			<titleStmt><title level='a'>Bio-Inspired Untethered Robot-Sensor Platform for Minimally Invasive Biomedical Sensing</title></titleStmt>
			<publicationStmt>
				<publisher>American Chemical Society</publisher>
				<date>12/20/2023</date>
			</publicationStmt>
			<sourceDesc>
				<bibl> 
					<idno type="par_id">10491497</idno>
					<idno type="doi">10.1021/acsami.3c13425</idno>
					<title level='j'>ACS Applied Materials &amp; Interfaces</title>
<idno>1944-8244</idno>
<biblScope unit="volume">15</biblScope>
<biblScope unit="issue">50</biblScope>					

					<author>Yizong Li</author><author>Amro Halwah</author><author>Shah R. Bhuiyan</author><author>Shanshan Yao</author>
				</bibl>
			</sourceDesc>
		</fileDesc>
		<profileDesc>
			<abstract><ab><![CDATA[Conventional catheter- or probe-based in vivo biomedical sensing is uncomfortable, inconvenient, and sometimes infeasible for longterm monitoring. Existing implantable sensors often require an invasive procedure for sensor placement. Untethered soft robots with the capability to deliver the sensor to the desired monitoring point hold great promise for minimally invasive biomedical sensing. Inspired by the locomotion modes of snakes, we present here a soft kirigami robot for sensor deployment and real-time wireless sensing. The locomotion mechanism of the soft robot is achieved by kirigami patterns that offer asymmetric tribological properties that mimic the skin of the snake. The robot exhibits good deployability, excellent load capacity (up to 150 times its own weight), high-speed locomotion (0.25 body length per step), and wide environmental adaptability with multimodal movements (obstacle crossing, locomotion in wet and dry conditions, climbing, and inverted crawling). When integrated with passive sensors, the versatile soft robot can locomote inside the human body, deliver the passive sensor to the desired location, and hold the sensor in place for real-timemonitoring in a minimally invasive manner. The proof-of-concept prototype demonstrates that the platform can perform real-time impedance monitoring for the diagnosis of gastroesophageal reflux disease.]]></ab></abstract>
		</profileDesc>
	</teiHeader>
	<text><body xmlns="http://www.tei-c.org/ns/1.0" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:xlink="http://www.w3.org/1999/xlink">
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="1.">INTRODUCTION</head><p>Close monitoring of health parameters facilitates early diagnosis and provides effective therapeutic guidance. <ref type="bibr">[1]</ref><ref type="bibr">[2]</ref><ref type="bibr">[3]</ref> The conventional catheter or probe-based monitoring is uncomfortable and inconvenient, making it infeasible for frequent monitoring. <ref type="bibr">4,</ref><ref type="bibr">5</ref> When used for continuous monitoring, the procedures are not risk-free and can result in inaccurate readings if patients' daily activities are affected by the presence of the invasive probe. <ref type="bibr">6</ref> Recently developed implantable sensors, used as an alternative, often require invasive procedures to place the sensors in desired locations. <ref type="bibr">3,</ref><ref type="bibr">[5]</ref><ref type="bibr">[6]</ref><ref type="bibr">[7]</ref><ref type="bibr">[8]</ref><ref type="bibr">[9]</ref><ref type="bibr">[10]</ref> Therefore, a minimally invasive method of biomedical monitoring is in urgent demand.</p><p>Bio-inspired untethered soft robots that possess the capability to deliver sensors precisely to target areas offer significant promise for advancing minimally invasive biomedical monitoring.. <ref type="bibr">[11]</ref><ref type="bibr">[12]</ref><ref type="bibr">[13]</ref> A wealth of inspiration has been drawn from various living organisms in the design of soft robots. <ref type="bibr">12,</ref><ref type="bibr">[14]</ref><ref type="bibr">[15]</ref><ref type="bibr">[16]</ref><ref type="bibr">[17]</ref><ref type="bibr">[18]</ref><ref type="bibr">[19]</ref><ref type="bibr">[20]</ref><ref type="bibr">[21]</ref><ref type="bibr">[22]</ref><ref type="bibr">[23]</ref><ref type="bibr">[24]</ref><ref type="bibr">[25]</ref><ref type="bibr">[26]</ref> Among the commonly observed biological structures, the design of soft robots has frequently incorporated leg structures. <ref type="bibr">27,</ref><ref type="bibr">28</ref> The multilegged soft robot demonstrated potential applications in drug delivery, temperature sensing, and tapping sensing. <ref type="bibr">27,</ref><ref type="bibr">28</ref> However, the multilegged soft robot has yet to explore the capability of vertical and inverted crawling, desired features for navigating the intricate structures inside the human body. <ref type="bibr">27</ref> Additionally, the wireless sensing circuit is complex, necessitating the incorporation of a rigid printed circuit board. <ref type="bibr">28</ref> The remarkable climbing abilities of geckos have garnered significant attention, leading to the development of several soft robots inspired by the gecko's footpad. <ref type="bibr">29,</ref><ref type="bibr">30</ref> Nevertheless, these robots have a limited climbing distance, as the adhesion between the robot footpad and tissue surface is weak and diminishes quickly, particularly when considering the viscous surface inside the human body. <ref type="bibr">29</ref> Snakes possess a remarkable and adaptable capability of limbless locomotion enabling them to navigate and thrive in various terrains, including deserts, forests, and mountains. <ref type="bibr">[31]</ref><ref type="bibr">[32]</ref><ref type="bibr">[33]</ref><ref type="bibr">[34]</ref><ref type="bibr">[35]</ref> The limbless locomotion strategy allows us to emulate snakelike movements through a single actuator. <ref type="bibr">36,</ref><ref type="bibr">37</ref> This singular actuation approach not only simplifies the control mechanisms but also facilitates size reduction, which is important for minimally invasive biomedical applications. Several soft robots have already been inspired by the locomotion mechanism of snakes and developed using the pneumatic actuation method. <ref type="bibr">17,</ref><ref type="bibr">36,</ref><ref type="bibr">37</ref> Nevertheless, these soft robots have faced limitations in terms of environmental adaptability, and the presence of tethered connections, which makes them less suitable for noninvasive or minimally invasive biomedical sensing. <ref type="bibr">18,</ref><ref type="bibr">38,</ref><ref type="bibr">39</ref> Moreover, high load capacity and agile maneuverability are crucial for efficient sensor delivery. <ref type="bibr">11,</ref><ref type="bibr">12,</ref><ref type="bibr">27</ref> It is still challenging for existing bio-inspired untethered soft robots to achieve all desired characteristics for biomedical sensing, including high-speed locomotion, wide environmental adaptability, and high load capacity (Table <ref type="table">S1</ref>). All of these are essential for biomedical sensing in the intricate and dynamically changing environments within the human body. <ref type="bibr">11,</ref><ref type="bibr">12,</ref><ref type="bibr">39</ref> To overcome these challenges, we report a bio-inspired magnetically actuated soft kirigami crawling robot with agile maneuverability and flexibility to operate in constricted spaces, as shown in Figure <ref type="figure">1</ref>. The soft robots are designed to mimic the functionality of snake skin. Under the actuation of a permanent magnet, the snake-inspired kirigami pattern undergoes a transformation from a two-dimensional (2D) to threedimensional (3D) configuration. The pop-up 3D pattern is directional and can provide anisotropic friction that enables forward motion. By optimizing the kirigami patterns and employing both magnetic force and torque, the robot can achieve good deployability, high-speed locomotion (0.25 body length (BL) per step), excellent load capacity (up to 150 times its own weight), and versatile locomotion modes (e.g., obstacle crossing, locomotion in wet and dry conditions, climbing, and inverted crawling). Toward real-world biomedical applications, an integrated soft robot-sensor system has been developed, which consists of a soft robot for locomotion and a passive impedance sensor for real-time wireless sensing. The integrated robot-sensor system was demonstrated for the wireless monitoring of gastroesophageal reflux disease (GERD) in a minimally invasive manner.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.">RESULTS AND DISCUSSION</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.1.">Design and Fabrication of the Soft Kirigami</head><p>Robot. The rectilinear locomotion mode stands out as an efficient means of movement for snakes when navigating in confined spaces, which aligns well with the requirement of minimally invasive monitoring. <ref type="bibr">40</ref> Herein, we adopted this specific locomotion mode as inspiration for the design of the soft robot. Specifically, during the rectilinear locomotion of snakes, part of the snake's body acts as an anchor point, facilitated by anisotropic scales. <ref type="bibr">37,</ref><ref type="bibr">[40]</ref><ref type="bibr">[41]</ref><ref type="bibr">[42]</ref><ref type="bibr">[43]</ref> This adaptation enables snakes to maintain a grip on the substrate and propel its body  forward without sliding backward. <ref type="bibr">[40]</ref><ref type="bibr">[41]</ref><ref type="bibr">[42]</ref><ref type="bibr">[43]</ref> Then, the rest of the snake's body moves forward due to its high stretchability and flexibility. <ref type="bibr">[40]</ref><ref type="bibr">[41]</ref><ref type="bibr">[42]</ref><ref type="bibr">[43]</ref> During movement, the snake skin is tuned to exert high friction for propelling, or low friction for sliding along the substrate. <ref type="bibr">44</ref> To achieve the asymmetric friction that propels the body to move forward, a kirigami skin was designed to mimic the snake skin. The kirigami skin is fabricated by mixing magnetic microparticles and polymer precursor, spin-coating the mixture to form a thin film, and laser cutting the cured thin film into designed kirigami patterns (Figures <ref type="figure">2</ref> and<ref type="figure">3</ref>). When stretching the kirigami skin, the kirigami pattern undergoes a transformation from 2D to 3D  configurations. The pattern pops up, which produces asymmetric friction (i.e., the forward friction is smaller than the backward friction, as shown in Figure <ref type="figure">S1</ref>). Then, the key is how to mechanically stretch the kirigami skin with noncontact magnetic fields. To impart such mechanical stimuli required for crawling, an anisotropic magnetization profile was introduced during the magnetization process. After laser patterning, the resulting kirigami skin was pre-stretched and magnetized in a Halbach array with a magnetic field of 1.2 T (Figure <ref type="figure">2</ref>). Then, the pre-stretch was released, resulting in an anisotropic magnetization profile, as shown in Figure <ref type="figure">3B</ref>. Once the magnetic field is applied again, the magnetic dipole tends to align with the magnetic field. The pop-up angle then introduces anisotropic friction. By exploiting this asymmetric friction induced by stretching and releasing the kirigami skin, the soft robot is inclined to move in the direction in which friction is larger.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.2.">Analysis of Robot Locomotion.</head><p>The kirigami robot could be actuated with a permanent magnet. Magnetic torque and magnetic force are applied to the kirigami robot due to the magnetic field generated by a permanent magnet. The torque aligns the soft robot with the applied magnetic field, forces the kirigami pattern to buckle out in the out-of-plane direction, and therefore raises the body of the soft robot. The magnetic force pulls the soft robot to move to the local maximum magnetic field. More details in the analysis of the magnetic torque and force are provided in the Supporting Information.</p><p>By mimicking the snake skin and the locomotion of the snake (Figure <ref type="figure">3C-E</ref>), the movement of the kirigami robot can be divided into two phases: stretching and recovery processes. Before the stretching process (without the magnetic field), the kirigami robot can be recognized as a thin film that can slide in both forward and backward directions. During the stretching process induced by the magnetic field, the kirigami cut buckles out of the plane, changing the friction property of the soft kirigami robot. The pop-up kirigami pattern results in a significant increase in the coefficient of friction (COF) in the backward direction. As illustrated in Figure <ref type="figure">4A-C</ref>, as the strain increases, the pop-up angle increases, accompanied by more obvious 3D deformations. Thinner kirigami skin leads to larger pop-up angles. As a result of the increased pop-up angle, the COF increases in both backward and forward directions, and the difference between backward COF and forward COF becomes larger (Figure <ref type="figure">4D-F</ref>). Among different kirigami patterns (i.e., trapezoidal, circular, and triangular patterns), trapezoidal patterns exhibit the largest pop-up angle and COF.</p><p>In the strain recovery process induced by turning off or reversing the magnetic field, the kirigami cut buckles back inward until the kirigami robot becomes a thin film. The asymmetric COF in the forward and backward directions also gradually recovered. The change in the friction properties affects the locomotion of the soft kirigami robot by shifting the position of the anchor point (Figures <ref type="figure">3C</ref> and<ref type="figure">S3</ref>). The location of the anchor point in the stretching process and recovery process can be calculated as <ref type="bibr">37</ref> A L</p><p>where A is the distance between the center of the soft kirigami robot and the anchor point, L represents the length of the soft robot, and &#956; f and &#956; b indicate the COF in the forward and backward directions, respectively. At the beginning of the stretching, the anchor point is located at the center of the soft kirigami robot. Therefore, the soft kirigami robot extends equally at the head and tail, as shown in Figure <ref type="figure">S4A</ref>. As the body of the soft kirigami robot extends, the kirigami pattern starts to buckle outward and the COF changes. The COF increases in both the forward and backward directions. The COF in the backward direction increases more rapidly than that in the forward direction. The anchor point starts to move back toward the tail of the kirigami robot. This backward shift of the anchor point leads to an asymmetric extension of the soft kirigami robot. The soft kirigami robot extends more at the head than it does at the tail (Figure <ref type="figure">S4A</ref>). When the robot begins to recover, which is controlled by turning off or reversing the magnetic field, the anchor point jumps to a position between the center and the head of the soft-body robot. The position of the anchor point leads to a greater contraction of the soft robot at its tail than at its head (Figure <ref type="figure">S4A</ref>). The total displacement (&#916;d) of the soft kirigami robot in one step can be calculated using the following expression 37</p><p>where L 0 is the initial length of the soft robot, and &#949; and &#949; max represent the strain of the soft robot and the strain corresponding to the maximum extension of the soft robot, respectively.</p><p>The kirigami robot with an asymmetric COF can be actuated by a permanent magnet. Two actuation modes are investigated, as presented in Figure <ref type="figure">3D</ref>,E. In the first on/off actuation mode (Figure <ref type="figure">3D</ref>), a bar magnet is put underneath the soft robot to make the soft robot stretch, and then the magnet is moved away from the kirigami robot to recover the tensile strain. Nevertheless, in the on/off mode, the applied magnetic field from the permanent magnet induces only a small strain (9%) on the kirigami robot. As a result, the difference in the COF between the forward and backward directions is small, and the anchor point is very close to the center of the kirigami robot. It is also implied that the moving speed can be optimized by adjusting the direction and magnitude of the applied magnetic field. From Figure <ref type="figure">S4A</ref> and eq 3, the total displacement in one step can be simplified as</p><p>or</p><p>Equations 4 and 5 imply that the moving speed can be optimized by minimizing &#916;d 1 -, &#916;d 2 -, &#916;d 3 +, and &#916;d 4 + and maximizing &#916;d 3 -, &#916;d 4 -, &#916;d 1 + and &#916;d 2 +.</p><p>To optimize the moving speed, another actuation mode, the reciprocating actuation mode, is introduced. As shown in Figures <ref type="figure">3E</ref> and<ref type="figure">S4B</ref>, the magnet is put below the kirigami robot and moved toward the direction of locomotion. As the magnet moves forward, the magnetic field applied to the kirigami robot gradually changes from parallel to the substrate to a certain angle to the substrate. Magnetic torque bends the head of the kirigami robot upward, and meanwhile, the magnetic force pulls the kirigami robot for stretching (Figure <ref type="figure">S5</ref>). The kirigami cut buckles outward to align with the magnetic flux and balance the magnetic force. The magnet bar is moved forward until the tail is on the verge of sliding when the anchor point is at the tail of the kirigami robot. As a result, &#916;d 1and &#916;d 2are equal to 0. Then, the magnet is moved backward, resulting in the recovery of the kirigami robot, as shown in Figure <ref type="figure">3E</ref>. The anchor point is located where the robot bends (Figure <ref type="figure">S4B</ref>). The closer the anchor point is to the head of the robot, the faster the moving speed of the robot can be achieved, as shown in eqs 4 and 5. The stretched length (i.e., &#916;d 1 + and &#916;d 2 +) and the anchor point during the recovery process are associated with the thickness of the robot and the applied magnetic flux density. Therefore, to improve the moving speed, the effects of thickness and applied magnetic flux density are investigated. As shown in Figure <ref type="figure">5A-C</ref>, the moving speed is increased at a reduced thickness of the kirigami robot except for a very large magnetic flux density. If the applied magnetic field is too large, kirigami robots would be attracted to the magnet by the strong magnetic force, resulting in a lower moving speed. Conversely, insufficient magnetic fields produce limited stretching of the kirigami robot, consequently impeding its locomotion. A suitable applied magnetic flux density is essential to achieve a high moving speed. When &#916;d 1and &#916;d 2are zero, &#916;d 1 + and &#916;d 2 + that the soft kirigami robot can achieve are determined by the forward COF. A larger forward COF at the tail leads to more stretching in the robot head before the tail to slide. Since the trapezoidal kirigami pattern possesses the largest forward COF, the soft robot with the trapezoidal kirigami pattern can achieve the maximum moving speed (0.25 BL/ step). Table <ref type="table">S1</ref> shows a comparison with existing magnetic crawlers in the literature in terms of moving speed, adaptability, and real-world applications. Because the reported magnetic robots were driven at different actuation frequencies, the moving speed is normalized to BL per step to have a fair comparison. The reported soft robot meets all desired requirements for biomedical sensing. The fast-moving speed and high loading capacity are important for steering sensors efficiently to the designated location. The agile maneuverability and adaptability to different environments and terrains are especially conducive for biomedical applications where the crawling environment is complex and diverse.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.3.">Robot Locomotion in Various Environments.</head><p>To demonstrate the snake-like versatility of the developed soft kirigami robot, locomotion experiments were performed in various environments (Movie S1). To imitate the complex topography in biological environments during biomedical sensing, the moving speed of the soft kirigami robot was tested on sandpaper of different roughnesses (Figure <ref type="figure">5D</ref>). The results indicate that the soft kirigami robot is able to move effectively on the sandpaper without a significant decrease of speed. Load capacity is an essential parameter in determining the potential applications of soft robots, such as cargo delivery. <ref type="bibr">20,</ref><ref type="bibr">27,</ref><ref type="bibr">29</ref> As shown in Figure <ref type="figure">5E</ref>,F, the soft kirigami robot is capable of dragging loads on its tail and carrying loads on top. The maximum load-carrying capacity of the robot was 150 times its weight, while the maximum load-dragging capacity was 90 times its weight.</p><p>As shown in Figures <ref type="figure">S4B</ref> and<ref type="figure">S5</ref>, the magnetic force applied to the soft crawling robot results in a normal force exerted by the substrate, which counterbalances the magnetic force. This generates sufficient friction to allow for the continuous movement of the soft kirigami robot over complex terrains. The capability of snakes to traverse gaps and crawl on tree trunks was mimicked by the soft kirigami robot (Figure <ref type="figure">6A-D</ref>). Figure <ref type="figure">6A</ref> shows that the soft kirigami robot can cross a 2 mm wide gap. The soft kirigami robot can move in multiple directions on surfaces with varying slopes, including perpendicular walls, as shown in Figure <ref type="figure">6B,</ref><ref type="figure">C</ref>. Additionally, the soft kirigami robot is able to overhang and move inversely on surfaces (Figure <ref type="figure">6D</ref>). The adaptability of the soft kirigami robot to various environments was further demonstrated by its ability to move in wet environments (Figures <ref type="figure">6E,H</ref> and<ref type="figure">7A</ref>). Despite a reduction in moving speed in wet environments, the soft kirigami robot is capable of locomoting in the water, as well as traversing a 0.05 mm thick viscous surface possessing a viscosity of 3000 cP. It is noteworthy that the viscosity of the experimental surface surpasses the typical viscosity of mucus within the human esophagus (130 cP). <ref type="bibr">45</ref> These capabilities demonstrate the soft kirigami robot's snake-like adaptability to complex environments, including dry environments, wet environments, undulated topography, and surfaces with different slopes. The soft Kirigami robot also exhibits superior steering capability and deployability, with the ability to be steered in all directions and flipped upside down, as shown in Figure <ref type="figure">6F</ref>,G. In summary, with adaptability, maneuverability, and load capacity, the soft kirigami robot shows great potential in real-world applications where locomotion in complex environments and good load capability are necessary. <ref type="bibr">46</ref> In the following section, an integrated robot-sensor platform for biomedical sensing is demonstrated.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.4.">Integrated Soft Robot-Sensor Platform and Application in the Diagnosis of GERD.</head><p>As one of the most common gastrointestinal diseases, GERD affects people of all ages, with a global prevalence of 2.5-33%. <ref type="bibr">[47]</ref><ref type="bibr">[48]</ref><ref type="bibr">[49]</ref> Acid regurgitation, heartburn, chest discomfort, odynophagia, burping, nausea, and vomiting are symptoms of GERD. <ref type="bibr">50</ref> These symptoms significantly lower the quality of life. As the disease worsens, GERD can also cause Barrett's esophagus and even esophageal cancer. <ref type="bibr">48</ref> Monitoring esophageal reflux is an effective way for diagnosing GERD. <ref type="bibr">48</ref> Traditionally, a catheter is delivered to the patient's esophagus through the patient's nose or mouth and then is held in place to measure esophageal acidity for 24 h. <ref type="bibr">47</ref> However, it is reported that 10% of patients refuse to take the monitoring and 4% of patients could not tolerate the discomfort caused by the catheter. <ref type="bibr">6,</ref><ref type="bibr">51</ref> Moreover, the discomfort of the catheter can change the patient's daily eating habits and routine, which can affect the observation results. <ref type="bibr">47</ref> Another way to monitor esophageal reflux is by using a wireless sensing capsule that can achieve wireless esophageal reflux monitoring by integrated sensors. This method offers greater comfort and fewer restrictions on daily activities compared to traditional catheter-based methods. <ref type="bibr">6,</ref><ref type="bibr">[52]</ref><ref type="bibr">[53]</ref><ref type="bibr">[54]</ref><ref type="bibr">[55]</ref> The wireless sensing capsule is typically introduced into the gastrointestinal tract either via catheter-based delivery or swallowing accompanied by natural peristaltic waves. <ref type="bibr">6,</ref><ref type="bibr">54,</ref><ref type="bibr">56,</ref><ref type="bibr">57</ref> Nevertheless, catheter-based delivery can elicit unfavorable side effects, including throat pain, cough, and esophageal discomfort. Capsule delivery relying on natural movements can lead to incomplete or inaccurate results due to the capsule's limited mobility, which may impede delivery to the designated area. <ref type="bibr">54,</ref><ref type="bibr">57</ref> Moreover, the constrained mobility and rigid body of the capsule increase the chance of being entrapped within the gastrointestinal tract, leading to blockage. Sometimes surgical interventions are necessitated to remove the entrapped capsule. <ref type="bibr">56,</ref><ref type="bibr">57</ref> Hence, a controllable noninvasive way to deliver sensors for GERD diagnosis is needed. Here, a swallowable soft device consisting of a soft kirigami robot and a passive sensor is developed (Figure <ref type="figure">1</ref>). As shown in Figure <ref type="figure">7</ref>, once swallowed, the soft kirigami robot can be deployed and steered to the desired sensing location on the esophageal wall, and the passive sensor can be secured to perform wireless monitoring of esophageal impedance, a crucial diagnostic parameter for GERD. <ref type="bibr">47,</ref><ref type="bibr">48,</ref><ref type="bibr">58</ref> The passive impedance sensor exploits capacitive sensing and passive inductive coupling. As shown in Figures <ref type="figure">1</ref> and<ref type="figure">7</ref>, the sensor consists of two parts: (1) A sensor tag part placed internally on the esophageal wall that contains a screen-printed interdigitated impedance sensor, a planar inductive coil, and a 100 pF surface mount capacitor. (2) An external reader with a planar inductive coil connected to the impedance analyzer. Due to its passive nature, no battery and complex circuitry are needed for the internal sensor tag, and no wired connection is necessary between the internal sensor tag and the external reader. The overall sensor tag size is 15 mm &#215; 10 mm, as shown in Figure <ref type="figure">S6A</ref>. When integrated with the soft kirigami robot that can be folded on top, the overall dimension of the robot-sensor platform is 18 mm long and 10 mm wide, which can be swallowed without difficulties. <ref type="bibr">7</ref> The impedance sensor is designed as an interdigitated electrode (IDE) coated with a thin layer of poly-(dimethylsiloxane) (PDMS)/BaTiO 3 composites. Introducing BaTiO 3 into the PDMS helps increase the dielectric permittivity, as shown in Figure <ref type="figure">S6B</ref>. A 20-turn coil with an inductance of L T is connected in parallel with the IDE sensor. A surface-mounted 100 pF capacitor (C 2 ) is also placed in parallel with the coil and the IDE sensor to reduce the resonant frequency of the sensor tag. To monitor the impedance change due to the esophageal reflux, the impedance analyzer is connected to the reader (i.e., a hand-wound planar coil). The sensor tag is separated from the reader while coupled by passive inductive coupling. The equivalent circuit is shown in Figure <ref type="figure">7B</ref>. The measured real part of the impedance (Re{Z R }) and resonant frequency (f R ) from the reader can be expressed as <ref type="bibr">[59]</ref><ref type="bibr">[60]</ref><ref type="bibr">[61]</ref> ( )</p><p>where R R is the resistance and L R is the inductance of the reader, and k is the coupling coefficient. Z T , f T , and Q T are the impedance, resonant frequency, and quality factor of the sensor tag, respectively. Since Q T &#8811; 1, f R can be simplified as</p><p>where C T is the capacitance of the sensor tag (C T = C 1 + C 2 ). From eq 6, the resonant frequency of the reader measured by the impedance analyzer is determined by the resonant frequency of the tag. When acid or nonacid reflux occurs, the medium surrounding the sensor tag changes from air to liquid. The overall capacitance of the IDE will increase due to the increased dielectric permittivity. As a result, the resonant frequency of the sensor tag will shift to a lower frequency. Due to the inductive coupling, such a resonant frequency shift can be reflected in the frequency shift measured from the external reader (Figure <ref type="figure">S7</ref>). More analysis of the IDE sensor and the coupling of the sensor and tag are provided in the Supporting Information.</p><p>To demonstrate the practical applications, the entire device, including the soft robot and the sensor tag, is designed to be biocompatible for in vivo monitoring. The biocompatibility of NdFeB/PDMS composites employed in the fabrication of the soft robot has been previously validated in cell viability tests for a duration of 24 h. <ref type="bibr">62</ref> In addition, a thin film of PDMS was coated to encapsulate the sensor tag, thereby ensuring the sensor tag's biocompatibility. We then conducted an ex vivo evaluation of the robot-sensor platform for monitoring GERD (Movie S2). A pig esophagus immersed in water to a depth of 30 mm was employed to emulate real-world conditions. It is noteworthy that the water depth used in this experiment exceeds the average lumen diameter of the human esophagus (20 mm). <ref type="bibr">63</ref> A hand-held magnet that can generate sufficient magnetic flux density (around 170 mT) at a distance of 30 mm was used to deploy, actuate, and hold the robot-sensor platform during the test. As shown in Figure <ref type="figure">7A</ref>, the integrated robot-sensor with the kirigami robot folded on top of the sensor can be swallowed. The kirigami robot can then be unfolded and steered by an external magnet toward the monitoring point (around 50 mm above the lower esophageal sphincter). <ref type="bibr">6</ref> After arriving at the monitoring point, the kirigami robot is flipped upside down to hold the sensor in place by an external permanent magnet for 24 h GERD monitoring. Note, in real applications, the robot-sensor device can be monitored and guided using ultrasound. <ref type="bibr">20,</ref><ref type="bibr">64</ref> The retention test demonstrates that the platform can be reliably attached to the esophagus by a permanent magnet with a slight sliding of only 2.5 mm (Figure <ref type="figure">S8</ref>). Notably, the robot-sensor platform did not exhibit any signs of detachment, damage, or degradation throughout the ex vivo experiment (Figure <ref type="figure">7A</ref>), retention test (Figure <ref type="figure">S8</ref>), and the subsequent 24 h stability test in an acidic environment of pH = 1 (Figure <ref type="figure">S9</ref>), enabling an uninterrupted 24 h GERD monitoring. The presence of permanent magnets may raise concerns about their impact on daily activities. This can be addressed by encapsulating the magnet's outer surfaces (except the surface in contact with the skin) with magnetic shield layers. After the shield layers were applied, the magnetic field from the side of the magnet covered with shield layers is minimized (Figure <ref type="figure">S10</ref>). Meanwhile, the magnetic field from the unshielded side of the magnet remains nearly unaltered, ensuring the retention of the robot-sensor platform.</p><p>Different types of liquids were used to mimic the gastroesophageal reflex, including water, milk, tea, and pH buffer solutions. As expected, when there is a reflex, the measured resonant frequency shifts to the lower frequency (Figure <ref type="figure">7C</ref>) and can be detected at a maximum distance of 40 mm, which is sufficient for GERD monitoring. <ref type="bibr">52,</ref><ref type="bibr">53</ref> The frequency shift is related to the amount of liquid on the IDE sensor, and a small amount of liquid can be detected by the sensor (Figure <ref type="figure">7D</ref>). Overall, the developed robot-sensor system is small, minimally invasive, and does not need a catheter or other types of wired connection for real-time GERD monitoring.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.">CONCLUSIONS</head><p>In this work, we have demonstrated a snake-inspired kirigamibased soft magnetic robot with agile maneuverability, good load capacity, and wide adaptability to different environments. The kirigami pattern is designed by mimicking snake skin to generate high friction to propel itself. The anisotropic magnetization profile required for magnetic actuation is introduced by pre-stretching during the magnetization process. It is found that the out-of-plane deformation of the kirigami pattern leads to the difference in the COF in the forward and backward directions, which plays a vital role in the moving speed of the kirigami robot. The applied magnetic field and thickness of the kirigami robot also affect the moving speed. The actuation method was optimized by employing both magnetic force and magnetic torque to improve the moving speed. The resulting kirigami robots are capable of steering in all directions and deployable, moving in water, crossing gaps, moving vertically, and moving on an inverted wall. Finally, a new robot-sensor platform was developed that consisted of a magnetic kirigami robot and a wireless passive sensor. The integrated robot-sensor platform paves the way for real-time wireless biomedical sensing in a minimally invasive manner. The main contributions of this work include: (1) The development of snake-inspired kirigami crawlers with a fastmoving speed, agile maneuverability, and multimodal movements for wide environmental adaptability, and (2) the development of a new untethered soft robot-sensor platform for minimally invasive biomedical sensing.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.">MATERIALS AND METHODS</head><p>4.1. Fabrication of the Soft Kirigami Robot. PDMS (Sylgard 184, Dow Corning) with a base-to-curing agent ratio of 10:1 (by weight) and NdFeB microparticles (MQFP 15-7, Neo Magnequench) were mixed by a planetary mixer (AR-100, Thinky) at a weight ratio of 1:2. The mixture was spin-coated onto a substrate to obtain a thin film followed by curing the thin film at 60 &#176;C for 12 h. The cured film was cut into designed kirigami patterns by using a laser cutter. A lab-made tensile strain stage was used to pre-stretch the patterned kirigami film. Then, the pre-stretched kirigami film was magnetized by a Halbach array with a central magnetic flux density of 1.2 T for 30 min. After magnetization, the kirigami film was removed from the strain stage, and the tensile strain was released.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.2.">Fabrication of the Sensor Tag and Integrated</head><p>Robot-Sensor Platform. The sensor tag consists of two parts: A screenprinted IDE and an inductive coil. To fabricate the IDE, polyimide (PI) tape (thickness: 0.06 mm) and toner transfer paper (Pulsar Professional fx) were used as the shadow mask and substrate, respectively. The PI tape was first laminated onto the toner transfer paper and then cut into a desired pattern using a mechanical plotter (CAMEO 4, Silhouette). The PI tape can be cut into the desired pattern without damaging the substrate (i.e., toner transfer paper). Conductive silver ink (Creative Materials Inc.) was screen-printed onto the toner transfer paper. After curing the silver ink at 60 &#176;C for 2 h, the mask (PI tape) was removed. A mixture of PDMS (with the based to the curing agent ratio of 10:1 by weight) and BaTiO 3 (Nanostructured &amp; Amorphous Materials, average particle size: 100-200 nm) at a weight ratio of 1:1 was spin-coated onto a substrate and then cured at 60 &#176;C for 12 h. Afterward, the screen-printed IDE was laminated on the PDMS/BaTiO 3 thin film with the silver ink side facing downward. The printed IDE was then transferred from the toner transfer paper to the PDMS/BaTiO 3 thin film by immersion in deionized (DI) water. Note that the top coating of the toner transfer paper is water-soluble and serves as a sacrificial layer for transfer. A 100 pF capacitor and a 20-turn hand-wound coil made from copper wires (36 AWG, Remington Industries) were connected in parallel with the IDE, respectively. Finally, PDMS (base-to-curing agent ratio of 10:1 by weight) was spin-coated over the sensor tag to encapsulate the sensor. To integrate the soft kirigami robot and the IDE sensor, both components were placed within a mold, where a 1 mm gap was maintained between the robot's tail and the sensor's head. A small amount of PDMS precursor with a base-to-curing agent weight ratio of 10:1 was dispensed into the gap. Following this, the assembly underwent a curing process at 60 &#176;C for 12 h to bond the robot and the sensor together. Due to the good elasticity of PDMS and strong bonding of PDMS to the robot and sensor, the bonded connection between the robot and the sensor is sufficiently strong, ensuring the successful execution of monitoring tasks.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.3.">Friction Measurement of the Soft Kirigami</head><p>Robot. The experimental setup for characterizing the COF of the kirigami robot in forward and backward directions is shown in Figure <ref type="figure">S11</ref>. The soft kirigami robot was fixed on a shelf to remain stretched. A materials testing system (858 Mini Bionix II, MTS mechanical tester) was connected to the soft kirigami robot via a thread through a pulley. During the testing process, the thread dragged the kirigami robot to move horizontally. The tensional force of the thread, which equals the friction exerted on the kirigami robot by the substrate, was recorded. Finally, the COF was calculated according to the recorded tensional force.</p><p>4.4. Magnetic Actuation Method. We employed a hand-held cylindrical magnet (DEX2, K&amp;J Magnetics) to control the movements of the soft kirigami robot in various environments (Figure <ref type="figure">6</ref>). For tasks requiring gap crossing, vertical crawling, inverted crawling, and subaquatic locomotion (Figure <ref type="figure">6A-E</ref>), the magnet was positioned at a distance of 5 mm directly beneath the soft kirigami robot. The longitudinal axis of the magnet was aligned with the soft kirigami robot's length, and its south pole was oriented in the direction of the desired locomotion. A reciprocating actuation mode, as delineated in Figure <ref type="figure">3E</ref>, was employed to facilitate the locomotion of the soft kirigami robot. To achieve directional control and steering of the soft kirigami robot, we placed the magnet 10 mm beneath the robot. The magnet's axis remained parallel to the length of the soft kirigami robot. The desired rotation of the soft kirigami robot was induced by rotating the magnet horizontally, prompting it to reorient itself toward the south pole of the magnet, as depicted in Figures <ref type="figure">6F</ref> and<ref type="figure">S12A</ref>. Additionally, for the flipping motion of the soft kirigami robot, the magnet was again positioned 10 mm beneath the robot with its axis maintaining alignment with the robot's length. The flipping action was achieved by rotating the magnet in plane, as shown in Figures <ref type="figure">6G</ref> and<ref type="figure">S12B</ref>.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.5.">Demonstrations with Ex Vivo Experiments.</head><p>For the evaluation of the integrated robot-sensor platform, an ex vivo experiment was conducted using a pig esophagus procured from Nebraska Scientific. To provide the requisite magnetic flux density (170 mT) for the deployment and actuation of the robot-sensor platform, as well as to maintain a distance of 30 mm between the platform and the magnet, a hand-held magnet (BX8 &#215; 8X8-N52, K&amp;J Magnetics) was employed throughout the ex vivo testing procedure. During the ex vivo test, the pig esophagus was thoroughly immersed in water with a depth of 30 mm. Subsequently, the robot-sensor platform was actuated by the magnet, following the methodology outlined in Section 4.4, in order to execute a series of predetermined tasks. Following the successful actuation of the platform, the IDE sensor was immobilized in place and an additional segment of pig esophagus was interposed between the IDE sensor and the reader before performing the gastroesophageal reflux test. To assess the wireless sensor for impedance monitoring, an impedance analyzer (E4990A, Keysight) was connected to the reader. Diverse liquid mediums, encompassing water, milk, tea, and pH buffer solutions, were poured onto the IDE sensor to simulate conditions akin to gastroesophageal reflux. Subsequently, the resulting alterations in resistance were meticulously recorded by the impedance analyzer.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>&#9632; ASSOCIATED CONTENT</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>* s&#305; Supporting Information</head><p>The Supporting Information is available free of charge at <ref type="url">https://pubs.acs.org/doi/10.1021/acsami.3c13425</ref>.</p><p>Analysis of magnetic torque and force; analysis of anisotropic frictional forces; size of the kirigami pattern; characterization of the IDE sensor and the coupling of the sensor and tag; anchor point locations for on-off actuation mode; comparison of two actuation modes; free body diagrams of the kirigami robot; experimental results for wireless impedance sensing when the distance between the IDE sensor and the reader is 40 mm; experimental results of the retention test; experimental results of the stability test; experimental results of the impact of magnetic shield layers on magnetic flux density; experimental setup for measuring the coefficient of friction; schematics of the actuation methods for steering and flipping; and table <ref type="table">of</ref>  </p></div><note xmlns="http://www.tei-c.org/ns/1.0" place="foot" xml:id="foot_0"><p>Downloaded via STONY BROOK UNIV SUNY on January 26, 2024 at 15:25:17 (UTC).See https://pubs.acs.org/sharingguidelines for options on how to legitimately share published articles.</p></note>
			<note xmlns="http://www.tei-c.org/ns/1.0" place="foot" xml:id="foot_1"><p>https://doi.org/10.1021/acsami.3c13425 ACS Appl. Mater. Interfaces 2023, 15, 58839-58849</p></note>
			<note xmlns="http://www.tei-c.org/ns/1.0" place="foot" xml:id="foot_2"><p>https://doi.org/10.1021/acsami.3c13425</p></note>