skip to main content


Title: PreQual: An automated pipeline for integrated preprocessing and quality assurance of diffusion weighted MRI images
Purpose

Diffusion weighted MRI imaging (DWI) is often subject to low signal‐to‐noise ratios (SNRs) and artifacts. Recent work has produced software tools that can correct individual problems, but these tools have not been combined with each other and with quality assurance (QA). A single integrated pipeline is proposed to perform DWI preprocessing with a spectrum of tools and produce an intuitive QA document.

Methods

The proposed pipeline, built around the FSL, MRTrix3, and ANTs software packages, performs DWI denoising; inter‐scan intensity normalization; susceptibility‐, eddy current‐, and motion‐induced artifact correction; and slice‐wise signal drop‐out imputation. To perform QA on the raw and preprocessed data and each preprocessing operation, the pipeline documents qualitative visualizations, quantitative plots, gradient verifications, and tensor goodness‐of‐fit and fractional anisotropy analyses.

Results

Raw DWI data were preprocessed and quality checked with the proposed pipeline and demonstrated improved SNRs; physiologic intensity ratios; corrected susceptibility‐, eddy current‐, and motion‐induced artifacts; imputed signal‐lost slices; and improved tensor fits. The pipeline identified incorrect gradient configurations and file‐type conversion errors and was shown to be effective on externally available datasets.

Conclusions

The proposed pipeline is a single integrated pipeline that combines established diffusion preprocessing tools from major MRI‐focused software packages with intuitive QA.

 
more » « less
PAR ID:
10450982
Author(s) / Creator(s):
 ;  ;  ;  ;  ;  ;  ;  ;  ;  ;  ;  ;  ;  ;  ;  ;  ;  
Publisher / Repository:
Wiley Blackwell (John Wiley & Sons)
Date Published:
Journal Name:
Magnetic Resonance in Medicine
Volume:
86
Issue:
1
ISSN:
0740-3194
Page Range / eLocation ID:
p. 456-470
Format(s):
Medium: X
Sponsoring Org:
National Science Foundation
More Like this
  1. Purpose

    To develop a physics‐guided deep learning (PG‐DL) reconstruction strategy based on a signal intensity informed multi‐coil (SIIM) encoding operator for highly‐accelerated simultaneous multislice (SMS) myocardial perfusion cardiac MRI (CMR).

    Methods

    First‐pass perfusion CMR acquires highly‐accelerated images with dynamically varying signal intensity/SNR following the administration of a gadolinium‐based contrast agent. Thus, using PG‐DL reconstruction with a conventional multi‐coil encoding operator leads to analogous signal intensity variations across different time‐frames at the network output, creating difficulties in generalization for varying SNR levels. We propose to use a SIIM encoding operator to capture the signal intensity/SNR variations across time‐frames in a reformulated encoding operator. This leads to a more uniform/flat contrast at the output of the PG‐DL network, facilitating generalizability across time‐frames. PG‐DL reconstruction with the proposed SIIM encoding operator is compared to PG‐DL with conventional encoding operator, split slice‐GRAPPA, locally low‐rank (LLR) regularized reconstruction, low‐rank plus sparse (L + S) reconstruction, and regularized ROCK‐SPIRiT.

    Results

    Results on highly accelerated free‐breathing first pass myocardial perfusion CMR at three‐fold SMS and four‐fold in‐plane acceleration show that the proposed method improves upon the reconstruction methods use for comparison. Substantial noise reduction is achieved compared to split slice‐GRAPPA, and aliasing artifacts reduction compared to LLR regularized reconstruction, L + S reconstruction and PG‐DL with conventional encoding. Furthermore, a qualitative reader study indicated that proposed method outperformed all methods.

    Conclusion

    PG‐DL reconstruction with the proposed SIIM encoding operator improves generalization across different time‐frames /SNRs in highly accelerated perfusion CMR.

     
    more » « less
  2. Purpose

    Isotropic diffusion encoding efficiently encodes additional microstructural information in combination with conventional linear diffusion encoding. However, the gradient‐intensive isotropic diffusion waveforms generate significant eddy currents, which cause image distortions. The purpose of this study is to present a method for designing isotropic diffusion encoding waveforms with intrinsic eddy current nulling.

    Methods

    Eddy current nulled gradient waveforms were designed using a constrained optimization waveform for a 3T GE Premier MRI system. Encoding waveforms were designed for a variety of eddy current null times and sequence timings to evaluate the achievable b‐value. Waveforms were also tested with both eddy current nulling and concomitant field compensation. Distortion reduction was tested in both phantoms and the in vivo human brain.

    Results

    The feasibility of isotropic diffusion encoding with intrinsic correction of eddy current distortion and signal bias from concomitant fields was demonstrated. The constrained optimization algorithm produced gradient waveforms with the specified eddy current null times. The reduction in the achievable diffusion weighting was dependent on the number of eddy current null times. A reduction in the eddy current–induced image distortions was observed in both phantoms and in vivo human subjects.

    Conclusion

    The proposed framework allows the design of isotropic diffusion‐encoding sequences with reduced image distortion.

     
    more » « less
  3. Abstract Purpose

    The aim of this work is to develop a method to solve the ill‐posed inverse problem of accelerated image reconstruction while correcting forward model imperfections in the context of subject motion during MRI examinations.

    Methods

    The proposed solution uses a Bayesian framework based on deep generative diffusion models to jointly estimate a motion‐free image and rigid motion estimates from subsampled and motion‐corrupt two‐dimensional (2D) k‐space data.

    Results

    We demonstrate the ability to reconstruct motion‐free images from accelerated two‐dimensional (2D) Cartesian and non‐Cartesian scans without any external reference signal. We show that our method improves over existing correction techniques on both simulated and prospectively accelerated data.

    Conclusion

    We propose a flexible framework for retrospective motion correction of accelerated MRI based on deep generative diffusion models, with potential application to other forward model corruptions.

     
    more » « less
  4. Purpose

    To mitigate a common artifact in spiral real‐time MRI, caused by aliasing of signal outside the desired FOV. This artifact frequently occurs in midsagittal speech real‐time MRI.

    Methods

    Simulations were performed to determine the likely origin of the artifact. Two methods to mitigate the artifact are proposed. The first approach, denoted as “large FOV” (LF), keeps an FOV that is large enough to include the artifact signal source during reconstruction. The second approach, denoted as “estimation‐subtraction” (ES), estimates the artifact signal source before subtracting a synthetic signal representing that source in multicoil k‐space raw data. Twenty‐five midsagittal speech‐production real‐time MRI data sets were used to evaluate both of the proposed methods. Reconstructions without and with corrections were evaluated by two expert readers using a 5‐level Likert scale assessing artifact severity. Reconstruction time was also compared.

    Results

    The origin of the artifact was found to be a combination of gradient nonlinearity and imperfect anti‐aliasing in spiral sampling. The LF and ES methods were both able to substantially reduce the artifact, with an averaged qualitative score improvement of 1.25 and 1.35 Likert levels for LF correction and ES correction, respectively. Average reconstruction time without correction, with LF correction, and with ES correction were 160.69 ± 1.56, 526.43 ± 5.17, and 171.47 ± 1.71 ms/frame.

    Conclusion

    Both proposed methods were able to reduce the spiral aliasing artifacts, with the ES‐reduction method being more effective and more time efficient.

     
    more » « less
  5. Abstract Introduction

    Golden retriever muscular dystrophy (GRMD) is a spontaneous X‐linked canine model of Duchenne muscular dystrophy that resembles the human condition. Muscle percentage index (MPI) is proposed as an imaging biomarker of disease severity in GRMD.

    Methods

    To assess MPI, we used MRI data acquired from nine GRMD samples using a 4.7 T small‐bore scanner. A machine learning approach was used with eight raw quantitative mapping of MRI data images (T1m, T2m, two Dixon maps, and four diffusion tensor imaging maps), three types of texture descriptors (local binary pattern, gray‐level co‐occurrence matrix, gray‐level run‐length matrix), and a gradient descriptor (histogram of oriented gradients).

    Results

    The confusion matrix, averaged over all samples, showed 93.5% of muscle pixels classified correctly. The classification, optimized in a leave‐one‐out cross‐validation, provided an average accuracy of 80% with a discrepancy in overestimation for young (8%) and old (20%) dogs.

    Discussion

    MPI could be useful for quantifying GRMD severity, but careful interpretation is needed for severe cases.

     
    more » « less