			<titleStmt><title level='a'>Modulating Cardiac Hemodynamics Using Tunable Soft Robotic Sleeves in a Porcine Model of HFpEF Physiology for Device Testing Applications</title></titleStmt>
			<publicationStmt>
				<publisher>Wiley</publisher>
				<date>02/01/2024</date>
			</publicationStmt>
			<sourceDesc>
				<bibl> 
					<idno type="par_id">10550533</idno>
					<idno type="doi">10.1002/adfm.202310085</idno>
					<title level='j'>Advanced Functional Materials</title>
<idno>1616-301X</idno>
<biblScope unit="volume">34</biblScope>
<biblScope unit="issue">8</biblScope>					

					<author>Luca Rosalia</author><author>Caglar Ozturk</author><author>Sophie X Wang</author><author>Diego Quevedo‐Moreno</author><author>Mossab Y Saeed</author><author>Adam Mauskapf</author><author>Ellen T Roche</author>
				</bibl>
			</sourceDesc>
		</fileDesc>
		<profileDesc>
			<abstract><ab><![CDATA[<title>Abstract</title> <p>Heart failure with preserved ejection fraction (HFpEF) is a major challenge in cardiovascular medicine, accounting for ≈50% of all cases of heart failure. Despite the ongoing efforts, no medical device has yet received FDA approval. This is largely due to the lack of an in vivo model of the HFpEF hemodynamics, resulting in the inability to evaluate device effectiveness in vivo prior to clinical trials. Here, the development of a highly tunable porcine model of HFpEF hemodynamics is described using implantable soft robotic sleeves, where controlled actuation of a left ventricular and an aortic sleeve can recapitulate changes in ventricular compliance and afterload associated with a broad spectrum of HFpEF hemodynamic phenotypes. The feasibility of the proposed model in preclinical testing is demonstrated by evaluating the hemodynamic response of the model post‐implantation of an interatrial shunt device, which is found to be consistent with findings from in silico studies and clinical trials. This work overcomes limitations of prior HFpEF models, such as low hemodynamic accuracy, high costs, and long development phases. The versatile and adjustable platform introduced can transform HFpEF device development, aiming to enhance the lives of the 32 million people affected globally.</p>]]></ab></abstract>
		</profileDesc>
	</teiHeader>
	<text><body xmlns="http://www.tei-c.org/ns/1.0" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:xlink="http://www.w3.org/1999/xlink">
<div xmlns="http://www.tei-c.org/ns/1.0"><head>INTRODUCTION</head><p>Heart failure with preserved ejection fraction (HFpEF) is a clinical syndrome in which patients have signs and symptoms of heart failure despite a normal left ventricular ejection fraction (LVEF; &#8805; 50%). <ref type="bibr">[1]</ref> HFpEF accounts for approximately 50% of all patients with heart failure, resulting in over 3 million people with this condition in the US and up to 32 million globally. <ref type="bibr">[2,</ref><ref type="bibr">3]</ref> Currently, pharmacological therapies have shown limited mortality benefits and no medical device has yet been approved by the FDA for the treatment of HFpEF. As a result, the mortality associated with HFpEF is higher than that of most cancers. <ref type="bibr">[4,</ref><ref type="bibr">5]</ref> HFpEF encompasses conditions with diverse etiologies, usually with multisystem involvement, entailing cardiac, pulmonary, vascular, metabolic, renal, and/or hepatic abnormalities. <ref type="bibr">[6,</ref><ref type="bibr">7]</ref> Our understanding of the pathophysiological processes that lead to these changes remains incomplete and hypotheses are constantly evolving. From a biomechanical standpoint, HFpEF is characterized by left ventricular (LV) stiffening or loss of compliance, hindering its ability to relax during diastole and reducing its capacity to fill. <ref type="bibr">[8]</ref> Loss of LV compliance can result from a variety of conditions, including metabolic disease, microvascular inflammation, restrictive, infiltrative, and hypertrophic cardiomyopathies, or can be secondary to pressure overload (or increased afterload) due to hypertension or aortic stenosis. <ref type="bibr">[9]</ref><ref type="bibr">[10]</ref><ref type="bibr">[11]</ref><ref type="bibr">[12]</ref><ref type="bibr">[13]</ref><ref type="bibr">[14]</ref><ref type="bibr">[15]</ref> In the context of pressure overload, concentric remodeling or thickening of the LV is thought to ensue as a compensatory mechanism to minimize changes in LV wall stress, governed by the law of Laplace. <ref type="bibr">[16]</ref> Such loss of LV compliance causes an upward shift of the end-diastolic pressure-volume relationship (EDPVR), which results in a drop in the LV filling volumes, or end-diastolic volume (EDV), and higher filling pressures or LV end-diastolic pressure (LVEDP). <ref type="bibr">[17]</ref> As a result, the amount of blood pumped during each heartbeat, or stroke volume (SV), is diminished, driving a drop in cardiac output (CO) and the inability of the heart to meet the metabolic demands of the body. Evidence of increased filling pressures (LVEDP, pulmonary capillary wedge pressure, or right atrial pressure) at rest or with exercise is required for a clinical diagnosis of HFpEF. <ref type="bibr">[18]</ref> Elevated LV filling pressures are transmitted retrogradely to the left atrium (LA) and the pulmonary circulation, resulting in a multitude of manifestations and complications of HFpEF. First, higher atrial pressures drive LA remodeling processes, which cause atrial fibrillation in 40-60% of the patients, increasing the risk of embolic stroke and rhythm abnormalities remarkably. <ref type="bibr">[19]</ref> Second, high pressures are transmitted to the pulmonary circulation, causing pulmonary hypertension (up to 80%) as well as symptoms of pulmonary edema and dyspnea on exertion, and eventually to the right heart, leading to right ventricular (RV) dysfunction (up to 50%) and other complications. <ref type="bibr">[20,</ref><ref type="bibr">21]</ref> Further, symptom management is often complicated by autonomic imbalance, characterized by the upregulation of sympathetic activity with withdrawal of the vagal tone, further increasing the risk for atrial fibrillation and chronotropic incompetence. <ref type="bibr">[22,</ref><ref type="bibr">23]</ref> First-line therapy for HFpEF includes sodium-glucose cotransporter type 2 (SGLT2) inhibitors. These drugs have been found to decrease the combined risk of cardiovascular death or the first heart failure hospitalization. Nevertheless, the exact mechanism by which they confer these benefits is not fully understood, and they have not been shown to significantly reduce all-cause mortality. <ref type="bibr">[24]</ref><ref type="bibr">[25]</ref><ref type="bibr">[26]</ref> Loop diuretics are also prescribed in patients with overt congestion and aerobic exercise is broadly recommended. <ref type="bibr">[2,</ref><ref type="bibr">27]</ref> In addition, clinical management aims to address any known underlying causes or complications of HFpEF. <ref type="bibr">[28]</ref> In tandem with the search for pharmacologic therapies that can improve the survival and quality of life of HFpEF patients, several of which are currently undergoing clinical trials, substantial efforts have been made toward the development of device-based solutions. <ref type="bibr">[29,</ref><ref type="bibr">30]</ref> Medical devices for HFpEF seek to restore adequate cardiac mechanics or hemodynamics through a variety of mechanisms. The primary classes of devices for HFpEF include atrial shunts, LV expanders, and mechanical circulatory support (MCS) devices.</p><p>Atrial shunts are designed to lower elevated LA pressures by shunting blood from the LA to lowerpressure structures, such as the right atrium (RA) or the coronary sinus. LA-to-RA interatrial shunts include the interatrial shunt device (IASD, Corvia Medical Inc), the V-Wave shunt (V-Wave Ltd), and the atrial flow regulator (AFR) device (Occlutech). These devices are made of a Nitinol frame or mesh of various geometries that creates a 6-10 mm opening between the LA and RA to shunt blood down its pressure gradient. <ref type="bibr">[31]</ref><ref type="bibr">[32]</ref><ref type="bibr">[33]</ref> Recently, studies have investigated the feasibility of device-free interatrial shunting. <ref type="bibr">[34]</ref> An alternative design by Edwards LifeSciences (Transcatheter Atrial Shunt system) aims to prevent RA overload by creating a 7-mm shunt between the LA and the coronary sinus. <ref type="bibr">[35]</ref> Another class of devices is that of LV expanders, which store mechanical energy during contraction -transferring it to the LV during diastole -seeking to augment LV filling capacity. Two main types of LV expanders have been designed to date, namely the ImCardia (CorAssist Inc.) device, which is an extracardiac expander designed for implantation on the epicardial surface, now withdrawn due to safety concerns, and the endocardial CORolla (CorAssist Inc.) device for transapical implantation. <ref type="bibr">[36]</ref> Finally, MCS strategies involve pumps to reduce LA pressures and increase CO. <ref type="bibr">[37]</ref> Examples of MCS devices include the PulseVad by Northern Development, which is a minimally invasive smart pump aiming to enable adaptive flow from the LA to the descending aorta, the left atrial assist device (LAAD) -a continuous-flow LAto-LV pump designed to be implanted at the level of the mitral valve, and the CoPulse device -a type of valveless pulsatile pneumatic pump made of a flexible polyurethane membrane that is designed for apical implantation. <ref type="bibr">[38,</ref><ref type="bibr">39]</ref> In addition, studies have been conducted to investigate the feasibility of traditional LVAD devices for use in HFpEF patients, although concerns remain over the risk of LA suction or inflow cannula obstruction due to concentric LV remodeling. <ref type="bibr">[40]</ref> These device-based solutions for HFpEF are all at different stages of development, with interatrial shunts having been recently marketed for use in the European Union, the CORolla LV expander undergoing clinical trials, and MCS devices in various preclinical testing phases. However, to date, none of these devices have received FDA approval. <ref type="bibr">[30]</ref> A major limitation to the development of these devices and their market approval in the US is the lack of an adequate in vivo model of the HFpEF hemodynamics. <ref type="bibr">[41,</ref><ref type="bibr">42]</ref> As a result, preclinical safety and functional evaluation of devicebased solutions for HFpEF can only be conducted on healthy animals, considerably hindering their reliability. <ref type="bibr">[36,</ref><ref type="bibr">43]</ref> Current large animal models of HFpEF suffer from a variety of limitations, including their limited hemodynamic fidelity, elevated costs, lengthy development times, and low throughput. These models can only rarely successfully re-create the elevations in filling pressures associated with HFpEF, require follow-ups ranging between 6-24 weeks, and are affected by mortality rates of 30% or higher. <ref type="bibr">[42]</ref> Most of these models aim to recapitulate the structural and functional changes associated with the disease by inducing pressure overload by means of an aortic cuff, band, or stent. <ref type="bibr">[44]</ref><ref type="bibr">[45]</ref><ref type="bibr">[46]</ref> Another method to induce pressure overload involves inducing renovascular hypertension via renal wrapping, renal clamping, renal embolization, or via the administration of deoxy-corticosterone acetate (DOCA) salt. <ref type="bibr">[47]</ref><ref type="bibr">[48]</ref><ref type="bibr">[49]</ref> High-fat diets are often leveraged as a supplemental method for these models, aiming to recapitulate changes in metabolic function associated with HFpEF. <ref type="bibr">[50]</ref> Another method currently under development involves reducing LV filling by the use of an intraventricular balloon; however, this technique is still lacking a postoperative functional evaluation and may induce systolic dysfunction, which is only rarely associated with HFpEF. <ref type="bibr">[42]</ref> Because of all these limitations, no such models are currently being used for the evaluation of medical devices for HFpEF.</p><p>In this work, we describe the development of a large animal model of HFpEF hemodynamics that can recapitulate acute changes in LV biomechanics and afterload associated with disease in a tunable and controllable manner. The model is enabled by two implantable soft robotic sleeves; an LV sleeve that re-creates the loss of LV compliance characteristic of HFpEF, and an aortic sleeve that can recapitulate changes in the afterload associated with aortic stenosis, often leading to HFpEF (Fig. <ref type="figure">1A</ref>). <ref type="bibr">[51]</ref> In previous studies, we demonstrated the ability of the LV sleeve to limit cardiac filling in an in vitro model of aortic stenosis and that of the aortic sleeve to recreate the hemodynamics of pressure overload in vitro and in vivo. <ref type="bibr">[52,</ref><ref type="bibr">53]</ref> Here, we redesign the LV sleeve for the development of the first porcine model of HFpEF capable of tuning pressure overload and ventricular compliance independently, in a facile, immediate, tunable manner, which allows us to re-create a broad spectrum of HFpEF hemodynamics. By investigating the hemodynamic response of the model post-implantation of an interatrial shunt device, we then demonstrate the applicability of this model for in-vivo evaluation of device-based solutions for HFpEF (Fig. <ref type="figure">1B</ref>). We believe that this proposed porcine model of HFpEF hemodynamics can establish a new standard for the development of device-based solutions for HFpEF, addressing an unmet need in cardiovascular medicine by supporting translational efforts towards the development of more effective treatment strategies for people with HFpEF. </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>MATERIALS AND METHODS</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Study design</head><p>This study involved the design of an implantable LV sleeve and the use of an aortic sleeve in a swine model (Yorkshire, male, ~38-45 kg) for controllable recapitulation of HFpEF hemodynamics. The sleeves were sized using our cardiac magnetic resonance imaging database on swine within weight range and manufactured using materials and techniques detailed below. The sleeves were implanted in n = 3 swine and actuated using hydraulic pressure. The model hemodynamics were evaluated for i) two distinct actuation levels of the aortic sleeve (AoS), ii) three distinct levels of the LV sleeve (LVS), and iii) six combined levels, leading to a total of eleven hemodynamic scenarios. In each case, actuation was carried out at a constant infusion rate and held until stable hemodynamics were reached. On a subset of swine (n = 2), medical intervention was then simulated by percutaneous creation of a device-free interatrial shunt and implantation of an interatrial shunt device. Invasive hemodynamic monitoring was performed through LV and </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Sleeve design and manufacturing</head><p>The LV sleeve was designed on SolidWorks (2019, Dassault Syst&#232;mes) using digital anatomies obtained from cardiac magnetic resonance images swine studies. The outer surface of the heart was offset by 10 mm to generate the sleeve geometry, which was then divided into four circumferential quadrants (each approximately 90 degrees apart), two for each ventricle. These quadrants were flattened to a plane to create the contours of the molds for manufacturing of the two Thermoplastic Polyurethane (TPU) pockets for LV actuation. The design of the two molds was obtained by extrusion of the relative quadrant by the same offset (10 mm), and each mold was then 3D-printed using a rigid photopolymer (Veroblue, Stratasys) with an inkjet-based Objet30 3D-printer (Stratasys).</p><p>For each of the two LV molds, two TPU sheets (TPU, HTM 8001-M 80A shore polyether film, 0.012" thick, American Polyfilm) were vacuum-formed (Dental Vacuum Former, Yescom) to the shape of the molds. Each pair of TPU sheets was then heat-sealed at 320F for 8 seconds on a heat-press transfer machine using laser-cut negative acrylic molds to create enclosed and inflatable geometries. The edges of the two inflatable pockets were then heat-sealed to a 200-Denier TPU-coated fabric (Oxford fabric, Seattle fabrics Inc.), which was designed to be wrapped around the entire cardiac anatomy. Adjustable Velcro straps were added to the RV edges of the fabric to enable fastening. Two small openings were created through the fabric on one side of each of the two pockets to connect soft tubes (latex rubber 1/16" inner diameter 1/8" outer diameter tubing, McMaster-Carr) as actuation lines through PVC connectors (polycarbonate plastic double-barbed tube fitting for 1/16" tube ID, McMaster-Carr). The soft robotic aortic sleeve was designed and manufactured using a similar approach, previously described by our group. <ref type="bibr">[53]</ref> </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Surgical procedures and sleeve implantation</head><p>All animal procedures were approved by the Institutional Animal Care and Use Committee of our institute. In vivo studies were conducted on n = 3 Yorkshire swine housed in the Massachusetts Institute of Technology Department of Comparative Medicine Swine Facility.</p><p>After induction of anesthesia and endotracheal intubation, median sternotomy was performed using an oscillating saw. Dissection of the aortopulmonary window allowed for implantation of the aortic sleeve around the ascending aorta approximately 1-2 cm distal to the native aortic valve.</p><p>The sleeve was tightened and secured using sutures. The LV sleeve was then fastened around the heart, taking care to orient the inflatable pockets around the LV, and tightened. The sleeve was secured with silk stay sutures to the pericardium to prevent downward translation during actuation.</p><p>Hemodynamic monitoring devices were placed using both endovascular and open-surgical access approaches. A vascular flow probe (ME 13 PXN, Transonic Systems Inc.) was placed around the descending thoracic aorta, and two pressure-volume (PV) catheters (Transonic Systems Inc.) were used to perform PV loop characterization with one 5F straight catheter inserted through femoral arterial access into the aortic arch and a 5F pigtail catheter placed transapically into the LV. Vital signs were monitored for the entire duration of the study.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>In vivo sleeve actuation</head><p>Actuation of the sleeves was performed using a syringe pump (70-3007 PHD ULTRA Syringe Pump Infuse/Withdraw, Harvard Apparatus). Three actuation volumes were used for the LV sleeve [mild (L1) = 40-45 mL, moderate (L2) = 50-55 mL, severe (L3) = 60-65 mL] and two for the aortic sleeve [moderate (A1) = 3.25-3.5 mL, severe (A2) = 4-4.25 mL]. The period of actuation was approximately 40 and 5 seconds for the LV and aortic sleeve, respectively. A stopcock was used to maintain a given actuation level of each sleeve throughout the duration of the study. The sleeves were actuated in isolation (A1-2, L1-L3) and in a combined fashion (A1L1, A1L2, A1L3, A2L1, A2L2, A2L3), leading to a total of 11 actuation modalities. The evaluation of interatrial shunting as an interventional approach for HFpEF was conducted on two representative cases, referred to as moderate (A1L2) and severe (A2L3).</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Interatrial shunting</head><p>Intervention was simulated in n = 2 swine. To this end, a device-free interatrial shunt was created percutaneously in one animal and a 6-mm AFR (Occlutech) was implanted in another animal. To create the interatrial shunt, ultrasound-guided access to the right femoral vein was obtained using the Seldinger technique. Under fluoroscopic guidance (OEC 9800), an 0.035" Bentson wire (Cook Medical) was then advanced into the RA. An 8.5Fr transseptal sheath (SJM Swartz, Abbott Laboratories) was then positioned in the RA and then across the interatrial septum into the LA. The wire was then placed into the left inferior pulmonary vein, and an 8-mm balloon (EverCross, Medtronic) was used to dilate the shunt. Once the atrial septal defect was created, deployment of the 6-mm AFR was performed under fluoroscopic guidance, using the manufacturer's delivery system and pusher. Shunt patency was confirmed using epicardial color Doppler echocardiography (Philips EPiQ CVx with X5-1 transducer).</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Data acquisition</head><p>LV and aortic hemodynamics were monitored throughout the study following placement of the 5F PV catheters (Transonic, Inc.) in the LV and ascending arch, and the vascular flow probe on the descending thoracic aorta (Transonic, Inc.). The PV catheters were connected to the ADV500 PV system (Transonic Inc.) for measurements of LV pressures (LVP), LV volumes (LVV), and aortic pressures (AoP). SV estimates of each animal were provided as inputs to the PV system for calibration, as obtained using trans-epicardial echocardiography (B-mode) on the EpiQ CVx system (Philips) prior to implantation of the sleeves. The vascular flow probe was connected to a two-channel flowmeter console (400 series, Transonic Inc.). Both the PV and flowmeter consoles were, in turn, connected to an eight-channel Powerlab system (ADInstruments) for real-time data acquisition with a 400-Hz sampling rate and recording. During the experiments, data were continuously monitored via LabChart software (Pro v8.1.16, ADInstruments). The default 50-Hz bandstop filter was applied to all signals. After the experiments, data were exported and processed in Matlab R2020a (MathWorks).</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Hemodynamic evaluation</head><p>Clinical metrics for the assessment of diastolic dysfunction and pressure overload were calculated for hemodynamic evaluation of the proposed model. The LVPmax, LVEDP, EDV, and end-systolic volumes (ESV) were measured directly from the LVP and LVV tracings. Equations 1-3 were used for calculation of the SV, LVEF, and CO:</p><p>&#119862;&#119874; = &#119878;&#119881; * &#119867;&#119877; (3) where HR indicates the heart rate in beats per minute (bpm).</p><p>For the assessment of pressure overload, the mean transaortic pressure gradient (dPmean) was calculated as the mean difference between the LVP and the AoP during cardiac ejection, each measured by the corresponding PV catheter.</p><p>The indexed effective orifice area (iEOA) is calculated as in Equation <ref type="formula">4</ref>, using the Gorlin equation: <ref type="bibr">[54]</ref> iEOA = Q BSA 51.6 &#8730;dP mean (4)   where Q is the flow through the aortic valve measured using the PV catheter in the LV and BSA indicates the body surface area of the animal.</p><p>The valvulo-arterial impedance (ZVA) was measured using Equation <ref type="formula">5</ref>: <ref type="bibr">[55]</ref> Z VA = LVP max SV BSA (5)   In Equations 4-5, the BSA (in m 2 ) was estimated from the body weight (BW) (in kg) of the swine, using the Kelley formula (Equation <ref type="formula">6</ref>) <ref type="bibr">[56]</ref> : BSA = 0.0734 BW 0.656 (6).</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Data analysis</head><p>All data were processed and analyzed on Matlab R2020a (Mathworks). Steady-state average and standard deviation values were obtained over ten consecutive heartbeats for each condition. Statistical significance between actuation levels was determined with respect to using two-tailed t-tests with a 95% confidence interval (P&lt;0.05) on Matlab R2020a.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>RESULTS</head><p>Left ventricular sleeve can modulate cardiac compliance to re-create HFpEF hemodynamics in a porcine model.</p><p>We designed an LV sleeve that can be implanted around the epicardial surface of the heart in a swine model to modulate cardiac filling function. The sleeve (Fig. <ref type="figure">2A</ref>) is composed of two inflatable pockets made of TPU (HTM 8001-M 80A shore polyether film, 0.012 inches in thickness; American Polyfilm Inc.) that expand under pressure and of a 200-Denier inelastic TPU-coated fabric (Oxford fabric, Seattle fabrics Inc.) that directs the expansion of the TPU pockets towards the LV and enables anchoring of the sleeve onto the epicardial surface. The sleeve is designed to be implanted so that the inflatable pockets are in contact with the outer wall of the LV, whereas the rest of the fabric wraps around the RV. Velcro straps were used to secure the fabric tightly around the epicardial wall. The TPU pockets were designed to conform with the LV anatomy; they consist of a double layer of TPU sheets that are first vacuum formed to 3D printed (Veroblue, Objet 30 3D printer, Stratasys) molds, and then heat-sealed at their edges to create a fully enclosed compartment. The pockets were then heat-sealed to the fabric and connected to their respective actuation line (latex rubber 1/16-inch inner diameter and 1/8-inch outer diameter tubing; McMaster-Carr). Actuation of the LV sleeve results in a drop in LV compliance indicated as an upward shift of the end-diastolic pressure-volume relationship (EDPVR) in the LV PV loop (Fig. <ref type="figure">2A</ref>).</p><p>In a total of n = 3 swine, the LV sleeve was actuated at various volumes, using a syringe pump over a period of approximately 40 seconds (see Materials and Methods section). Fig. <ref type="figure">2</ref> B-C show representative changes in LVP and LVV, respectively, achieved at moderate and severe actuation compared to baseline. Notably, overlapping AoP tracings (Fig. <ref type="figure">2B</ref>) show that actuation of the LV sleeve alone does not have a significant effect on aortic hemodynamics. The dotted lines in Fig. <ref type="figure">2B</ref> highlight a progressive increase in the filling pressures, whereas Fig. <ref type="figure">2C</ref> shows a corresponding drop in the filling volumes or maximum LVV. The progression of the LV PV loop from baseline to severe actuation corroborates these changes and confirms the expected upward shift of the EDPVR, as an indication of limited LV filling induced by actuation of the LV sleeve (Fig. <ref type="figure">2D</ref>). Fig. <ref type="figure">2</ref> E-F illustrates the evolution of these hemodynamic changes averaged across the n = 3 trials, during moderate and severe actuation of the LV sleeve. From a baseline value of LVEDPBL = 13.3 &#61617; 1.2 mmHg, the LVEDP increased to 18.9 &#61617; 1.7 mmHg and 21.9 &#61617; 1.8 mmHg at moderate and severe actuation, respectively. Correspondingly, actuation caused a drop in the EDV equal to &#61508;EDVM = -7.1 &#61617; 1.6 % and &#61508;EDVS = -11.0 &#61617; 1.6 %. Overall, actuation of the LV sleeve alone was shown to recapitulate changes in LV compliance, which is the primary hemodynamic hallmark of HFpEF. 3T 6T 9T Time 20 40 60 80 100 Pressure (mmHg) 0 50 100 LVV (mL) 0 50 100 LVP (mmHg) 0 20 40 Time (s) -10 -5 0 EDV (%) 0 20 40 Time (s) 15 20 25 LVEDP (mmHg) A C B 3T 6T 9T Time 20 40 60 80 LVV (mL) BL Moderate Severe EDPVR D E F BL Severe Moderate Severe BL Moderate Severe LV sleeve Primary effect Inelastic fabric Inflatable pockets [mmHg] [mL] EDPVR &#8593; Compliance &#8595; LVP AoP</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Aortic sleeve can recapitulate pressure overload driving HFpEF</head><p>Being able to induce changes in the afterload in conjunction with LV filling modulation is critical to recapitulate a broader spectrum of HFpEF hemodynamic phenotypes. To this end, we integrated our in vivo model of HFpEF with a soft robotic aortic sleeve that can be controlled to re-create elevations in afterload often responsible for or associated with HFpEF. Here, we describe the changes in the LV and aortic hemodynamics induced by the aortic sleeve only, before characterizing the hemodynamics resulting from combining both sleeves in the following section. The aortic sleeve is made of three inflatable TPU pockets that expand under pressure to compress the ascending aorta, inducing pressure overload in a controllable fashion (Fig. <ref type="figure">3A</ref>). Analogous to the LV sleeve, the three pockets are connected to an actuation line and to an inelastic fabric layer that directs the expansion of the pockets inwards and secures the sleeve around the aortic anatomy (Fig. <ref type="figure">3A</ref>). The materials and manufacturing workflow, involving vacuum forming and heat sealing, of the aortic sleeve are the same as those of the LV sleeve. The aortic sleeve was actuated using the same syringe pump over a period of 5 seconds at two distinct actuation levels, namely moderate and severe actuation (see Materials and Methods section).</p><p>Fig. <ref type="figure">3 B-C</ref> shows representative changes in LVP and AoP waveforms induced by actuation of the aortic sleeve. Fig. <ref type="figure">3B</ref> highlights these changes from baseline to moderate and severe actuation in a discrete fashion, whereas the evolution of the waveforms over time is visualized in Fig. <ref type="figure">3C</ref>. Consistently, these graphs show that actuation of the aortic sleeve leads to elevations in the systolic LVP and a reduction of the AoP, establishing a pressure gradient between the two. Evaluation of hemodynamic parameters used for clinical assessment of aortic stenosis showed a progressive drop in the indexed effective orifice area (iEOABL = 1.09 &#61617; 0.19 cm/m 2 , iEOAM = 0.71 &#61617; 0.09 cm/m 2 , iEOAS = 0.20 &#61617; 0.05 cm/m 2 ; Fig. <ref type="figure">3D</ref>) caused by controlled aortic constriction. This led to an increase in the mean transaortic pressure gradient (dP mean_BL = 2.7 &#61617; 1.3 mmHg, dP mean_M = 36.4 &#61617; 3.2 mmHg, dPmean_S = 54.6 &#61617; 3.7 mmHg; Fig. <ref type="figure">3E</ref>) and in elevations in the valvulo-arterial impedance (ZVA_BL = 1.9 &#61617; 0.1 mmHg m 2 /mL, ZVA_M = 3.3 &#61617; 0.6 mmHg m 2 /mL, ZVA_S = 5.5 &#61617; 0.5 mmHg m 2 /mL; Fig. <ref type="figure">3F</ref>).</p><p>Representative changes in cardiac function during actuation of the aortic sleeve are shown in Fig. <ref type="figure">3</ref> G-I. As predicted, controlled aortic constriction led to a rightward shift of the LV PV loop, with an increase in systolic LVP and a drop in the SV (Fig. <ref type="figure">3G</ref>). Quantitatively, moderate and severe actuation led to changes in LVP and SV equal to &#61508;LVPmax_M = 47.6 &#61617; 6.0 % and &#61508;LVPmax_S = 82.4 &#61617; 11.8 % (Fig. <ref type="figure">3H</ref>) and &#61508;SVM = -29.7 &#61617; 7.0 % and &#61508;SVS = -43.9 &#61617; 5.9 % (Fig. <ref type="figure">3I</ref>). </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Soft robotic sleeves can be modulated to recapitulate a spectrum of HFpEF hemodynamics</head><p>Through the combination of the LV and aortic sleeves, our porcine model of HFpEF can re-create various hemodynamic phenotypes of HFpEF, which will allow us to capture a broad spectrum of HFpEF etiologies and disease severities. The illustration in Fig. <ref type="figure">4A</ref> shows representative combinations of the LV and aortic sleeve actuation schemes that can be leveraged to recapitulate various HFpEF hemodynamics. When the aortic sleeve is off, actuation of the LV sleeve can capture the hemodynamics of limited LV filling associated with HFpEF in the absence of pressure overload, for example, relevant for patients with restrictive, infiltrative, or hypertrophic cardiomyopathies. Conversely, actuation of both sleeves enables afterload and LV compliance modulation for the recapitulation of the hemodynamics of HFpEF caused by (or comorbid with) aortic stenosis or hypertension. The schematic in Fig. <ref type="figure">4A</ref> also shows that the actuation levels of the sleeves can be modulated individually to recreate various severities of disease.</p><p>Fig. <ref type="figure">4</ref> B-D illustrates hemodynamic changes induced by actuation of the aortic sleeve only (AoS), then followed by actuation of both sleeves (Ao+LVS). This allowed us to investigate the ability of our model to simulate the hemodynamic progression of patients with HFpEF secondary to pressure overload in a longitudinal manner. Fig. <ref type="figure">4 B-C</ref> shows that actuation of the LV following actuation of the aortic sleeve can cause an increase in filling pressures and a drop in filling volume, even in the context of pressure overload. Further, Fig. <ref type="figure">4B</ref> highlights that the systolic LVP may drop slightly due to actuation of the LV sleeve, which is a consequence of the decrease in filling volumes (or preload). Similarly, actuation of the LV sleeve induces a considerable drop in both EDV and ESV, as seen in Fig. <ref type="figure">4C</ref>. Overall, the SV is reduced in Ao+LVS compared to baseline, which is consistent with the effects of limited cardiac filling associated with HFpEF. These changes can be visualized in Fig. <ref type="figure">4D</ref>, which shows the progression of PV loops from baseline to pressure overload only (AoS), and finally to both pressure overload and limited cardiac filling (Ao+LVS).</p><p>The graphs in Fig. <ref type="figure">4 E-G</ref> showcase the broad spectrum of hemodynamic characteristics that could be obtained by selective and independent actuation of the LV and aortic sleeves. Changes in dPmean in Fig. <ref type="figure">4E</ref> show that the degree of pressure overload can be reliably controlled by the aortic sleeve (dPmean_Off_L1 = 4.9 &#61617; 0.4 mmHg, dPmean_A1L1 = 21.1 &#61617; 1.8 mmHg, dPmean_A2L1 = 45.3 &#61617; 3.7 mmHg, P &lt; 0.001; Fig. <ref type="figure">4E</ref>) and that progressively reducing filling capacity through actuation of the LV sleeve does not cause significant alterations to it (dPmean_A2L3 = 45.7 &#61617; 3.6 mmHg; Fig. <ref type="figure">4E</ref>). Actuation of the LV sleeve resulted in an increase in the LVEDP (LVEDPOff_L1 = 14.2 &#61617; 1.3 mmHg, LVEDPOff_L3 = 22.3 &#61617; 1.7 mmHg, P &lt; 0.01; Fig. <ref type="figure">4F</ref>). When combined with pressure overload, and for the same actuation level of the LV sleeve, these changes become slightly more pronounced (LVEDPA2L1 = 15.5 &#61617; 1.7 mmHg, LVEDPA2L3 = 25.6 &#61617; 2.9 mmHg, P &lt; 0.01; Fig. <ref type="figure">4F</ref>). Similarly, a reduction in end-diastolic volumes ranging from 2-10% can be achieved, depending on the level of actuation (&#61508;EDVOff_L1 = -2.4 &#61617; 1.5 %, &#61508;EDVOff_L3 = -10.9 &#61617; 4.5 %, P &lt; 0.05; Fig. <ref type="figure">4F</ref>). Because pressure overload is also associated with an increase in EDV (Fig. <ref type="figure">4F</ref>) when the LV sleeve is used in tandem with the aortic sleeve, its actuation level can be adjusted to obtain a desired change in EDV (&#61508;EDVOFF_L1 = -2.4 &#61617; 1.5 %, &#61508;EDVA2L2 = -2.9 &#61617; 3.2 %; Fig. <ref type="figure">4F</ref>). These data demonstrate that by varying the actuation levels independently for each sleeve, this model can re-create a variety of hemodynamic profiles of relevance for studies of HFpEF and device testing. </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>In vivo HFpEF hemodynamic model enables high-fidelity testing of device-based solutions</head><p>Fine control over the hemodynamic characteristics of HFpEF in a preclinical model enables highfidelity testing of medical devices for this condition. To demonstrate feasibility in device testing, we simulated HFpEF intervention by creating a device-free atrial septal defect (ASD) and by implantation of an interatrial shunt device (Fig. <ref type="figure">5A</ref>). Details of the interventional approaches can be found in the Materials and Methods section. Fluoroscopy and color Doppler imaging confirmed the correct positioning and patency of the interatrial shunt device (Fig. <ref type="figure">5B</ref>, Supplementary Video 1). Competence of the aortic, mitral, and tricuspid valves was confirmed via color Doppler imaging (Supplementary Video 2).</p><p>The hemodynamic effects of interatrial shunting on LV unloading can be visualized via the LVP and AoP waveforms and LV PV loops in Fig. <ref type="figure">5</ref> C-D. Both graphs show that shunting restores physiologic levels of filling pressures and that it causes a reduction in the systolic LVP. This occurs because interatrial shunting results in a drop in the preload, which causes the LV PV loop to move down its EDPVR curve, leading to a reduction in both peak systolic and end-diastolic pressures (Fig. <ref type="figure">5D</ref>).</p><p>The drop in systolic LVP causes a decrease in the dPmean, which is more prominent when severe disease (both pressure overload and limited filling) is simulated (dPmean_S = 45.7 &#61617; 3.6 mmHg, dPmean_S+SH = 40.8 &#61617; 0.4 mmHg; Fig. <ref type="figure">5E</ref>) compared to moderate disease (dPmean_M = 20.8 &#61617; 3.7 mmHg, dPmean_M+SH = 22.4 &#61617; 1.0 mmHg; Fig. <ref type="figure">5E</ref>). LA-to-RA shunting restored levels of the LVEDP close to baseline (LVEDPBL = 13.1 &#61617; 2.4 mmHg, LVEDPS = 25.6 &#61617; 2.9 mmHg, LVEDPS+SH = 17.1 &#61617; 0.2 mmHg; Fig. <ref type="figure">5F</ref>). Further, varying the level of disease severity was shown to influence the LVEDP measured after intervention, where higher severity levels led to slightly elevated values of LVEDP (LVEDPM+SH = 14.4 &#61617; 1.1 mmHg, LVEDPS+SH = 17.1 &#61617; 0.2 mmHg; Fig. <ref type="figure">5F</ref>). Analogously, the reduction in EDV induced by shunting was slightly more prominent for mild disease (&#61508;EDVM = -5.4 &#61617; 2.9 %, &#61508;EDVM+SH = -11.5 &#61617; 4.0 %; Fig. <ref type="figure">5G</ref>) than for the simulated severe condition (&#61508;EDVS = -7.3 &#61617; 3.2 %, &#61508;EDVS+SH = -10.4 &#61617; 2.1 %; Fig. <ref type="figure">5G</ref>).</p><p>Our model was able to predict changes in other metrics of cardiac function, including SV, LVEF, and CO, associated with intervention (Fig. <ref type="figure">5 H-J</ref>). Due to a reduction in both the EDV and the ESV, we measured further drops in SV caused by shunting both when simulating mild (&#61508;SVM = -16.0 &#61617; 9.2 %, &#61508;SVM+SH = -25.9 &#61617; 0.9 %; Fig. <ref type="figure">5H</ref>) and severe (&#61508;SVS = -19.5 &#61617; 11.5 %, &#61508;SVS+SH = -28.2 &#61617; 2.7 %; Fig. <ref type="figure">5H</ref>) HFpEF hemodynamics. This led to a slight increase in the LVEF due to shunting for both disease levels (LVEFM = 45.2 &#61617; 3.7 %, LVEFM+SH = 49.2 &#61617; 5.5 % LVEFS = 44.1 &#61617; 3.4 %, LVEFS+SH = 46.3 &#61617; 1.9 %; Fig. <ref type="figure">5I</ref>). Finally, we observed an increase, albeit modest, in CO (COM = 4.5 &#61617; 0.4 L/min, COM+SH = 4.6 &#61617; 0.2 L/min, COS = 4.5 &#61617; 0.8 L/min, COS+SH = 4.6 &#61617; 0.1 L/min; Fig. <ref type="figure">5J</ref>). This is likely due to chronotropic compensation, as apparent from elevations in heart rate induced by shunting (Fig. <ref type="figure">5C</ref>). </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>DISCUSSION</head><p>This work presents the first large animal model of HFpEF hemodynamics for use in preclinical testing of medical devices for HFpEF. The proposed model is based on implantable soft robotic sleeves that can be controllably actuated to finely re-create changes in preload and afterload associated with (or leading to) HFpEF.</p><p>We describe the design, manufacturing, and development of a soft robotic LV sleeve to be implanted around the heart in a swine model to modulate LV compliance and filling capacity (Fig. <ref type="figure">2A</ref>). The sleeve is made of pre-formed actuatable TPU pockets in contact with the epicardial surface of the LV. When pressurized, these pockets inflate to mechanically limit LV filling, leading to a controlled increase in LVEDP and a drop in the EDV and SV (Fig. <ref type="figure">2 D-F</ref>), which represent the main hemodynamic signature of HFpEF. Elevations in the afterload (associated with up to 90% of cases of HFpEF) <ref type="bibr">[57]</ref> were recapitulated using controlled aortic constriction using a soft robotic aortic sleeve (Fig. <ref type="figure">3A</ref>). Analogous to the LV sleeve, the aortic sleeve is made of expandable TPU pockets attached to an inelastic fabric layer for implantation around the wall of the ascending aorta. By calculating diagnostic metrics of aortic stenosis (Fig. <ref type="figure">3 D-F</ref>) and cardiac function (Fig. <ref type="figure">3</ref> G-I), we showed that actuation of the aortic sleeve can recapitulate various degrees of pressure overload that are clinically relevant for HFpEF.</p><p>In a swine model, we implanted both the LV and aortic sleeves and showed that they can be actuated in isolation or in combination to recapitulate a broad spectrum of HFpEF hemodynamic phenotypes (Fig. <ref type="figure">4A</ref>). For each of the simulated conditions, we measured LVP, AoP (Fig. <ref type="figure">4B</ref>), LVV (Fig. <ref type="figure">4C</ref>), LV PV loops (Fig. <ref type="figure">4D</ref>), dPmean (Fig. <ref type="figure">4E</ref>), LVEDP (Fig. <ref type="figure">4F</ref>), and EDV (Fig. <ref type="figure">4G</ref>) to investigate the effects of selective and progressive sleeve actuation on cardiac hemodynamics and confirmed that the changes in the hemodynamics obtained with our model are consistent with the clinical literature of HFpEF. <ref type="bibr">[17]</ref> We showed that actuation of the LV sleeve did not affect changes in the afterload induced by the aortic sleeve (Fig. <ref type="figure">4E</ref>). Conversely, we observed an effect, albeit minor, of aortic sleeve actuation on the modulation of LVEDP and EDV induced by the LV sleeve. Particularly in the context of acute onset of pressure overload, severe elevations in the afterload may cause a rightward shift of the LV PV loop (Fig. <ref type="figure">4D</ref>) characterized by an increase in the LVEDP and EDV. As a result, actuation of the aortic sleeve may augment any increase in LVEDP induced by the LV sleeve (Fig. <ref type="figure">4F</ref>), while attenuating associated drops in EDV (Fig. <ref type="figure">4G</ref>). In this work, we used two representative actuation volumes for the aortic sleeve and three volumes for the LV sleeve, which led to a total of eleven conditions being simulated; however, the volumes of each sleeve could be adjusted and modulated further to achieve additional hemodynamic states.</p><p>HFpEF represents a huge burden on the US healthcare system, with associated medical costs projected to exceed $25 million in 2030. <ref type="bibr">[58]</ref> The efforts made towards the development of medical devices that can enhance cardiac mechanics in HFpEF patients have been largely hindered by the absence of adequate hemodynamic in vivo models, leading to the lack of FDA-approved device-based solutions for this condition. Most of the preclinical functional evaluation of HFpEF devices report testing on healthy in vivo models and unanimously emphasize the inability to simulate HFpEF hemodynamics as the primary limitation of their studies. <ref type="bibr">[36,</ref><ref type="bibr">43]</ref> There is a broad variety of devices for HFpEF that are currently under development that would likely benefit from a reliable hemodynamic model of HFpEF. As an example of utility in acute evaluation of medical interventions for HFpEF, we demonstrated that our model could predict the hemodynamics associated with implant-free interatrial shunting and of an interatrial shunt device currently in clinical trials and seeking FDA approval (Fig. <ref type="figure">5A-B</ref>). Although evaluating the efficacy of these interventions is not the purpose of this work, we showed that our model could recapitulate the hemodynamic changes associated with these treatment strategies, as predicted by computational models and human studies. <ref type="bibr">[30,</ref><ref type="bibr">34,</ref><ref type="bibr">59,</ref><ref type="bibr">60]</ref> Particularly, we observed that shunting induced a shift of LV hemodynamics down the EDPVR curve due to a drop in preload (Fig. <ref type="figure">5D</ref>), which led to a decrease in LVEDP (Fig. <ref type="figure">5F</ref>), EDV (Fig. <ref type="figure">5G</ref>), and SV (Fig. <ref type="figure">5H</ref>). Further, diminished preload caused by shunting resulted in decreased systolic LVP (Fig. <ref type="figure">5B</ref>) and dPmean (Fig. <ref type="figure">5E</ref>), likely governed by the Frank-Starling mechanism.</p><p>There are several advantages associated with this model. First, by relying on soft robotic actuatable sleeves, this model can recapitulate the hemodynamics of HFpEF in a controllable and highly tunable manner. The effect of each sleeve can be independently modulated by varying their respective actuation volume (or pressure). Having two degrees of freedom allows the models to mimic a wide variety of hemodynamic changes associated with HFpEF, from mild to severe filling dysfunction and pressure overload. As a result, our platform can capture the remarkably heterogeneous spectrum of HFpEF hemodynamics, which are observed clinically. This lends our model to medical device evaluation for several hemodynamic subphenotypes of HFpEF and disease severities. Ultimately, this may lead to more reliable methods for device testing and the identification of subphenotype-or patient-specific device-based solutions for HFpEF. Finally, by acutely recreating the HFpEF hemodynamics, this model does not suffer from the prohibitive costs, lengthy development time, and limited throughput of previously reported large animal models of HFpEF induced by chronic pressure overload and/or metabolic disease.</p><p>One of the limitations of this approach involves its inability to predict changes in the long-term progression of HFpEF as affected by medical intervention. This is because this platform recapitulates the HFpEF hemodynamics using externally applied forces rather than biological processes. As a result, the effects of medical devices on remodeling processes and associated chronic hemodynamic changes cannot be comprehensively captured by this model. Secondly, although we demonstrated the ability to predict the acute hemodynamic effects of medical intervention by creating an LA-to-RA shunt, testing of extracardiac strategies, such as certain types of mechanical circulatory support strategies, may require some adjustments to the design of the LV sleeve, which may depend on the specific type and implantation approach of a given device. Future studies are warranted to further improve the clinical relevance of our proposed approach by exploring further the spectrum of medical devices for HFpEF that can be evaluated using this platform. Through close partnerships with industry, we also aim to conduct additional studies to evaluate the efficacy of device-based solutions under development for HFpEF more comprehensively, by increasing the power of our trials and by recapitulating specific hemodynamic phenotypes of HFpEF and severities of disease.</p><p>In future work, we aim to further enhance the ability of the model to recapitulate human disease. We will do so by embedding flexible sensors in the to enable adequate pre-tensioning around the aortic wall and epicardial surface of the heart and integrating closed-loop feedback to mimic patient-specific hemodynamic states. Further, we will seek to broaden the range of clinical scenarios that can be recapitulated by this model by combining LV compliance modulation, as achieved by LV sleeve actuation, with hemodynamic states commonly associated with HFpEF, such as systemic and pulmonary hypertension and right heart dysfunction. For example, the LV and aortic sleeve designs could be modified to modulate RV compliance and pulmonary vascular resistance, simulating a broader range of hemodynamic states. Since these conditions play a critical role in the pathophysiology of HFpEF and are each associated with a unique regulatory response, it is crucial to accurately represent them. This will ensure a comprehensive evaluation of intervention effectiveness across a wider range of scenarios addressed in this study.</p><p>By describing a method that can recreate a broad spectrum of clinically relevant HFpEF hemodynamics in a highly controllable manner, this work can have a profound impact on current strategies for HFpEF device development and evaluation. This study could bridge the gap between in silico/in vitro testing and human trials for device-based solutions of HFpEF by overcoming the elevated costs and need for substantial resources for preclinical testing in a large animal model of disease, and by providing a better alternative to testing in healthy hemodynamic conditions, that is more reliable and tunable, hence more clinically relevant. We believe that this work represents a paradigm-shifting application of the use of soft robotics in in vivo disease modeling and device testing, which is a step forward towards addressing the lack of adequate treatment for HFpEF and alleviating its burden on healthcare and on the lives of the 32 million people currently suffering from this condition.</p></div></body>