located. The vacuum chamber is a 6-way DN160 CF cross, which houses the ion-source assembly, 90 DC quadrupole deflector and a Faraday cup. UHV conditions are achieved by a turbomolecular pump attached to the bottom flange of the cross. The pressure is monitored with a cold cathode gauge, and is typically in the range of 1 &#194; 10 &#192;7 to 1 &#194; 10 &#192;8 mbar. A schematic drawing of the setup is shown in Fig. <ref type="figure">1</ref>.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.1.">Laser scanning</head><p>Ions are produced by a pulsed 349 nm neodymium-doped yttrium lithium fluoride (Nd:YLF) UV Spectra Physics Explorer 349 120. A series of mirrors and lenses focus the laser beam into the vacuum chamber. The UV laser can be pulsed with a variable frequency ranging from 1 Hz to 5 kHz, with a nominal energy of 120 mJ per pulse, and a pulse width &lt;5 ns <ref type="bibr">[29]</ref>. A visible laser is included in the setup to ease the alignment process.</p><p>The UV laser beam is focused onto the target by a plano-convex lens with a 750 mm focal length, which is mounted onto a motorized stage oriented parallel to the beam path, so that the location of the focal plane may be remotely adjusted to optimize ablation at the target surface. After passing through the focusing lens, the laser beam is reflected into the chamber and onto the target by a mirror on a computer-controlled motorized mount. Motorized mirror mounts are used to selectively ablate materials off of a fixed target (Fig. <ref type="figure">1 (l)</ref>). Two different motorized mirror mounts were used in this study. The first is a KS1-Z8 mount from Thorlabs with Z-812 series DC motor actuators. The second is a Physik Instrumente (PI) mount with N-472 linear actuators and E&#192;871 servo controllers. The Thorlabs mount has a travel range of 12 mm, a minimum incremental movement of 0.2 mm, and a bidirectional repeatability of &lt; 1:5 mm. The PI mirror has a travel range of 13 mm, a minimum incremental movement of 0.05 mm, and a unidirectional repeatability of 0.2 mm.</p><p>The mirror motors are controlled by a LabVIEW program that steps the laser spot in a rasterized grid across the target's surface, while recording the ion current measured at each point in the grid with the Faraday cup (Fig. <ref type="figure">1</ref> (p)). In this manner, one and two dimensional scans of a target may be produced.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.2.">Ion source and ion optics</head><p>The ion-source assembly, shown in Fig. <ref type="figure">2</ref>, enables the mounting of custom targets in a stainless steel target-holder. Ions produced by ablation are accelerated by a series of five extraction electrodes, after which they are deflected by a 90 quadrupole bender to either a Faraday cup or another device such as a mass spectrometer. The 90 quadrupole bender is formed by 25.4 mm diameter stainless steel rods, which form a square with a center-to-center spacing of 59.1 mm between adjacent rods. The quadrupole bender's widely spaced electrodes allow surface scans over ranges as large as 50 mm in width.</p><p>DC potentials are applied to the ion source, quadrupole bender and Faraday cup with Rohde and Schwarz HMC804 power supplies, which can output up to 32V per channel. In the current work, only the Faraday cup has been used since the mass spectrometer is still under development. The ion current produced by the ion source is typically measured with an Agilent 34465A digital multimeter connected to the Faraday cup. For direct measurement of the ion transport efficiency through the ion-source assembly and the quadrupole bender, the ion currents leaving the target and arriving at the Faraday cup were measured simultaneously using two Keithley 6485 picoammeters. For measurement of the ion transport efficiency, the LAS had been upgraded by replacing the diverging and converging lenses with a Thorlabs BE10-UVB beam expander to create a smaller laser diameter on the target. A cylindrical Cu only target was used to investigate the homogeneity of the ion transport efficiency across the target surface with a 2D scan.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.">Target scanning in 2D</head><p>The actuated mirror moves the laser beam in two directions across the target surface. The directions of movement are to first The beam is then focused onto a target with a f &#188; 750 mm UVFS plano-convex lens (j), which is mounted onto a motorized stage parallel to the optical axis. A knife-edge (i) and power meter (k) are used to measure the laser beam profile and energy. The beam is reflected onto the target by a UV-enhanced mirror mounted in a motorized kinematic mirror mount (l), which allows for precise manipulation of the laser spot on the target. The laser beam enters into the vacuum chamber through a UV transparent window (m), to strike the target (n). Once ions are produced from the target, they are accelerated by a series of ring electrodes. The ions then pass through a 90 quadrupole bender (o), to either a Faraday cup (p), or through an einzel lens (q) to another device. For measurements of ion transport efficiency, the diverging and converging lenses were replaced with a Thorlabs BE10-UVB beam expander (s). order perpendicular to each other. The system is calibrated so that the distance actuated by the mirror stages can be converted to distance travelled on the target surface. The inner diameter of the stainless steel target holder is used as a reference to construct the target coordinate system from a 2D scan of the target. This coordinate system can be described simply with the following transformation &#240;x 0 ; y 0 &#222; &#188; &#240;S x x; S y y&#222;;</p><p>where (x, y) are the coordinates of the mirror stages, (S x , S y ) are scaling factors, and (x 0 , y 0 ) are the target coordinates.</p><p>A multi-element target consisting of gold, niobium and copper (AueNbeCu) was scanned in 2D by the LAS. Scans of the AueNbeCu target are used here to demonstrate a measurement of the scaling factors described by Eq. ( <ref type="formula">1</ref>). The AueNbeCu target is formed by a hollow niobium cylinder, with a rectangular copper rod in its center held in place by stainless steel set screws. A 0.1 mm thick piece of gold foil covers one half of the niobium face. The target is shown mounted into the target holder in Fig. <ref type="figure">3</ref>. These metals have been chosen such that ions from the target span a significant mass range, for the benefit of mass-spectrometer calibration.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.1.">Determination of scaling factors</head><p>The scaling factors were calculated for each mirror, by creating a 2D scan of the target and fitting an ellipse to the diameter of the target holder. Two examples are shown in Fig. <ref type="figure">4</ref> for the PI and Thorlabs mirror mounts on the top and bottom respectively. The ellipse was fit to a total of 30 points which were selected manually at the interface of the target and the target holder. The diameter of the target holder was measured as D TH &#188; 7.8 (1) mm, S x and S y were calculated as the ratio of D TH with the best-fit values for the semiminor and semi-major axes respectively. The measured scaling factors for both mirrors are summarized in Table <ref type="table">1</ref>, where the smallest step size on the target surface for each mirror is also presented. As discussed in Section 4, the spatial resolution is limited by the diameter of the laser spot on the target, and not the smallest step size. Thus, both mirror mounts are able to reconstruct the physical features of the target with similar performance. Small differences in the mean ion current between the two scans are not caused by the use of different motorized mirror mounts, but by slight differences in the optical setup, as the scans presented here were taken at different stages of the development of the technique. Additionally, the surface of the target was scanned repeatedly over the development period, and differences in the amount of damage accumulated on the target surface could result in differences in the measured ion current.</p><p>The large spacing between the quadrupole electrodes allows for scanning over ranges as large as 50 mm. To demonstrate this, the focusing lens, which is mounted onto a motorized stage, was adjusted so that the laser beam was focused on the surface of the final accelerating electrode. A 2D scan of the full electrode is shown on top in Fig. <ref type="figure">5</ref>. By fitting an ellipse to the inner diameter, the scaling factors were measured as S x &#188; 19.0 (3) and S y &#188; 12.8 (2) for the PI mirror. The scan of the full electrode was taken over the period of a day with step sizes of 475 (8) mm and 320 (5) mm in x and y respectively. A detailed scan of one of the bolts securing the electrode is shown at the bottom of Fig. <ref type="figure">5</ref> with a 10-fold smaller step size. The scaling factors were found to remain consistent between measurements, but the system needs to be re-calibrated if there are changes to the optical setup.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.2.">Measurement of laser spot step size</head><p>An optical microscope was used to directly measure the distance between ablated spots, to confirm that the scaling factors accurately predict the laser spot step size on the target surface. A thin Cu foil was scanned over the full target range using the PI mirror  mount, with a laser beam energy of 18.1 (2) mJ per pulse at 500 Hz and a fixed step size of 20 mm in mirror coordinates in both the x and y directions. Since the laser emitted pulses while the spot was being moved on the target surface, there were ablation craters as well as tracks that connect them. To include scans at a different laser beam energy, two opposite quarters of the target were scanned a second time with a higher laser beam energy of 28.8 (3) mJ The average x and y distances between the center positions of 4 elliptical ablation craters, which are indicated with red dashed lines in Fig. <ref type="figure">6</ref>, were measured at &#194; 20 magnification with an optical microscope. The ablation craters are elliptical in shape, with the average of the four major and minor axes measured as 68 (5) mm and 39 (2) mm respectively. The distance in x between ablation points was measured as Dx &#188; 441.6 (6) mm, and the distance in y was measured as Dy &#188; 305.1 <ref type="bibr">(9)</ref> mm. This measured distance is in agreement with the predicted distances travelled in x and y of 438 <ref type="bibr">(6)</ref> mm and 302 (4) mm, respectively, calculated using the scaling factors S x &#188; 21.9 (3) and S y &#188; 15.1 (2). It was found that the first step taken in y, after the mirror had reversed the scanning direction, was shorter than the subsequent steps in y. Thus there is a shift in y between adjacent scan lines. For reference, the difference in the y position of adjacent craters was measured as 66 (2) mm.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.">Spatial resolution</head><p>The spatial resolution at which surface features can be resolved with the LAS system is limited by either the diameter of the ablated crater, or the smallest step size that the laser beam can move on the </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Table 1</head><p>The measured scaling factors for x and y for the PI and Thorlabs actuated mirrors, obtained through the fit of an ellipse as demonstrated in Fig. <ref type="figure">4</ref>.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Mirror Mount</head><p>Scaling  target surface. The diameter of the ablated crater can be measured directly with an optical microscope, for comparison with the smallest achievable steps calculated in Table <ref type="table">1</ref>. Alternatively, the diameter of the ablated crater can be estimated theoretically with the following method.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.1.">Ablation diameter</head><p>The diameter of the ablated crater is related to the focused 1/e 2 beam diameter, D f , which can be calculated from a measurement of the laser beam width before the focusing lens, using the relationship <ref type="bibr">[30,</ref><ref type="bibr">31]</ref>.</p><p>where f is the lens' focal length, l the laser wavelength, D is the 1/e 2 diameter of the beam as it enters the lens and M 2 is the beam quality parameter. It is important to note that material is only ablated where the laser spot fluence is higher than the ablation threshold of the material being ablated. The relationship between the ablated crater diameter, D a , and the threshold fluence, F th , is given by [32e34].</p><p>where the peak fluence, F 0 , is</p><p>and E T is the total beam energy per pulse. It is also important to note that the absorptivity of the irradiated zone changes after multiple pulses. The threshold fluence after N pulses, F th (N) is related to the single-shot threshold fluence by a power law <ref type="bibr">[35]</ref>.</p><p>F th &#240;N&#222; &#188; F th &#240;1&#222;N z&#192;1 ;</p><p>(5</p><p>where z is known as the incubation coefficient, and F th (1) is the fluence threshold of the first laser shot on a fresh target. Since the threshold fluence changes, D a will vary depending on the number of times a particular spot has been ablated, and damage accumulation will affect the spatial resolution for repeated pulses. The profile of the laser beam before focusing was measured by mounting a razor blade on a micrometer stage and moving the blade in steps of 0.25 mm across the beam, while measuring the average laser beam energy that passes the blade for 10 s at each step. The laser beam energy was measured after the focusing lens and before reflection by the motorized mirror. The resulting profile was fit with an integrated Gaussian function, from which the 1/e 2 diameter D &#188; 3.9 (1) mm was extracted.</p><p>The M 2 beam quality parameter was specified for the laser as M 2 &lt; 1.3 <ref type="bibr">[29]</ref>. The M 2 parameter was also measured, by deflecting the laser beam after focusing away from the motorized mirror mount, and taking knife-edge measurements of the beam intensity profile near and around the focus, as outlined in Ref. <ref type="bibr">[31]</ref>. The beam quality was measured as M 2 &#188; 1.32 <ref type="bibr">(7)</ref>, which agrees with the laser specifications. Thus D f was calculated using Eq. ( <ref type="formula">2</ref>), with f &#188; 750 <ref type="bibr">(8)</ref> mm, l &#188; 349 nm, and M 2 &#188; 1.32 <ref type="bibr">(7)</ref> as D f &#188; 113 ( <ref type="formula">7</ref>) mm, with a peak fluence of 0.9 J/cm 2 . The focus of the laser beam on the target surface was optimized prior to this profile measurement by adjusting the position of the focusing lens, and maximizing the ion current produced from the steel target holder.</p><p>To calculate the diameter of the ablated crater with Eq. ( <ref type="formula">3</ref>), F th <ref type="bibr">(1)</ref> and z must be known. The incubation parameter z is sensitive to the pulse duration, and will typically have a value between 0.8 and 0.9 for femtosecond and nanosecond pulses respectively <ref type="bibr">[36]</ref>. However, the single-shot fluence threshold F th (1) can vary by orders of magnitude depending on the wavelength [37e39], pulse duration and repetition rate <ref type="bibr">[40]</ref>, material composition and beam diameter <ref type="bibr">[41,</ref><ref type="bibr">42]</ref>. Thus it is difficult to choose a value for F th (1) when calculating the ablated crater diameter for this experiment.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.2.">Scanning metal junctions</head><p>The metals of the multi-metal targets can be distinguished by the amplitude of emitted ion current produced at constant average laser beam power. This fact is utilized to perform a measurement of the spatial resolution for the setup. When the laser spot was scanned across the junction of two metals, the ion current appeared as a smeared step function. The width of this transition, D t , was measured by fitting an integrated Gaussian function to the resulting distribution, from which the spatial resolution was extracted as the 1/e 2 diameter. Both motorized mirror mounts were capable of repeatable increments smaller than 4 mm on the target surface, which is significantly smaller than the focused laser spot diameter on the target. Thus the resolution was dominated by the laser spot diameter on the target, and both mirror actuator types were found to have similar spatial resolutions.</p><p>The spatial resolution was measured with the AueNbeCu target as well as an AleSieAu target. The AleSieAu target consisted of a 0.4 mm thick piece of aluminum foil and 0.1 mm thick gold foil on top of a silicon wafer, such that the upper portion of the face is covered with gold and the lower portion is covered by aluminum, with a 2 mm gap of silicon in between. This target provides straight edges of gold and aluminum that span the diameter of the target, allowing for more scans of metal transitions to be made before needing to replace the target. For the experiments presented here, the ion current was averaged over a time window of 0.25 s for each scan position, and the standard deviation was calculated and is presented as the statistical uncertainty. The laser repetition rate was fixed at 1000 Hz and measurements of the laser beam energy were taken at the location just before the beam reflects from the motorized mirror. The following measurements were performed with the PI mirror mount by scanning targets in the y direction, thus the scaling factor S y &#188; 14.9 (2) is used to calculate the results. The junction between the steel target-holder and the niobium component of the AueNbeCu target was scanned, and the resulting profile fit with an integrated Gaussian function. The 1/e 2 full width of the transition was extracted as D t &#188; 50 (3) mm, with the laser beam energy per pulse measured as 33 (1) mJ. The AleSieAu target was then installed such that the aluminum and gold edges were perpendicular to the y-direction. The junction of silicon and gold foil was scanned, and the width extracted as D t &#188; 47 (7) mm, with the laser beam energy measured as 42 (1) mJ. These scans are displayed on the top and bottom of Fig. <ref type="figure">7</ref> for stainless steel-niobium and silicon-gold foil interfaces, respectively. Under the described experimental conditions no ablation was observed on the Al target. Using eqs. ( <ref type="formula">3</ref>)e( <ref type="formula">5</ref>), the threshold fluence for stainless steel was estimated. With D a &#188; 50 (3) mm, E T &#188; 33 (1)mJ and D f &#188; 113 <ref type="bibr">(7)</ref> mm, the single shot threshold fluence could be expressed as F th &#240;1&#222; &#188; 0:4</p><p>J/cm 2 . Using the laser repetition rate, f r , the measurement time, t m , the scan step size, Ds, and the beam diameter, the number of overlapping pulses was estimated as N z f r &#194; t m &#194; D f /Ds z 1000 &#194; 0.25 &#194; 113/2 z 14 000. Thus the single-shot threshold fluence was calculated as F th (1) ~1.2 J/cm 2 , assuming an incubation coefficient of z &#188; 0.9 <ref type="bibr">[36]</ref>. It could be argued that the choice of D f as the overlap length to be divided by Ds is somewhat arbitrary, and that N falls within a range of values. However, the value of F th (1) is not strongly dependent on N, provided that N $ O&#240;1000&#222; pulses. For example, if D a had been chosen as the overlap length instead, then N z 6250 and F th (1) ~1.0 J/cm 2 . Taking a range of values for N from 6250 to 14 000 results in an estimate for the single-shot threshold fluence of F th (1) ~1.0 &#192; 1.2 J/cm 2 . This result is comparable to the results of a study where, for a comparable beam diameter, the threshold fluence after 100 pulses was found to be F th (100) ~0.8 J/cm 2 for stainless steel <ref type="bibr">[41]</ref>. Assuming the same incubation coefficient, this corresponds to a single-shot threshold fluence of F th (1) ~1.3 J/cm 2 .</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="5.">Ion transport efficiency</head><p>The ion transport efficiency, T eff , is measured as the ratio of the ion current arriving at the Faraday cup, I FC , to the ion current leaving the target holder, I TH , T eff &#188; I FC I TH :</p><p>However, the process of laser ablation does not only release singly charged ions, but electrons, neutrals and multiply-charged ions as well <ref type="bibr">[43]</ref>. Thus, Eq. ( <ref type="formula">6</ref>) is only an accurate measurement of the ion transport efficiency if the electrons are re-captured by the target-holder, and the distribution of ionisation states in the ablated material is dominated by &#254; 1. To prevent electrons escaping the target holder, a positive DC bias of 32V was applied to the target holder. A larger DC bias than what was used for the previous experiments was applied to the quadrupole bender electrodes to increase the bending efficiency, with potentials of &#192;200 V and 40 V applied to the quadrupole electrodes nearest to the Faraday cup.</p><p>The laser beam energy and DC biases applied to the ion source were adjusted to maximise the transport efficiency, while ensuring that the potential gradient between the target holder and the first extraction electrode was sufficient to re-capture electrons on the target holder. Replacing the diverging and converging lenses with the Thorlabs beam expander increased the diameter of laser beam as it enters the focusing lens from 3.9 (1) mm to 9.6 (1) mm, thus reducing the calculated spot diameter on the target to D f &#188; 45</p><p>(1) mm. The laser repetition rate was fixed at 500 Hz for the experiment and the laser beam energy measured as 18.1 (1) mJ per pulse, which calculates to a peak fluence of 2.3 J/cm 2 . The laser beam energy was measured by placing the energy meter at the target's location, but without the target in place. The Cu target was scanned with a step size of 109 (2) mm in x and 75 (1) mm in y. The current was measured simultaneously at the target holder and Faraday cup for 8 s at each point of the scan.</p><p>For 5741 scan points across the surface of the Cu target, the average current measured leaving the target holder was 121 <ref type="bibr">(17)</ref> nA with the uncertainty taken as the standard deviation. The average current measured arriving at the Faraday cup for the same scan points was 2.6 (2) nA. From Eq. ( <ref type="formula">6</ref>), the average ion transport efficiency across the target surface was calculated to be 2.2 (7)%. A plot of the calculated efficiency for each scan point is shown in Fig. <ref type="figure">8</ref>. The efficiency is fairly homogeneous across the surface of the target, with the left side slightly less efficient than the right. This small gradient could be caused by the difference in proximity to the bending quadrupole electrodes.</p><p>For the ablation of Cu with a 308 nm laser spot at a fluence of 3 J/ cm 2 , a study has shown that the charge state distribution has a Cu 1&#254; /Cu 2&#254; ratio of 15.2 <ref type="bibr">[44]</ref>. There, the average kinetic energy of ablated Cu ions was O $ &#240;100&#222; eV and the angular distribution of ablated yield was found to have a Full Width at Half Maximum (FWHM) of 36 . This study had similar ablation conditions to the system presented in this work, thus singly charged Cu ions are expected to dominate the ablated ions, and ions are expected to have similar angular and kinetic energy distributions. In the system presented here, the aperture of the final electrode subtends an angle of $ 15 . Assuming a Guassian angular distribution, ~65% of ablated ions are expected to be lost by angular acceptance due to collision with the extraction electrodes, primarily the first extraction electrode.</p><p>Further ion loss is expected to result from inefficient bending of the quadrupole bender. The ion transport efficiency has been investigated using SIMION [45] simulations based on published ion distributions <ref type="bibr">[24,</ref><ref type="bibr">28,</ref><ref type="bibr">44]</ref>, and gave results comparable to the measured ion transport efficiency. Further simulations were conducted with ions leaving normal to the target surface with energies from 0 eV to 80 eV, with an increment of 1 eV between consecutive ions. Simulation results indicated that the widely spaced electrodes of the quadrupole bender, with the same potentials applied to the quadrupole electrodes as in the experiment, caused the bender to effectively act as a kinetic energy filter. In simulations, the quadrupole bender was unable to sufficiently bend Cu 1&#254; ions with a kinetic energy greater than ~20 eV towards the Faraday Cup for collection. This cut-off in kinetic energy was independent of the ion's origin on the target surface within a range of ~2 eV, and was significantly lower than the expected average kinetic energy of Cu 1&#254; ions reported in literature. The combined effects of ion loss due to the acceptance of the ion source geometry and inefficient bending are thought to result in the observed low ion transport efficiency. However, it shall be noted that the focus of this work has been to demonstrate the feasibility of a LAS with a large ablation range of up to 50 mm. A bender with smaller inter-electrode spacing and a biased aperture will improve the ion transmission efficiency, but at the cost of a reduced scanning range.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="6.">Conclusions</head><p>The use of high-precision motorized kinematic mirror mounts in laser ablation is a simple and cost-effective solution for positioning the laser beam on a target, without the need to move the target inside a vacuum chamber. A 2D scan of the ion current may be used to establish a coordinate system on the surface of the target, by a simple transformation of the mirror mount motor positions. Once a coordinate system is established, different target materials may be selected and the ablated ions may be guided to another device such as a mass spectrometer. Multi-element targets can be designed from 1 mm diameter stock materials and selectively ablated for mass spectrometer calibration. The establishment of a coordinate system also allows for the monitoring and management of damage accumulation on the target surface.</p><p>The PI and Thorlabs mirror mounts are capable of laser spot displacements less than 4 mm on the target surface. However, the spatial resolution of the presented LAS system is limited by the focused laser beam diameter on the target, which is one of the many factors affecting the diameter of ablated craters. In principle, the diameter of an ablated crater is also dependent on the material and sensitive to the total optical energy deposited on the target. Systems may be designed to leverage this for a specific application. By measuring the transitions between metals, the ablation spot diameter was shown to be as low as 40e50 mm for this particular device. The calculated laser beam diameter on the target was reduced by z40% by expanding the beam from 3.9 (1) mm to 9.6 (1) mm before focusing. The spatial resolution and resolving power of the LAS could be further improved by decreasing the focal length of the focusing lens.</p><p>The ion transport efficiency of the ion-source assembly and quadrupole bender was measured as 2.2 (7)% for Cu, with a slight gradient in the observed in the x direction. This deviation from homogeneity is likely due to a difference in proximity between the ion trajectories and the quadrupole electrodes.</p></div></body>