Title: A feasibility study of biodegradable magnesium-aluminum-zinc-calcium-manganese (AZXM) alloys for tracheal stent application

Airway obstruction conditions are relatively rarely observed in clinical settings but nevertheless, extremely challenging to handle, especially when observed in pediatric patients. Several surgical procedures, including tracheal resection, end-to-end tracheal anastomosis, and tracheoplasty, have been developed and practised of late, to treat airway obstruction. However, the clinical outcome is typically not satisfactory due to airway restenosis conditions that develop following surgery. Various types of stents are currently available for airway stenting ranging from non-degradable silicone tubes and bio-inert metallic stents (bare or coated with polymer matrix) to hybrid silicone tubes strengthened by metallic cores, but none of the stents provides the satisfactory long-term effectiveness. Therefore, there is a significant clinical need for a biodegradable airway stent that would maintain airway patency and totally degrade over time after meeting the desired objectives. The present study aims to investigate biodegradable magnesium-aluminum-zinc-calcium-manganese (AZXM) alloy as a potential tracheal stent. The new AZXM alloy was fabricated by partially replacing aliminum in commercial AZ31 alloy with calcium. The present study demonstrates that calcium preferentially segregates along the grain boundaries as intermetallic phases (Mg2Ca) and is homogeneously distributed in the magnesium matrix. The extruded AZXM alloy showed less pitting, higher corrosion resistance in Hank's Balanced Salt Solution (HBSS) compared to the as-cast and solution-treated AZXM alloys and exhibited optimized mechanical properties. In vitro cytotoxicity evaluation using human trachea epithelial cells demonstrated excellent cyto-compatibility of AZXM alloys compared to pure Mg and commercial AZ31 validated by a very preliminary rabbit in vivo tracheal model study. Preliminary results show that the approach to use biodegradable AZXM alloys as a tracheal stent is indeed promising, although further alloy processing is required to improve the ductility needed followed by a more exhaustive in vivo study to demonstrate full viability for stent applications.

 
more » « less
PAR ID:
10547222
Author(s) / Creator(s):
 ;  ;  ;  
Publisher / Repository:
SAGE Publications
Date Published:
Journal Name:
Journal of Biomaterials Applications
Volume:
33
Issue:
8
ISSN:
0885-3282
Format(s):
Medium: X Size: p. 1080-1093
Size(s):
p. 1080-1093
Sponsoring Org:
National Science Foundation
More Like this
  1. Abstract

    Tracheal stenting currently using non‐degradable stents is commonplace for treatment of trauma, prolonged intubation related adult airway obstructions, and pediatric patients‐associated tracheal stenosis conditions. Degradable tracheal stent placement will avoid complications of stent removal and restenosis. Widespread reports exist on degradable magnesium alloys success for orthopedic and cardiovascular applications but none to date for intra tracheal use. This research explores the use of pure Mg, AZ31, and Mg‐3Y alloys for degradable tracheal stent assessment.In vitroevaluation of magnesium, prototype stents in a bioreactor simulate the airway environment and corrosion. Micro‐CT imaging and biocompatibility evaluation helped assess the 24‐week degradation of intraluminal alloy stents following implantation in a rat trachealin vivobypass model. Histological analysis indicate tissue response of the harvested stented trachea segments after each time point. Corrosion studies for each alloy indicate significant differences between the simulated and controlin vitroconditions. AZ31 exhibited the lowest volume loss of 6.8% in saline, while pure Mg displayed the lowest volume loss of 4.6% in simulated airway fluid (SAF), both at 1‐week time points. Significant differences in percentage of total volume lost after 6 months were determined between the alloys over time. MgY alloy displayed the slowest corrosion losing only 15.1% volume after 24 weeks of immersion. Additionally,in vitromagnesium alloy corrosion was not significantly different from the percentage of total volume lostin vivoat 1‐week time point. The study demonstrates promise of magnesium alloys for intraluminal tracheal stent application albeit viability of a clinically translatable model warrants further studies. © 2018 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 107B: 1844–1853, 2019.

     
    more » « less
  2. Abstract

    For degradable magnesium (Mg) alloy‐based stents it would be desirable to delay early corrosion to maintain mechanical strength. Similarly, early after stent placement reduced thrombogenicity is an important feature, while chronically, endothelial cell adhesion and vessel integration are desirable. In this study, surface eroding polymers of amino‐grafted poly(1,3‐trimethylene carbonate) (PTMC‐NH2) and PTMC‐NH2combined with sulfobetaine bearing polymer PSB (PTMC‐NHCO‐PSB) are developed, and these polymers are covalently attached onto 6‐phosphonohexanoic acid (PHA)‐coated AZ31 Mg alloy surfaces in sequence. In vitro degradation testing in ovine plasma shows PTMC, PTMC‐NH2, and PTMC‐NHCO‐PSB cast films experience a gradual thickness and mass loss with maintenance of smooth surfaces, confirming surface erosion behavior. The PTMC‐NH2polymer is firmly bound to the PHA‐modified AZ31 surface and demonstrates a resistance to peeling. PTMC, PTMC‐NH2, and PTMC‐NHCO‐PSB coated AZ31 have a lower corrosion rate versus polylactide‐co‐glycolide coated and untreated AZ31. PTMC‐NHCO‐PSB coated AZ31 inhibits platelet deposition and smooth muscle cell adhesion and growth, but after 2‐week immersion in plasma, this surface supports endothelial cell adhesion and growth. These results suggest PTMC‐NHCO‐PSB surface eroding coating offers a means of controlling corrosion while providing a temporally varying bio‐functionality for biodegradable vascular stent applications.

     
    more » « less
  3. Polymeric coatings can provide temporary stability to bioresorbable metallic stents at the initial stage of deployment by alleviating rapid degradation and providing better interaction with surrounding vasculature. To understand this interfacing biocompatibility, this study explored the endothelial-cytocompatibility of polymer-coated magnesium (Mg) alloys under static and dynamic conditions compared to that of non-coated Mg alloy surfaces. Poly (carbonate urethane) urea (PCUU) and poly (lactic-co-glycolic acid) (PLGA) were coated on Mg alloys (WE43, AZ31, ZWEKL, ZWEKC) and 316L stainless steel (316L SS, control sample), which were embedded into a microfluidic device to simulate a vascular environment with dynamic flow. The results from attachment and viability tests showed that more cells were attached on the polymer-coated Mg alloys than on non-coated Mg alloys in both static and dynamic conditions. In particular, the attachment and viability on PCUU-coated surfaces were significantly higher than that of PLGA-coated surfaces of WE43 and ZWEKC in both static and dynamic conditions, and of AZ31 in dynamic conditions (P<0.05). The elementary distribution map showed that there were relatively higher Carbon weight percentages and lower Mg weight percentages on PCUU-coated alloys than PLGA-coated alloys. Various levels of pittings were observed underneath the polymer coatings, and the pittings were more severe on the surface of Mg alloys that corroded rapidly. Polymer coatings are recommended to be applied on Mg alloys with relatively low corrosion rates, or after pre-stabilizing the substrate. PCUU-coating has more selective potential to enhance the biocompatibility and mitigate the endothelium damage of Mg alloy stenting. 
    more » « less
  4. Magnesium (Mg)-based alloys have the potential for bone repair due to their properties of biodegradation, biocompatibility, and structural stability, which can eliminate the requirement for a second surgery for the removal of the implant. Nevertheless, uncontrolled degradation rate and possible cytotoxicity of the corrosion products at the implant sites are known current challenges for clinical applications. In this study, we assessed in vitro cytotoxicity of different concentrations (0 to 50 mM) of possible corrosion products in the form of magnesium oxide (MgO) and magnesium hydroxide (Mg(OH)2) nanoparticles (NPs) in human fetal osteoblast (hFOB) 1.19 cells. We measured cell proliferation, adhesion, migration, and cytotoxicity using a real-time, label-free, non-invasive electric cell-substrate impedance sensing (ECIS) system. Our results suggest that 1 mM concentrations of MgO/Mg(OH)2 NPs are tolerable in hFOB 1.19 cells. Based on our findings, we propose the development of innovative biodegradable Mg-based alloys for further in vivo animal testing and clinical trials in orthopedics. 
    more » « less
  5. Magnesium (Mg) alloys are being investigated as a biodegradable metallic biomaterial because of their mechanical property profile, which is similar to the human bone. However, implants based on Mg alloys are corroded quickly in the body before the bone fracture is fully healed. Therefore, we aimed to reduce the corrosion rate of Mg using a double protective layer. We used a magnesium-aluminum-zinc alloy (AZ91) and treated its surface with micro-arc oxidation (MAO) technique to first form an intermediate layer. Next, a bioceramic nanocomposite composed of diopside, bredigite, and fluoridated hydroxyapatite (FHA) was coated on the surface of MAO treated AZ91 using the electrophoretic deposition (EPD) technique. Our in vivo results showed a significant enhancement in the bioactivity of the nanocomposite coated AZ91 implant compared to the uncoated control implant. Implantation of the uncoated AZ91 caused a significant release of hydrogen bubbles around the implant, which was reduced when the nanocomposite coated implants were used. Using histology, this reduction in the corrosion rate of the coated implants resulted in an improved new bone formation and reduced inflammation in the interface of the implants and the surrounding tissue. Hence, our strategy using a MAO/EPD of a bioceramic nanocomposite coating (i.e., diopside-bredigite-FHA) can significantly reduce the corrosion rate and improve the bioactivity of the biodegradable AZ91 Mg implant. 
    more » « less