<?xml-model href='http://www.tei-c.org/release/xml/tei/custom/schema/relaxng/tei_all.rng' schematypens='http://relaxng.org/ns/structure/1.0'?><TEI xmlns="http://www.tei-c.org/ns/1.0">
	<teiHeader>
		<fileDesc>
			<titleStmt><title level='a'>The MUSE beamline calorimeter</title></titleStmt>
			<publicationStmt>
				<publisher>Elsevier</publisher>
				<date>07/03/2025</date>
			</publicationStmt>
			<sourceDesc>
				<bibl> 
					<idno type="par_id">10616514</idno>
					<idno type="doi">10.1016/j.nima.2025.170754</idno>
					<title level='j'>Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment</title>
<idno>0168-9002</idno>
<biblScope unit="volume">1080</biblScope>
<biblScope unit="issue">A</biblScope>					

					<author>W Lin</author><author>T Rostomyan</author><author>R Gilman</author><author>S Strauch</author><author>C Meier</author><author>C Nestler</author><author>M Ali</author><author>H Atac</author><author>JC Bernauer</author><author>WJ Briscoe</author><author>A Christopher Ndukwe</author><author>EW Cline</author><author>K Deiters</author><author>S Dogra</author><author>EJ Downie</author><author>Z Duan</author><author>IP Fernando</author><author>A Flannery</author><author>D Ghosal</author><author>A Golossanov</author><author>J Guo</author><author>NS Ifat</author><author>Y Ilieva</author><author>M Kohl</author><author>I Lavrukhin</author><author>L Li</author><author>W Lorenzon</author><author>P Mohanmurthy</author><author>SJ Nazeer</author><author>M Nicol</author><author>T Patel</author><author>A Prosnyakov</author><author>RD Ransome</author><author>R Ratvasky</author><author>H Reid</author><author>PE Reimer</author><author>R Richards</author><author>G Ron</author><author>OM Ruimi</author><author>K Salamone</author><author>N Sparveris</author><author>N Wuerfel</author><author>DA Yaari</author>
				</bibl>
			</sourceDesc>
		</fileDesc>
		<profileDesc>
			<abstract><ab><![CDATA[]]></ab></abstract>
		</profileDesc>
	</teiHeader>
	<text><body xmlns="http://www.tei-c.org/ns/1.0" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:xlink="http://www.w3.org/1999/xlink">
<div xmlns="http://www.tei-c.org/ns/1.0"><head>A B S T R A C T</head><p>The MUon Scattering Experiment (MUSE) was motivated by the proton radius puzzle arising from the discrepancy between muonic hydrogen spectroscopy and electron-proton measurements. The MUSE physics goals also include testing lepton universality, precisely measuring two-photon exchange contribution, and testing radiative corrections. MUSE addresses these physics goals through simultaneous measurement of high precision cross sections for electron-proton and muon-proton scattering using a mixed-species beam. The experiment will run at both positive and negative beam polarities. Measuring precise cross sections requires understanding both the incident beam energy and the radiative corrections. For this purpose, a lead-glass calorimeter was installed at the end of the beam line in the MUSE detector system. In this article we discuss the detector specifications, calibration and performance. We demonstrate that the detector performance is well reproduced by simulation, and meets experimental requirements. </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="1.">Introduction</head><p>The Proton Radius Puzzle was the observation of a significantly smaller proton charge radius determined in muonic hydrogen spectroscopy than had been determined from atomic hydrogen spectroscopy and electron proton scattering <ref type="bibr">[1,</ref><ref type="bibr">2]</ref>. This observation led to strong interest in the proton radius and related physics of potential new forces, lepton universality, and radiative corrections, including two-photon exchange and polarizability. Possible explanations for the puzzle generally fall into the categories of new forces, novel aspects of conventional physics, or issues in experimental extractions of the radius. While there are new electronic measurements of the radius in agreement with the muonic spectroscopy result, the lack of overall agreement among the new results suggests that there are ongoing experimental issues in determining the radius <ref type="bibr">[3]</ref>.</p><p>The MUSE experiment, motivated by a lack of high-precision muonscattering data that address these physics issues, is working towards a sub-percent level, high-precision experiment that simultaneously measures elastic electron-proton scattering and muon-proton scattering. MUSE is located at the Paul Scherrer Institute in Villigen, Switzerland. The High-Intensity Proton Accelerator facility uses a cyclotron to produce an approximately 2 mA, 590 MeV proton beam. The beam passes through the graphite-wheel M target to produce electrons, muons and pions that are transported by the PiM1 channel to MUSE. MUSE has run at three momentum settings, of 115 MeV/ , 160 MeV/ and 210 MeV/ , at selectable positive or negative beam polarity <ref type="bibr">[4]</ref>.</p><p>MUSE comprises beamline and scattering detectors. Fig. <ref type="figure">1</ref> is a drawing of the full detector setup. The first two detectors in the beam are a two-plane plastic scintillator Beam Hodoscope (BH) <ref type="bibr">[5]</ref> and a four-plane Gas Electron Multiplier (GEM) detector. These two detectors provide incoming particle timing, species identification, and tracking. Downstream of the GEM detector are the veto scintillator detector (VETO) and the trapezoidal target chamber <ref type="bibr">[6]</ref>. The annular VETO detector rejects events with beam particle trajectories that hit the target chamber walls, such as decays in flight and beam halo. On both sides of the target chamber, the Straw-Tube Tracker (STT) and Scattered Particle Scintillator (SPS) are responsible for scattered event tracking and timing. Downstream of the target chamber the Beam Monitor (BM) Fig. <ref type="figure">2</ref>. Radiative corrections for scattering as a function of the minimum electron momentum at one MUSE kinematic setting, calculated using the ESEPP event generator <ref type="bibr">[7]</ref>. The radiative correction factor indicates the difference between the experimental and Born cross sections: &#8725; 0 = 1 + . See Ref. <ref type="bibr">[8]</ref> for details. and calorimeter (CALO) help monitor the beam, measure its energy, and identify events with high energy forward-going photons.</p><p>Extracting precise Born cross sections and the proton radius from the experimental scattering cross sections requires radiative corrections to account for the contribution of higher-order processes to the scattering cross section. The leading-order correction with additional particles in the final state is the Bremsstrahlung correction, in which the charged leptons or protons emit real photons when accelerated. This correction generates a radiative tail, a continuous distribution in momentum for the nominally two-body elastic scattering &#179; reaction. This correction is most significant for electrons, as it is suppressed for muons and pions due to their higher mass.</p><p>Most electron-scattering experiments use magnetic spectrometers, and the measured spectrum and limited momentum acceptance of the experiments allow the Bremsstrahlung correction to be calculated with good precision. MUSE, however, runs with a large solid angle, non-magnetic spectrometer. As a result, MUSE measurements integrate over a wide range of outgoing electron momenta, including most of the radiative tail, to determine the cross section. The experimental uncertainty is sensitive to the limits of integration over the radiative tail, determined by detector thresholds on particle energy loss. MUSE employs two strategies that allow radiative corrections to be studied with the experimental data, with the goal of limiting the associated uncertainty from these corrections. One strategy is to use a low detector hardware threshold along with a software energy loss cut in the analysis. This procedure allows the cut and its uncertainties to be checked with simulations. A second strategy is to suppress the Bremsstrahlung correction by removing events from the analysis with high-energy, forward-going photons. This limits the initial-state radiation, largely removing the low-energy tail of the momentum distribution, due to the dominance of the initial-state over the final-state Bremsstrahlung correction -the initial-state radiation moves the vertex to lower 2 and thus higher cross section.</p><p>Removing events with high-energy photons is done with the leadglass calorimeter installed at the most downstream position of the experiment. Fig. <ref type="figure">2</ref>, reproduced from Ref. <ref type="bibr">[8]</ref>, shows an example of the dependence of the radiative correction on the minimum electron momentum in the integration. The red curve represents the radiative correction for scattering when there is no suppression of highenergy photons. The green curves are the radiative correction including calorimeter energy cuts in the data analysis. With the photon energy cuts, the corrections are less sensitive to the electron momentum threshold in the MUSE threshold region, where integration begins. Thus, the sensitivity and the resulting uncertainties are reduced. In particular, with a photon energy cut of greater than 40% of the beam momentum, the correction is small and nearly linear in the region of the cut, limiting this important experimental systematic correction and uncertainty <ref type="bibr">[8]</ref>. The variation of cross section with these radiative correction cuts is a strong test for MUSE that can demonstrate the quality of the applied radiative corrections, and that they are well understood.</p><p>The MUSE calorimeter specifications were developed based on Geant4 simulations with ESEPP radiative corrections <ref type="bibr">[7]</ref> shown in Fig. <ref type="figure">2</ref>. The cross section can be increased or decreased by offsets in the calorimeter light output calibration and by detector resolution, depending on the slope of the flux and the variation in resolution near threshold. With the goal of limiting changes in the absolute cross section from these resolution and offsets to the 0.1% level, we observed in the simulation that the calorimeter light output scale needed to be determined to approximately 2 MeV in the cut region, and corrections could be performed at an acceptable level for a resolution near threshold of approximately 5%/ &#8730; , with in GeV. In the following sections, we discuss the hardware details, calibration procedures and performance of the calorimeter. We show that detector simulations agree well with the experimental data.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.">Detector description</head><p>The calorimeter detector consists of an 8 &#215; 8 array of SF5 lead-glass bars, <ref type="foot">9</ref> on loan from the A2 experiment @ MAMI <ref type="bibr">[9]</ref>. The radiation length and Moli&#232;re radius of lead-glass are 1.265 cm and 2.578 cm, respectively <ref type="bibr">[10]</ref>. Fig. <ref type="figure">3</ref> shows a schematic drawing of a calorimeter bar and the associated photomultiplier. The bars are 24 radiation lengths long along the beam direction. Each lead-glass bar is individually wrapped with aluminized Mylar foil and black shrink tubing for optical isolation. The detector signals are read out by Hamamatsu R1355 photo-multiplier tubes (PMTs), which produce a negative anode signal.</p><p>When particles directly hit the center of a detector bar, that bar absorbs most of the particle's kinetic energy, with a small fraction of the energy spreading to neighboring bars. However, particles can hit edges and corners of the bars, depositing significant energy in neighbors. Hence, most of the energy of an incoming particle is easily captured by summing the energy of the bar with maximum energy with the energies of the 8 surrounding neighbors. The energy response and analysis will be discussed more in Section 4.</p><p>Fig. <ref type="figure">4</ref> shows a schematic drawing of the readout electronics. The signals propagate through delay lines before entering the Mesytec MCFD-16 constant fraction discriminators (CFDs) <ref type="bibr">[12]</ref>. Fig. <ref type="figure">5</ref> shows typical bar signals in an oscilloscope, after the delay, before being input to the CFDs. Data are for a 210 MeV/ negative polarity beam, comprised of approximately 30% 's, 5% 's, and 65% 's, with 's ( 's) generating the largest (smallest) signals. The CFDs generate logical timing signals that are digitized by TRB3 TDCs <ref type="bibr">[13]</ref>. The CFDs also send copies of the analog signals to Mesytec MQDC-32 Charge to Digital Converters (QDCs) <ref type="bibr">[14]</ref> for integration with 12-bit precision. While the QDC information is sufficient to determine light output in the detector for each event, the few MHz beam rate leads to light output from randomly coincident beam particles. The TDC information helps to distinguish in-time clusters from a scattering event from clusters from randomly coincident beam particles that are also read out in the same event.</p><p>Fig. <ref type="figure">6</ref> shows pictures of the calorimeter detector viewed from different directions. The front face of the calorimeter is located 138.5 cm from the center of the target. The detector covers an area of about 33 cm by 33 cm, or an angular range of about &#177;6.8 &#235; in horizontal and vertical directions, which is sufficient to capture most of the forwardgoing photons. Fig. <ref type="figure">7</ref> shows the simulated photon distribution at the front face of the calorimeter. The left panel shows the full photon distribution. The right panel shows the distribution of photons with &lt; 0.4 0 . With this calorimeter design, most of the high-energyphoton events will be removed in the analysis, while the lower energy photon events will be retained.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.">Calibration procedures</head><p>The goal of the calibration procedure is to convert the signal sizes in each detector channel, which are proportional to the light generated by electrons in the bars, to an energy, so that the incident electron energies can be determined. The calibration requires both gain matching the   calorimeter channels and determining the light output scale. Gain is the proportionality between the output QDC values and the particle signal size in the detector. We calibrate the gain for each channel to ensure the detector light output derived is consistent for all channels. It is convenient to use both cosmic rays and beam particles for this purpose.</p><p>Cosmic rays are always available and illuminate the calorimeter bars more uniformly than does the beam. However, light from cosmic rays is not collected with the same efficiency as for beam particles due to the different trajectories of the particles, so determining an absolute light output scale is more straightforward with beam particles of known energy. We focus on the response of the calorimeter to electrons in the beam, which generate similar signals in the calorimeter to photons of the same energy. Thus, we initially gain match with cosmics, and then we calibrate the absolute energy scale with beam.</p><p>The gain is calibrated in two steps. First, a hardware gain match adjusts PMT HV. Second, the QDC response of each bar is fine tuned in software. During the hardware gain matching, we adjust PMT high voltages to approximately match the positions of the peaks in the QDC spectra of each bar. The Hamamatsu R1355 PMT has 10 dynode stages. The relationship between applied high voltage and gain was found to be gain gain =</p><p>where was determined empirically to be approximately 8, with some variation between bars. Following this relation, the high voltage applied to each bar of the detector is adjusted iteratively until the QDC peak positions of the cosmic events for each bar are roughly matched at around QDC channel 500. This setting optimizes the calorimeter performance while keeping the QDC spectra from overflowing at higher beam momentum settings. At the highest beam momentum setting used for production data, 210 MeV/ , the QDC peaks are at about channel 3100, with a long tail reaching to channel 3840, near the end of the 12-bit QDC range. After the hardware gain match, in the software gain matching step, we fine tune the gain match by fitting the QDC peak of the cosmic signal for each bar, determining the factor needed to scale the peak position to the common QDC channel of 500. An example of the QDC spectra of the central bars after the gain matching calibration is shown in Fig. <ref type="figure">8</ref>.</p><p>Finally, to check the uniformity of the gain matching, the cluster energies are determined as a function of position in the calorimeter. Fig. <ref type="figure">9</ref> shows the result of this study, where the cluster light output sum vs. central bar location for 110 MeV/c beam data are plotted. From the figure, the average light output of clusters with a 9-bar sum is roughly the same independent of their position. To better understand the distribution, we compare the data to simulation. The simulation reproduces the behavior of the light output peak vs. bar position with maximum light output. More details about the calorimeter simulation and data digitization will be discussed in later sections. </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.">Detector performance</head><p>For each event, the sum of light output in the calorimeter is determined. First, the bar with the highest QDC value is found. Second, the light output of that bar and the eight surrounding neighbors are summed. Third, a search is performed for additional clusters away from the first cluster found. TDC information is checked for each cluster to determine if it is in-time, so that energy cuts should be applied. Fig. <ref type="figure">10</ref> shows the 9-bar QDC sum for the calorimeter at 160 MeV/ beam momentum. Data were taken with a beam particle trigger at approximately 180 kHz beam rate, leading to an accidental coincidence rate within the 200-ns QDC gate of approximately 4%. Incident particle species are identified using timing information, not from the calorimeter, but from a different detector, the BH, with respect to the accelerator RF <ref type="bibr">[15]</ref>. The bottom of Fig. <ref type="figure">10</ref> shows the TDC distribution of the calorimeter for events with light output greater than half of the beam energy. The TDC spectrum consists of two components, an intime peak from hits corresponding to particles triggering the readout of the data acquisition (DAQ) system, and random background from additional beam particles close in time to the triggering particle. The in-time peak is at approximately 40 ns and can be seen to be about two orders of magnitude larger than the random background. The background can be seen to be at a different level before vs. after the in-time peak. This arises from a dead-time effect. After the in-time peak, all additional beam particles can be recorded, except for those striking the same calorimeter bars close in time, due to an electronic dead time of about 25 ns. Earlier-in-time random particles are suppressed, as they would cause the DAQ system to read out, unless the system is still processing a previous event. Also, the 19.75 ns beam RF structure leads to time structure in the random background events, which is partially washed out by the multiple particle types at different phases of the RF convoluted with 10-ns FPGA clocks in the trigger.</p><p>To distinguish the photon events from the leptons and pions, the BM detector in front of the calorimeter is used. Photons are not expected to leave a signal in the thin scintillators of the BM, while electrons, muons Nuclear Inst. and Methods in Physics Research, A 1080 (2025) 170754 Fig. <ref type="figure">10</ref>. Top: Distribution of calorimeter light output 9-bar sums for beam particle events taken at momentum of 160 MeV/ . The distributions from different particle types are normalized to the same peak height as the distribution for all events. Bottom: TDC times relative to the event trigger for events with light output greater than half of the beam energy. The in-time peak is at H 40 ns. and pions will deposit energy in the detector. In the following sections, the calorimeter detector performance will be shown using electrons, which generate similar signals in the calorimeter to photons of the same energy.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.1.">Energy response</head><p>Beam data were used to study and calibrate the calorimeter absolute light output response as shown in Fig. <ref type="figure">11</ref>. The PiM1 beam line is designed to operate at momenta above about 100 MeV/ <ref type="bibr">[4]</ref>. To map out the detector response in a wide range of energy, data were taken at beam momenta from 20 MeV/ to 230 MeV/ , and simulations were used to estimate the energy of the particles entering the calorimeter. The measured energy sums for electrons, such as shown in Fig. <ref type="figure">10</ref>, were fit with Gaussians to determine the mean light output and its width, shown in Fig. <ref type="figure">11</ref>. The momentum scan result shows that the light output in the calorimeter as a function of the beam energy is very close to linear. The measured light output as a function of energy was fit with a linear function to check linearity and to determine parameters to be used in the simulation to model the detector light output (detailed in Section 4.3). The light output resolution was fit with</p><p>, which is commonly used to describe the resolution of calorimeters <ref type="bibr">[16]</ref>. In this fit, is the stochastic term that is governed by the electromagnetic shower fluctuations in the material, is the systematic term that reflects the uniformity of the detector and how well the detector is calibrated, and is the noise term from electronic readout when measuring the energies. Typical light output resolutions for lead glass calorimeters are about 5%&#8725; &#8730; &#8725;GeV <ref type="bibr">[16]</ref>. For the MUSE calorimeter, the parameters from the fit in Fig. <ref type="figure">11</ref> indicates that the resolution from the stochastic term is about 5.17%&#8725; &#8730; &#8725;GeV. About 6.22% resolution arises from the calibration term, while less than 1% is from the electronics.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.2.">Detector timing</head><p>Light output in the calorimeter from randomly coincident beam particles can be identified by two techniques. The BM hodoscope immediately upstream of the calorimeter localizes incoming charged particles at the cm level in position and at the 100 ps level in time. The light output from these beam particles is known from the beam momentum. The MUSE calorimeter also reads out timing information for the bars to help distinguish multiple hits in the detector, from both charged particles and photons. Fig. <ref type="figure">12</ref> shows the example of calorimeter timing at 160 MeV/ for one of the four central bars of the calorimeter. The top plot shows the detector time relative to the accelerator RF (modulo the RF period). The distribution shows three distinct peaks for the three particle types in the beam. The bottom plot shows time of flight from the beam hodoscope to the calorimeter, over a flight path of approximately 2.2 m. The electron, muon and pion peaks are also observed. The combination of the RF time and the time of flight identifies the particle species independent of the BH and also suppresses randomly coincident beam particles.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.3.">Simulation</head><p>The MUSE setup is simulated with Geant4. The simulation records hits and saves the events to a ROOT output tree. The simulated data is digitized and run through the same analysis as experimental data to allow a direct comparison of simulation to data. For the calorimeter, the simulation records Cherenkov light production in the detector bars. There are two modes of simulation that can be selected: a fast simulation that integrates a light yield over the path in the bar for &gt; &#8725; , and a more detailed simulation that creates optical photons at each reaction step and tracks those that reach the PMTs. The good agreement between data and the fast simulation, shown in Fig. <ref type="figure">14</ref>, indicates that it is not necessary to run the simulation in the much more computationally intensive photon-counting mode. Since the number of photons emitted by incident particles is proportional to (1 -1 2 2 ), for each reaction step in the simulation, we can find the relation for the number of photo-electrons read out in an event by the calorimeter as</p><p>where is the step length, is the index of refraction, and is scaling factor tuned to match the simulation to data. Integrating for all the steps in the bar determines the total number of photo-electrons generated by particles in that bar.</p><p>The digitization converts the simulated light output of the calorimeter bar to detector QDC channels. The pedestal of each QDC spectrum is modeled with a Gaussian distribution with the same width as the data, which is found during the calibration. If the bar has energy deposited in it, the energy is multiplied by a factor to more precisely model the gain of detector and then randomized with a Gaussian, where the width of the distribution is equal to</p><p>(</p><p>We tuned the values of parameter (H 4.46% GeV 1&#8725;2 ) and (H 0.066% GeV) to match simulated QDC spectra to the measured spectra. The parameter values are similar to the parameter and in the resolution fit of the light output scan calibration shown in Fig. <ref type="figure">11</ref>. Because the simulation has some resolution effects already when calculating the light output for each reaction steps in the lead-glass bar, the additional resolution needed at the digitization is smaller than the fit values from the data.</p><p>The digitized, simulated data are processed through the same analysis as the experimental data. Fig. <ref type="figure">13</ref> compares data and simulation for the spectrum of a single bar and for the 9-bar light output sum from the calorimeter. The data and simulation agree well, with a small mismatch  at the mean value. This difference is due to a mismatch in the energy to QDC conversion between data and simulation for some channels, along with small differences in the air gap between bars.</p><p>Figs. 14 compares the 9-bar-sum light output response and resolution of data and simulation at different momentum settings. Both data and simulation show similar linear relationships in the light output response, and similar light output resolution. While there is some disagreement at higher momenta, in the region where event cuts will be applied for radiative corrections (40% of the beam energy), the differences are small and the agreement is better than our 2-MeV requirement. Note that the energy to QDC channel conversion presented here is based on a calibration at one momentum setting, 110 MeV/ -no tuning was done to adjust the light output response of the simulation to match the data.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="5.">Reconstructed photons</head><p>Fig. <ref type="figure">15</ref> presents an example of the light output distribution of the reconstructed photons in the calorimeter for scattering data and preliminary simulation. The default radiative correction cut at &lt; 0.4 0 , where 0 is the beam electron momentum, is indicated by the dashed line. The scattering data and simulation are blinded independently based on scattering angles at the current stage of the analysis <ref type="bibr">[17]</ref>, which prevent the exact agreement between the two, but similar features are observed in both. The two distributions are similar in shape for the high energy photon events, with simulation being slightly narrower in width. In addition to blinding effect, the slight mismatch in resolution is expected as the simulation is at a preliminary stage with the calorimeter energy calibration based on one momentum setting only, and as the scattered-particle scintillator energy cuts have not been precisely matched in simulation and in data. The difference in the lower energy events is due to the difference in threshold setting between data and simulation. Both data and simulation show a prominent peak from the high-energy photons emitted by some scattering electrons, as expected. The radiative correction cut will remove these events, reducing the experiment's sensitivity to radiative corrections. The agreement between the calibrated data and simulation indicates that the calorimeter performance is sufficient to obtain the needed experimental uncertainties, discussed in Section 1 and in relation to Fig. <ref type="figure">14</ref>. Further tuning in the simulation calibration, including calibration using data taken at multiple energies and studying possible time dependence of the calorimeter response will improve the agreement to be better than requirements.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="6.">Summary</head><p>In order to have radiative corrections under control for the MUSE scattering experiment, a lead-glass calorimeter detector was built to capture high-energy photons from initial-state radiation. The construction of the system was described in Section 2, and calibration of the system was described in Section 3. Performance    output response and resolution sufficient to identify and remove highenergy photon events consistent with the requirements of agreement better than 2 MeV, as discussed in Section 1. Thus the calorimeter will allow MUSE to test and control radiative corrections at the needed level. </p></div><note xmlns="http://www.tei-c.org/ns/1.0" place="foot" n="9" xml:id="foot_0"><p>The unused optical cables were used in A2 to couple in test signals for calibration. These unused optical cables were sealed into a light-tight box located on top of the bar array.</p></note>
		</body>
		</text>
</TEI>
