Title: Application of magnetically actuated self-clearing catheter for rapid in situ blood clot clearance in hemorrhagic stroke treatment
Abstract Maintaining the patency of indwelling drainage devices is critical in preventing further complications following an intraventricular hemorrhage (IVH) and other chronic disease management. Surgeons often use drainage devices to remove blood and cerebrospinal fluid but these catheters frequently become occluded with hematoma. Using an implantable magnetic microactuator, we created a self-clearing catheter that can generate large enough forces to break down obstructive blood clots by applying time-varying magnetic fields. In a blood-circulating model, our self-clearing catheters demonstrated a > 7x longer functionality than traditional catheters (211 vs. 27 min) and maintained a low pressure for longer periods (239 vs. 79 min). Using a porcine IVH model, the self-clearing catheters showed a greater survival rate than control catheters (86% vs. 0%) over the course of 6 weeks. The treated animals also had significantly smaller ventricle sizes 1 week after implantation compared to the control animals with traditional catheters. Our results suggest that these magnetic microactuator-embedded smart catheters can expedite the removal of blood from the ventricles and potentially improve the outcomes of critical patients suffering from often deadly IVH.  more » « less
Award ID(s):
1944480
PAR ID:
10395595
Author(s) / Creator(s):
; ; ; ; ; ; ; ; ; ;
Date Published:
Journal Name:
Nature Communications
Volume:
13
Issue:
1
ISSN:
2041-1723
Format(s):
Medium: X
Sponsoring Org:
National Science Foundation
More Like this
  1. Miniature Magnetic Rotating Swimmers (MMRSs) are emerging as a promising technology to improve minimally invasive vascular and cardiac surgeries. Currently, these procedures are typically performed using catheters, which are thin flexible tubes inserted into blood vessels. However, catheters rub against artery walls and can dislodge fat deposits, which can lead to complications such as stroke. In contrast, MMRSs are untethered, wirelessly controlled devices actuated by an external magnetic field. Their compact size could allow them to navigate the bloodstream of a patient and reach treatment areas without the risks associated with catheter use. Surgical tasks, such as tissue cutting or ablation, require significant power, but due to their miniature size, MMRSs lack the capacity to store sufficient onboard energy. This paper studies an MMRS that can be heated wirelessly via induction. The method allows transferring enough power to denature proteins. 
    more » « less
  2. Abstract Recent micro-CT scans have demonstrated a much larger magnetic nanoparticle distribution volume in tumors after localized heating than those without heating, suggesting possible heating-induced nanoparticle migration. In this study, a theoretical simulation was performed on tumors injected with magnetic nanoparticles to evaluate the extent to which the nanoparticle redistribution affects the temperature elevation and thermal dosage required to cause permanent thermal damage to PC3 tumors. 0.1 cc of a commercially available ferrofluid containing magnetic nanoparticles was injected directly to the center of PC3 tumors. The control group consisted of four PC3 tumors resected after the intratumoral injection, while the experimental group consisted of another four PC3 tumors injected with ferrofluid and resected after 25 min of local heating. The micro-CT scan generated tumor model was attached to a mouse body model. The blood perfusion rates in the mouse body and PC3 tumor were first extracted based on the experimental data of average mouse surface temperatures using an infrared camera. A previously determined relationship between nanoparticle concentration and nanoparticle-induced volumetric heat generation rate was implemented into the theoretical simulation. Simulation results showed that the average steady-state temperature elevation in the tumors of the control group is higher than that in the experimental group where the nanoparticles are more spreading from the tumor center to the tumor periphery (control group: 70.6±4.7 °C versus experimental group: 69.2±2.6 °C). Further, we assessed heating time needed to cause permanent thermal damage to the entire tumor, based on the nanoparticle distribution in each tumor. The more spreading of nanoparticles to tumor periphery in the experimental group resulted in a much longer heating time than that in the control group. The modified thermal damage model by Dr. John Pearce led to almost the same temperature elevation distribution; however, the required heating time was at least 24% shorter than that using the traditional Arrhenius integral, despite the initial time delay. The results from this study suggest that in future simulation, the heating time needed when considering dynamic nanoparticle migration during heating is probably between 19 and 29 min based on the Pearce model. In conclusion, the study demonstrates the importance of including dynamic nanoparticle spreading during heating and accurate thermal damage model into theoretical simulation of temperature elevations in tumors to determine thermal dosage needed in magnetic nanoparticle hyperthermia design. 
    more » « less
  3. Abstract The development of inexpensive and portable point‐of‐care devices for measuring nutritional physiological parameters from blood (e.g., glucose, ketones) has accelerated our understanding and assessment of real‐time variation in human health, but these have infrequently been tested or implemented in wild animals, especially in relation to other key biological or fitness‐related traits. Here we used point‐of‐care devices to measure blood levels of glucose, ketones, uric acid, and triglycerides in free‐ranging house finches (Haemorhous mexicanus)—a common songbird in North America that has been well‐studied in the context of urbanization, nutrition, health, and sexual selection—during winter and examined (1) repeatability of these methods for evaluating blood levels in these wild passerines, (2) intercorrelations among these measurements within individuals, (3) how blood nutritional‐physiology metrics related to a bird's body condition, habitat of origin (urban vs. suburban), poxvirus infection, and sex; and (4) if the expression of male sexually selected plumage coloration was linked to any of the nutritional‐physiological metrics. All blood‐nutritional parameters were repeatable. Also, there was significant positive covariation between concentrations of circulating triglycerides and glucose and triglycerides and uric acid. Urban finches had higher blood glucose concentrations than suburban finches, and pox‐infected individuals had lower blood triglyceride concentrations than uninfected ones. Last, redder males had higher blood glucose, but lower uric acid levels. These results demonstrate that point‐of‐care devices can be useful, inexpensive ways of measuring real‐time variation in the nutritional physiology of wild birds. 
    more » « less
  4. This study investigates the efficacy of an untethered magnetic robot (UMR) for wireless mechanical and hybrid blood clot removal in ex vivo tissue environments. By integrating x-ray-guided wireless manipulation with UMRs, we aim to address challenges associated with precise and controlled blood clot intervention. The untethered nature and size of these robots enhance maneuverability and accessibility within complex vascular networks, potentially improving clot removal efficiency. We explore mechanical fragmentation, chemical lysis, and hybrid dissolution techniques that combine mechanical fragmentation with chemical lysis, highlighting their potential for targeted and efficient blood clot removal. Through experimental validation using an ex vivo endovascular thrombosis model within the iliac artery of a sheep, we demonstrate direct revascularization of a 13-mm-long, 1-day-old blood clot positioned inside the left common iliac artery. This was achieved by deploying a UMR into the abdominal aorta within 15 min. Additionally, both mechanical fragmentation and hybrid dissolution achieve a greater volume rate of change compared to no intervention (control) and chemical lysis alone. Mechanical fragmentation exhibits clot removal with a median of 0.87 mm3/min and a range of 2.81 mm3/min, while the hybrid approach demonstrates slower but more consistent clot removal, with a median of 0.45 mm3/min and a range of 0.23 mm3/min. 
    more » « less
  5. Syed, Zainulabeuddin (Ed.)
    Abstract We evaluated miRNA and mRNA expression differences in head tissues between avid-biting vs. reluctant-biting Aedes albopictus (Skuse) females from a single population over a 20-min timescale. We found no differences in miRNA expression between avid vs. reluctant biters, indicating that translational modulation of blood-feeding behavior occurs on a longer timescale than mRNA transcription. In contrast, we detected 19 differentially expressed mRNAs. Of the 19 differentially expressed genes at the mRNA level between avid-biting vs. reluctant-biting A. albopictus, 9 are implicated in olfaction, consistent with the well-documented role of olfaction in mosquito host-seeking. Additionally, several of the genes that we identified as differentially expressed in association with phenotypic variation in biting behavior share similar functions with or are inferred orthologues of, genes associated with evolutionary variation in biting behaviors of Wyeomyia smithii (Coq.) and Culex pipiens (Lin.). A future goal is to determine whether these genes are involved in the evolutionary transition from a biting to a non-biting life history. 
    more » « less