Title: Comparison of gaseous ubiquitin ion structures obtained from a solid and solution matrix using ion mobility spectrometry/mass spectrometry
Rationale

Examining surface protein conformations, and especially achieving this with spatial resolution, is an important goal. The recently discovered ionization processes offer spatial‐resolution measurements similar to matrix‐assisted laser desorption/ionization (MALDI) and produce charge states similar to electrospray ionization (ESI) extending higher‐mass protein applications directly from surfaces on high‐performance mass spectrometers. Studying a well‐interrogated protein by ion mobility spectrometry‐mass spectrometry (IMS‐MS) to access effects on structures using a solidvs.solvent matrix may provide insights.

Methods

Ubiquitin was studied by IMS‐MS using new ionization processes with commercial and homebuilt ion sources and instruments (Waters SYNAPT G2(S)) and homebuilt 2 m drift‐tube instrument; MS™ sources). Mass‐to‐charge and drift‐time (td)‐measurements are compared for ubiquitin ions obtained byinletandvacuumionization using laserspray ionization (LSI), matrix‐ (MAI) and solvent‐assisted ionization (SAI), respectively, and compared with those from ESI under conditions that are most comparable.

Results

Using the same solution conditions with SYNAPT G2(S) instruments, td‐distributions of various ubiquitin charge states from MAI, LSI, and SAI are similar to those from ESI using a variety of solvents, matrices, extraction voltages, a laser, and temperature only, showing subtle differences in more compact features within the elongated distribution of structures. However, on a homebuilt drift‐tube instrument, within the elongated distribution of structures, both similar and different td‐distributions are observed for ubiquitin ions obtained by MAI and ESI. MAI‐generated ions are frequently narrower in their td‐distributions.

Conclusions

Direct comparisons between ESI and the new ionization methods operational directly from surfaces suggest that the protein in its solution structure prior to exposure to the ionization event is either captured (frozen out) at the time of crystallization, or that the protein in the solid matrix is associated with sufficient solvent to maintain the solution structure, or, alternatively, that the observed structures are those related to what occurs in the gas phase with ESI‐ or MAI‐generated ions and not with the solution structures.

 
more » « less
NSF-PAR ID:
10453315
Author(s) / Creator(s):
 ;  ;  ;  ;  ;  ;  ;  ;  ;  
Publisher / Repository:
Wiley Blackwell (John Wiley & Sons)
Date Published:
Journal Name:
Rapid Communications in Mass Spectrometry
Volume:
35
Issue:
S1
ISSN:
0951-4198
Format(s):
Medium: X
Sponsoring Org:
National Science Foundation
More Like this
  1. Rationale

    The developments of new ionization technologies based on processes previously unknown to mass spectrometry (MS) have gained significant momentum. Herein we address the importance of understanding these unique ionization processes, demonstrate the new capabilities currently unmet by other methods, and outline their considerable analytical potential.

    Methods

    Theinletandvacuumionization methods of solvent‐assisted ionization (SAI), matrix‐assisted ionization (MAI), and laserspray ionization can be used with commercial and dedicated ion sources producing ions from atmospheric or vacuum conditions for analyses of a variety of materials including drugs, lipids, and proteins introduced from well plates, pipet tips and plate surfaces with and without a laser using solid or solvent matrices. Mass spectrometers from various vendors are employed.

    Results

    Results are presented highlighting strengths relative to ionization methods of electrospray ionization (ESI) and matrix‐assisted laser desorption/ionization. We demonstrate the utility of multi‐ionization platforms encompassing MAI, SAI, and ESI and enabling detection of what otherwise is missed, especially when directly analyzing mixtures. Unmatched robustness is achieved with dedicated vacuum MAI sources with mechanical introduction of the sample to the sub‐atmospheric pressure (vacuumMAI). Simplicity and use of a wide array of matrices are attained using a conduit (inletionization), preferably heated, with sample introduction from atmospheric pressure. Tissue, whole blood, urine (including mouse, chicken, and human origin), bacteria strains and chemical on‐probe reactions are analyzed directly and, especially in the case ofvacuumionization, without concern of carryover or instrument contamination.

    Conclusions

    Examples are provided highlighting the exceptional analytical capabilities associated with the novel ionization processes in MS that reduce operational complexity while increasing speed and robustness, achieving mass spectra with low background for improved sensitivity, suggesting the potential of this simple ionization technology to drive MS into areas currently underserved, such as clinical and medical applications.

     
    more » « less
  2. Rationale

    Successful coupling of a multi‐ionization automated platform with commercially available mass spectrometers provides improved coverage of compounds in complex mixtures through implementation of new and traditional ionization methods. The versatility of the automated platform is demonstrated through coupling with mass spectrometers from two different vendors. Standards and complex biological samples were acquired using electrospray ionization (ESI), solvent‐assisted ionization (SAI) and matrix‐assisted ionization (MAI).

    Methods

    The MS™ prototype automated platform samples from 96‐ or 384‐well plates as well as surfaces. The platform interfaces with Thermo Fisher Scientific mass spectrometers by replacement of the IonMax source, and on Waters mass spectrometers with additional minor source inlet modifications. The sample is transferred to the ionization region using a fused‐silica or metal capillary which is cleaned between acquisitions using solvents. For ESI and SAI, typically 1 μL of sample solution is drawn into the capillary tube and for ESI slowly dispensed near the inlet of the mass spectrometer with voltage placed on the delivering syringe barrel to which the tubing is attached, while for SAI the sample delivery tubing inserts into the inlet without the need for high voltage. For MAI, typically, 0.2 μL of matrix solution is drawn into the syringe before drawing 0.1 μL of the sample solution and dispensing to dry before insertion into the inlet.

    Results

    A comparison study of a mixture of angiotensin I, verapamil, crystal violet, and atrazine representative of peptides, drugs, dyes, and herbicides using SAI, MAI, and ESI shows large differences in ionization efficiency of the various components. Solutions of a mixture of erythromycin and azithromycin in wells of a 384‐microtiter well plate were mass analyzed at the rate ofca1 min per sample using MAI and ESI. In addition, we report the analysis of bacterial extracts using automated MAI and ESI methods. Finally, the ability to perform surface analysis with the automated platform is also demonstrated by directly analyzing dyes separated on a thin‐layer chromatography (TLC) plate and compounds extracted from the surface of a beef liver tissue section.

    Conclusions

    The prototype multi‐ionization automated platform offers solid matrix introduction used with MAI, as well as solution introduction using either ESI or SAI. The combination of ionization methods extends the types of compounds which are efficiently ionized and is especially valuable with complex mixtures as demonstrated for bacterial extracts. While coupling of the automated multi‐ionization platform to Thermo and Waters mass spectrometers is demonstrated, it should be possible to interface it with most commercial mass spectrometers.

     
    more » « less
  3. Rationale

    In droplet‐assisted ionization (DAI), intact molecular ions are generated from molecules in aerosol droplets by passing the droplets through a temperature‐controlled capillary inlet. Ion formation is explored through the effects of analyte mass flow, droplet solvent composition, and capillary temperature on ion signal intensity.

    Methods

    A Waters SYNAPT G2‐S is adapted for DAI by reconfiguring the inlet with a temperature‐controlled capillary. Droplets are generated by atomization of a solution containing analyte and then sampled through the inlet. If desired, solvent can be removed from the droplets prior to analysis by sending the aerosol through a series of diffusion dryers. Size distributions of the dried aerosols allow the mass flow of analyte into the inlet to be determined.

    Results

    Analyte signal intensities are orders of magnitude higher from droplets containing a protic solvent (water) than an aprotic solvent (acetonitrile). The highest signal intensities for DAI are obtained with inlet temperatures above 500°C, though the optimum temperature is analyte dependent. At elevated temperatures, droplets are thought to undergo rapid solvent evaporation and bursting to produce ions. The lowest signal intensities are generally obtained in the 100–350°C range, where slow solvent evaporation is thought to inhibit ion formation. As the temperature decreases from 100°C down to 25°C, the signal intensity increases significantly. When 3‐nitrobenzonitrile, a common matrix for solid‐state matrix‐assisted ionization (MAI), is added to droplets consisting of 50/50 v/v water and acetonitrile, the matrix enhances ion formation to produce a signal intensity comparable to DAI in 100% water.

    Conclusions

    The results are consistent with other inlet ionization techniques, suggesting that similar ion formation mechanisms are operative. Optimized ion yields (the combined effects of ionization probability and ion transmission) for DAI are currently in the 10−5to 10−6range, which is sufficient for many aerosol applications.

     
    more » « less
  4. Rationale

    New ionization processes have been developed for biological mass spectrometry (MS) in which the matrix lifts the nonvolatile analyte into the gas phase as ions without any additional energy input. We rationalized that additional fundamental knowledge is needed to assess analytical utility for the field of synthetic polymers and additives.

    Methods

    Different mass spectrometers (Thermo Orbitrap (Q‐)Exactive (Focus); Waters SYNAPT G2(S)) were employed. The formation of multiply charged polymer ions upon exposure of the matrix/analyte(/salt) sample to sub‐atmospheric pressure directly from the solid state and surfaces facilitates the use of advanced mass spectrometers for detection of polymeric materials including consumer products (e.g., gum).

    Results

    Astonishingly, using nothing more than a small molecule matrix compound (e.g., 2‐methyl‐2‐nitropropane‐1,3‐diol or 3‐nitrobenzonitrile) and a salt (e.g., mono‐ or divalent cation(s)), such samples upon exposure to sub‐atmospheric pressure transfer nonvolatile polymersandnonvolatile salts into the gas phase as multiply charged ions. These successes contradict the conventional understanding of ionization in MS, because can nonvolatile polymers be lifted in the gas phase as ions not only by as little as a volatile matrix but also by the salt required for ionizing the analyte through noncovalent metal cation adduction(s). Prototypevacuummatrix‐assisted ionization (vMAI) and automated sources using a contactless approach are demonstrated for direct analyses of synthetic polymers and plasticizers, minimizing the risk of contamination using direct sample introduction into the mass spectrometer vacuum.

    Conclusions

    Direct ionization methods from surfaces without the need of high voltage, a laser, or even applied heat are demonstrated for characterization of detailed materials using (ultra)high‐resolution and accurate mass measurements enabled by the multiply charged ions extending the mass range of high‐performance mass spectrometers and use of a split probe sample introduction device. Our vision is that, with further development of fundamentals and dedicated sources, both spatial‐ and temporal‐resolution measurements are within reach if sensitivity is addressed for decreasing sample‐size measurements.

     
    more » « less
  5. null (Ed.)
    Ion mobility spectrometry (IMS) mass spectrometry (MS) centers on the ability to separate gaseous structures by size, charge, shape, and followed by mass-to-charge (m/z). For oligomeric structures, improved separation is hypothesized to be related to the ability to extend structures through repulsive forces between cations electrostatically bonded to the oligomers. Here we show the ability to separate differently branched multiply charged ions of star-branched poly(ethylene glycol) oligomers (up to 2000 Da) regardless of whether formed by electrospray ionization (ESI) charged solution droplets or from charged solid particles produced directly from a surface by matrix-assisted ionization. Detailed structural characterization of isomers of the star-branched compositions was first established using a home-built high-resolution ESI IMS-MS instrument. The doubly charged ions have well-resolved drift times, achieving separation of isomers and also allowing differentiation of star-branched versus linear oligomers. An IMS-MS “snapshot” approach allows visualization of architectural dispersity and (im)purity of samples in a straightforward manner. Analyses capabilities are shown for different cations and ionization methods using commercially available traveling wave IMS-MS instruments. Analyses directly from surfaces using the new ionization processes are, because of the multiply charging, not only associated with the benefits of improved gas-phase separations, relative to that of ions produced by matrix-assisted laser desorption/ionization, but also provide the potential for spatially resolved measurements relative to ESI and other ionization methods. 
    more » « less