			<titleStmt><title level='a'>Quantitative Stimulated Raman Scattering Microscopy: Promises and Pitfalls</title></titleStmt>
			<publicationStmt>
				<publisher></publisher>
				<date>06/13/2022</date>
			</publicationStmt>
			<sourceDesc>
				<bibl> 
					<idno type="par_id">10413776</idno>
					<idno type="doi">10.1146/annurev-anchem-061020-015110</idno>
					<title level='j'>Annual Review of Analytical Chemistry</title>
<idno>1936-1327</idno>
<biblScope unit="volume">15</biblScope>
<biblScope unit="issue">1</biblScope>					

					<author>Bryce Manifold</author><author>Dan Fu</author>
				</bibl>
			</sourceDesc>
		</fileDesc>
		<profileDesc>
			<abstract><ab><![CDATA[Since its first demonstration, stimulated Raman scattering (SRS) microscopy has become a powerful chemical imaging tool that shows promise in numerous biological and biomedical applications. The spectroscopic capability of SRS enables identification and tracking of specific molecules or classes of molecules, often without labeling. SRS microscopy also has the hallmark advantage of signal strength that is directly proportional to molecular concentration, allowing for in situ quantitative analysis of chemical composition of heterogeneous samples with submicron spatial resolution and subminute temporal resolution. However, it is important to recognize that quantification through SRS microscopy requires assumptions regarding both system and sample. Such assumptions are often taken axiomatically, which may lead to erroneous conclusions without proper validation. In this review, we focus on the tacitly accepted, yet complex, quantitative aspect of SRS microscopy. We discuss the various approaches to quantitative analysis, examples of such approaches, challenges in different systems, and potential solutions. Through our examination of published literature, we conclude that a scrupulous approach to experimental design can further expand the powerful and incisive quantitative capabilities of SRS microscopy.]]></ab></abstract>
		</profileDesc>
	</teiHeader>
	<text><body xmlns="http://www.tei-c.org/ns/1.0" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:xlink="http://www.w3.org/1999/xlink">
<div xmlns="http://www.tei-c.org/ns/1.0"><head>INTRODUCTION</head><p>This contrasts with so-called spontaneous Raman scattering, where a single (typically continuous wave) laser source focused on a sample will cause spontaneous emission of photons of different wavelengths (i.e., energies) corresponding to vibrational energy differences with respect to the original laser source. While spontaneous Raman scattering affords broad and readily interpretable spectral information about a sample, the probability of spontaneously emitted Raman photons is typically quite low with differential scattering cross-sections in the ~10 -30 cm 2 regime (2-4). Thus, if microscopy is desired via spontaneous Raman, acquisition times are often egregiously slow given the typical excitation wavelengths and maximum laser powers appropriate for biological samples. Coherent Raman scattering microscopies offer significantly faster and more sensitive Raman imaging via coherent nonlinear optical interactions, typically at the expense of broad spectral coverage and some spectral specificity <ref type="bibr">(5,</ref><ref type="bibr">6)</ref>.</p><p>CARS microscopy has seen broad use in biophotonics research since its first demonstration but has largely fallen to the wayside of SRS microscopy <ref type="bibr">(7)</ref><ref type="bibr">(8)</ref><ref type="bibr">(9)</ref><ref type="bibr">(10)</ref><ref type="bibr">(11)</ref><ref type="bibr">(12)</ref><ref type="bibr">(13)</ref>. This is due largely to the shortcomings of CARS microscopy with respect to SRS microscopy. Specifically, CARS microscopy, though technically easier to implement and detect signal, suffers from a significant nonresonant background effect and spectral distortion (with respect to the spontaneous Raman spectrum). This nonresonant background and resultant spectral distortions (e.g., wavenumber shifts, peak broadening, peak intensity changes) make quantitative chemical analysis of CARS microscopy images difficult <ref type="bibr">(14,</ref><ref type="bibr">15)</ref>. SRS microcopy, in contrast, does not suffer from the nonresonant background issue. SRS signal arises from energy transfer between the two laser sources and the molecular vibration, with spectral features nearly identical to spontaneous Raman except for slight peak broadening related to the pulse duration and bandwidth of the lasers <ref type="bibr">(16)</ref>. Additionally, SRS microscopy offers the distinct advantage of signal strength that is linearly dependent on sample concentration. This affords SRS microscopy with fast, sensitive, label-free, and quantitatively intuitive detection of chemically specific imaging contrast of samples. Moreover, SRS microscopy is well suited for biological samples because Raman scattering of water vibrations is relatively weak and the typically used near infrared picosecond laser sources cause minimal photodamage due to heating and multiphoton ionization. Indeed, SRS microscopy has found its largest use in biophotonics as a label-free yet chemically specific alternative to more ubiquitously adopted fluorescence microscopy techniques. While fluorescence microscopy is a valuable tool with a relatively low barrier to utilize, the use of exogenous fluorophores has some commonly cited shortcomings <ref type="bibr">(6,</ref><ref type="bibr">12,</ref><ref type="bibr">(18)</ref><ref type="bibr">(19)</ref><ref type="bibr">(20)</ref><ref type="bibr">(21)</ref>, including label specificity, cytotoxicity, photobleaching, and color cross talk. SRS microscopy has found utility in a wide variety of biological applications, such as imaging pharmacokinetics <ref type="bibr">(22)</ref><ref type="bibr">(23)</ref><ref type="bibr">(24)</ref><ref type="bibr">(25)</ref><ref type="bibr">(26)</ref>, cellular and physiological metabolism <ref type="bibr">(27)</ref><ref type="bibr">(28)</ref><ref type="bibr">(29)</ref><ref type="bibr">(30)</ref><ref type="bibr">(31)</ref>, antibiotic susceptibility and resistance <ref type="bibr">(32,</ref><ref type="bibr">33)</ref>, and histopathological diagnosis <ref type="bibr">(34)</ref><ref type="bibr">(35)</ref><ref type="bibr">(36)</ref><ref type="bibr">(37)</ref>. Although such experiments could also be performed using fluorescent labels, the value of a label-free method that also provides quantitative insight is highly desirable for modern biological studies.</p><p>The growing popularity of SRS microscopy has led to an explosive growth of publications in recent years. There are abundant literature reviews dedicated to the discussion of coherent Raman scattering microscopy <ref type="bibr">(9)</ref><ref type="bibr">(10)</ref><ref type="bibr">(11)</ref><ref type="bibr">(12)</ref><ref type="bibr">(38)</ref><ref type="bibr">(39)</ref><ref type="bibr">(40)</ref><ref type="bibr">(41)</ref><ref type="bibr">(42)</ref><ref type="bibr">(43)</ref><ref type="bibr">(44)</ref><ref type="bibr">(45)</ref>. Instead of summarizing recent advances in SRS instrumentation and applications, we take a different approach and focus on an underappreciated aspect: quantification in SRS microscopy. With the combination of spatial information and vibrational chemical information acquired in SRS, even complex and temporally dynamic samples can be observed with quantitative insight. However, there are many pitfalls associated with quantification owing to complications such as sample scattering and non-Raman background. The problem is analogous to one in fluorescence microscopy: while it is simple to compare the intensity of two fluorescent specimens, quantitation of concentration or other molecular properties is error prone and requires a thorough understanding of limitations and the careful use of proper methods <ref type="bibr">(46)</ref><ref type="bibr">(47)</ref><ref type="bibr">(48)</ref>. A detailed discussion of SRS-based quantitative analysis is missing in the literature. In this article, we focus on the common quantitative metrics afforded by SRS microscopy. The review is organized based on the five main categories of quantitative measurements in SRS microscopy based on a Web of Science search for "quantitative/quantify/quantification" + "SRS":</p><p>1. Morphological measurements such as size, shape, or count of objects in a field of view; 2. Relative concentration measurements, where SRS signal strength is taken as a proxy for molecular concentration; 3. Absolute concentration measurements, where SRS signal is calibrated with respect to an analyte such that exact spatially resolved concentrations can be calculated; 4. Ratiometric measurements, where SRS signals from different vibrational peaks are proportionally compared to provide compositional insight; and 5. Spectrally dynamic measurements, where a vibrational peak (typically of a sensor molecule) shifts in wavenumber in response to changes in the microenvironment.</p><p>The relative prevalence of these five categories based on the search is shown in Figure <ref type="figure">1a</ref>.</p><p>We explore these five categories of quantitative SRS microscopy and discuss their respective prevalence, utility, and potential pitfalls or shortcomings. For each category, we highlight only a few example applications to demonstrate the relevant method for quantitative analysis. Representative "morphological" figure adapted with permission from Reference <ref type="bibr">(67)</ref>; copyright 2016 American Chemical Society. Representative "absolute" figure adapted with permission from Reference (83); copyright 2020 American Chemical Society. Representative "ratiometric" figure adapted with permission from Reference (104); copyright 2020 Ivyspring International Publisher.. Representative "relative" figure adapted with permission from Reference <ref type="bibr">(32)</ref>; copyright 2018 American Chemical Society.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>PRINCIPLES OF QUANTITATIVE STIMULATED RAMAN SCATTERING MICROSCOPY AND POTENTIAL CHALLENGES</head><p>Specific implementations of SRS microscopy have been extensively discussed in previous articles and reviews <ref type="bibr">(1,</ref><ref type="bibr">6,</ref><ref type="bibr">42,</ref><ref type="bibr">49)</ref>. Briefly, pulsed laser sources produce synchronized pump and Stokes beams. The Stokes pulse train is amplitude modulated at &gt;1 MHz frequency, as shown in In principle, given the Raman cross-section, the concentration can be directly calculated. In practice, a calibration curve of SRS intensity versus concentration is typically obtained for quantitative analysis. The procedure is similar to the classical spectrophotometric analysis using Beer-Lambert's law, except that the cuvette length is a hidden parameter (typically ~1 &#181;m) that depends on the imaging condition (including beam size, objective numerical aperture, and condenser numerical aperture). When the sample and the calibration solutions are measured under the same imaging condition, the concentration of the unknown sample can be simply calculated as</p><p>where Sunknown is the signal strength of the unknown signal, S0 is the signal strength of the calibration sample, and C0 is the concentration of the calibration sample.</p><p>Although this seems straightforward, many pitfalls can obfuscate quantitative interpretation in imaging complex samples. The most notable challenge is light scattering. Light scattering is a ubiquitous process in optical imaging of biological tissues and other heterogeneous materials.</p><p>Scattering decreases the amount of light reaching the objective focus, thus decreasing SRS signal intensity. It also reduces the amount of light collected by the photodiode, which further decreases the measured signal. Because the amount of light scattering cannot be directly measured, it is challenging to quantify concentration based on measured SRS intensity. This is due to calibration samples likely not having the same light scattering-induced signal intensity loss. To circumvent this challenge, either relative comparisons are made on similar samples, or an internal standard is necessary to correct for light scattering.</p><p>In imaging monolayer cell cultures, light scattering is typically negligible. However, there are other important considerations when absolute concentration is to be quantified. The first one is the non-Raman background. Non-Raman-induced pump intensity change is detected by the lock-in amplifier. This extraneous background may be due to transient absorption, photothermal process, and cross-phase modulation <ref type="bibr">(50)</ref><ref type="bibr">(51)</ref><ref type="bibr">(52)</ref>. These parasitic processes can contribute interfering background with respect to the SRS signal. This is particularly relevant when imaging molecules at low concentrations. In practice, any SRS imaging that is not targeting dominant species such as proteins, lipids, water, and nucleic acids should be scrutinized for background contribution. Various background removal methods have been developed to overcome this challenge <ref type="bibr">(38,</ref><ref type="bibr">51,</ref><ref type="bibr">53,</ref><ref type="bibr">54)</ref>. The second consideration is cross talk. Like fluorescence, Raman peaks of different molecules may overlap with one another. Thus, separation of their contributions is necessary for quantitative analysis. Cross talk is particularly poignant for SRS signal from carbon-hydrogen (C-H) or carbon-deuterium (C-D) stretching, which is commonly used in SRS imaging. The third consideration is the nonuniform imaging intensity across the field of view. Due to chromatic and spatial aberration of the pump and Stokes beams at the objective focus, it is common that the intensity at the edge of the field of view is weaker than that at the center. Such field variation must be corrected before applying any calibration curve obtained from solutions <ref type="bibr">(55)</ref>. A related but more subtle point is sample-dependent optical aberration that may influence quantification accuracy. Calibration is typically done in solutions, which introduce different optical aberrations than biological cells and tissue. This difference in optical aberration changes the effective laser intensity at focus and focal volume. Consequently, applying calibration of solutions to cells or tissues may bias the result. This effect is rarely considered but could lead to significant difference in signal intensity with high numerical aperture (NA) objectives and a large refractive index mismatch <ref type="bibr">(56,</ref><ref type="bibr">57)</ref>.</p><p>Lastly, we must acknowledge the distinction between effective concentration and genuine concentration. In SRS imaging where absolute concentration measurements are made (by creating appropriate calibrations and accounting for the above pitfalls), it is ultimately a measurement of average concentration within the focal volume (i.e., effective concentration).</p><p>When the measured molecule is contained within an area smaller than the focal spot size, the actual local concentration can be much larger than the effective concentration. This is an important consideration when measuring molecules in lipid droplets, vesicles, or other organelles <ref type="bibr">(58,</ref><ref type="bibr">59)</ref>.</p><p>To ensure robust quantitative analysis, it is necessary to consider the aforementioned challenges and design proper calibration or other quantitative metrics that are not solely intensity dependent. In the following sections, we discuss a few major types of SRS-based quantitative analysis and provide some example applications.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>MORPHOLOGICAL ANALYSIS</head><p>The first quantitative metric, morphological measurement, is not strictly unique to SRS microscopy and is, perhaps, the most straightforward and familiar. These measurements include qualities such as shape, size, number, orientation, spatial relation to other observed objects, or dynamics. Indeed, any microscopy with an appropriate contrast will provide morphological or spatial information about a sample. More unique to SRS, however, is the origin, quality, and temporal resolution of such acquired spatial information.</p><p>Specifically, the origin of the spatial information is still inherently chemical. Because the observed signal is ideally related only to the SRS process, any images formed can be thought of as a chemical map showing the presence of molecules vibrationally resonant with the chosen laser frequencies. Further, the relative intensity of pixels (i.e., the spatial contrast) is directly related to the concentration of resonant molecules at a given location within a field of view.</p><p>Regarding the quality of spatial information, there are a few important factors for SRS microscopy. First, the spatial resolution of images acquired with SRS microscopy is limited by the laser wavelengths in accordance with Abbe's limit of diffraction. The lateral resolution (with a high NA objective) can be approximated as where &#955;pump and &#955;Stokes are the two wavelengths used for SRS, and NA is the numerical aperture of the objective <ref type="bibr">(60,</ref><ref type="bibr">61)</ref>. As an example, a system using a pump wavelength at 800 nm and Stokes wavelength at 1,040 nm for C-H SRS imaging using a 1.0 NA objective should have a spatial resolution of ~343 nm.</p><p>Quantitative metrics based solely on morphological characteristics have found many uses, particularly in lipid droplet analysis <ref type="bibr">(25,</ref><ref type="bibr">(62)</ref><ref type="bibr">(63)</ref><ref type="bibr">(64)</ref><ref type="bibr">(65)</ref><ref type="bibr">(66)</ref>. Cao et al. <ref type="bibr">(67)</ref> demonstrate an excellent use of morphological measurements in their work quantifying lipid droplets within cells in a microfluidic platform. They measure the number, size, distribution, and intensity of lipid droplets within thousands of cells. They manage the vast numbers of spatially segmented lipid droplets for analysis in a workflow depicted in Figure <ref type="figure">2a</ref>. By leveraging the chemical difference between lipids (where CH2 signal is strong) and the rest of the cell (mostly CH3 signal), they create segmentation maps for the lipids within cells in a field of view and then parameterize the lipid droplets based on an intensity threshold. The quantitative parameters they derive demonstrate the capabilities of SRS microscopy to produce relevant phenotypic markers in a label-free manner.</p><p>In addition to size, shape, and number, SRS microscopy can also monitor lipid droplets over time within live cells <ref type="bibr">(29)</ref>. Time-lapse imaging provides dynamic information about the lipids and can be used to characterize lipid droplet function and potential disease states <ref type="bibr">(68)</ref>. Because morphological metrics are independent of concentration, they are typically not affected by the challenges mentioned in the previous section. However, one potential pitfall is the segmentation method. As the metrics are strongly based on the effective pixel area designated as a lipid droplet or drug particle, it is important to distinguish signal from background. Improper thresholding in this regard would lead to potentially erroneous observations in size, location, or effective intensity of the area. To mitigate this risk, it is important to choose molecular targets that are strong and distinct from the background or other signal sources for the sample. In the above examples, this is achieved by targeting CH2 bonds (strong and abundant, chiefly in lipid droplets) and carbon-carbon double bonds in the drug (strong and highly specific to the drug). However, for other samples, features may be less distinct within a heterogeneous mixture or weak in Raman signal. Even for lipid droplets, smaller droplets may be obscured by the limited spatial resolution of SRS, leading to erroneous results if the thresholding was done improperly <ref type="bibr">(29)</ref>. We note that contrast can be improved with deuterium labeling or exogenous vibrational labels with distinct spectral features and minimal Raman background <ref type="bibr">(28,</ref><ref type="bibr">(70)</ref><ref type="bibr">(71)</ref><ref type="bibr">(72)</ref><ref type="bibr">(73)</ref> or by utilizing machine learning or deep learning to enhance visual contrast <ref type="bibr">(74)</ref><ref type="bibr">(75)</ref><ref type="bibr">(76)</ref><ref type="bibr">(77)</ref><ref type="bibr">(78)</ref>.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>RELATIVE CONCENTRATION MEASUREMENTS</head><p>Relative concentration measurements comprise the largest proportion of studies that utilize SRS microscopy for quantitative measurements (Figure <ref type="figure">1a</ref>). This is perhaps not surprising given the relative ease by which relative measurements can provide quantitative information without the complications and validations of more precise concentration measurements. Here we take relative concentration to mean any measurement or visualization of molecular presence via SRS microscopy that does not report an exact concentration for the molecule. This is often discussed in publications as SRS intensity. SRS intensity can be depicted as either normalized or nonnormalized and is typically reported as arbitrary units (a.u.). These relative measurements are then used to draw quantitative comparisons between different samples or samples at different conditions. Such comparisons still rely on the linear dependence between SRS signal and molecular concentration. However, the conclusions being drawn do not require exact concentration knowledge.</p><p>Relative measurements can take on a variety of tasks in SRS imaging. One example is timelapse measurements that provide chemical insight into the dynamics of biological systems. For example, Hong et al. <ref type="bibr">(32)</ref> demonstrate antibiotic susceptibility testing via SRS metabolic imaging. They show that antibiotic susceptibility can be predicted by monitoring glucose uptake within a single cell cycle. By targeting the C-D vibrational peak of deuterated glucose, they visualize uptake within live cells over the course of hours, as seen in Figure <ref type="figure">3a</ref>. As the bacteria cells take up deuterated glucose, SRS signal at the corresponding 2,178 cm -1 peak increases.</p><p>When the strains are additionally treated with vancomycin, the susceptible line shows a marked decrease in deuterated glucose uptake while the resistant line remains the same. Although exact concentrations are not calculated here, the relative SRS intensity provides insight into uptake dynamics and thus antibiotic susceptibility. These examples of relative metrics highlight the value of SRS microscopy's inherent quantitative nature while avoiding the litany of problems associated with measurements of exact concentrations. Where scattering and background signal will obfuscate direct quantification, relative measurements are valid so long as comparison among sufficiently similar samples is performed under uniform conditions. Ultimately, relative metrics trade the precision of quantitative information for an ease of understanding and avoidance of particularly difficult, if not impossible, experimental calibrations. This can be particularly appropriate for methods involving machine learning for further mining data (79-81). It is worth noting, however, that relative measurements make it difficult to compare results between labs, or even results from the same lab but obtained at different times where imaging conditions may be different. Reported results may be relevant more within the paradigm of one study rather than broadly applicable.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>ABSOLUTE CONCENTRATION MEASUREMENTS</head><p>In contrast to relative measurements, absolute measurements provide exact numbers for concentration of observed molecular species in SRS imaging. They are important for comparison with other analytical techniques or deeper mechanistic understanding. Applications of absolute measurements are often shown in imaging of cultured monolayer cells, where scattering is negligible.</p><p>As discussed earlier, attaining absolute concentrations numbers from SRS images is, unfortunately, rarely as easy as making a calibration curve and then imaging the sample. In spatially and spectrally complex samples, interfering species, background, aberration, absorption, and scattering can all distort SRS signal with respect to a calibration curve based on solution images. Thus, significant care must be taken toward verifying the concentration numbers reported are accurate.</p><p>An example of absolute concentration measurement is Fu et al.'s (82) demonstration of intracellular quantification of tyrosine kinase inhibitors (TKIs). TKIs are drugs used in the treatment of specific types of cancers to inhibit cancer cell growth and proliferation. In the study, solution spectra of two TKIs, imatinib and nilotinib, were first acquired and then used to calculate intracellular drug concentrations of treated cells. The concentrations are calculated after spectral unmixing and background subtraction (Figure <ref type="figure">4b</ref>). Specifically, compared to extracellular drug concentration, the drugs were shown to exhibit over 1,000-fold enrichment in the lysosomes of the cells while cytosolic drug presence was below the sensitivity for the system.</p><p>The enrichment of imatinib agrees with the expected results from the lysosomotropic effect.</p><p>Disagreement in nilotinib enrichment can be attributed to its precipitation in lysosomes. Recent reports have also shown similar results, but without the same absolute concentration measurements <ref type="bibr">(22)</ref>. Here, the impact and potential of measured concentrations over simple relative uptake over time are clear: The drug uptake for given dosage concentrations can be tracked quantitatively to better understand drug disposition across many drugs and cell types.</p><p>Figure <ref type="figure">4</ref> Absolute concentration measurements with stimulated Raman scattering (SRS) microscopy. (a) Spectra of lysosome-sequestered imatinib and nilotinib and solution spectra of the same drugs i and ii, respectively ), time-lapse observed intracellular concentration of imatinib and nilotinib (iii and iv, respectively), and representative images of BaF3 cells treated with imatinib and nilotinib (v and vi respectively). Panel adapted permission from Reference 82; copyright 2014 Nature Publishing Group. (b) Spontaneous and stimulated Raman spectra of the deuterated glutamine (i) and the linear response of its SRS signal with respect to concentration (ii). Representative carbon-deuterium SRS images with noted deuterated glutamine-tagged mHtt-97Q protein in intracellular protein aggregates (iii). Panel adapted with permission from Reference 83; copyright 2020 American Chemical Society.</p><p>One important note regarding label-free imaging for concentration measurements such as Fu et al.'s (82) demonstration is that hyperspectral imaging is imperative for proper calibration. This is due to background signal that interferes with the target molecule's spectrum. Various spectral unmixing techniques have been developed and reported to this end <ref type="bibr">(41,</ref><ref type="bibr">80,</ref><ref type="bibr">81)</ref>. Unmixing techniques allow the background to be accounted for, thus enabling more precise concentration measurements.</p><p>Another approach to mitigate background contribution is to target deuterated signal molecules or other exogenous Raman vibrational reporters by working in the so-called cell silent region (~1,800-2,800 cm -1 ), where endogenous biomolecules have no vibrational contributions.</p><p>For example, Miao &amp; Wei (83) demonstrate the ability to characterize polyglutamine protein aggregates within cells and provide concentration measurements of the different constituent proteins within aggregates. They show significant aggregation of the proteins with deuterated glutamine levels in the millimolar range. In combination with size and ratiometric measurements of the aggregates, they also show the composition and steady concentration of nontoxic proteins within aggregates, suggesting a toxic protein scavenging functionality for the aggregates. It is worth noting that in deuterated samples, non-Raman background contribution may still contribute significantly, and either hyperspectral SRS or frequency modulation SRS may be needed to provide quantitative measurements of concentrations <ref type="bibr">(53)</ref>.</p><p>Absolute concentration measurement from SRS imaging is a somewhat underdeveloped branch of quantitative SRS measurements, likely due to the difficulty of obtaining controlled and verifiable quantitative information. However, subcellular concentration measurements within live and dynamic systems are significant and can provide unique mechanistic insight into various small molecules within cells or tissue. The demonstrations so far have been promising but ultimately remain limited by the challenges associated with attaining precise chemical information in complex samples. For example, Fu et al. <ref type="bibr">(59)</ref>  have not yet been reported.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>RATIOMETRIC ANALYSIS</head><p>Ratiometric measurements are a category of quantitative measurements based on the ratio of two vibrational peaks that provide quantitative information about the chemical composition of one molecular species or the relative abundance of two species. Ratiometric measurements can provide information about sample composition <ref type="bibr">(63,</ref><ref type="bibr">(83)</ref><ref type="bibr">(84)</ref><ref type="bibr">(85)</ref><ref type="bibr">(86)</ref><ref type="bibr">(87)</ref>, diagnostic histology <ref type="bibr">(35,</ref><ref type="bibr">(88)</ref><ref type="bibr">(89)</ref><ref type="bibr">(90)</ref><ref type="bibr">(91)</ref>, chemical reaction dynamics (92), or biological metabolism and homeostasis <ref type="bibr">(63,</ref><ref type="bibr">(93)</ref><ref type="bibr">(94)</ref><ref type="bibr">(95)</ref><ref type="bibr">(96)</ref>. The main advantage of ratiometric measurements is that the quantitative information is not susceptible to distortion due to absorption and scattering because they affect both peaks equally and thus cancel out in the ratio. However, background signals often have a different ratio that affects quantification when they have a non-negligible contribution to the overall signal. In this case, background subtraction is necessary before ratiometric analysis. To this end, spectral coverage and pulse duration play a significant role in the relative strength of SRS to background signal when broadband pulses are used. Spectral coverage refers to the observable vibrational bandwidth for a given SRS setup based on the laser pulse wavelengths, duration, and chirp (in the case of spectral-focusing SRS). If the peaks to be measured are far apart, yet are to be imaged either simultaneously <ref type="bibr">(75,</ref><ref type="bibr">97,</ref><ref type="bibr">98)</ref> or quickly in sequence <ref type="bibr">(24,</ref><ref type="bibr">49,</ref><ref type="bibr">99)</ref>, large spectral coverage is necessary to cover both peaks. Methods to broaden spectral coverage have also demonstrated significant background reduction (53, 100-102).   Ratiometric imaging can also be used for metabolic imaging, where dynamic compositional information relates to cellular uptake and growth. For example, Du et al. <ref type="bibr">(104)</ref> show that transcriptome data-mined phenotypic metabolic susceptibilities could be visualized with SRS by measuring the ratio between lipid and protein synthesis for different metastatic melanoma cell lines, as illustrated in Figure <ref type="figure">5b</ref>. They note that the hypothesized correlation between differentiation and metabolism is treatment-targetable through the elevated fatty acid synthesis as visualized in the elevated CH2/CH3 (lipid/protein) signal ratio. This demonstrates the ability of ratiometric imaging to quantitatively measure metabolism among different cancer cell lines and elucidate potential cancer treatment targets. It is worth nothing that deuterium labeling SRS imaging has also be widely used in ratiometric measurements to study metabolism <ref type="bibr">(28,</ref><ref type="bibr">98,</ref><ref type="bibr">105)</ref>.</p><p>In ratiometric imaging, when the excitation wavelengths remain unchanged, we can assume that the scattering-induced signal loss for both SRS peaks is the same, thus obviating the significant challenges in acquiring absolute chemical measurements in tissue. This is likely why it is highly favored in SRS imaging of tissue. Ratiometric SRS imaging offers the practical quantitative information necessary to study and understand relevant chemical compositions of systems. Ultimately, ratiometric imaging provides a wealth of quantitative information with similar levels of ease as with relative concentration but with the added benefit of providing absolute metrics that can be compared across experiments.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>SPECTRALLY DYNAMIC MEASUREMENTS</head><p>Spectrally dynamic measurements are dependent on the shift of a vibrational peak of a molecule due to the molecule undergoing some chemical or physical change. This, in effect, utilizes the molecule as a sensor of the local environment through SRS microscopy. The spectral shift can be measured through either hyperspectral SRS imaging or ratiometric imaging. However, in these measurements, the information sought is not about the peak intensity or the concentration, but Figure <ref type="figure">6</ref> Spectrally dynamic measurements with SRS microscopy. (a) The Raman dye molecule (i), mitokyne, and its SRS spectra (ii) showing the shifting peak in response to different pH. The SRS intensity ratio of 2,230/2,216 cm -1 with respect to pH (iii). Representative SRS images (iv) of HeLa cells treated with mitokyne at different controlled pH values at 2,216 cm -1 , 2,230 cm -1 , and their ratio. Panel adapted with permission from Reference 106; copyright 2021 American Chemical Society. (b) SRS difference spectra of liquid water at various temperatures (i). The calibration of the spectral intensity ratios and temperatures (ii). Temperature map images (iii) based on SRS signal of an A549 cell being heated with an infrared laser at different positions.</p><p>Thermal decay curves (iv) with respect to distance for each field of view. Panel adapted with permission from Reference 107; copyright 2020 American Chemical Society. Abbreviations: a.u., arbitrary unit; SRS, stimulated Raman scattering.</p><p>Aside from imaging the microenvironment using reporter molecules, more exotic measurements may also be attained through spectrally dynamic SRS measurements. For Besides small molecules for pH and temperature sensing, there have also been recent reports of using Raman reporters for gas molecule sensing <ref type="bibr">(108)</ref> and mapping water solvation within cells <ref type="bibr">(109,</ref><ref type="bibr">110)</ref>, showcasing the potential diversity of quantitative measurements possible through spectral shifts. The unique advantage of this type of measurement is that the spectral shape or peak position can be determined much more reliably and accurately than intensity.</p><p>Though the examples are few so far for this type of quantitative SRS imaging, we expect significant growth in spectrally dynamic sensor molecule-based SRS imaging.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>CONCLUSIONS AND OUTLOOK</head><p>SRS microscopy has established itself as a useful tool for quantitative chemical imaging.</p><p>Through the wide variety of quantitative metrics, we expect SRS microscopy to continue growing in use across a broad range of biological and biomedical applications. While the challenges associated with quantitative measurements can be onerous, we have discussed potential strategies and solutions to maximize the capabilities of SRS microscopy. We have also pointed out types of quantitative SRS imaging approaches that obviate the challenges of light scattering, such as ratiometric SRS imaging and imaging of Raman reporter molecules that are sensitive to their local environment. Although relative concentration metrics are still most common owing to their ease of use, we point out that the more accurate and robust ratiometric measurements offer desirable quantitative information that can be compared across experiments or even across different labs. Sample preparation and system configuration are much less demanding because scattering and aberration contributions are largely removed. However, for ratiometric measurements to work, it is necessary to have large and reproducible SRS spectral changes due to either the compositional variation of one molecular species or relative abundance change of two molecular species. Moreover, many of the ratiometric examples shown here used two distinct Raman peaks. It is common to have more subtle spectral changes of a group of Raman peaks (often with overlapping features). In those cases, chemometric approaches are necessary to quantitatively disentangle the spectral changes (85).</p><p>Determining the absolute, spatially resolved concentration of molecular species in living cells and tissues with SRS is highly desired but largely unfulfilled owing to many challenges. In a few limited cases (mostly cells in monolayer culture), absolute concentration is shown to offer unique insights into the molecular processes that are largely unattainable with relative concentration measurements. However, these measurements cannot be directly translated to tissue due to light scattering. A universal approach to scattering correction using internal standards is needed to enable quantitative concentration measurements in tissue. Toward that end, a recent approach using water as that internal standard offers hope for quantitative SRS imaging in tissue <ref type="bibr">(111)</ref>.</p><p>Regardless of the quantitative metrics used, it is important to stress that rigorous  </p></div><note xmlns="http://www.tei-c.org/ns/1.0" place="foot" xml:id="foot_0"><p>AC15_FuFig06.pdf 1 2/17/22 2:55 PM</p></note>