			<titleStmt><title level='a'>Perfluorocarbon nanodroplet size, acoustic vaporization, and inertial cavitation affected by lipid shell composition &lt;i&gt;in vitro&lt;/i&gt;</title></titleStmt>
			<publicationStmt>
				<publisher></publisher>
				<date>10/01/2022</date>
			</publicationStmt>
			<sourceDesc>
				<bibl> 
					<idno type="par_id">10430034</idno>
					<idno type="doi">10.1121/10.0014934</idno>
					<title level='j'>The Journal of the Acoustical Society of America</title>
<idno>0001-4966</idno>
<biblScope unit="volume">152</biblScope>
<biblScope unit="issue">4</biblScope>					

					<author>Phoebe J. Welch</author><author>David S. Li</author><author>Craig R. Forest</author><author>Lilo D. Pozzo</author><author>Chengzhi Shi</author>
				</bibl>
			</sourceDesc>
		</fileDesc>
		<profileDesc>
			<abstract><ab><![CDATA[Perfluorocarbon nanodroplets (PFCnDs) are ultrasound contrast agents that phase-transition from liquid nanodroplets to gas microbubbles when activated by laser irradiation or insonated with an ultrasound pulse. The dynamics of PFCnDs can vary drastically depending on the nanodroplet composition, including the lipid shell properties. In this paper, we investigate the effect of varying the ratio of PEGylated to non-PEGylated phospholipids in the outer shell of PFCnDs on the acoustic nanodroplet vaporization (liquid to gas phase transition) and inertial cavitation (rapid collapse of the vaporized nanodroplets) dynamics in vitro when insonated with focused ultrasound. Nanodroplets with a high concentration of PEGylated lipids had larger diameters and exhibited greater variance in size distribution compared to nanodroplets with lower proportions of PEGylated lipids in the lipid shell. PFCnDs with a lipid shell composed of 50:50 PEGylated to non-PEGylated lipids yielded the highest B-mode image intensity and duration, as well as the greatest pressure difference between acoustic droplet vaporization onset and inertial cavitation onset. We demonstrate that slight changes in lipid shell composition of PFCnDs can significantly impact droplet phase transitioning and inertial cavitation dynamics. These findings can help guide researchers to fabricate PFCnDs with optimized compositions for their specific applications.]]></ab></abstract>
		</profileDesc>
	</teiHeader>
	<text><body xmlns="http://www.tei-c.org/ns/1.0" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:xlink="http://www.w3.org/1999/xlink">
<div xmlns="http://www.tei-c.org/ns/1.0"><head>I. INTRODUCTION</head><p>Lipid shelled microbubbles have been used as Food and Drug Administration (FDA) approved ultrasound contrast agents for several decades <ref type="bibr">(Schneider 1999;</ref><ref type="bibr">Goertz et al., 2007;</ref><ref type="bibr">Faez et al., 2011)</ref>. These gaseous structures ranging in size from 1 to 5 lm in diameter can be injected into the bloodstream and provide contrast for imaging the blood and tissue environment within the body. Furthermore, these microbubbles scatter ultrasound nonlinearly, which enables improved resolution and high contrast images with less tissue background signal <ref type="bibr">(Goertz et al., 2003)</ref>. The use of microbubbles has enabled super-resolution ultrasound imaging of microvasculature in deep tissue structures, including the brain, muscles, and kidney <ref type="bibr">(O'Reilly and Hynynen, 2013;</ref><ref type="bibr">Errico et al., 2015;</ref><ref type="bibr">Song et al., 2018;</ref><ref type="bibr">Yu et al., 2018)</ref>. Microbubbles can also be utilized for ultrasound-mediated drug delivery to blood clots and endothelial cells <ref type="bibr">(Lentacker et al., 2009;</ref><ref type="bibr">Lensen et al., 2011)</ref>. However, there are several drawbacks to microbubble contrast agents. Their size limits them to the bloodstream, and they cannot extravasate into tissue via inter-endothelial gaps in blood vessels <ref type="bibr">(Matsunaga et al., 2012)</ref>. Microbubbles are also unstable, only lasting minutes in the body before dissolution <ref type="bibr">(Li et al., 2018)</ref>. In response to these issues, perfluorocarbon nanodroplets (PFCnDs) have been developed composed of similar materials, a lipid shell and perfluorocarbon core, but the nanodroplet core is liquid at room temperature, either because the perfluorocarbon core has a high bulk vaporization temperature or because the nanodroplets are in a superheated state <ref type="bibr">(Kripfgans et al., 2000;</ref><ref type="bibr">de Gracia Lux et al., 2017)</ref>. After exposure to an ultrasound pulse with sufficient energy, these nanodroplets phase-transition into gas microbubbles and can be ultrasonically imaged using nonlinear imaging techniques for improved contrast in the liquid or tissue environment <ref type="bibr">(Sheeran et al., 2011;</ref><ref type="bibr">Liu et al., 2017;</ref><ref type="bibr">Jing et al., 2020)</ref>. Depending on environmental conditions and nanodroplet composition, such as the choice of perfluorocarbon used and temperature of the nanodroplet suspension, the PFCnDs can recondense back to a liquid state and undergo the expansion-recondensation cycle for repeated imaging purposes or can remain as gas microbubbles and be eliminated from the body within minutes to hours <ref type="bibr">(Hannah et al., 2016;</ref><ref type="bibr">Yu et al., 2016;</ref><ref type="bibr">Yoon et al., 2018)</ref>. Due to their small size (&lt;300 nm), they can extravasate into the leaky vasculature of tumors, which have large inter-endothelial gaps ranging from 380 nm to 2 lm <ref type="bibr">(Rapoport et al., 2011;</ref><ref type="bibr">Yin et al., 2012;</ref><ref type="bibr">Achmad et al., 2018)</ref>. PFCnDs also exhibit significantly better stability in circulation; compared to microbubbles whose lifespan is minutes long, PFCnDs with perfluoropentane (PFP) or perfluorohexane (PFH) cores remain stable in their liquid state for hours to days in vitro and in vivo <ref type="bibr">(Williams et al., 2013;</ref><ref type="bibr">Mountford and Borden, 2016;</ref><ref type="bibr">Li et al., 2018;</ref><ref type="bibr">Ji et al., 2022)</ref>. PFCnDs can be utilized for targeted drug delivery and are superior for selective tissue ablation as they facilitate targeted, deep tissue heating without prefocal thermal delivery and damage to skin that occurs when using microbubbles <ref type="bibr">(Rapoport et al., 2011;</ref><ref type="bibr">Moyer et al., 2015;</ref><ref type="bibr">Cao et al., 2018;</ref><ref type="bibr">Goel et al., 2021)</ref>. Furthermore, due to their nanoscale size, functionalized PFCnDs can enter specific cell types via endocytosis for localized cell and tissue ablation; microbubbles are neither small nor stable enough for such applications <ref type="bibr">(Ishijima et al., 2019)</ref>.</p><p>The process of PFCnDs transitioning from liquid nanodroplets to gaseous microbubbles from an ultrasound pulse is termed acoustic droplet vaporization (ADV). The pressure at which the nanodroplets undergo ADV can vary significantly based on several nanodroplet properties: nanodroplet diameter, core composition, and shell composition <ref type="bibr">(Kawabata et al., 2005;</ref><ref type="bibr">Aliabouzar et al., 2019;</ref><ref type="bibr">Yarmoska et al., 2019)</ref>. These nanodroplet properties also affect inertial cavitation (IC) of PFCnDs, which is unstable bubble collapse with a broadband noise signature <ref type="bibr">(Kang et al., 2014)</ref>. So long as the core bulk boiling point of PFCnDs is lower than the temperature of its surrounding environment, ADV and IC of PFCnDs are serially linked to one another, with ADV occurring first to form gas microbubbles before IC occurs in those microbubbles, although ADV and IC pressure thresholds can overlap <ref type="bibr">(Schad and Hynynen, 2010;</ref><ref type="bibr">Wu et al., 2021)</ref>. Preventing IC from occurring while using PFCnDs and microbubbles as ultrasound contrast agents is particularly crucial for in vivo imaging due to its potential to cause significant damage to surrounding tissues; therefore, having a large pressure difference between ADV onset and IC onset is of importance to medical imaging researchers.</p><p>Previous studies have demonstrated that nanodroplet size, stability, and image contrast are heavily reliant on their lipid shell composition. <ref type="bibr">Mountford et al. (2015)</ref> studied PFCnDs composed of phospholipid shells with acyl lengths ranging from C14 to C24 and noted that the energy required to induce phase transitioning linearly trended with increasing acyl length chain <ref type="bibr">(Mountford et al., 2015)</ref>. <ref type="bibr">Yarmoska et al. (2019)</ref> showed that increasing the ratio of PEGylated to non-PEGylated lipids in photoacoustic PFCnDs yielded smaller nanodroplets with smaller standard deviations and stronger photoacoustic signals compared to PFCnDs with higher ratios of non-PEGylated to PEGylated lipids <ref type="bibr">(Yarmoska et al., 2019)</ref>. <ref type="bibr">Chattaraj et al. (2016)</ref> noted that different combinations of saturated and unsaturated phospholipids in combination with cholesterol affected the Bmode acoustic intensity of PFH-core PFCnDs, possibly due to the clustering and phases of the lipid shells <ref type="bibr">(Chattaraj et al., 2016)</ref>. From these papers, slight changes in lipid shell composition, like the carbon chain lengths or the ratio of different phospholipids and surfactants, can significantly impact size distribution and ultrasound image contrast. However, there is limited research on how the lipid shell impacts ultrasonically induced phase transitioning of PFCnDs and whether the shell composition influences the IC threshold.</p><p>In this work, we investigated the effect of varying lipid shell composition of PFCnDs on nanodroplet size, ADV and IC onset, and the ultrasound intensity and duration of phasetransitioned PFCnDs by varying the ratio of PEGylated and non-PEGylated lipids, which are commonly used for fabricating lipid shelled ultrasound contrast agents. These nanodroplets had cores composed of either PFP, PFH, or a combination of the two perfluorocarbons. These perfluorocarbons were selected to encourage recondensation of the phase-transitioned nanodroplets; using perfluorocarbons with lower boiling points like perfluorobutane (PFB) (boiling point &#188; -2 C) would result in microbubbles unable to recondense in the water bath heated to 37 C. PFCnDs fabricated via spontaneous nucleation were suspended in agarose hydrogels and insonated using a focused ultrasound transducer. A linear array transducer, synchronized to the focused ultrasound transducer, captured B-mode images and radio frequency (RF) data of the insonated PFCnDs, which were used to study the vaporization and cavitation dynamics of the nanodroplets, respectively.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>II. MATERIALS AND METHODS</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>A. Nanodroplet fabrication and size-exclusion techniques</head><p>Nanodroplets were fabricated using a spontaneous nucleation method as previously described by <ref type="bibr">Li et al. (2018)</ref>. This fabrication method was used as it can consistently create small (&lt;250 nm diameter), uniform PFCnDs with no need for specialized equipment or harmful chemicals like chloroform. Lipid stock solutions were created with varying molar ratios of 1,2-dipalmitoyl-sn-glycero-3-phosphocholine (DPPC) and N-(carbonyl-methoxypolyethylyeneglycol 2000)-1,2-distearoylsn-glycero-3-phosphoethanolamine (DSPE-PEG 2000 ) (NOF America Corp., White Plains, NY) dissolved in 190 proof ethanol. These lipid stocks were composed of 90:10 DPPC:DSPE-PEG 2000 , 50:50 DPPC:DSPE-PEG 2000 , or 10:90 DPPC:DSPE-PEG 2000 . The final concentration of lipids in these solutions was 2 lM. One milliliter of each stock solution was added to a microcentrifuge tube, and perfluorocarbonseither PFP (FluoroMed L.P., Round Rock, TX), PFH (FluoroMed L.P.), or a combination of the two-were added to the lipid solutions until the solution was saturated with perfluorocarbon, as characterized by a noticeable pellet of undissolved perfluorocarbon collected at the bottom of the microcentrifuge tube. The ratios of PFP:PFH used in these experiments were as follows: 100:0, 90:10, 70:30, 50:50, and 0:100. We ultimately want to create nanodroplets that will phase-transition under diagnostically safe acoustic pressures [mechanical index (MI) below 1.9] while also being capable of recondensing back into a liquid state, so the nanodroplets fabricated tended to have a higher ratio of PFP to PFH, and 30:70 and 10:90 PFP:PFH nanodroplets were not included in this study <ref type="bibr">(S &#184;en et al., 2015)</ref>. A 2% solution of DiI (a lipophilic fluorescent tracer; MilliporeSigma, St. Louis, MO) dissolved in ethanol was added to the lipid/ perfluorocarbon solution. DiI was used in all experiments as these PFCnDs with this composition will later be used in cell and tissue imaging studies. The microcentrifuge tube was vortex mixed and sonicated until the solution was cloudy. At this point, the solution was left for 20 min for additional undissolved perfluorocarbon to fall out of solution. 50 ll of this lipid/PFC/DiI solution were added to 100 ll of lipid stock solution in microcentrifuge tubes. Then 850 ll of a 7:2:1 water:propylene glycol:glycerol (MilliporeSigma) solution was added to the diluted lipid/PFC solutions. The rapid addition of this hydrophilic miscible solution caused the dissolved lipids and PFC to spontaneously nucleate into very small droplets with the lipid encapsulating the PFC "core" to form a stable dispersion in a process called spontaneous emulsification. From here, all samples were centrifuged for 80 min at 6000 &#194; g, the supernatant was removed, and the pellet of nanodroplets was resuspended in phosphate buffered saline (PBS) (Corning Inc., Corning, NY).</p><p>To test the optimal filtration technique, we created several batches of nanodroplets made from the same lipid shell stock through the spontaneous nucleation method. These solutions were then either passed through a 450 nm mesh filter, centrifugated, or left undisturbed. The size and concentration of these nanodroplet solutions were measured the same day as the ultrasound experiments, which was typically $24 h after fabrication. These were the four conditions we tested:</p><p>&#8226; Control: No changes to nanodroplet solution after resuspending in PBS. &#8226; Size-exclusion centrifugation: The nanodroplet solution was spun down at 300 &#194; g for 5 min to allow the large nanodroplets to settle to the bottom of the tubes. The supernatant was recovered, and this process was repeated three times. &#8226; Gravity filtration: A 450 nm mesh filter (Corning) was attached to a syringe. The nanodroplet solution was loaded into the syringe and allowed to pass through the filter via gravity. &#8226; Pressure filtration: A 450 nm mesh filter was attached to a syringe, and after the nanodroplet solution was loaded, the syringe plunger was depressed slowly until the entire nanodroplet solution passed through the filter.</p><p>On the day of experiments ($24 h post-fabrication), the concentration and size of the nanodroplets were measured using a NanoSight300 (Malvern Panalytical, Worcestershire, UK). The concentration for each PFCnD stock was noted and used to create appropriate dilution factors such that the same quantity of each type of PFCnDs was suspended in agarose gels. PFCnD solutions were stored at 4 C until use.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>B. Ultrasound imaging and calibration</head><p>All experiments took place in a water bath heated to 37 C. Nanodroplets were suspended in a 1% agarose gel (MilliporeSigma) to a final concentration of 1 &#194; 10 8 nanodroplets/ml (4.2 &#194; 10 &#192;5 %, v/v) and loaded into thin-walled plastic containers. A focused single-element transducer (H-101, Sonic Concepts Inc., Bothell, WA) operating at its third harmonic frequency of 3.5 MHz (determined by performing a parametric frequency sweep between 3 and 4 MHz) was aligned with the nanodroplet gel sample. The single-element transducer was connected to a function generator (Agilent E4422B, Santa Clara, CA) and amplified with an RF power amplifier (325 LA, E&amp;I Ltd., Rochester, NY). A coupling cone was placed on the single-element transducer to align the focus of the transducer to the nanodroplet sample. This single-element transducer setup was used to initiate droplet vaporization in the nanodroplet sample. Orthogonal to the nanodroplet sample was a linear array transducer (L7-4v, Philips, Bothell, WA) operating at its center frequency of 5 MHz and transmitting plane waves at a frame rate of 3500 Hz, which captured B-mode images of the vaporized nanodroplets above the focal spot of the focused ultrasound transducer <ref type="bibr">[Fig. 1(a)</ref>]. Data were acquired using a Research Ultrasound system (Vantage 256, Verasonics, Inc., Kirkland, WA) running a custom MATLAB script that synchronized the focused ultrasound pulse with the captured B-mode image. Thirty preactivation B-mode image frames were collected as background signal, and then the single-element transducer was triggered by the Vantage 256 to emit a 20-cycle burst sine wave at a specified pressure output to initiate ADV in the nanodroplet-loaded agarose gel, followed by 400 post-ADV B-mode frames. A 20-cycle burst was selected to keep the transmission duty cycle well below 10%. The 20cycle burst was also selected to simultaneously phasetransition a high proportion of PFCnDs while preventing IC, as longer bursts (20&#254; cycles) have greater chances of phase-transitioning more nanodroplets at lower insonation pressures while also risking IC at lower pressures <ref type="bibr">(Chen et al., 2003;</ref><ref type="bibr">Reznik et al., 2011;</ref><ref type="bibr">Guo et al., 2013;</ref><ref type="bibr">Wu et al., 2021)</ref>. This activation and imaging sequence was repeated 20 times per pressure output, starting at the lowest pressure (2 MPa) output and gradually ramping the pressure amplitude from 2 to 9.5 MPa PNP (MI &#188; 1-5). Data were stored as both raw RF data (for IC detection) and B-mode image frames (for ADV analysis). The singleelement transducer pressure output was calibrated prior to experiments using an HGL-0200 capsule hydrophone (ONDA Corp., Sunnyvale, CA), with RF data collected in LabVIEW 2019 (National Instruments, Austin, TX) and analyzed in MATLAB 2020a (MathWorks, Natick, MA). Voltage to pressure calibration of the single-element transducer was performed with the coupling cone and a thin ($1 mm) layer of 1% agarose gel in front of the transducer to simulate experimental conditions and account for any attenuation caused by the coupling cone and gel setup.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>C. ADV characterization</head><p>ADV was quantified by measuring the mean echo power (MEP) of the focal spot within the imaging window, as described by <ref type="bibr">Fabiilli et al. (2009)</ref> and used in later applications. The MEP is calculated by summing the squared amplitude of all pixel intensities within an imaging window and then normalized by the window area as follows:</p><p>where m is the frame number with dimensions M, N, and A is the amplitude at pixel i, j. The MEP of ultrasound frames just after insonation <ref type="bibr">[Fig. 1(c)</ref>] will be significantly higher than frames preceding the activation event <ref type="bibr">[Fig. 1(b)</ref>] if the activating ultrasound sequence was powerful enough to induce ADV. The MEP would also significantly decrease post-insonation after a short period if PFCnDs either recondensed or dissolved into the surrounding scaffold <ref type="bibr">[Fig. 1(d)</ref>]. We calculated the average MEP of 20 post-activation frames, ignoring the two frames immediately after the activating ultrasound sequence to prevent signal from the focused ultrasound transducer from interfering with our calculations.</p><p>The MEP calculations were used to analyze the duration and intensity of vaporized nanodroplets. MEP duration was characterized as the number of post-activation frames where the MEP was elevated above a threshold value, 20 &#194; MEP Avg, pre-activation , where MEP Avg, pre-activation is the average MEP value of the 30 pre-activation frames before the vaporization pulse is triggered. This threshold was determined by measuring the average and standard deviation of the pre-activation frames and selecting a multiplier that would ensure the MEP signal measured in the post-activation frames was caused by phase-transitioning PFCnDs and not background noise. Usually, the MEP of B-mode images of ADV was several orders of magnitude higher than frames containing no ADV, so increasing or decreasing the multiplier for the threshold value slightly (e.g., 65) does not significantly affect the measurement. MEP intensity comparisons between different groups of PFCnDs were made by determining the maximum MEP induced by the activation pulse per activation and imaging sequence.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>D. IC characterization</head><p>IC was determined by analyzing the RF data received by the imaging transducer immediately after the phasetransitioning ultrasound pulse. The RF data collected by the four elements of the linear array transducer located directly above the focal spot of the single-element transducer was analyzed by taking the fast Fourier transform and analyzing the signal amplitude between 4 and 6 MHz. This frequency window was selected because it did not contain any of the second harmonic signal coming from the nanodroplets caused by the single-element transducer ($7 MHz) and could be used as a region to characterize the noise floor. Since IC is typically measured by broadband acoustic emission, we could use this frequency region to gauge if there was an elevated noise floor after the activation ultrasound pulse. Transducer elements were specifically selected to be located above the focus spot of the activation pulse, as determined by analyzing the B-mode image data. The signal from the lowest activation pulse pressure was used as the noise floor baseline measurement. A threshold was empirically selected in a similar manner as used in <ref type="bibr">Fabiilli et al. (2009)</ref> to distinguish an IC event from background noise caused by the insonating focused ultrasound transducer. The threshold for IC detection was set as The threshold for IC detection is quite low so that the IC analysis is very sensitive to any potential IC events in the nanodroplet samples.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>E. Statistical analysis</head><p>ADV and IC data were plotted using MATLAB, and sigmoidal best-fit curves were fitted to the scatterplot data. The custom equation for the sigmoidal curve is y &#188; a &#254; &#240;b &#192; a&#222;= &#189;1 &#254; &#240;x=c&#222; d , where x is the pressure value; y is either the cavitation probability or ADV intensity; a and b are the maximum and minimum y-values, respectively; c is the xvalue at the y-midpoint; and d is the slope at c <ref type="bibr">(Fabiilli et al., 2009)</ref>. These sigmoidal curves were used to determine the onset of ADV and IC. ADV onset was characterized by the appearance of ultrasound contrast immediately after the insonating pulse by the focused ultrasound transducer, caused by the phase transitioning of the PFCnDs. This appearance of ultrasound contrast correlates well with the xvalue at the y-midpoint of the sigmoidal curve. IC onset is defined as the 50% crossing of the sigmoid fit on the cavitation curve <ref type="bibr">(Li et al., 2018)</ref>. Box plots and statistical analysis were conducted in R Studio. One-way analysis of variance (ANOVA) and Tukey's t-test were used to determine statistical significance in variances between different nanodroplet compositions.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>III. RESULTS AND DISCUSSION</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>A. Nanodroplet filtration</head><p>Nanodroplet composition impacts the size distribution of the resulting nanodroplets, which can significantly influence ADV and IC thresholds <ref type="bibr">(Ferri et al., 2021)</ref>. If there are large nanodroplets present in the nanodroplet suspension, ADV can be observed at relatively low pressures as nanodroplets with larger diameters require lower pressures to vaporize than those with smaller diameters due to superharmonic focusing <ref type="bibr">(Kripfgans et al., 2004;</ref><ref type="bibr">Shpak et al., 2014;</ref><ref type="bibr">Aliabouzar et al., 2019)</ref>. To eliminate particularly large nanodroplets, centrifugation is commonly used to separate large, coalesced nanodroplets from the rest of the sample <ref type="bibr">(Mercado et al., 2016)</ref>. However, this size-exclusion technique has not been extensively compared to other separation methods, such as mesh filtration. We sought to optimize a simple but effective nanodroplet filtration method to create nanodroplet solutions of uniform, monodisperse nanodroplets by passing our nanodroplet suspensions through a 450 nm mesh filter. We selected 450 nm mesh filters for our filtration studies knowing that mesh filters are not perfectly monodisperse, and if we used filters with an average pore size closer to our initial average nanodroplet diameter ($220 nm), we would risk excluding a much higher proportion of nanodroplets and, thus, significantly reduce the quantity of nanodroplets in our solutions <ref type="bibr">(Ullmann et al., 2019)</ref>. The control sample of nanodroplets, which did not undergo any size-exclusion separation technique, had the largest mean diameter and standard deviation in diameter distribution amongst the samples [Figs. <ref type="figure">2(a</ref> To eliminate the risk of including large (&gt;450 nm) nanodroplets in our study, all nanodroplet samples were pressure filtered before being used for ultrasound imaging in subsequent experiments.</p><p>Nanodroplets passed through a 450 nm mesh filter had significantly smaller diameters than non-filtered and centrifugated nanodroplets. At surface level, these findings seem obvious, but in reality, the results are somewhat surprising because it has been hypothesized that pushing lipid shelled PFCnDs through a mesh filter can cause some nanodroplets to break open and coalesce with nearby droplets <ref type="bibr">(Paproski et al., 2016)</ref>. This theory is still possible, though perhaps the large nanodroplets formed via coalescing are less stable than the smaller ones and either evaporate or coalesce and sediment within the 24 h period between fabricating the PFCnDs and measuring their size and concentration. The more likely reason that mesh filtered PFCnDs tend to have smaller average diameters is because any nanodroplets greater than the mesh pore size (in this case, 450 nm) are excluded from size measurements and do not contribute to the diameter calculations by the NanoSight 300. The mesh filtered nanodroplets may have a slightly larger standard deviation compared to the centrifugated and control nanodroplets either because some nanodroplets are fractured by the filter and the lipid shell fragments form very small (&lt;100 nm) liposomes containing no perfluorocarbons or because some of the nanodroplets have excess lipid shell that is shed as they pass through the filter and form liposomes <ref type="bibr">(Borden et al., 2005)</ref>.</p><p>Next, we investigated how the lipid shell composition impacts nanodroplet size distribution. PFCnDs with a PFH core were passed through a 450 nm mesh filter attached to a 1 ml syringe with the plunger gently applied to eliminate larger nanodroplets. Based on three different batches of PFCnDs, nanodroplets with a 10:90 ratio of DPPC:DSPE-PEG 2000 were significantly larger than 50:50 and 90:10 lipid shelled nanodroplets <ref type="bibr">[Fig. 3(a)</ref>]. There was no significant difference in size between the 50:50 and 90:10 lipid shelled PFCnDs. No significant difference in nanodroplet diameter is observed when PFCnDs are grouped by core composition <ref type="bibr">[Fig. 3(b)</ref>], indicating that the lipid shell, rather than core composition, influences nanodroplet size distributions. </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>B. ADV intensity vs shell composition</head><p>Ultrasound B-mode image intensity caused by the phase transitioning of the PFCnDs is critical for medical imaging applications, as PFCnDs that produce a stronger acoustic signal are easier to locate in vivo. The intensity of the resulting image contrast produced after ADV was quantified by calculating the MEP in the focal region <ref type="bibr">[Fig. 4(a)</ref>]. Once the insonating pressure from the single-element transducer reached a certain threshold, a noticeable bubble cloud would appear in the transducer's focal region. As the transducer increased above this pressure, the bubble cloud intensity increased, indicating a greater proportion of nanodroplets were phase-transitioned in the focal region and yielding a higher MEP. This trend continued until the insonating pressure reached 7-7.5 MPa peak negative pressure (PNP) (MI &#188; 3.75-4), at which point the maximum number of PFCnDs were phase-transitioned during each insonating pulse. Analyzed across three experimental periods with separate batches of nanodroplets, the 50:50 lipid shelled PFCnDs exhibited significantly stronger ADV intensities than the 10:90 and 90:10 lipid shelled nanodroplets <ref type="bibr">[Fig. 4(b)</ref>]. The 90:10 and 10:90 non-PEGylated:PEGylated PFCnDs had comparable MEP values.</p><p>These insonating pressures exceed the recommended pressures used in medical ultrasound and are required in this study due to the agarose hydrogel environment whose matrix suppresses some nanodroplet expansion (as opposed to an all-liquid environment more commonly used in nanodroplet studies). The ADV threshold can be easily modified by changing the core composition to contain lower boiling point perfluorocarbons, such as PFB, and increasing the number of cycles per insonating pulse.</p><p>The increased B-mode intensity in nanodroplets with a 50:50 lipid shell ratio of non-PEGylated:PEGylated lipids could be caused by several factors. It is possible that these 50:50 nanodroplets have improved vaporization efficiency compared to other nanodroplet compositions, so a larger proportion of nanodroplets expand with each insonating pulse. The cause of this improved vaporization efficiency may be the packing structure and distribution of the two phospholipids in the nanodroplet shell. In one study of multi-component phospholipid micelles by <ref type="bibr">Viitala et al. (2019)</ref>, the authors noted that increasing the ratio of DSPE-PEG 2000 in the DPPC:DSPE-PEG 2000 liposomes from $10% DSPE-PEG 2000 to 50% DSPE-PEG 2000 caused a shape change from bicelles to slightly elongated micelles <ref type="bibr">(Viitala et al., 2019)</ref>. DPPC:DSPE-PEG 2000 lipid shelled nanodroplets have small domains of only DPPC or DSPE-PEG 2000 , creating a solid-liquid ordered phase coexistence, as detailed in <ref type="bibr">Chattaraj et al. (2016)</ref>. The authors also noted that increasing the PEG concentration in these droplets from 3% mol to 20% mol caused a significant increase in acoustic signal, likely because PEGylated lipids are typically included in lipid shelled nanodroplets for increased steric stabilization and perhaps because this lipid shell organization, with distinct regions of DPPC and DSPE-PEG 2000 , causes nanodroplets to phase-transition well without IC or dissolution into the surrounding medium. Our results agree with Chattaraj et al. in that increasing the ratio of PEGylated lipids from 10% mol to 50% mol yielded a significantly stronger acoustic response post-ultrasound insonation. However, the addition of too much PEGylated lipid could create steric hindrance issues and/or create lipid shells too stiff for efficient nanodroplet expansion, hence, why our 10:90 DPPC:DSPE-PEG 2000 nanodroplets had lower acoustic signal overall.</p><p>Another potential cause of stronger acoustic intensity is that the 50:50 lipid shelled nanodroplets yield larger gaseous microbubbles after insonation compared to the other two compositions. PFCnDs will typically yield microbubbles that are 3-5 times larger in diameter than their liquid, condensed form, with smaller nanodroplets forming proportionally smaller microbubbles due to increased Laplace pressure <ref type="bibr">(Sheeran et al., 2011)</ref>. The interfacial tension differences between the lipid shell and perfluorocarbon core may impact the Laplace pressure and shell stiffness, which could affect ADV conversion efficiency and the responsiveness of the resulting bubbles for the lipid shelled PFCnDs. Thus, the 50:50 lipid shelled nanodroplets may stably expand to a larger microbubble diameter than the other compositions, or a greater proportion of the 50:50 lipid shelled nanodroplets expand with each insonating ultrasound pulse, providing greater ultrasound contrast. Confirmation of these theories would require an experimental setup similar to that in <ref type="bibr">Seda et al. (2015)</ref>, in which PFCnDs were placed above a microscope objective, and the resulting bubble clouds postinsonation were optically observed <ref type="bibr">(Seda et al., 2015)</ref>. Further research on the ideal ratio of PEGylated to non-PEGylated lipids in PFCnDs must be conducted to determine the ideal shell ratio, but among the nanodroplet compositions used in this study, 50:50 lipid shelled PFCnDs would be the ideal ultrasound contrast agent to use in imaging applications where stark image contrast is required.</p><p>Although using nanodroplets with a higher PEGylated lipid ratio may yield improved ultrasound contrast, one area of concern is potential bioeffects caused by lipid shell composition. PEG is a common additive to a variety of nanomaterials used to increase in vivo stability <ref type="bibr">(Niidome et al., 2006;</ref><ref type="bibr">Diaz et al., 2018)</ref>. Studies investigating the use of PFCnDs in vivo have used a variety of PEG ratios in lipid shells, even up to 90% PEGylated lipids, with no reported ill effect <ref type="bibr">(Xiang et al., 2019;</ref><ref type="bibr">Yarmoska et al., 2019;</ref><ref type="bibr">Sheng et al., 2021)</ref>. While using PFCnDs with various PEG ratios appears to be safe for one-time, short-term experiments, long-term and repeated administration of PFCnDs, especially those with higher PEG ratios, has not been thoroughly investigated and could potentially lead to nanodroplet build-up within certain organs, increased clearance from the circulatory system, and cytotoxicity. Researchers studying PEGylated liposomes for drug delivery have noted that repeated injections of PEGylated liposomes at certain concentrations lead to accelerated blood clearance after the first injection due to anti-PEG IgM-mediated activation <ref type="bibr">(Ishida et al., 2007;</ref><ref type="bibr">Ichihara et al., 2010;</ref><ref type="bibr">Suzuki et al., 2012)</ref>. Additionally, high PEG ratios can prevent nanoparticle uptake into cells, which may be an issue if the desired use of these PFCnDs is to use them for targeted intracellular uptake <ref type="bibr">(Verhoef and Anchordoquy, 2013;</ref><ref type="bibr">Pozzi et al., 2014)</ref>. Researchers should consider these potential effects that may result in vivo when creating PFCnDs and design the lipid shell composition appropriately based on the desired application. Researchers should also use caution if utilizing these PFCnDs for long durations or repeated injections as there may be unknown cytotoxic effects or increased immune activation that result from repeated use.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>C. ADV duration vs shell composition</head><p>The duration for which vaporized PFCnDs remained in their gaseous state was calculated by measuring the number of frames in which the MEP in ultrasound frames post-ADV was elevated above the threshold value 20 &#194; MEP Avg, pre-activation . PFCnDs composed of a 10:90 DPPC:DSPE-PEG 2000 lipid shell had the briefest duration in the gaseous state across all nanodroplet batches and core compositions <ref type="bibr">[Fig. 5(a)</ref>]. There was no significant difference in ADV duration between the 50:50 and 90:10 lipid shelled PFCnDs. Post-insonation peaks in MEP were observed in the same nanodroplet samples after multiple insonation events, even in nanodroplets with boiling points below the environment temperature, suggesting nanodroplet recondensation <ref type="bibr">[Fig. 5(b)</ref>].</p><p>The disappearing ultrasound signal after phase transitioning can be attributed to several factors. The perfluorocarbon core composition can significantly impact whether nanodroplets can recondense or dissolve into the surrounding environment. Nanodroplet recondensation has been observed in nanodroplets composed of both PFP and PFH cores <ref type="bibr">(Hannah et al., 2016;</ref><ref type="bibr">Ishijima et al., 2016;</ref><ref type="bibr">Luke et al., 2016;</ref><ref type="bibr">Yu et al., 2016;</ref><ref type="bibr">Yoon et al., 2018)</ref>, so it is possible that many of the nanodroplets within the PFCnD samples tested in these experiments are recondensing to a liquid state after insonation, as observed in <ref type="bibr">Fig. 5(b)</ref>. The sustained signal intensity across all subsequent insonating pulses, even at very high pressures, supports this claim. Furthermore, the agarose matrix in which the PFCnDs are suspended can encourage recondensation rather than fragmentation, whereas in flow tube phantoms, there tends to be a loss in ultrasound signal as the liquid nanodroplet suspension is repeatedly subjected to high intensity ultrasound pulses <ref type="bibr">(Yoo et al., 2018)</ref>. Differences in interfacial properties of the lipid shelled PFCnDs could significantly impact the recondensation of the vaporized bubbles, with the 10:90 lipid shelled nanodroplets exhibiting the greatest interfacial forces and the 50:50 lipid shelled nanodroplets possessing weaker interfacial forces. The high proportion of PEGylated lipids in the 10:90 PFCnDs could also be the cause of rapidly disappearing acoustic signal as the PEG creates stiffer encapsulating shells, encouraging fracturing and/or rapid recondensation of the nanodroplets. Last, some PFCnDs, especially those composed of a core with a lower vaporization temperature threshold, may fragment post-phase transitioning due to being in an environment above the vaporization point of the core and possessing an unstable lipid shell, despite their suspension in agarose gel <ref type="bibr">(de Jong et al., 2000;</ref><ref type="bibr">Huynh et al., 2015)</ref>.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>D. Pressure differential between ADV onset and IC onset</head><p>We also investigated the onset of ADV and IC in all PFCnD samples. IC was characterized by a significant increase in the noise floor directly after insonation by the single-element transducer compared to the noise floor of the lowest insonation pressure used (2 MPa). From here, the pressure difference between ADV onset (insonation pressure at which nanodroplets begin to phase-transition) and IC onset (characterized as 50% cavitation probability) was calculated across all samples. PFCnDs with a 50:50 lipid shell ratio had the largest pressure difference between when ADV was noticeable and when IC surpassed the vaporization threshold compared to 10:90 and 90:10 lipid shelled nanodroplets, as exhibited by the representative plots in Figs. 6(a)-6(c). This trend was observed across all nanodroplet samples of various sizes and core composition, and the pressure differential was considerably larger in nanodroplets with smaller diameters <ref type="bibr">Fig. 6(d)</ref>].</p><p>The differential between ADV and IC onset is an important metric for both imaging and therapeutic applications. For ultrasonic imaging of these contrast agents in vivo, IC can be detrimental to surrounding cells and tissues; inducing ADV without triggering IC, therefore, is critical. Creating PFCnDs with the lowest possible ADV threshold is crucial in diagnostic imaging to enable phase transitioning without using insonating pressures above the FDA's MI threshold for in vivo applications. Contrary to this, damaging tissues and cells via IC can be the goal of using ultrasonically triggerable PFCnDs, so nanodroplets with a lower IC threshold are ideal, and the pressure difference between ADV onset and IC onset is negligible. Based on these considerations, 50:50 lipid shelled PFCnDs would be ideal contrast agents for theranostic applications; these droplets exhibited the greatest ADV-to-IC onset pressure differential and highest contrast to noise and could, thus, facilitate both ultrasound imaging (ADV) and ultrasoundinduced drug released or localized tissue ablation (IC). However, the perfluorocarbon cores used in this paper yielded PFCnDs with ADV thresholds above the FDA's MI limit. Changing the perfluorocarbon core to include perfluorocarbons with lower vaporization temperatures (e.g., PFB) can bring the ADV threshold to biologically safe pressures.</p><p>It is important to consider that the window between ADV and IC is quite narrow, even for the 50:50 lipid shelled PFCnDs. This narrow pressure differential has been observed in other works and has implications for utilizing PFCnDs safely for in vivo applications <ref type="bibr">(Wu et al., 2021)</ref>. Our findings demonstrate that the ADV-to-IC pressure differential is larger for PFCnDs with smaller diameters, but one problem with small nanodroplets (diameters &lt;200 nm) is that they require very high insonation pressures exceeding MI limits compared to larger nanodroplets. Researchers have developed ways to lower and control both ADV and IC thresholds by using unique transducer configurations or ultrasound standing waves, but future work should investigate whether these techniques can be used to create a greater ADV-to-IC differential, whether these techniques can be used to insonate and phase-transition nanodroplets with sub-200 nm diameters, and whether these techniques are applicable in vivo <ref type="bibr">(Guo et al., 2017;</ref><ref type="bibr">Xu et al., 2018)</ref>.</p><p>It is important to note that all experiments in this paper were conducted in tissue-mimicking phantoms within a temperature-controlled water bath. We opted to conduct experiments in this well-controlled environment, rather than in vivo, so that any differences in acoustic behavior were attributable to the PFCnDs instead of slight changes to the experimental setup (e.g., differences between animals, changes in tissue stiffness, etc.). Future work should include testing these PFCnDs with different lipid shell compositions in vivo or in environments that better represent in vivo conditions, such as excised tissue, to validate that these differences in nanodroplet behavior are still exhibited in vivo. Additionally, as mentioned in Sec. III B, special care should be taken if using these PFCnDs for long-term studies with repeated injections to understand potential bioeffects caused by the presence of these nanodroplets in vivo.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>IV. CONCLUSION</head><p>We investigated the influences of the lipid shell composition on the size distribution, ultrasound characteristics, and vaporization dynamics of PFCnDs. Our results suggest that passing nanodroplets through a mesh filter yields nanodroplet suspensions with smaller average diameters by excluding large (450&#254; nm diameter) nanodroplets. Furthermore, lipid shell PFCnDs with a high proportion of PEGylated lipids have larger diameters, though the size distribution of nanodroplets may be more dependent on the fabrication method than lipid shell composition. PFCnDs with a 50:50 DPPC:DSPE-PEG 2000 lipid shell composition created the strongest ultrasound contrast, had the longest duration in the phasetransitioned state, and had the greatest pressure difference between ADV onset and IC onset among all nanodroplet samples used in this study. Based on these results, the 50:50 lipid shelled PFCnDs are ideal candidates for theranostic applications and ultrasound imaging in general due to their monodispersity, high contrast-to-noise ratio, and large ADV-to-IC pressure differential. 90:10 and 10:90 lipid shelled PFCnDs are both excellent candidates for tissue ablation and localized drug delivery. Altering the perfluorocarbon core can influence the duration in which the PFCnD remains gaseous as well as the vaporization threshold of the nanodroplets. The findings in this work can help guide researchers to fabricate PFCnDs with the desired ultrasonic properties for a variety of applications.</p></div><note xmlns="http://www.tei-c.org/ns/1.0" place="foot" xml:id="foot_0"><p>https://doi.org/10.1121/10.0014934</p></note>