			<titleStmt><title level='a'>Apparatus for the room temperature measurement of low field Nernst and magneto-Seebeck coefficients</title></titleStmt>
			<publicationStmt>
				<publisher>AIP Publishing</publisher>
				<date>08/01/2024</date>
			</publicationStmt>
			<sourceDesc>
				<bibl> 
					<idno type="par_id">10534757</idno>
					<idno type="doi">10.1063/5.0222406</idno>
					<title level='j'>Review of Scientific Instruments</title>
<idno>0034-6748</idno>
<biblScope unit="volume">95</biblScope>
<biblScope unit="issue">8</biblScope>					

					<author>C E Porter</author><author>C M Crawford</author><author>E S Toberer</author>
				</bibl>
			</sourceDesc>
		</fileDesc>
		<profileDesc>
			<abstract><ab><![CDATA[Nernst coefficient measurements are a classic approach to investigate charge carrier scattering in both metals and semiconductors. However, such measurements are not commonly performed, despite the potential to inform material design strategies in applications such as thermoelectricity. As dedicated instruments are extremely scarce, we present here a room temperature apparatus to measure the low field Nernst coefficient (and magneto-Seebeck coefficient) in bulk polycrystalline samples. This apparatus is specifically designed to promote accurate and facile use, with the expectation that such an instrument will make Nernst measurements de rigueur. In this apparatus, sample loading and electrical contacts are all pressure-based and alignment is automatic. Extremely stable thermal control (10 mK of fluctuation when ΔT = 1K) is achieved from actively cooled thermoelectric modules that operate as heaters or Peltier coolers. Magneto-Seebeck measurements are integrated into the system to correct for residual probe offsets. Data from the apparatus are provided on bulk polycrystalline samples of bismuth, InSb, and SnTe, including raw data to illustrate the process of calculating the Nernst coefficient. Finally, we review how Nernst measurements, in concert with Seebeck, Hall, and electrical resistivity, can be analyzed via the Boltzmann equation in the relaxation time approximation to self-consistently predict the Fermi level, effective mass, and energy-dependent relaxation time.]]></ab></abstract>
		</profileDesc>
	</teiHeader>
	<text><body xmlns="http://www.tei-c.org/ns/1.0" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:xlink="http://www.w3.org/1999/xlink">
<div xmlns="http://www.tei-c.org/ns/1.0"><head>I. INTRODUCTION</head><p>The Nernst effect is the induced transverse electric field Ey in response to a longitudinal temperature gradient &#8711;xT and an applied magnetic field Bz [Fig. <ref type="figure">1(a)</ref>]. <ref type="bibr">[1]</ref><ref type="bibr">[2]</ref><ref type="bibr">[3]</ref> Analogous to the Hall effect, where the electric field is E Hall = RH(J &#215; B), the Nernst electric field is defined as</p><p>Here, N is the Nernst coefficient, which is typically on the order of nV/KT to V/KT for most conductive materials. <ref type="bibr">[4]</ref><ref type="bibr">[5]</ref><ref type="bibr">[6]</ref> Nernst measurements have an impressive history of providing insight into the energy dependence of charge carrier transport in materials ranging from semimetals to classic thermoelectrics to superconductors. In particular, Nernst measurements have resolved the energy dependence of charge carrier scattering and, thus, the dominant scattering mechanisms. <ref type="bibr">3,</ref><ref type="bibr">[7]</ref><ref type="bibr">[8]</ref><ref type="bibr">[9]</ref> The Nernst effect resembles the Hall effect in that both measurements utilize the Lorentz force to generate a transverse voltage in a magnetic field. These two measurements provide unique insights into material properties due to their distinct origins. This is a result of the current in the Hall effect being electrically driven (df /dE) vs thermally driven in the Nernst effect (df /dT). In the Hall effect, the charge carriers that contribute most to the induced voltage reside at the Fermi level (EF). The charge carriers participating in the Nernst effect are those residing slightly above and slightly below EF. <ref type="bibr">10</ref> As shown in Fig. <ref type="figure">1(b)</ref>, df dT = 0 at EF; hence, carriers with energy equal to EF do not contribute to this particular effect.</p><p>The Nernst effect also has similarities to the Seebeck effect, in that an applied temperature gradient drives a thermal current of holes and electrons. As such, both measurements probe the same population of carriers, namely, those residing slightly above and below the Fermi level [Fig. 1(b)]. <ref type="bibr">10,</ref><ref type="bibr">11</ref> However, the longitudinal See- beck effect is an electrical field that is induced along the same axis as the applied temperature gradient, whereas the Nernst signal <ref type="bibr">12</ref> is perpendicular to the temperature gradient. <ref type="bibr">1,</ref><ref type="bibr">10,</ref><ref type="bibr">13</ref> Furthermore, Seebeck coefficient measurements are generally sample geometryindependent, whereas Nernst measurements are sensitive to the sample geometry. Rearranging Eq. ( <ref type="formula">1</ref>) for the measured transverse Nernst voltage and adopting the coordinate scheme in Fig. <ref type="figure">1</ref>(a),</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Review of</head><note type="other">Scientific Instruments</note><p>We see that the sample geometry (lylx) has a significant impact on the magnitude of the measured voltage. Nernst measurements yield information beyond what is offered by electrical resistivity, Hall effect, and Seebeck coefficient measurements. These traditional measurements provide information about the majority carrier type, carrier concentration, and Fermi level position. The Nernst coefficient is incredibly sensitive to scattering, and the polarity of the coefficient yields information on the dominant scattering mechanism. <ref type="bibr">7,</ref><ref type="bibr">[13]</ref><ref type="bibr">[14]</ref><ref type="bibr">[15]</ref> This is in contrast with the Hall and Seebeck coefficients, whose polarity typically reflects the polarity of the primary charged carrier type. Within the single parabolic band regime, Nernst measurements can be analyzed using the method of four coefficients to determine the dominant carrier scattering mechanism. <ref type="bibr">3,</ref><ref type="bibr">4,</ref><ref type="bibr">6,</ref><ref type="bibr">8,</ref><ref type="bibr">9</ref> When analyzing materials where the single parabolic band approximation does not apply, Nernst measurements can be analyzed with the general Boltzmann transport equation within the relaxation time approximation, as discussed in Sec. IV.</p><p>Nernst measurements to date are frequently conducted at cryogenic temperatures to probe exotic phenomena such as superconductivity. <ref type="bibr">2,</ref><ref type="bibr">5,</ref><ref type="bibr">[16]</ref><ref type="bibr">[17]</ref><ref type="bibr">[18]</ref> For conventional metals and semiconductors, the Nernst signal is also typically measured at temperatures below 300 K. <ref type="bibr">[19]</ref><ref type="bibr">[20]</ref><ref type="bibr">[21]</ref> Practically, Nernst is an excellent phenomenon to measure at low T because the signal often strengthens with decreasing T. Whereas the magnitude of the Seebeck coefficient often decays monotonically toward zero upon cooling from room temperature, the low field [B 1, : mobility (m 2 /V s), B: magnetic field (T)] Nernst coefficient is generally observed to peak between 0 K and room temperature. <ref type="bibr">5,</ref><ref type="bibr">18,</ref><ref type="bibr">22</ref> This is because the electron relaxation time &#964; typically grows with decreasing temperature, leading to a greater buildup of carriers contributing to the Nernst voltage [Fig. <ref type="figure">1(c)]</ref>. For example, the Nernst coefficient of bismuth increases from 0 to 100 K <ref type="bibr">22</ref> and then decreases at higher temperatures. <ref type="bibr">22,</ref><ref type="bibr">23</ref> At 100 K, the N of bismuth is -183 V/KT, but by 300 K, it has plummeted in magnitude to -14 V/KT. <ref type="bibr">23</ref> This is still a sizable signal as far as N goes, and in fact, bismuth possesses one of the largest Nernst signals known to date. <ref type="bibr">16</ref> Cryogenic Nernst measurements are generally performed on very small, thin film or bulk samples in a measurement system such as a Quantum Design PPMS. <ref type="bibr">8,</ref><ref type="bibr">22</ref> The measurement typically requires the wire bonding of eight or more delicate contacts to the sample-four for reading T, two for measuring Vy, and at least two more for supplying heat, typically via a small resistive heater. The Nernst coefficient is derived from either (i) dV/dB at a constant T or (ii) dV/dT at a constant B. Systems performing the former are more common and more accurate. <ref type="bibr">5,</ref><ref type="bibr">8</ref> For example, Wang et al. found that method (i) yielded Nernst coefficient data accurate to 0.1 V/KT, and method (ii) produced less accurate data, even when the temperature was swept at a slow rate of 1 K/min. The authors attributed this loss of accuracy to the long relaxation time of the temperature gradient within the crystal. <ref type="bibr">5</ref> In addition, Nernst measurements performed in cryostats face challenges at room temperature regarding thermometry sensitivity and longer equilibration times since the sample temperature is much closer to the heat bath temperature.</p><p>Room temperature Nernst measurements are not nearly as commonly reported as Seebeck, Hall, or resistivity measurements. Most measurements of the Nernst coefficient above 300 K are conducted using a cryostat. <ref type="bibr">19,</ref><ref type="bibr">24</ref> Furthermore, we are aware of one system designed specifically for high temperature measurements. <ref type="bibr">25</ref> We posit that this scarcity is due to two separate reasons. First,</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Review of Scientific Instruments</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>ARTICLE</head><p>pubs.aip.org/aip/rsi since the Nernst signal scales with electronic mobility, <ref type="bibr">16</ref> the induced Nernst voltage is often too small to be measured in low mobility samples, especially above room temperature when mobility is limited by acoustic phonon scattering. Second, while Hall and Seebeck effect instruments <ref type="bibr">[26]</ref><ref type="bibr">[27]</ref><ref type="bibr">[28]</ref> have experienced rapid advancements in usability and accuracy [i.e., pressure contacts instead of tedious wire bonding, using the differential method (small T) in lieu of the integral method (large T) for measuring the Seebeck coefficient], Nernst has not experienced such a renaissance yet.</p><p>Here, we demonstrate a near-room temperature Nernst coefficient measurement system that is designed for both accuracy and usability. To maximize usability, the system was designed to be simple to load, with self-aligning pressure contacts for measuring Vy. Furthermore, the cylindrical sample geometry is compatible with other standard thermoelectric instruments and requires no specific machining of the sample. All terms in Eq. ( <ref type="formula">2</ref>) were considered in the context of maximizing accuracy. Examples include (i) incorporating Peltier coolers to deliver an extremely stable T, (ii) precise pressure contacts to minimize geometry uncertainty, and (iii) the incorporation of magneto-Seebeck measurements to correct for any residual probe offsets. In the following, we detail these specific design choices, provide example measurements on a variety of materials, and discuss the analysis of Nernst coefficient data.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>II. INSTRUMENTATION</head><p>An apparatus to measure the low field Nernst coefficient N must supply a thermal gradient dTdx and magnetic field Bz to the sample and measure the induced transverse voltage Vy [Eq. ( <ref type="formula">2</ref> Python is used to write settings to and read data from each hardware component, improving the ease of data collection as well as minimizing the chance of human error in calculating the Nernst coefficient. Given that the Nernst signal rarely exceeds a few V, <ref type="bibr">5,</ref><ref type="bibr">29</ref> careful instrumentation design is required to achieve high sensitivity and minimize uncertainty. In the following sections, we describe our design to minimize the uncertainty that each hardware component introduces to the calculation of the Nernst coefficient.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>A. Thermal control</head><p>Our thermal design criteria are (i) ability to apply a stable T of at least 1 K in a temperature regime near or below room temperature without a cryostat, (ii) achieving excellent thermal contact between the sample and sample holder such that heat flow is only along one dimension, and (iii) accurately controlling and measuring the temperature of each block.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="1.">Heating/cooling blocks</head><p>Our thermoelectric setup operates between 0 and 40 &#9675; C with mirrored Peltier coolers. The sample is clamped between two  heater/cooler blocks (Fig. <ref type="figure">2</ref>), which are machined from copper to efficiently conduct heat to or away from the sample. Each copper block (5 &#215; 20 &#215; 32 mm) has a square recess (0.5 mm deep, not visible in Fig. <ref type="figure">2</ref>) machined to receive the cold side of a Peltier module (CUI Devices CP60133, Q max 12.2 W at 27 &#9675; C). For a sample of &#8764;12 mm diameter and &#8764;5 mm tall [an elongated version of the disk shaped samples commonly produced from hot pressing and spark plasma sintering, Fig. <ref type="figure">1</ref>(a), lx = 5.0 mm and ly = 12 mm] with a thermal conductivity of 3.0 W m -1 K -1 , the heat requirement at room temperature to deliver a T of 5 K across the sample is only 0.34 W from Fourier's law. It takes us less than 1 min to establish a T of 5 K at room temperature for a variety of materials with thermal conductivities ranging from 2 to 10 W m -1 K -1 . The hot side of the Peltier is heat sunk to an aluminum water-cooled block (Fig. <ref type="figure">2</ref>). Thermal grease is applied to either side of the Peltier module to improve contact between the interfaces. The bottom aluminum heat sink block is 55 mm wide (along the z-direction, Fig. <ref type="figure">2</ref>) and is directly clamped between the pole caps of the magnet [Fig. <ref type="figure">4(a)</ref>]. The upper heat sink block is 50 mm wide, which allows the lower block to remain clamped while the upper block can be easily removed to load samples.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Review of Scientific Instruments</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.">Excellent thermal contact</head><p>Minimizing the thermal contact resistance between the sample and heater/cooler blocks is paramount for estimating the thermal gradient within the sample. The primary source of this thermal contact resistance is expected to be due to the microscopic gaps between the surface of the pellet and the copper stage of the Nernst instrument. This is addressed by a combination of compression and filling interfacial gaps by placing a thin piece of graphoil between the two surfaces before applying compression. Threaded guide rods pass through the aluminum blocks [Fig. <ref type="figure">2(b)</ref>] and serve two functions: (1) to precisely locate the top block on top of the sample, and (2) to ensure excellent thermal contact with the sample; when the nuts are tightened on the installed rods, the blocks firmly press against the sample [Fig. <ref type="figure">2(b)</ref>]. This design thus encourages uniform thermal contact resistance. Coupled with the copper blocks serving as effective heat spreaders, the expectation is that the boundary conditions on the sample are uniform laterally and that a 1D heat flow approximation can be made. This one-dimensional heat flow approximation is successfully employed in uniaxial Seebeck measurement systems. <ref type="bibr">28</ref> </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.">Accurate control and measurement</head><p>Measurement and control of the thermal gradient are performed via a multi-channel thermoelectric controller (TEC) from Meerstetter Engineering with bipolar DC output to the Peltier coolers. The TEC contains PID controllers, which dictate the current output to the independent Peltier modules. The copper heater/cooler blocks sandwich the Peltier cooler to the aluminum heat sink block; nylon screws are used to clamp the blocks to avoid thermally shorting the copper block to the aluminum heat sink block. Within the copper block is a machined hole that sits 1.25 mm beneath the sample and receives the thermocouple bead. Our (type T) thermocouples are made of spot welded, sheathed copper and constantan wires. Thermocouples should be made from extremely fine wires (e.g., California Fine Wire) as a matter of best practice. We employ an exposed junction thermocouple for fast response time. The temperature we measure at the copper block is the temperature used to drive the desired T across the sample, as described in Sec. III. Measuring the temperature at the copper blocks is an accurate proxy for sample temperature due to the small Seebeck coefficient and colossal thermal conductivity of copper. To demonstrate this, we measure the Seebeck effect using copper wires also embedded in the copper blocks [Fig. <ref type="figure">3(b)</ref>, <ref type="figure">V -</ref> ], and we achieve accurate values (Fig. <ref type="figure">7</ref>). By measuring the sample temperature indirectly, we free up space on the sample for transverse voltage probes as well as avoid unwanted heat transfer between the sample and the thermocouple (i.e., the "hot" or "cold finger" effect).</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>B. Voltage measurement</head><p>Our electrical design achieves accurate voltage measurements without compromising usability. For accurate measurements, the probes must be highly electrically conductive, possess a low Seebeck coefficient, be non-magnetic, form point contacts on the sample, and be located perpendicular to the temperature gradient and magnetic field. Furthermore, we seek usability. Our design goals for enhancing instrument usability include ease of contacting the sample with probes, automatic alignment along the x-axis to avoid spurious thermally induced voltages (discussed in detail in Sec. III C), and probes that are mechanically stable and chemically inert. In brief, we use a 3D printed voltage probe ring with embedded brass set screws to form point contacts on the sample (Fig. <ref type="figure">3</ref>) at a prescribed height and alignment.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="1.">Geometry constraints for electrical contacts</head><p>The geometry constraints that must be satisfied to measure the Nernst voltage warrant further detail. In general, voltage probes should form point contacts with the sample when performing an electrical measurement. Point contact geometry is satisfied by using small (M2) brass flat-tipped set screws that press into the side of the sample (Fig. <ref type="figure">3</ref>). The diameter of the set screw tip is much smaller than the height of the sample. Specific to Nernst, the electrical contacts must be co-planar with the yz plane (i.e., perpendicular to the temperature gradient along the x-direction). To ensure that the set screws are co-planar, we designed and 3D printed a retainer ring that pins the two electrical contacts to the same height (in the x-direction). The 3D printed ring is electrically insulative to avoid shorting the two contacts and can be readily manufactured. The set screws thread into heat-set threaded inserts, chosen to be of the same</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Review of Scientific Instruments</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>ARTICLE</head><p>pubs.aip.org/aip/rsi material as the screws to avoid thermal EMFs at the thread/screw interface. The inserts are easily installed (only once) with a soldering iron and allow for easy insertion/removal of the set screws. While not implemented here, higher temperatures could be probed through the use of rings based on fluoropolymers, machineable glass ceramics, or boron nitride.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.">Complete electrical circuit: Sample to nanovoltmeter</head><p>Next, we describe the electrical connections between the sample and the nanovoltmeter [Fig. <ref type="figure">4(b)</ref>] and the process of making electrical contact with the sample prior to performing each measurement. Starting from the sample, one end of the brass set screw is pressed into the sample to ensure excellent electrical contact. The other end of the set screw is equipped with a brass M2 nut that clamps a copper ring terminal crimped to a copper wire (Fig. <ref type="figure">3</ref>). Each copper wire is joined to a LEMO cable [obeying the Gerlach convention for voltage polarity, Fig. <ref type="figure">1(a)</ref>] emanating from the nanovoltmeter via a set of terminal blocks. While we have done our best to minimize the number of electrical junctions, especially those consisting of dissimilar metals-these unavoidable junctions can lead to unwanted thermal offset EMFs, which will be discussed in Sec. III.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.">User experience</head><p>The process of making electrical contact with the sample for a measurement is quite easy. The sample is placed in the recessed portion of the copper stage [Fig. <ref type="figure">3(c)</ref>] and the 3D printed voltage retainer ring is placed around the sample. The keyed design of the voltage ring ensures that it can only be installed one way. Three locating pins (two in the front and one in the back) interface with corresponding holes on the copper block [Fig. <ref type="figure">3(b)</ref>]. The pins ensure that the voltage probes are located at the same point every time, that the measured voltage is perpendicular to the plane of the electric field, and that the correct polarity is measured. Once the voltage ring is firmly seated, the set screws are fastened until they make good contact with the sample, and the nuts are tightened. The top heater/cooler block is installed along the guide rods and clamped down. After the Nernst measurement is complete (Sec. III), the nuts are loosened and the screws are backed out. The ring and its accoutrements (Fig. <ref type="figure">3</ref>) are indefinitely reusable until a component breaks, and the replacement of any component is inexpensive in both time and labor. The sample contact making process thus takes less than a minute, which is extraordinary in the world of electrical and especially thermomagnetic measurements. In addition, the thermocouples are permanently embedded in the copper blocks directly under the sample [Fig. <ref type="figure">3(c)]</ref>.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>C. Magnet</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="1.">Magnet considerations</head><p>Some of the considerations for the magnet used for a Nernst measurement include the desired field strength, the ability to vary the field strength, and the pole gap, which dictates the distance to fit the sample holder assembly. While the particular method to deliver T varies across Nernst instruments, all designs are united by the constraint that the sample holder must fit within the poles of a magnet or the bore of a superconducting coil. Regarding the required field strength, for a low field Nernst measurement [B 1, recall : mobility (m 2 /V s), B: magnetic field (T)], a magnet that can achieve 1 T is quite sufficient for most thermoelectric materials. An electromagnet has the advantage of a relatively larger pole gap and ease of reversing field direction as compared to a permanent magnet. <ref type="bibr">26</ref> The ability to control the field in steps allows for more accurate measurements than transient measurements at the expense of equilibration time.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.">Calibration</head><p>To calibrate our magnet, we installed a gaussmeter where the sample sits. We kept all of the wires and sample holder components in place during our calibration to get an accurate environment for what our sample experiences. While we specifically chose materials with low magnetic susceptibility, we wanted to ensure that our gaussmeter experienced the same electromagnetic environment as our sample during measurement. We then applied a DC current to the magnet at linear increments up to 50 amps and reversed the polarity, recording the field at each step. The resulting dataset is linear (r = 0.9998) and shown in Fig. <ref type="figure">S3</ref>. Our maximum field is achieved at 55 amps, which corresponds to 0.935 T. In practice, we typically sweep up to 50 amps.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>D. Comparison to other Nernst instruments</head><p>Having introduced our embodiment, we compare and contrast the design to other reports in the field. In this discussion, we will focus on (i) electrical contacts, (ii) sample geometry, (iii) heat flow geometry and thermal management, and (iv) thermometry.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="1.">Electrical contacts</head><p>Concerning electrical contacts, the literature to date has focused on wire-bonding or other adhesion-based approaches to attach contacts to the sample. <ref type="bibr">6,</ref><ref type="bibr">8,</ref><ref type="bibr">22,</ref><ref type="bibr">25,</ref><ref type="bibr">29,</ref><ref type="bibr">30</ref> One notable exception is the on-chip lithographic lead fabrication employed by Wu et al. <ref type="bibr">29</ref> Our approach focuses on mechanical removable contacts constrained within a ring to improve alignment, ease of application, and resiliency [Fig. <ref type="figure">3(a)</ref>].</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.">Sample geometry</head><p>The sample geometry in the literature varies widely, with thin films, <ref type="bibr">6,</ref><ref type="bibr">21,</ref><ref type="bibr">29</ref> parallelepiped bars, <ref type="bibr">8,</ref><ref type="bibr">22,</ref><ref type="bibr">24,</ref><ref type="bibr">25</ref> and single crystal measurements <ref type="bibr">30</ref> reported by various groups. Thin-film samples have the advantage of lithographic techniques to pattern geometries to maximize the Nernst voltage, avoid unwanted spurious voltages, <ref type="bibr">6</ref> and even fabricate on-chip electrical leads for the sample. <ref type="bibr">29</ref> Within the context of bulk materials, we have found that the cylindrical geometry is appealing due to the limited modifications required for a uniaxially pressed sample. Through the use of our voltage probe ring [Fig. <ref type="figure">3(a)</ref>], we are able to easily measure the voltage Vy and calculate ly as simply the diameter of the sample.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.">Heat flow geometry and thermal management</head><p>Establishing a well-defined thermal gradient in a bulk material is challenging, and multiple sample holder geometries are found in the literature. For example, a sandwich geometry is sometimes used, with heaters on the y-z faces [i.e., faces normal to the heat flow direction, Fig. <ref type="figure">1(a)</ref>]. <ref type="bibr">8,</ref><ref type="bibr">30</ref> This is attractive as it facilitates pseudo-one dimensional heat flow. However, other approaches apply heat in a Review of Scientific Instruments ARTICLE pubs.aip.org/aip/rsi "bridge" geometry, with heat being applied and removed at the ends of the x-y face. <ref type="bibr">6,</ref><ref type="bibr">22,</ref><ref type="bibr">25</ref> This approach is effective for thin film samples but may lead to errors in the temperature gradient in thicker samples due to thermal gradients in the z-direction. Here, we have utilized the sandwich geometry with active cooling and heating; this has several benefits: (1) active cooling at one end of the sample allows us to focus on using a geometry that optimizes signal and accuracy rather than worrying about dissipating sufficient heat flux, (2) supplying the temperature gradient across the faces of the sample (sandwich geometry) clearly defines the lx portion of &#8711;xT (accurately determining lx can be a challenge <ref type="bibr">25</ref> ), and (3) the use of active cooling on the cold side lends itself to an extremely controllable and stable T. This last point is in contrast to the vast majority of systems that simply use heat baths to passively remove heat, rendering control of T difficult.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.">Thermometry</head><p>In the Nernst literature, there seems to be a broad consensus that accurate thermometry is challenging. <ref type="bibr">8,</ref><ref type="bibr">25,</ref><ref type="bibr">29</ref> Correctly determining the temperature gradient at the sample requires decisions concerning thermocouple placement and environment. Specific concerns include thermal contact resistance, associated hot/cold finger effects, and estimation of lx. Wu et al. resolved these conflicts in their Nernst measurement device by fabricating on-chip thermometers from indium oxide. <ref type="bibr">29</ref> This is an innovative approach to optimize thermal contact between the thermometer and the sample; however, it requires lithography (not applicable to bulk materials), and systematic errors are introduced when translating the current to a temperature using a linear approximation. <ref type="bibr">29</ref> Universally, thermal contact resistance between the sample and thermometer leads to errors in the T used to estimate &#8711;xT; <ref type="bibr">30</ref> our system embeds the thermocouples in copper blocks to increase the contact area. <ref type="bibr">31</ref> Another concern is hot/cold finger effects; this is mitigated in the Wu design as well as in designs where the thermocouple is attached to a copper plate thermally sunk to the sample <ref type="bibr">8</ref> or the thermocouple is threaded through the primary stage heater. <ref type="bibr">25</ref> </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>III. MEASUREMENT</head><p>Having discussed our hardware embodiment, we move on to our measurement protocol. The Nernst measurement consists of multiple voltage readings in a constant temperature gradient as the magnetic field is swept from a negative to a positive field. In this section, we describe our measurement procedure to calculate the Nernst coefficient with minimum uncertainty and error. Along the way, we correct for assumptions and unwanted voltages from competing transport phenomena. We also discuss what types of materials are good candidates for Nernst measurements and their ideal geometry to generate the largest possible signal.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>A. Materials studied</head><p>We provide demonstration measurements for three different materials with diverse electronic properties. Degenerate p-type SnTe, intrinsic n-type InSb, and semimetal Bi are all synthesized using typical solid state techniques and uniaxial hot pressing to form dense polycrystalline pellets for electronic measurements. The dimensions of the samples are lx = 6.327 mm, ly = 12.9 mm (SnTe); lx = 8.150 mm, ly = 12.9 mm (InSb); and lx = 7.160 mm, ly = 13.0 mm (Bi). Knowledge of these dimensions is required to calculate the Nernst coefficient [Eq. ( <ref type="formula">2</ref>)]. The dimension ly is set by the diameter of the hot press die, and therefore, this dimension is similar between samples.</p><p>These materials were selected for their relevance to the thermoelectric community as well as their excellent room temperature mobility. At 50 &#9675; C, Hall mobility values measured on our high temperature Hall effect apparatus <ref type="bibr">26</ref> are the following: 466 cm 2 /V s (SnTe), -786 cm 2 /V s (InSb), and -3170 cm 2 /V s (Bi). While these are high electronic mobility values, they still fall under the low magnetic field regime up to 1 T. Hall carrier concentrations at 50 &#9675; C are 1.09 &#215; 10 20 cm -3 (SnTe), -1.58 &#215; 10 17 cm -3 (InSb), and -1.27 &#215; 10 19 cm -3 (Bi). We calculated these carrier concentrations from RH = <ref type="bibr">1</ref> ne , where e is the fundamental charge, n is the Hall carrier concentration, and RH is the measured Hall coefficient. Mobility values were derived using the Drude theory 1 &#961; = ne, where &#961; is the measured electrical resistivity. Resistivity values measured at 50 &#9675; C are 50.45 m-cm (InSb), 0.13 m-cm (SnTe), and 0.16 m-cm (Bi). Seebeck coefficient values are shown in Fig. <ref type="figure">7</ref>.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>B. Measurement sequence</head><p>The low field Nernst coefficient (N) measurement is carried out in a thermal gradient applied along the x-axis of the sample in a variable magnetic field (z-axis).</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="1.">Set temperature gradient</head><p>First, a T of 1-5 K (higher magnitude for low mobility samples to drive a detectable voltage) is applied via the thermoelectric controller [Fig. <ref type="figure">5(a)</ref>]. A desired average temperature and T are input to our software program (all done in Python), and the upper and lower temperatures are automatically calculated and sent to the controllers [i.e., an input of T = 20 &#9675; C, T = 5 K, will drive the top block to 22.5 &#9675; C and the bottom block to 17.5 &#9675; C, Fig. <ref type="figure">5(a)</ref>]. The two Peltier coolers are driven to the required set points and equilibrated for 45 s. Our 1D heat flow modeling in COMSOL suggests that for a 6 mm long sample with a thermal conductivity of 3 W m -1 K -1 (i.e., SnTe), the temperature gradient will be established within 10 s (Fig. <ref type="figure">S1</ref>). Since this is calculated with many approximations (dT is set as a boundary condition at sample edges rather than at the copper blocks in actuality), we allow ample extra time for equilibration. Our measurements show temperature fluctuations on the order of mK (Fig. <ref type="figure">S5</ref>). Nernst measurements are fundamentally different from Seebeck measurements in that the Seebeck measurement is resilient against non-linear temperature profiles when the sample is homogeneous; this is not the case for Nernst measurements.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.">Step the magnet field</head><p>Once the T has equilibrated for 45 s, current is sent to the magnet to drive a field of -1 T [Fig. <ref type="figure">5(b)</ref>]. Once the field has settled, the induced transverse sample voltage is measured by the nanovoltmeter [Fig. <ref type="figure">5(c)</ref>]. Then, the magnetic field is incremented in user-specified linear steps up to +1 T, and the nanovoltmeter measures voltage along the way [Figs. 5(b) and 5(c)]. We note that we wait for the magnetic field to stabilize at each step, sacrificing acquisition time in order to maximize accuracy. To verify that our field sweep rate is not too rapid, we performed additional measurements at a significantly reduced rate and achieved the same value N (Fig. <ref type="figure">S6</ref>). </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Review of Scientific Instruments</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.">Measure voltage</head><p>Under low magnetic fields, the Nernst voltage is a linear function of the field. <ref type="bibr">5</ref> From the field-dependent voltage measurements taken at a constant T (Fig. <ref type="figure">5</ref>), we are able to calculate the low field Nernst coefficient,</p><p>By measuring the voltage at different field values, we are able to calculate V/B. This slope method has the benefit of effectively canceling out any Seebeck voltages or other offset voltages that contribute to our Nernst measurement, assuming the Seebeck offset is an even function of the field (Fig. <ref type="figure">6</ref>). Any thermal EMFs that may arise due to temperature differences along the wires will be canceled out via our differential measurement.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.">Reverse temperature gradient</head><p>Our instrument's design allows us to easily reverse the direction of the temperature gradient. By reversing the direction of T and taking an average (paying attention to polarity) of the V/B for the forward and reverse temperature gradient measurements, we effectively cancel out thermal EMF offsets. This is akin to the DC reversal or delta method commonly performed in Hall and resistivity measurements. <ref type="bibr">26,</ref><ref type="bibr">32,</ref><ref type="bibr">33</ref> The voltage data recorded for the forward and reverse T are shown in Fig. <ref type="figure">5(d</ref>) and, as expected, have opposite slopes. Of note, at zero field, some offset is present in the data. This offset will be discussed in detail in Subsection III B 5.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="5.">Treatment of offset voltages</head><p>At B = 0, the measured Nernst voltage should be 0 V in theory [Eq. ( <ref type="formula">1</ref>)]. However, this is not the case in practice, 8 as evidenced herein [Figs. <ref type="figure">5(c)</ref> and <ref type="figure">5(d)]</ref>. In this subsection, we will explain why these offsets are not overly problematic.</p><p>The measurement of the Nernst voltage includes not only the desired voltage but also the Seebeck voltage due to any misalignment in the voltage probes and any other thermal EMFs or noise in the circuit. Considering Fig. <ref type="figure">5(c</ref>) at B = 0 (minutes 5.5 and 15.5), the voltages are -7.3 and -8.9 V for the forward and reverse temperature gradients. Treating the voltage offset as a combination of two linear terms, one of which depends on the sign of the temperature gradient, we conclude that the majority of the offset (-8.1 V) is independent of the temperature gradient. The Seebeck voltage is only 0.8 V; this provides direct insight into the probe offsets.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Review of Scientific Instruments</head><p>The corresponding probe offset along the x-direction is l offset =</p><p>V offset E S = 0.8 V 2.4&#215;10 -3 V/m = 0.33 mm, where ES refers to the Seebeck electric field. For S SnTe = 5 V/K [Fig. <ref type="figure">7</ref></p><p>. This l offset is significantly smaller than the height of the pellet (lx &#8776; 6 mm), highlighting the excellent alignment of the two probes.</p><p>To further investigate the role of probe misalignment, we took additional measurements at various T and saw that the relationship between dVdB and T is linear and nearly pierces the origin (Fig. <ref type="figure">S7</ref>). This indicates that our measurement procedure is effective at canceling out thermal offset voltages. In conclusion, our measurement procedure of extracting the Nernst coefficient from the slope of dVdB effectively cancels out both thermal EMF offsets and Seebeck offset voltages.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>C. Magneto-Seebeck coefficient measurement</head><p>Measuring the magneto-Seebeck coefficient can be useful for understanding charge carrier transport in a given material in addition to the Nernst coefficient. Our measurement protocol consists of (i) moving the voltage probes from a transverse (Nernst) to a longitudinal configuration along the sample [Fig. <ref type="figure">7(a)</ref>] and (ii) performing dVdT measurements at different B fields. Each copper block [Fig. <ref type="figure">2(a)</ref>] has a small drilled hole directly under the sample. To measure the magneto-Seebeck voltage, we insert copper wires into these holes [Fig. <ref type="figure">3(b)</ref>] and connect them to the Keithley 2182 nanovoltmeter. In this way, we measure the voltage longitudinally, taking advantage of the high electrical conductivity and low Seebeck coefficient of copper.</p><p>While both the Nernst and the magneto-Seebeck measurements use a differential method to eliminate offsets and calculate the desired coefficient, the protocol differs in what is swept and what is held constant. For the Nernst measurement, we apply a constant T and sweep the field (Fig. <ref type="figure">5</ref>), and for the magneto-Seebeck measurement, we hold the field constant and sweep T. In Fig. <ref type="figure">7</ref>, each data point represents a slope V/T at a given field. The procedure of sweeping T and recording V/T to obtain the Seebeck coefficient is quite common and considered to be more accurate than other methods. <ref type="bibr">22,</ref><ref type="bibr">27,</ref><ref type="bibr">28</ref> As was the case for measured Nernst voltage vs field [Fig. <ref type="figure">5(d)]</ref>, non-linearity in V/T for Seebeck may suggest poor contacts, a poor signal, or both. <ref type="bibr">22,</ref><ref type="bibr">27,</ref><ref type="bibr">28</ref> In the absence of a magnetic field, misalignment of the Nernst probes will induce a small Seebeck voltage due to the ensuing Tx; as discussed above, this can be eliminated through a differential [V/B, Eq. ( <ref type="formula">3</ref>)] Nernst measurement. In the case where the Seebeck response is an even function of the magnetic field, a linear regression of the resulting curve will continue to preserve the true Nernst value, provided that both positive and negative magnetic field values are sampled evenly [Fig. <ref type="figure">6(b)</ref>]. However, if the magneto-Seebeck response is not an even function, this approach will not work, and its contribution cannot be simply removed from the Nernst voltage by calculating V/B. As real samples may show anisotropic transport, the magneto-Seebeck response may not be a perfectly even function, and we generally recommend determining the magneto-Seebeck response when analyzing experimental Nernst data.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>D. Sample property considerations</head><p>The need to measure nV to V signals distinguishes the Nernst effect from more common electronic transport measurements, which typically involve measuring voltages that are orders of magnitude higher, such as electrical resistivity, Seebeck, and Hall. <ref type="bibr">6</ref> The difficulty in measuring petite Nernst signals is exacerbated in samples with low electronic mobility. In this section, we provide an overview of the fundamental physics that contributes to a detectable Nernst signal and, therefore, what material properties must be in place for a detectable Nernst measurement.</p><p>Several formalisms describe the Nernst effect and aid the reader in understanding what material properties are required for a detectable Nernst signal. <ref type="bibr">1,</ref><ref type="bibr">[3]</ref><ref type="bibr">[4]</ref><ref type="bibr">[5]</ref><ref type="bibr">8,</ref><ref type="bibr">16</ref> The solution of the Boltzmann equation (independent of single parabolic band assumptions) yields the following relationship between Nernst and mobility: N &#8733; &#964; 0 m * . <ref type="bibr">3,</ref><ref type="bibr">16</ref> Therefore, high electronic mobility is necessary to generate a detectable Nernst signal since is proportional to &#964; 0 m * . In practice, measuring the Nernst coefficient on samples with mobility values &lt;50 cm 2 V -1 s -1 can be challenging. We add that this heuristic approach depends on the energy dependence of scattering. In short, thermoelectric materials suffering</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Review of Scientific Instruments</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>ARTICLE</head><p>pubs.aip.org/aip/rsi from low mobility due to extensive defect engineering to decrease their thermal conductivity might not be good candidates for Nernst measurements.</p><p>Low thermal conductivity, on the other hand, is a desirable feature. This lowers the heat requirement to impose a dT across the sample. It also increases the amount of time required for equilibration. Beyond low total thermal conductivity, specifically the ratio of lattice thermal conductivity (&#954;L) to electronic thermal conductivity (&#954; el ), it is important to encourage isothermal conditions. High &#954;L relative to &#954; el helps avoid adiabatic conditions, wherein the phonons act as a "shorting fluid" if a temperature gradient tries to develop along the y-direction to cause a thermal Hall effect. <ref type="bibr">5</ref> Finally, we consider the case of bipolar materials. While bipolar materials are not desirable for Seebeck-based thermoelectric applications, they actually lead to a larger Nernst signal than materials with one dominant carrier type. For example, the large Nernst signal in kagome superconductor CsV 3 Sb5 is attributed to its ambipolar nature and the presence of hole-like and electron-like bands at critical points in the Brillouin zone. <ref type="bibr">17,</ref><ref type="bibr">34</ref> This is due to the heightened voltage generated by electrons and holes traveling in the same direction (from the thermal gradient) but curling in opposite directions (in a magnetic field). <ref type="bibr">16,</ref><ref type="bibr">17</ref> This differs from the Hall effect, where charge carriers travel in opposite directions (from the electrically generated field) and curl in opposite directions, effectively ending up on the same side of the material and canceling each other out.</p><p>In summary, high electronic mobility is a prerequisite for a good Nernst measurement. In addition, low thermal conductivity and ambipolarity are beneficial material properties for Nernst measurements. Examples of materials with high Nernst coefficients are bismuth, InSb, and NbSe 2 , all of which possess excellent electronic mobility. <ref type="bibr">8,</ref><ref type="bibr">16,</ref><ref type="bibr">22</ref> </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>E. Sample geometry</head><p>The ideal sample geometry for a Nernst measurement should be considered from two perspectives: (1) maintaining isothermal conditions, and (2) maximizing the measurable Nernst voltage. Under assumed isothermal conditions, the only temperature gradient present in the sample during a Nernst measurement is along the x-axis. However, a temperature gradient in a magnetic field can give rise to an additional temperature gradient (&#8711;yT) generated by the thermal Hall effect. If &#8711;yT develops, then adiabatic, not isothermal (&#8711;yT = 0) Nernst is measured. That is, an additional voltage develops along the y-axis due to the Seebeck effect (Vy = TySyy). In reality, some medium between pure isothermal or adiabatic conditions is likely to be met in most Nernst measurement systems.</p><p>We acknowledge that our relatively short, squat pellet geometry differs from the long, skinny samples often utilized for Nernst measurements. <ref type="bibr">8,</ref><ref type="bibr">25</ref> This difference is enabled by our active cooling (Peltier cooler), which allows for a squat geometry along lx while maintaining a large T. In addition, embedding thermocouple wires in the copper blocks avoids needing space for multiple contacts to the sample along lx [Fig. <ref type="figure">3(c)</ref>]. For our pellet geometry [Fig. <ref type="figure">3(b)</ref>], adiabatic conditions are avoided by minimizing the pellet height to ensure maximum pellet surface area is in contact with the copper heat spreaders. While a short sample is valuable in terms of maintaining isothermal conditions in y, this places further demands on the overall thermal circuit in x. In particular, maintaining a large &#8711;T becomes challenging as more heat is required to maintain a temperature difference across a thin sample. Furthermore, the thermal contact resistance between the copper blocks and the sample may begin to represent a nontrivial fraction of the overall thermal resistance. For researchers studying the Nernst effect in long (along the axis of temperature gradient), thin samples, isothermal assumptions may not be accurate. <ref type="bibr">30</ref> Conversion from the low field adiabatic to the isothermal Nernst coefficient may be difficult and require measurement of other galvano-thermomagnetic phenomena to apply the Heurlinger relations. <ref type="bibr">35</ref> Now that we have considered how maximizing sample surface area in contact with the heating/cooling blocks encourages an isothermal regime, we turn to sample geometries that generate the highest Nernst voltage. As shown in Eq. ( <ref type="formula">2</ref>), the Nernst voltage Vy scales with l y l x . The ly dimension [Fig. <ref type="figure">1(a)</ref>] is limited by the diameter of the die from which it was pressed. Therefore, maximizing this dimension is difficult without changing the internal dimensions of the hot press die. However, lx is the pellet thickness [Fig. <ref type="figure">1(a)</ref>], and this can be easily manipulated by sanding/polishing the faces of the pellet to reduce its thickness. Minimizing lx for a given ly should increase the measured Nernst voltage [Eq. ( <ref type="formula">2</ref>)]. Our results in Fig. <ref type="figure">S4</ref> of the supplementary material show that increasing the ratio ly/lx did, in fact, increase the measured Nernst voltage. The Nernst coefficient (Fig. <ref type="figure">S4</ref>) is unchanged as expected since N is an intrinsic property.</p><p>We have considered the sample geometry now from two different perspectives: (1) maintaining isothermal measurement conditions and (2) maximizing the Nernst voltage (Fig. <ref type="figure">S4</ref>). As a reminder, a sample that is thin along the direction of the temperature gradient (lx) encourages isothermal conditions by maintaining a greater sample surface area in contact with the copper heat blocks. <ref type="bibr">30</ref> The ideal geometry to maximize the Nernst voltage also benefits from a thin lx since Vy &#8733; l y l x [Eq. (3)]. Therefore, we use the minimal thickness lx [Fig. <ref type="figure">1(a)</ref>] pellet to maximize the Nernst voltage. For our system, this corresponds with a sample at least 5.1 mm thick because our plastic voltage ring assembly [Fig. <ref type="figure">3(b)</ref>] is 5 mm thick.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>F. Measurement uncertainty</head><p>The origins of uncertainty in our Nernst coefficient measurements could arise from the measurement hardware (nanovoltmeter, thermocouples, magnet, thermal equilibration between sample holder and sample) or from imprecise sample geometry measurement [Eq. ( <ref type="formula">3</ref>)]. Our efforts to quantify this unavoidable uncertainty consist of performing a root-square-sum (RSS) [Eq. ( <ref type="formula">4</ref>)] on the primary sources of uncertainty.</p><p>Inspection of Eq. ( <ref type="formula">3</ref>) reveals the primary sources of uncertainty in a Nernst measurement. The sources we used, which correspond to the standard deviation &#963;i in Eq. ( <ref type="formula">4</ref>), are (i) repeated measurements of the Nernst coefficient itself, (ii) uncertainty in lx, and (iii) uncertainty in ly. The Nernst coefficient (V/KT) measurement embodies error from thermometry, the voltmeter, and the magnet. To quantify the uncertainty in the Nernst measurement, we repeated 26 Nernst coefficient measurements on the same sample of SnTe at T = 20 &#9675; C Review of Scientific Instruments ARTICLE pubs.aip.org/aip/rsi over the course of a week. The average value was 2.014 V/KT, and the sample standard deviation was 0.023 V/KT. For the sample geometry (ii) and (iii), we determined the actual tolerance range, which was 0.050 mm for lx (sample height) and 0.10 mm for ly (sample diameter). The total standard deviation (uncertainty) is</p><p>Using Eq. ( <ref type="formula">4</ref>), we calculate a &#963; total of 0.052 V/KT for N = 2.014 V/KT. This agrees well with the experiment-the standard deviation taken from repeated measurements on the same sample (over the span of several days) is &#963; = 0.023 /KT.</p><p>A final note-the lack of a calibration standard for Nernst coefficient measurements makes absolute error analysis difficult in reference to a calibrated standard. However, we are most interested with trends within a given experimental campaign. As such, we are most concerned with the precision rather than error minimization of our measurement collection sequence.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>IV. FUNDAMENTAL THEORY AND ANALYSIS</head><p>In the Introduction, we discussed the physical origin and the significance of the Nernst effect measurement. We stated that within the single parabolic band regime, Nernst measurements can be analyzed using the method of four coefficients to determine the dominant carrier scattering mechanism in a conductive material. In this section, we describe in more detail how this is done, starting with the macroscopic picture and then diving into the underlying transport physics that can be used to analyze Nernst data.</p><p>On a macroscopic level, the coupled current equations explain the relation between measurable electronic and heat currents (J e and J q ) and the force and field vectors that generate these currents. Here, we will focus on the electronic current J e , Je = &#963; &#8901; E&#945; &#8901; &#8711;T.</p><p>(</p><p>Writing out the full tensor and asserting that there is a magnetic field applied along the z-direction, <ref type="bibr">36</ref> we obtain Jx Jy Jz = &#963;xx &#963;xy 0 &#945;xx &#945;xy 0 &#963;yx &#963;yy 0 &#945;yx &#945;yy 0 0 0 &#963;zz 0 0 &#945;zz Ex Ey Ez -&#8711;xT -&#8711;yT -&#8711;zT .</p><p>The electric and thermoelectric conductivity tensors &#963; and &#945; have on-diagonal components that can be probed via longitudinal measurements (i.e., resistivity, Seebeck), and the off-diagonal components (&#963;ij, &#945;ij, i &#8800; j) are zero in the absence of a magnetic field (for an isotropic material; notwithstanding interesting cases such as materials exhibiting axis-dependent conduction polarity <ref type="bibr">30</ref> ). Before applying Eq. ( <ref type="formula">5</ref>) to Nernst measurements, we consider the simpler case of a Seebeck measurement. Here, a temperature gradient is applied only along the x-axis of a material, and there is no applied magnetic field. Under an open circuit boundary condition (J x = 0), the first row of Eq. ( <ref type="formula">5</ref>) becomes</p><p>The material property that describes the drift of electrons due to a temperature gradient (thermoelectric conductivity 10 ) is &#945;xx. The more familiar electrical conductivity associated with an applied electric field is denoted by &#963;xx. Thus, an electric field is generated to balance the diffusion of charge carriers due to the temperature gradient. Equation ( <ref type="formula">7</ref>) can be rearranged to yield the longitudinal Seebeck coefficient Sxx,</p><p>Thus, knowledge of only &#945;xx and &#963;xx is sufficient to predict the Seebeck effect. The quantity &#945; xx &#963; xx is known as the Seebeck coefficient Sxx = E x &#8711; x T (note on sign convention <ref type="bibr">37</ref> ). Moving on to the Nernst effect, the off-diagonal terms &#963;xy and &#945;xy become non-zero when we apply a magnetic field. The magnetic field causes a Lorentz deflection of the thermally generated Ex, and an off-diagonal electrical component &#963;yxEx is balanced by an off-diagonal thermoelectric (Peltier) component &#945;yx(-&#8711;xT). Now, in addition to J x , we must consider J y ,</p><p>J x is no longer equal to the expression in Eq. ( <ref type="formula">7</ref>), but off-diagonal terms are now included due to the magnetic field applied, </p><p>By assuming that &#963; 2 xy &#963; 2 xx (low magnetic field), we can rearrange Eq. (11),</p><p>Multiplying the first term of Eq. ( <ref type="formula">12</ref>) by &#945;xx&#945;xx leads to the further simplified equation,</p><p>The Nernst signal Ey/&#8711;xT is sometimes referred to in the literature as Syx or the Nernst thermopower, to distinguish it from the longitudinal thermopower or Seebeck coefficient. <ref type="bibr">15</ref> It is tempting to draw the conclusion upon inspection of Eq. ( <ref type="formula">13</ref>) that a large Seebeck coefficient (Sxx) will lead to a large Nernst signal, but in fact, we should draw our attention to the terms in parentheses. The second term in</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Review of</head><note type="other">Scientific Instruments ARTICLE pubs</note><p>.aip.org/aip/rsi parentheses in Eq. ( <ref type="formula">13</ref>) is &#963;xy/&#963;xx; this term can be related to the Hall angle, &#952;H. The magnitude of &#952;H can be considered a measure of how strongly a drifting charge carrier is deflected in a transverse magnetic field due to the Lorentz force. The deflection can be viewed as the sum of two orthogonal vectors: &#963;xx x + &#963;xy&#375;. The inner angle therein is thus given by &#963;xy/&#963;xx &#8801; tan(&#952;H). Similarly, the thermally driven carriers deflect in the magnetic field. The first term in parentheses on Eq. ( <ref type="formula">13</ref>) is &#945;xy/&#945;xx &#8801; tan(&#952;&#945;). <ref type="bibr">5</ref> Thus, the Nernst coefficient is a fight between the thermoelectric term tan(&#952;&#945;) and the Hall term tan(&#952;H). Now that we have considered what is measured macroscopically when performing a Nernst measurement, we consider the solutions to the Boltzmann transport using the transport distribution function, or TDF. The TDF in the case of periodic crystalline materials is <ref type="bibr">10,</ref><ref type="bibr">11,</ref><ref type="bibr">36,</ref><ref type="bibr">38</ref> </p><p>where E is the energy, g(E) is the density of states, v(E) is the group velocity of charge carriers along the direction of transport, and &#964;(E) is the electron relaxation time. If many bands or non-parabolic bands contribute to electron transport, then a summation must be performed instead of the simple Eq. ( <ref type="formula">14</ref>). <ref type="bibr">10,</ref><ref type="bibr">38</ref> The transport coefficients are given by solutions to the Boltzmann equation. We note that &#931;(E) could be a more complicated summation than the simple case of Eq. ( <ref type="formula">14</ref>). The longitudinal electric (&#963;) and thermoelectric (&#945;) conductivities are given below in terms of the TDF. For brevity, we omit the subscript on longitudinal [on-diagonal components of Eq. ( <ref type="formula">6</ref>)] coefficients. For an isotropic compound, &#963; = &#963;xx = &#963;yy = &#963;zz,</p><p>We can provide Eq. ( <ref type="formula">16</ref>) in terms of @f 0 /@E as well</p><p>In this form [Eq. ( <ref type="formula">17</ref>)], it is even more explicit that the carriers slightly above and below but not exactly at the Fermi contribute to thermoelectric conductivity (&#945;).</p><p>The transverse conductivities are as follows: where we pick up a magnetic (perpendicular to the electric field) contribution of B, where = e&#964;/m * b (electronic mobility), B is the magnetic field, and m * b is the inertial (band) mass (which is equal to the density of state mass in a single band model),</p><p>Under low field conditions, B 1, and the formula simplifies to</p><p>Making the same consideration as above for low-field conditions, the transverse thermoelectric conductivity is</p><p>Equations ( <ref type="formula">15</ref>)-( <ref type="formula">20</ref>) can be used to write an expression for the Nernst coefficient [Eq. (11)] in terms of the TDF. Similarly, expressions for the electrical conductivity [Eq. ( <ref type="formula">15</ref>)] Seebeck coefficient (&#945;/&#963;) and Hall coefficient (&#963;xy/B &#963; 2 xx ) can be developed. In practice, these expressions are considered under the constraints and boundary conditions of various models, including single or multiband Kane, bipolar, or single or multiband parabolic models. Employing a parabolic or Kane band model is quite common. <ref type="bibr">4,</ref><ref type="bibr">7,</ref><ref type="bibr">24,</ref><ref type="bibr">39</ref> In the next section, we focus on the single parabolic band (SPB) model.</p><p>The SPB model offers a way to solve for the underlying physical parameters that govern a material's electronic properties. The four fundamental electronic transport parameters (&#964;, r, EF, m * ) are visible in Eqs. ( <ref type="formula">21</ref>)-( <ref type="formula">24</ref>), although some do not contain all four. Along with the parabolic band approximations, an additional assumption is generally made during analysis. This assumption is that scattering time follows a power law: &#964; = &#964; 0 E r-0.5 , where &#964; is the carrier relaxation time, &#964; 0 is the relaxation time prefactor, and r is the scattering exponent. The scattering exponent r typically adopts a value between 0 (acoustic phonon scattering) and 2 (ionized impurity scattering). <ref type="bibr">3</ref>  </p><p>where Fr refers to the Fermi integral: &#8747; &#8734; 0 &#949; r d&#949; 1+Exp(&#949;-&#951;) . We provide the equations for Boltzmann solutions to equations for RH, S, and &#963; as follows: <ref type="bibr">3</ref> Electric conductivity: (</p><p>Within the SPB and power law scattering approximations, the four material parameters encompassed by the TDF (m * , &#964;, r, and EF) can be calculated by solving a system of four equations [Eqs. ( <ref type="formula">21</ref>)- (24)] with four measurable transport coefficient values as Review of Scientific Instruments ARTICLE pubs.aip.org/aip/rsi input. Measuring S, RH, &#963;, and N and solving the system of equations is deemed the method of four coefficients. <ref type="bibr">3,</ref><ref type="bibr">7,</ref><ref type="bibr">21,</ref><ref type="bibr">40</ref> The method of four coefficients has been employed over the past couple of decades by researchers deeply interested in the material limits to electron mobility across a breadth of fields, including thermoelectrics <ref type="bibr">4,</ref><ref type="bibr">7,</ref><ref type="bibr">8,</ref><ref type="bibr">24,</ref><ref type="bibr">41</ref> and transparent conducting oxides. <ref type="bibr">21,</ref><ref type="bibr">40,</ref><ref type="bibr">42</ref> Without measuring Nernst, an assumption about one of the variables is required. This can lead to an inaccurate calculation of the fundamental material parameters. <ref type="bibr">3,</ref><ref type="bibr">43,</ref><ref type="bibr">44</ref> In practice, the value of r is often assumed to be 0, which can lead to incorrect calculations of the TDF. <ref type="bibr">43,</ref><ref type="bibr">44</ref> The Nernst effect is incredibly sensitive to scattering type, <ref type="bibr">3,</ref><ref type="bibr">6,</ref><ref type="bibr">15</ref> and measuring N can lead to more accurate solutions to the TDF.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>SUPPLEMENTARY MATERIAL</head><p>See the supplementary material for figures on simulated heat flow, magnet calibration, sample geometry optimization, and additional measurements showing slower magnetic field sweep and the error in voltage readings. The data used to generate the figures in the manuscript as well as the supplement are provided in .xlsx format.</p></div></body>