			<titleStmt><title level='a'>Dual-modal photoacoustic and ultrasound microscopy using optically-transparent and high-NA PVDF transducer</title></titleStmt>
			<publicationStmt>
				<publisher></publisher>
				<date>03/03/2022</date>
			</publicationStmt>
			<sourceDesc>
				<bibl> 
					<idno type="par_id">10392055</idno>
					<idno type="doi">10.1117/12.2607808</idno>
					<title level='j'>Proceedings of the SPIE</title>
<idno></idno>
<biblScope unit="volume">11960</biblScope>
<biblScope unit="issue"></biblScope>					

					<author>Cheng Fang</author><author>Jun Zou</author><author>Alexander A. Oraevsky</author><author>Lihong V. Wang</author>
				</bibl>
			</sourceDesc>
		</fileDesc>
		<profileDesc>
			<abstract><ab><![CDATA[As two important acoustic imaging modalities, photoacoustic microscopy (PAM) and ultrasound microscopy (UM) provide complementary functional and structural information on biological tissues. At penetration depths beyond the optical diffraction limit, the spatial resolution of both PAM and UM is determined acoustically by the receiving (focused) ultrasound transducer. To obtain good dual-modal imaging resolution, the transducer should have a high center frequency with a wide bandwidth. More importantly, a high acoustic numerical aperture (NA) should be maintained for receiving the PA and ultrasound signals, which is however difficult to achieve as the light blockage by the transducer is more likely to occur at higher NAs. In this paper, we report the dual-modal PAM and UM based on an optically-transparent focused PVDF transducer with a high NA of 0.64. The transducer has an acoustic center frequency and bandwidth of 36 MHz and 44 MHz, respectively. The acoustic focal diameter and zone are 37.8 μm and 210 μm, respectively. With a central transparent window, the excitation laser pulses can directly pass through the transducer to illuminate the target. This allows a high acoustic NA to be obtained without light blockage. For demonstration, co-registered 3D and 2D PAM/UM imaging experiments have been conducted on a twisted wire target in both water and chicken breast tissue, and in-vivo on a mouse tail, respectively. The imaging results show that high acoustic resolution and sensitivity can be achieved to resolve the target at different depths with a simple and compact dual-modal imaging setup.]]></ab></abstract>
		</profileDesc>
	</teiHeader>
	<text><body xmlns="http://www.tei-c.org/ns/1.0" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:xlink="http://www.w3.org/1999/xlink">
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="1.">INTRODUCTION</head><p>As a hybrid imaging modality, photoacoustic microscopy (PAM) combines optical absorption contrasts with acoustic detection of laser-generated ultrasound signals (i.e., photoacoustic (PA) signals) <ref type="bibr">[1]</ref>  <ref type="bibr">[2]</ref>. Because the light for PA excitation will be highly scattered beyond the optical diffraction limit in tissues, the spatial resolution of PAM will be mainly determined acoustically by the receiving (focused) ultrasound transducer. PAM operating in this regime is termed as acoustic-resolution PAM (AR-PAM). To obtain good AR-PAM resolution, the acoustic focal spot and zone of the receiving transducer need to be minimized. To achieve this, the transducer should operate at high frequencies with a wide bandwidth, and also have a high acoustic numerical aperture (NA). However, this causes a fundamental challenge in the (AR-)PAM system design, as the excitation light can be easily blocked by the transducer. Although a transmission mode (where the optical excitation and acoustic detection are arranged on the opposite sides of the target) can be adopted to circumvent this issue, it is not practical for imaging thicker targets oftentimes encountered under in-vivo conditions.</p><p>To address this issue, a dark-field setup <ref type="bibr">[3]</ref> and opto-acoustic combiner <ref type="bibr">[4]</ref>  <ref type="bibr">[5]</ref> have been developed to separate the optical and acoustic paths, which however make the PA scanning head complex (Figs. <ref type="figure">1 (a</ref>) and (b)). Alternatively, focused hollow transducers <ref type="bibr">[6]</ref> and planar transparent transducers <ref type="bibr">[7]</ref> [8] <ref type="bibr">[9]</ref> allow the laser to directly pass through. Nevertheless, with the center part removed or without a well-defined acoustic focal point, their acoustic sensitivity and focusing capability are compromised, which make them unsuitable for AR-PAM. Recently, AR-PAM based on an optically-transparent focused transducer has been reported <ref type="bibr">[10]</ref>  <ref type="bibr">[11]</ref>(Fig. <ref type="figure">1 (c)</ref>). Different from the hollow transducers, the transparent transducer still maintains a full acoustic receiving aperture, while allowing the excitation light to transmit through the transparent window. Nevertheless, limited by the narrow bandwidth caused by the hard piezoelectric material and non-ideal backing, optimal imaging resolution has not been achieved. Previously, we reported an optically-transparent focused polyvinylidene fluoride (PVDF) transducer laminated onto a concave glass lens (Fig. <ref type="figure">1 (d</ref>)) <ref type="bibr">[12]</ref>. However, due to the remaining issues in the fabrication process, only a low (acoustic) NA of 0.23 was obtained, which is not sufficient to provide the needed lateral resolution for AR-PAM. In this paper, we report the dual-modal (AR-)PAM and UM based on a new optically-transparent focused PVDF transducer.</p><p>Owing to the improved fabrication process, the new transducer has a much higher (acoustic) NA of 0.64, and also higher center frequency and wider bandwidth than the previous low-NA PVDF transducer (together with other comparable transducers) <ref type="bibr">[13]</ref>. For demonstration, a dual-modal PAM and UM setup has been built, and imaging experiments have been conducted on a twisted wire target in water and chicken breast tissue, and in-vivo on a mouse tail. The imaging results show that high acoustic resolution and sensitivity can be achieved with a simple and compact setup to resolve the target at different depths beyond the optical diffraction limit. The optical transmittance of the central transparent region is around 60% at 532 nm <ref type="bibr">[12]</ref>. In order to build the high-NA transducer with a well-defined acoustic focal point, two major improvements have been made in the fabrication process. First, a two-step molding process using one low-NA (~0.49) and then one high-NA (~0.64) lens pairs is employed to address the wrinkling issue on the outer edge of the previous transducer <ref type="bibr">[12]</ref>, which is the main limiting factor to achieve high NA. Second, after the bonding, a degassing procedure is performed in a vacuum chamber to remove any trapped gas bubbles underneath the PVDF film before the epoxy is fully cured, which is the secondary limiting factor to achieve high NA. With the above two improvements, a larger and better-formed spherical PVDF film is obtained to enable the higher NA and also better acoustic performance not achieved in the previous work. </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.">TRANSDUCER DESIGN AND CONSTRUCTION</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.">TRANSDUCER TESTING AND CHARACTERIZATION</head><p>A combined PA and pulse-echo ultrasound testing setup was built to characterize the acoustic performance of the high-NA transparent PVDF transducer by using a razor blade in water (whose edge was coated with black ink) as the target (Fig. <ref type="figure">3</ref>). The front spherical surface of the transducer was immersed in water and the target was placed onto a computercontrolled two-axis motor stage and a manually-adjusted height stage. For PA testing, the light source is a 532-nm pulsed laser (Elforlight, UK) triggered by a function generator with a 1-kHz repetition frequency. The output laser beam was expanded, passed through a beam splitter, reflected by a stationary mirror, and focused by a lens with 10-cm focal length.</p><p>The loosely focused laser beam propagated through the transducer with effective optical focal length around -16 mm from air to water and was incident onto the target. A CCD camera was used to monitor the excitation region through the reverse path of the optical illumination. Laser pulse energy after the transparent transducer was measured to be 35 &#956;J, and the laser beam diameter on the top surface of the target was around 1 mm, which corresponds to a laser intensity of 4.5 mJ/cm 2 , far below the ANSI limit of 20 mJ/cm 2 . The transducer was connected to a pulser-receiver, which sends electrical voltage pulses to the transducer (for pulse-echo ultrasound testing and imaging), and also amplify the ultrasound and PA signals received by the transducer. The pulse-echo ultrasound axial resolution was quantified by the minimal FWHM (full width at half maximum) of the signal envelope in the depth direction, which was calculated by the absolute Hilbert transformation. The acoustic focal spot was characterized by the minimal FWHM of the amplitude profile in the lateral direction, and the acoustic depth of focus was estimated by the height range where the on-axis peak amplitude dropped to half of its maximal value <ref type="bibr">[14]</ref>. To characterize the sensitivity of the high-NA transparent PVDF transducer, a needle hydrophone (Precision Acoustics, UK) with sensitivity (or conversion factor) 60 mV/MPa served as the reference. The characterized main acoustic parameters of the high-NA transparent PVDF transducer are listed in Table <ref type="table">1</ref>.  </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.">IMAGING EXPERIMENTS AND RESULTS</head><p>The experimental setup (Fig. <ref type="figure">3</ref>) was also used to demonstrate the PA and ultrasound imaging capability of the high-NA transparent PVDF transducer. Because the laser illumination spot is much larger than the acoustic focal spot of the transducer, the imaging resolution is determined acoustically by the focal spot of the transducer. The high-NA transparent PVDF transducer collected both PA and pulseecho ultrasound signals from the target (Fig. <ref type="figure">4</ref>). The reconstructed ultrasound image served as a reference for evaluating the PA image. The pulser-receiver was synchronously triggered by the function generator to drive the transducer for sending ultrasound pulses to the target. One trigger from the function generator initiated simultaneous PA and ultrasound data acquisition. It should be noted that the single triggering scheme will not cause the mixing of PA and ultrasound signals. Upon excitation, the ultrasound signal goes through a round trip (from the target to the transducer) and reaches the transducer, while the PA signal arrives after a single trip. Therefore, both the PA and ultrasound signals are completely separated in the time domain and therefore can be received without any mixing. A twisted metal wire with black plastic coating (total diameter &#981; 0.5 mm) was used as the imaging target, which was immersed in water and embedded inside chicken breast tissue, respectively (Fig. <ref type="figure">5</ref>). The depth of the wire ranged from 1 to 3 mm. A total area of 8 &#215; 8 mm 2 (marked by the white dashed square in Fig. <ref type="figure">5</ref>) was scanned with a step size of 100 &#956;m along the two orthogonal directions, and one scan took around 0.5 hr. At each location, the data acquisition was repeated 40 times and the acquired PA &amp; ultrasound signals were averaged to improve the signal-to-noise ratio (SNR) or contrast-to-noise ratio (CNR). To improve the depth of view and reduce the imaging time, the PA and ultrasound data were acquired at five different depths (1.0, 1.5, 2.0, 2.5, and 3.0 mm), which covered the total depth range of the black wire within about 2.5 hrs.   </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="5.">SUMMARY AND CONCLUSION</head><p>In summary, a dual-modal PAM and UM based on an optically-transparent focused PVDF transducer has been demonstrated. With a high NA, a wide acoustic bandwidth, and good optical transparency, the transducer can provide a small acoustic focal spot without light blockage, thereby enabling a simple PAM and UM setup with good acoustic resolution and sensitivity. Such capabilities can be useful for the development of compact PAM and UM systems for handheld, wearable, and even endoscopic imaging applications.</p></div></body>