			<titleStmt><title level='a'>Encyclopedia of Analytical Chemistry: Applications, Theory, and Instrumentation</title></titleStmt>
			<publicationStmt>
				<publisher>Wiley</publisher>
				<date>10/01/2023</date>
			</publicationStmt>
			<sourceDesc>
				<bibl> 
					<idno type="par_id">10635274</idno>
					<idno type="doi">10.1002/9780470027318.a9811</idno>
					
					<author>Yongling Ai</author><author>Ivan Joel Praneeth</author><author>Pengyi Zhao</author><author>Md Tanim Al Hassan</author><author>Arjun Sharma</author><author>Gracelynne Hao</author><author>Zhenghao Zhou</author><author>Hao Chen</author><author>Robert A Meyers</author>
				</bibl>
			</sourceDesc>
		</fileDesc>
		<profileDesc>
			<abstract><ab><![CDATA[In the past decades, mass spectrometry (MS)-based absolute quantitation has become a powerful tool to facilitate advancement in both biomedical and clinical fields. The main advantage of MS-based quantitation is the accurate and simultaneous quantitation of multiple analytes with high sensitivity. However, a reference material (e.g., authentic target analytes, or isotope-labeled peptide and protein standards) is indispensable for those quantitation methods to build a calibration curve for each analyte, which might be not available or difficult to synthesize especially for some drug molecules with complicated structure or proteins/peptides with post-translational modifications (PTMs). To tackle this issue, our group recently developed a standard-free absolute quantitation approach based on the combination of electrochemistry (EC) and MS, involving the coulometric measurement of analyte oxidation/reduction current as well as mass spectrometric measurement of the redox reaction yield. We named the method as coulometric mass spectrometry (CMS). Although the combined EC/MS techniques is a quickly growing research field in analytical chemistry with numerous distinct applications, the quantitation centered application has not been systematically reported. In this review, the principle, apparatus, and various applications of this emerging EC/MS hybrid quantitation method are presented and discussed in detail. An overview of the capabilities of CMS for absolute quantitation across a variety of modalities including small molecule metabolites, peptides, proteins, large molecule biotherapeutics, and PTMs as well as the role of CMS for drug impurity quantitation is provided. Simultaneously, the limitations of the method, which need further optimization and improvement, as well as promising future applications in drug development and proteomics are presented and evaluated.]]></ab></abstract>
		</profileDesc>
	</teiHeader>
	<text><body xmlns="http://www.tei-c.org/ns/1.0" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:xlink="http://www.w3.org/1999/xlink">
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="1.">Introduction</head><p>Quantification of analytes is extremely important in analytical chemistry. Numerous analytes are quantified on a day-to-day basis in various industry settings to answer questions regarding the analytes in the samples. For instance, in the pharmaceutical industry, complex drug molecules are synthesized from simple starting materials in multiple steps, and quantitation of the drug analytes is required to monitor each step in the in-process testing and the final purity of the drug material. <ref type="bibr">[1]</ref> In the biotechnology industry, a plethora of pharmaceutically active proteins <ref type="bibr">[2]</ref> and peptides <ref type="bibr">[3]</ref> and their impurities are made, and quantification is needed to monitor their production. In hospitals and clinical laboratories, body fluids are analyzed to diagnose disease severity and patients' conditions. <ref type="bibr">[4]</ref> For quantitation, commercially available standards or in-house synthesized standards are traditionally compared to the analyte in the sample using various analytical instruments. In the pharmaceutical industry, the common techniques such as high performance liquid chromatography (HPLC) with ultraviolet visible (UV-Vis) light detector, <ref type="bibr">[5]</ref> charged aerosol detector (CAD), <ref type="bibr">[6]</ref> evaporative light scattering detector (ELSD), <ref type="bibr">[7]</ref> or refractory index detector (RI), <ref type="bibr">[8]</ref> and gas chromatography (GC) <ref type="bibr">[9]</ref> have been widely used. In the biotechnology industry, enzyme-linked immunosorbent assay (ELISA), <ref type="bibr">[10]</ref> Western blotting, <ref type="bibr">[11]</ref> gel electrophoresis, <ref type="bibr">[12,</ref><ref type="bibr">13]</ref> and size exclusion chromatography <ref type="bibr">[14]</ref> are popular, whereas immunoassay-based instruments <ref type="bibr">[15]</ref> and spectrophotometers <ref type="bibr">[16]</ref> are commonly used in the hospital and clinical settings. Although these instruments are user friendly, relatively inexpensive, and are routinely used for quantitation, limitations do exist. For example, the UV-Vis detector can only detect molecules with chromophores. Other detectors such as CAD, ELSD, and RI can analyze molecules without chromophores but ultimately lack sensitivity. <ref type="bibr">[17]</ref> Nowadays MS has emerged as a leading technique for accurate and sensitive analyte quantification and characterization. The m/z value, MS/MS capabilities, <ref type="bibr">[18]</ref> and high resolution of mass spectrometers <ref type="bibr">[19]</ref> help in the identification and confirmation of the target compound structure with great selectivity and sensitivity. MS is capable of analyzing compounds with a wide range of molecular weights such as small drug molecules, peptides, <ref type="bibr">[20]</ref> proteins, <ref type="bibr">[20]</ref> lipids, <ref type="bibr">[21]</ref> polymers, <ref type="bibr">[22]</ref> DNA, <ref type="bibr">[23]</ref> and RNA. <ref type="bibr">[24]</ref> A variety of mass spectrometers with different types of ionization sources and mass analyzers are used based on the requirements of the analysis.</p><p>The MS-based quantification of proteins can broadly be categorized into two categories: absolute quantification and relative quantification. Both isotope labelling strategies and label free strategies <ref type="bibr">[25]</ref><ref type="bibr">[26]</ref><ref type="bibr">[27]</ref><ref type="bibr">[28]</ref><ref type="bibr">[29]</ref><ref type="bibr">[30]</ref> are mainly used. Relative quantitation methods provide information regarding the protein abundance ratio or relative change between two or more samples by comparing the amount of a surrogate peptide from a single protein with an isotope labeled standard peptide.</p><p>Usually, relative quantitation methods in proteomics are used to help to understand the dynamics of the whole proteome. In discovery-based proteomics, for example, a broad analysis of the proteome is carried out to look for quantitative differences in proteins using relative quantitation. <ref type="bibr">[31]</ref> In contrast, absolute quantitation methods provide the actual concentration of a specific peptide or a protein in a sample, proteins and post translationally modified proteins, <ref type="bibr">[32]</ref><ref type="bibr">[33]</ref><ref type="bibr">[34]</ref> and proteins within protein complexes can be accurately quantified using absolute quantification technique. <ref type="bibr">[35]</ref> Comparison of concentrations between laboratories is possible using absolute quantification technique. <ref type="bibr">[36]</ref> In relative quantitation using isotope-labelled strategies, a stable isotope-labeled peptide that is chemically identical to the target peptide in the sample is used. Therefore, the two peptides behave identically during liquid chromatography/mass spectrometry (LC/MS) analysis. Given that MS can recognize the mass difference between the labeled and unlabeled forms of a peptide, quantification is achieved by comparing their respective ion signal intensities. <ref type="bibr">[37,</ref><ref type="bibr">38]</ref> Various methods of isotope labelling are available. Some of the most popular methods include isotopecoded affinity tags (ICAT), <ref type="bibr">[39]</ref> stable isotope labeling by amino acids in cell culture (SILAC), <ref type="bibr">[40]</ref> isobaric tags for relative and absolute quantitation (iTRAQ), <ref type="bibr">[41]</ref> metal element chelated tags (MECT), <ref type="bibr">[42]</ref> and isotope-coded protein labeling (ICPL) <ref type="bibr">[43]</ref>. Because differentially labelled peptides can be resolved with mass spectrometry and the labelled samples are pooled before LC/MS, simultaneous analysis of multiple samples can be performed in one run by a multiplexing strategy, and the differences due to sample losses during sample preparation and changing ionization efficiency between injections can be circumvented. Although multiple samples can be compared in a single run using isotope labelled strategy, labelling procedures are often timeconsuming, and the cost of labelling reagents are high. <ref type="bibr">[44]</ref> Label-free methods for relative quantitation are available. Each sample is separately prepared and then subjected to individual LC-MS/MS. <ref type="bibr">[45]</ref> Protein quantification is generally based on ion intensity changes or the spectral counting of identified proteins after MS/MS analysis. <ref type="bibr">[46]</ref> Although a relative quantification method allows for the quantitative profiling of tens of thousands of peptides from thousands of proteins within a single experiment and provides a broad scope of the proteome, obtaining an absolute concentration via absolute quantification is often desired. Absolute quantification provides a far more precise description of molecular events than relative quantification.</p><p>In absolute quantitation using isotope labelled strategies, the isotope labelled standards exhibit similar chromatographic behavior but exhibit differences in mass and possibly isotopic behavior. The extracted peak areas or peak heights from the isotope labelled peptide and peptide from the sample are compared for absolute quantitation. <ref type="bibr">[47]</ref> Three types of isotope labelled strategies are used for quantitation in the bottom-up methodology: absolute quantification (AQUA), quantification conCATamer (QconCAT) and protein standard absolute quantification (PSAQ). AQUA is a variation of the isotope dilution mass spectrometry strategy used for quantifying small molecules. <ref type="bibr">[48]</ref> The AQUA peptides are added in a known quantity to the sample preparation right before LC/MS analysis. Both the AQUA peptide and the native peptide are measured by LC/MS. The analyte and the standard are identical except for the masses. The ratio of the peak intensities is used to measure the absolute amount of the analyte peptide. <ref type="bibr">[49]</ref> The AQUA peptide strategy is attractive. However, custom synthesis of peptides might be required to quantify atypical proteins. <ref type="bibr">[47]</ref> Typically, one surrogate peptide in a protein is used to quantify a protein of interest using AQUA strategy as quantifying multiple surrogate peptides from a protein would require purchasing multiple peptides. <ref type="bibr">[50]</ref> The QconCAT method can address this issue of the need to purchase multiple peptides to quantify multiple peptides in a protein. QconCATs are proteins encoded by synthetic genes that are concatamers of peptide internal standards with usually more than one target peptide per protein. Multiple peptides in a protein can be quantified using peptide standards resulting from digestion of a single QconCAT protein standard. The native proteins and QconCAT proteins are trypsin digested and the digested native and QconCAT peptides are compared for quantification. <ref type="bibr">[51]</ref> The QconCAT strategy also circumvents the protein digestion efficiency issue with the AQUA peptides strategy. In the AQUA peptide strategy, since the AQUA peptides are added after sample protein digestion, the digestion efficiency of the sample protein for incomplete digestion must be monitored closely. However, with the QconCAT strategy, the QconCAT proteins and sample proteins are digested simultaneously and the digestion efficiency of the QconCAT proteins and the sample proteins is similar. <ref type="bibr">[47]</ref> The QconCAT method can circumvent the two aforementioned issues with AQUA strategy, which are protein digestion efficiency and the use of multiple peptides for quantitation. Protein standard absolute quantification (PSAQ) methodology is used in situations where the sample preparation is extensive. In PSAQ methodology, the DNA sequence to code for the PSAQ protein is cloned into a plasmid. The protein is expressed in an expression medium such as E. coli.. Usually, arginine and lysine residues are labelled so that the tryptic digested peptides show a consistent mass difference compared to the native peptide. This method eliminates the differences between the sample and the standard in all the prefractionation steps. <ref type="bibr">[52]</ref> For traditional absolute quantitation, a calibration curve using isotope-labeled standard or peptide standard is necessary. <ref type="bibr">[53]</ref> Commercially available standards can be expensive. In some cases, some peptides and small molecules are not commercially available and in-house synthesis can be time-consuming or complicated. The most common way to obtain peptides is solid phase chemical synthesis. It requires covalently attaching the first amino acid to an insoluble support.</p><p>Then, subsequent amino acids are added by a series of coupling and deprotection steps. <ref type="bibr">[3]</ref> This kind of synthesis is inconvenient and time consuming. In all of the aforementioned techniques, there is no single strategy or technique that can absolutely quantify analytes without the use of a calibration curve or standards. A technique which can quantify without standards will be ideal and useful. Herein, we review a standard and label-free strategy "Coulometric Mass Spectrometry" (CMS), for the absolute quantification of analytes wherein LC/MS is combined with electrochemistry (EC). This technique uses the electroactive property of the analyte to quantify the analyte of interest. Using this technique, a wide variety of molecules were quantified, including small molecules with electroactive functional groups such as dopamine (DA) and norepinephrine (NE). <ref type="bibr">[54,</ref><ref type="bibr">55]</ref> Small molecules metabolites with electroactive functional groups have been quantified <ref type="bibr">[56]</ref>. Additionally, peptides and proteins with electroactive amino acid residues such as tyrosine (Y) <ref type="bibr">[57,</ref><ref type="bibr">58]</ref> and tryptophan (W) <ref type="bibr">[59]</ref> were quantified and demonstrated for high throughput analysis. Post-translationally modified peptides such as phosphorylated peptides were quantified. <ref type="bibr">[58]</ref> Proteins are digested and the surrogate peptides containing electroactive amino acids are measured to obtain the concentration of the precursor protein. <ref type="bibr">[57]</ref> Drug impurities such as nitrosamines or host cell proteins (HCPs) were also quantified. <ref type="bibr">[60]</ref> </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.">CMS principle and instrumentation</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.1.">Principle of CMS</head><p>The combined technique of EC with MS, EC/MS, has many applications, including drug metabolism study, protein structural analysis, and electrochemical reaction mechanism elucidation. <ref type="bibr">[61]</ref><ref type="bibr">[62]</ref><ref type="bibr">[63]</ref><ref type="bibr">[64]</ref><ref type="bibr">[65]</ref><ref type="bibr">[66]</ref><ref type="bibr">[67]</ref> The method can also be used to observe elusive reaction intermediates <ref type="bibr">[68]</ref><ref type="bibr">[69]</ref><ref type="bibr">[70]</ref><ref type="bibr">[71]</ref><ref type="bibr">[72]</ref><ref type="bibr">[73]</ref><ref type="bibr">[74]</ref><ref type="bibr">[75]</ref><ref type="bibr">[76]</ref><ref type="bibr">[77]</ref> and for electrosynthetic reaction screening. <ref type="bibr">[78]</ref><ref type="bibr">[79]</ref><ref type="bibr">[80]</ref> As a new application of EC/MS techniques, the unprecedented advantage of CMS is the capability of absolute quantitation without using any standards or establishing a calibration curve. Although the traditional coulometric approach could also realize standard free quantitation based on Faraday's law, the complete electrochemical oxidation or reduction conversion is often realistically difficult to achieve. Since MS is integrated after EC, it enables the measurement of the oxidation or reduction conversion yield and is ideal for monitoring the reaction products to further confirm and validate the reaction mechanism and electron transfer. Theoretically, CMS quantitation can be applicabe for analytes which contain an electroactive functional group with known electrochemical activities. Before the coulometric measurement, the redox potential can be optimized with cyclic voltammetry (CV) and selected carefully to improve the electrochemical response and avoid over-oxidation/reduction. The applied working electrode (WE) potential needs to be optimized and controlled for the best sensitivity and selectivity in terms of electric current signal, since over-oxidation/reduction may give rise to a competitive side reaction. This could impact the accuracy of CMS quantitation if the number of electrons per molecule involved in the side reaction is different from that of the expected oxidation (or reduction) reaction. Once the electrochemical reaction is triggered, the total electric charge responsible for oxidizing/reducing analyte substance in coulombs in the redox reaction, Q, is directly proportional to the quantity of the oxidized/reduced substance: Q=nzF, where n is the moles of analyte, Z is the number of electrons transferred per molecule for the redox reaction, and F is the Faraday constant (9.65 &#215; 10 4 C/mol). Q can be directly measured from the integration of Faradaic current over time. The moles of the analyte that is oxidized or reduced can be calculated as n=Q/zF. On the other hand, electroactive species shows reduced intensities in the acquired MS spectra upon oxidation/reduction, and the relative MS intensity change upon redox reaction, &#916;i, can reflect the redox conversion yield. From the acquired MS spectra before and after electrolysis, &#916;i can be measured in two ways: the relative change of the target analyte peak intensity (relative to a reference peak) or the relative change of the target analyte peak area in the extracted ion chromatogram (EIC) upon electrolysis, as illustrated in Scheme 1. Thus, the amount of analyte converted, in combination with the conversion yield, can be used to calculate the total amount of analyte. In other words, Total amount of analyte =(amount of the oxidized/reduced analyte)/(the oxidation/ reduction yield) =(Q/zF)/&#916;i =Q/(zF&#916;i) Scheme 1. Schematic showing the process of CMS including WE potential optimization by CV, coulometric measurement of converted analytes and MS measurement of redox reaction yields.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.2.">Instrumentation</head><p>As shown in Scheme 2a, the original prototype of CMS is a coupled liquid chromatography/electrochemistry/mass spectrometry (LC/EC/MS) apparatus. <ref type="bibr">[54]</ref> A liquid chromatography system equipped with a C18 column was used for separation, allowing the CMS quantitation of an individual analyte in a mixture sample. An electrochemical thin-layer flow cell equipped with a Magic Diamond (boron-doped diamond) disc electrode (i.d., 8 mm) or a glassy carbon disc (i.d., 3 mm and 6 mm) as the WE used for the oxidation process. A positive potential (ranging from + 1.0 to + 1.3 V) was applied to the WE electrode for oxidation of the LC-separated target compounds. The eluate flowing out of the cell was either collected and subsequently analyzed using nano-electrospray ionization mass spectrometry (nanoESI-MS, Scheme 2a) or CV directly monitored online using home-built liquid sample desorption spray ionization mass spectrometry <ref type="bibr">[81]</ref> (LS-DESI-MS, Scheme 2b). The oxidation yield was calculated based on the EIC peak area or peak intensity change relative to an internal standard before and after electrolysis.</p><p>In addition to DESI, a commercial heated electrospray ionization (HESI) source was used later for convenience <ref type="bibr">[58]</ref>. For compounds that are already purified, a LC column is not necessary and can be removed (i.e., flow-through analysis), thus shortening the analysis time for each injected sample <ref type="bibr">[55]</ref>. Using an LC auto-sampler, multiple samples can be injected sequentially, via flow injection (FI), to expand the application of CMS for high throughput quantitation and fast screening analysis. As shown in Scheme 2c, the whole injection sequence for one sample took 7.5 min where 0-2 min was a blank solvent injection for cleaning purpose followed by a 2 min analyte injection in the "cell-off" mode. The electrochemical cell was turned on for 1.5 min before the second injection in the "cell on" mode due to the charging current decay. Moreover, to further simplify the setup, a versatile three-electrode system screen printed on the polymer support acting both as the coulometry platform for electrochemical oxidation and the sample loading tip for spray ionization was reported <ref type="bibr">[82]</ref>. The combination of CMS with direct ambient sampling methods facilitated the quantitation speed and simplicity with low cost, which showed the potential of CMS for rapid on-site testing. b) Reprinted and slightly modified with permission from Ref <ref type="bibr">[54]</ref>, Ref <ref type="bibr">[83]</ref>, Ref <ref type="bibr">[55]</ref> and Ref <ref type="bibr">[82]</ref>.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.">Applications</head><p>With rapid advances in instrumentation and technologies, mass spectrometers reporting the absolute quantity of target analytes have gained increasing popularity given its a wide range of applications such as drug discovery and development, biomarker verification/validation, and biological process exploration. However, MS-based quantitation strategies often rely on a reference material like isotope labeled standards, which is expensive, sometimes not available, or very time-consuming to synthesize. As an emerging and novel quantitative strategy, CMS has demonstrated its capabilities for absolute quantitation across different modalities such as small</p><p>molecule metabolites, peptides, proteins with PTMs and large molecule biotherapeutics (Scheme 3). These applications showed promising results of CMS and its great potential to be used in different fields.</p><p>Scheme 3. Summary of current applications and advances of CMS absolute quantitation of small organic molecules, peptides, proteins and drug impurities</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.1.">Quantitation of small organic molecules</head><p>Small molecule quantitation by MS has been widely carried out, mainly relying on structural similarities or isotope-labeled compounds as the internal standard due to MS signal fluctuations and the ion suppression effect from the matrices. In addition, MS-based absolute quantitation approaches require the construction of a calibration curve. Although LC coupled with electrochemical detection has been well established for absolute quantitation of electrochemically active analytes such as carbohydrates, antibiotics, and neurotransmitters based on the generated electric current response which is proportional to the quantity of analytes, authentic target compounds and the calibration curve are still needed. The traditional coulometric approach has</p><p>CMS Standard Free Quantitation Y/W/C containing peptides A&#946; peptide fragments Phosphopeptides Deamidated peptides and intermediate Endogenous metabolites Neurotransmitter N-Nitrosamines HCPs Kai proteins mAbs</p><p>been known for the direct quantitation of electroactive compounds based on Faraday's law. The absolute quantity of the analytes could be easily calculated if the targets undergo complete oxidation or reduction. However, full redox conversion is often difficult to achieve and control for multiple reasons. Therefore, the first prototype CMS experiment proposed that the absolute quantitation of analytes under partial electrolysis is still possible, if the conversion yield can be measured, for example, using mass spectrometry. <ref type="bibr">[54]</ref> In preliminary trials, several electroactive small organic molecules such as dopamine (DA), norepinephrine (NE), rutin (RN), and urine acid (UA), were first tested and quantified using CMS. The obtained quantitation error was small and ranged from -2.6 to + 4.6%.</p><p>a) b) Reprinted and slightly modified with permission from Ref <ref type="bibr">[54]</ref> and Ref <ref type="bibr">[55]</ref>.</p><p>Figure <ref type="figure">1</ref> shows the electrochemical oxidation reactions of these tested compounds. For example, neurotransmitter DA and NE as well as RN, a glycoside of the flavonoid quercetin containing two glucose units (structure shown in Figure <ref type="figure">1a</ref>), are known to undergo a two-electron transfer electrochemical oxidation to produce quinone products. Figure <ref type="figure">2</ref> illustrates the CMS quantitation result of rutin, as an example. The protonated rutin was seen at m/z 611 (Figure <ref type="figure">2a</ref>).</p><p>After electrolysis (Figure <ref type="figure">2b</ref>), a peak at m/z 609 was observed, corresponding to the oxidized product rutin quinone (RQ). Figure <ref type="figure">2c</ref>, d shows the EIC (m/z 611, the protonated rutin observed at 1.4 min) of 50 &#956;M rutin with an injection volume of 6 &#956;L (injected amount: 300 pmol) at the mobile flow rate of 300 &#956;L/min with the cell being turned off and on (applied potential:</p><p>respectively. The EIC peak area shown in Figure <ref type="figure">2d</ref> was smaller by 2.5%, in comparison with that of the peak shown in Figure <ref type="figure">2c</ref>, indicating that the oxidation yield for rutin was 2.5%. On the other hand, a sharp electric oxidation current was detected after cell was turned on, as shown in Figure <ref type="figure">2f</ref>-h (Figure <ref type="figure">2e</ref> shows the background current diagram for blank solvent sample as a contrast).</p><p>Based on the integration of the current peak area (Figure <ref type="figure">2f</ref>-h), the amount of the oxidized rutin was calculated to be 7.3 pmol on average. Therefore, the measured amount of rutin was 7.3 pmol/2.5% = 292 pmol, which was close to the injection amount of 300 pmol with the measurement error to be - 2.6%.</p><p>It's noteworthy that the integration of MS with traditional coulometric approach not only allows the measurement of oxidation yield, but also enables monitoring of the electrochemical reaction products and confirms the number of electrons per molecule that are involved in the redox conversion. In terms of the oxidation yield measurement, it could be also performed with offline MS analysis (Scheme 2a). For example, DA and NE were collected after flowing through LC and the electrochemical cell before and after electrolysis, spiked with arginine as a reference compound and injected for nano-ESI-MS analysis. The peak intensity ratio (relative to arginine) change provided the oxidation yield. Uric acid, another electroactive small molecule, was also quantified based on a two-electron electrochemical oxidation to generate a diimine product (Figure <ref type="figure">1d</ref>).</p><p>Elevated UA level in blood can trigger gout, aggregate in joints, and is related to kidney disease.</p><p>Therefore, quantitative analysis of UA level in urine is meaningful and was also demonstrated by Reprinted with permission from Ref <ref type="bibr">[54]</ref>.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>CMS. [54]</head><p>With using LC, CMS absolute quantitation is applicable for mixture sample. As an example, 5-HT (structure shown in Figure <ref type="figure">1e</ref>) and DA were quantified individually in a mixture sample following HILIC separation. <ref type="bibr">[55]</ref> To further improve the application of CMS to accommodate the high throughput feature and fast screen analysis in drug discovery and development process, an improved version of CMS setup for fast quantitation of purified electroactive compounds was demonstrated with a 7.5 min cycle time for each analyte. In this experiment, a flow injection (FI) analysis sequence was developed using a LC auto-sampler and resulted in a significant decrease of analysis time for each injected, allowing sequential multiple sample quantitation. As illustrated in Scheme 2c, an injection sequence consisted of three steps: 1) a blank solvent (ACN/H2O/FA, 50:50:0.1 by volume) injection to check if there was any sample carryover, 2) an analyte solution injection when the electrochemical cell was turned off (i.e., the "cell off" mode), and 3) an analyte solution injection when the electrochemical cell was turned on (i.e., the "cell on" mode). The running time for each injected sample was 2 min, but the electrochemical cell was turned on 1.5min before step 3) so that the charging current could fade away and not interfere with the analyte oxidation current. Because of its simplification in sample processing, cost, and ease, this standard-free absolute quantitation platform is ideal for onsite analysis. In addition, the versatile polymer tip can be fabricated with slight modification to selectively extract the analyte and oxidize the target analyte in the presence of biological matrixes. For example, bilirubin, an important metabolite marker for indicating various liver and gallbladder diseases, was spiked in urine and successfully quantified with CPSI-MS platform due to the specific DODMAC/MWCNT modification on the WE. However, for real-world applications, the sample matrixes might be more complicated and thus sample pretreatment steps might be necessary before the measurement to remove electrochemically active interferences. Alternatively, introducing target-specific elements such as enzymes, antibodies, aptamers, lectins, or molecularly imprinted polymer cavities for the modification of WE would significantly increase the selectivity of EC reaction as well as the quantitation accuracy. </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.2.">Peptide quantitation</head><p>In addition to the quantitation of small molecules, CMS has been explored to quantify peptides without using peptide standards or isotope-labeled peptide standards. Among the 20 naturally occurring amino acids, four are oxidizable: cysteine</p><p>and methionine (M), listed in the order of their increasing oxidation potential. Glutathione (GSH), a thiol-containing peptide, was first tested with CMS based on the electrochemical oxidation conversion of thiols into disulfides (Figure <ref type="figure">4a</ref>). CMS quantitation of GSH is similar to small molecules, as shown in Figure <ref type="figure">5</ref>. After electrolysis, a peak at m/z 307 was observed, corresponding to + 2 ion of the oxidized glutathione product GSSG (Figure <ref type="figure">5d</ref>), as expected. The integrated current area and the measured oxidation yield were combined to successfully quantify the amount of native GSH with a measurement error of 4.6%. Another oxidizable residue, tyrosine, with a relatively low oxidation potential and relatively high abundance in proteins was investigated and showed a reproducible and stable electric current for all the examined tyrosine-containing peptides. Under an appropriate oxidation potential (e.g., 1.00-1.05 V vs. Ag/AgCl), the phenol side chain of a peptide tyrosine residue can be oxidized into a semi-quinone by losses of two electrons and two protons (Figure <ref type="figure">4b</ref>), in which a mass shift of 2 Da occurs, which can be readily monitored by MS analysis. For example, GGYR was used as a test sample. Tyrosine-containing peptide GGYR can undergo oxidation by giving up two electrons. In fact, the protonated GGYR was found at m/z 452 prior to electrolysis (Figure <ref type="figure">6a</ref>). A new peak at m/z 450 was seen during electrolysis (Figure <ref type="figure">6b</ref>), which corresponded to +1 ion of the oxidized GGYR product, GGY'R (Y' stands for the oxidized tyrosine residue). The EIC of 50 &#956;M GGYR with an injection volume of 6 &#956;L was shown in Figures <ref type="figure">6c</ref> and <ref type="figure">d</ref>. The applied potentials were 0 V and +1.0 V. When compared to the peak in Figure <ref type="figure">6c</ref>, the integrated area for the peak in Figure <ref type="figure">6d</ref> was 5.2% smaller, showing that GGYR had a 5.2% oxidation yield (averaged in triplicate). The electric current seen in Figure <ref type="figure">6f</ref> was likely the result of GGYR oxidation because there was no electric current peak seen for the solvent in contrast (Figure <ref type="figure">6e</ref>). The average concentration of the oxidized GGYR was determined to be 14.9 pmol based on the integration of the present peak (Figure <ref type="figure">6f</ref>). Therefore, the measured amount of GGYR was 285 pmol, which was near to the injection amount of 300 pmol with a measurement error of -5.0%. Figure <ref type="figure">6</ref>. ESI-MS spectra of GGYR when the applied potential was a) 0 V and b) +1.0 V. The peak of the oxidized product GGY'R was clearly seen at m/z 450 in b). EIC of GGYR was recorded when the applied potential was c) 0 V and d) +1.0 V. Electric current responses were shown due to the oxidation of e) a blank solvent and f) GGYR peptide. Reprinted with permission from Ref <ref type="bibr">[58]</ref>.</p><p>In addition, the oxidized product (m/z 450) was subjected to MS/MS analysis to ensure that it was indeed generated via the electrochemical oxidation of tyrosine in the GGYR (Figure <ref type="figure">7</ref>).</p><p>In this experiment, a syringe pump was used to introduce the GGYR sample into an</p><p>2 3 4 5 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0 1 2 3 Current (&#181;A) 0.6 0.5 0.4 0.3 0.2 0.1 0.0 0 1 2 3 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0 1 2 3 Current (&#181;A) 0.6 0.5 0.4 0.3 0.2 0.1 0.0 445 450 455 m/z 0 50 100 0 50 100 Relative Abundance 452.22 453.23 454.23 451.22 450.21 453.23 454.23 451.22 450.21 x20 [GGYR+H] + [GGYR+H] + 452.22 450 455 m/z 100 0 100 Relative Abundance 452.22 453.23 454.23 451.22 450.21 453.23 454.23 451.22 450.21 x20 [GGYR+H] + [GGYR+H] + 452.22 // // 0 1 2 3 4 5 6 7 8 9 10 Time (min) 0 50 100 0 50 100 Relative A bundance 0 1 2 3 4 5 6 7 8 9 10 Time (min) 0 50 100 0 50 100 Relative A bundance 0 1 2 3 4 5 6 7 8 9 10 Time (min) 0 50 100 0 50 100 Relative A bundance 0 1 2 0.1 0.2 0.3 0.4 0.5 0.6 B A B Current (&#181;A) Time (min) 0.6 0.5 0.4 0.3 0.2 0.1 0 3 6 0 1 2 3 4 5 6 0.1 0.2 0.3 0.4 0.5 0.6 B A B 0 3 6 Time (min) Current (&#181;A) 0.6 0.5 0.4 0.3 0.2 0.1 a) Cell off b) Cell on c) Cell off Area=6.8E9 d) Cell on Area=6.4E9 e) blank f) GGYR Area 3.0&#215;10 -6 C [GGY'R+H] +</p><p>electrochemical flow cell for oxidation at +1.0 V potential. The final product was collected and reionized using nESI, and the resulting oxidation product, GGY'R was detected at m/z 450. To compare with the intact peptide GGYR ion at m/z 452, collision-induced dissociation (CID) data of m/z 450 was recorded. As seen in Figure <ref type="figure">7</ref>, the oxidized peptide ion (m/z 450) produced the fragment ion y1 (m/z 175) and y2' (m/z 336) while the intact peptide (m/z 452) produced y1 (m/z 175) and y2 (m/z 338), demonstrating the occurrence of the oxidation to the tyrosine residue where 2 Da mass shift took place. Further proof that oxidation does in fact occur to the side chain of tyrosine in this peptide can be seen in the fragment ion of m/z 344 that results from dissociation of m/z 450 by loss of the oxidized tyrosine side chain O=C6H4=CH2, as illustrated in Figure <ref type="figure">7a</ref>. To demonstrate how CMS is also applicable to peptides containing more than one tyrosine residue, angiotensinogen (1-14), a known tetra decapeptide (TDP) with the sequence</p><p>50 150 250 350 450 m/z 0 50 100 452.22 175.12 278.11 338.18 &#215;10 50 150 250 350 450 m/z 0 50 100 344.17 175.12 201.10 450.21 276.10 336.17</p><p>y 2 ' [GGY'R+H] + y 1 b 3 ' b 2 m/z 115.05</p><p>Relative Abundance Relative Abundance DRVYIHPFHLLVYS, was studied. It was found that oxidation occurred dominantly to one tyrosine of the peptide, due to a short residence time in the electrochemical flow cell with using a high sample flow rate. Oxidation of two tyrosine residues could also take place. In such cases, CMS quantitation of peptide containing more than one tyrosine residue is still possible based on the assumption that the intensities of the different oxidized products represent a good approximation for the concentrations of the different oxidized product forms. For instance, prior to electrolysis, the +4 charged TDP ion was identified at m/z 440.49 (Figure <ref type="figure">8a</ref>). A peak at m/z 439.98 was found after electrolysis (Figure <ref type="figure">8b</ref>), matching to the +4 ion of oxidized TDP product originating from the oxidation of one of the tyrosines. A second, smaller peak at m/z 439.98 resulted from the oxidation of both tyrosines (Figure <ref type="figure">8b</ref>), with an intensity ratio of 0.12:1 between m/z 439.48 and m/z 439.98. Figures <ref type="figure">8c</ref> and <ref type="figure">d</ref> displayed the EIC of m/z 440.49, the +4 charged TDP ion, as a result of the injection of 6 &#181;L of 50 &#181;M TDP (300 pmol in total) at 0 V and +1.05 V (vs. Ag/AgCl), respectively. The integrated area of the peak depicted in Figure <ref type="figure">8d</ref> was 2.6% less than that of the peak shown in Figure <ref type="figure">8c</ref>, indicating that the oxidation yield for TDP was 2.6%. Figure <ref type="figure">8f</ref> showed the TDP oxidation current peak (Figure <ref type="figure">8e</ref> showed the background current pattern for blank solvent sample at +1.05 V potential). The total electric charge Q involved in the TDP oxidation, leading to the generation of m/z 439.98 via one tyrosine oxidation (2e -per mole peptide) and m/z 439.48 via two tyrosine oxidation (4e -per mole peptide), could be computed by integrating the current peak area. Taking into account the intensity ratio of 0.12:1 between m/z 439.48 and m/z 439.98 and the similarity of the structures of the two oxidation products, the molar ratio of the two oxidation products was roughly 0.12:1. Therefore, the total amount of oxidized peptide was calculated to be 7.2 pmol. Considering the oxidation yield of 2.6%, the amount of TDP we measured was 283 pmol. Compared to the injection quantity of 300 pmol, the quantitation error was -5.5%. This result demonstrated that CMS can successfully quantify peptides with multiple tyrosine residues. In addition to GGYR and TDP, several model peptides containing tyrosine residues were quantified with good accuracy and results are summarized in Table <ref type="table">1</ref>.</p><p>Figure <ref type="figure">8</ref>. ESI-MS spectra of TDP when the applied potential was a) 0 V and b) +1.05 V. The major peak of the oxidized product of TDP was seen at m/z 439.98 (+4 charged) in b). EIC of the +4 charged TDP ion was recorded when the applied potential was c) 0 V and d) +1.05 V. Electric current responses were shown e) due to the blank solvent and f) the oxidation of TDP peptide. Reprinted with permission from Ref <ref type="bibr">[58]</ref>. x5 439.98 0 1 2 3 4 5 6 7 8 9 10 11 0 50 100 0 50 100 Relative Abundance 0 1 2 3 4 5 6 7 8 9 10 11 12 0 50 100 0 50 100 Relative Abundance Relative Abundance 4 5 6 7 0.0 0.1 0.2 0.3 0.4 0.5 0.5 0.4 0.3 0.2 0.1 0.0 0 1 2 3 Current (&#181;A) 3.0 3.5 4.0 4.5 0.0 0.1 0.2 0.3 0.4 0.5 0.5 0.4 0.3 0.2 0.1 0.0 0 1 2 3 Current (&#181;A) c) Cell off Area =5.5E9 d) Cell on Area =5.4E9 e) blank f) TDP (DRVYIHPFHLLVYS) Area 1.5&#215;10 -6 C a) Cell off b) Cell on</p><p>8 Fibronectin adhesion-promoting peptide WQPPRARI 1023.2 4.3 9 gp 100 (25-33) human KVPRNQDWL 1155.3 1.5 10 BDC2.5 mimotope RTRPLWVRME 1343.6 -2.3 11 delta sleep inducing peptide WAGGDASGE 848.8 -4.5</p><p>To further extend the CMS application, we also investigated the possibility of using CMS for absolute quantitation of tryptophan-containing peptides. Tryptophan is known to be electroactive and has a relatively low oxidation potential (+1.01 V, vs NHE). <ref type="bibr">[84]</ref> The major oxidation products and pathways of tryptophan-containing peptides were proposed and validated by MS analysis. <ref type="bibr">[85]</ref> As depicted in Figure <ref type="figure">4c</ref>, the major oxidation products are the singly hydroxylated M+16 (2e -oxidation product) and the ketone product M+14 (4e -oxidation product).</p><p>Minor oxidation products include the doubly hydroxylated M+32 (4e -oxidation product) and the cleavage product R1W+14 (4e -oxidation product). Online CMS allows for simple monitoring of the different oxidation products. On the basis of the similar structures of these oxidation products, we hypothesized that the intensity ratios of the various oxidized products are approximately equivalent to their molar ratios, as shown in eq. 1-3. Assuming that the moles of the products M+16, M+14, M+32, and R1W+14 are n1, n2, n3, and n4, the total mole of the oxidized peptide is n = n1+n2+n3+n4. According to Faraday's Law, Q is derived from each of the four oxidation pathways (eq 4). From equations 1 through 4, the total amount of oxidized peptide n can be calculated using equation 5, where Q is obtained experimentally by integrating the Faradaic current over time and F is the Faraday constant. Upon obtaining n and the oxidation yield, the total amount of the peptide can be calculated as the ratio of the amount of oxidized peptide n and the oxidation yield &#916;i (i.e., n/&#916;i).</p><p>) &#119899; 3 &#119899; 1 = [&#119872; + 32] [&#119872; + 16] (2) &#119899; 4 &#119899; 1 = [&#119877; 1 &#119882; + 14] [&#119872; + 16] (3) &#119876; = &#119865;(2&#119899; 1 + 4&#119899; 2 + 4&#119899; 3 + 4&#119899; 4 ) (4) &#119899; = &#119899; 1 + &#119899; 2 + &#119899; 3 + &#119899; 4 = &#119876; 2&#119865; &#215; 1 + ([&#119872; + 14] + [&#119872; + 32] + [&#119877; 1 &#119882; + 14]) ([&#119872; + 16]) &#8260; 1 + 2 &#215; ([&#119872; + 14] + [&#119872; + 32] + [&#119877; 1 &#119882; + 14]) ([&#119872; + 16]) &#8260;</p><p>To testify this hypothesis, a tripeptide WGG was chosen to determine the viability of CMS peptide quantification based on tryptophan oxidation. No electric current peak was seen when the oxidation potential of +1.00 V was applied to a blank solvent sample (Figure <ref type="figure">9e</ref>). In contrast, a sharp electric current peak was found after the injection of WGG sample at the same oxidation potential, showing that the current peak resulted from the electrochemical oxidation of WGG (Figure <ref type="figure">9f</ref>). The protonated WGG was discovered at m/z 319.14 prior to electrolysis (Figure <ref type="figure">9a</ref>).</p><p>The oxidation products WGG+14, WGG+16, and WGG+32 were detected at m/z 333.12, 335.13, and 351.11, respectively, after electrolysis (Figure <ref type="figure">9b</ref>). Specifically, the ion intensities of WGG+14 and WGG+16 were significantly greater than those of WGG+32. The integrated EIC peak area of m/z 319.14 in Figure <ref type="figure">12d</ref> was 4.4% less than in Figure <ref type="figure">9c</ref>, indicating that WGG's oxidation yield was 4.4%. On the basis of eq <ref type="bibr">(5)</ref> and the integration of the current peak area as well as the ion intensity ratios of the oxidized products, the quantity of WGG that was oxidized was determined to be 3.8 pmol. In conjunction with the oxidation yield of 4.4%, the amount of WGG was measured to be 87.1 pmol, which was fairly close to the injection amount of 89.0 pmol (measurement error of 2.1%).</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Figure 9.</head><p>MS spectra of WGG a) when the cell was off and b) when the cell was turned on (applied potential: +1.00 V). The oxidation products of WGG were detected at m/z 333.12, 335.13, and 351.13. EICs of WGG were acquired c) when the cell was off and d) when the cell was turned on (applied potential: +1.00 V). Electric current diagrams were collected from (e) blank solvent and (f) the oxidation of WGG. Reprinted with permission from Ref <ref type="bibr">[59]</ref>.</p><p>Several other peptides such as WGG, WQPPRARI, KVPRNQDWL, RTRPLWVRME and WAGGDASGE were also successfully quantified, with the quantitation errors ranging from -4.5 to 4.3% in triplicate measurements (as summarized in Table <ref type="table">1</ref>).</p><p>After a demonstration of the feasibility of CMS for peptide quantitation, several important applications were addressed, including the absolute quantitation of peptides containing PTMs like phosphorylation <ref type="bibr">[58]</ref> and deamidation <ref type="bibr">[83]</ref> as well as the quantitation of amyloid beta (A&#946;) peptide fragments, <ref type="bibr">[59]</ref> an important risk factor having a central role in the onset and progression of Alzheimer's disease (AD). Phosphorylation is an important post-translation modification and has been widely researched in proteomics studies. Absolute quantitation of phosphopeptides often involves isotope-labeled peptides, which typically requires multiple-step synthesis with high cost.</p><p>To solve this problem, we first tested one phosphopeptide RRLIEDAEpYAARG, whose tyrosine reside was phosphorylated. Interestingly, such a peptide did not display an oxidation current upon oxidation probably due to the covered electroactive phenol group. Another phosphopeptide which was phosphorylated in its serine residue and contained one free tyrosine residue was also tested and successfully quantified with CMS. Therefore, we assume that absolute quantitation of phosphopeptides without using any standards is still possible if one free oxidizable residue exists. Scheme 5. Workflows showing absolute quantitation of a) multiple proteins in a protein mixture, b) HCPs in mAb, and c) mAb deamidation by CMS. Reprinted with permission from Ref[83] Another CMS application is the quantitation of antibody deamidation. Monoclonal antibodies (mAbs) and their related formats representing a large category of biopharmaceutical products are meeting an undeniable success as therapeutic agents in different fields such as oncology, immune disorders, and, more recently, the treatment of the SARS-CoV-2 virus. However, these recombinant therapeutic proteins contain a variety of chemical and physical modifications. Common examples include glycan structural differences, deamidation, oxidation, and glycation. Thus, to ensure patient safety, it is crucial to control and minimize this heterogeneity for product reliability and consistency across the entire product life cycle. Deamidation of asparagine (Asn) residues to Asp or isoAsp (Scheme 4) is a common chemical modification and degradation pathway that occurs during the manufacturing and storage of monoclonal antibodies (mAbs).[86, 87] Previous studies have shown that the degradation products as well as the succinimide intermediate can cause serious quality and safety issues such as aggregation, intermolecular crosslinking, activity loss, and immunogenicity. Therefore, mAb deamidation needs to be closely monitored. So far, absolute quantitation of deamidation is rare and absolute quantitation of the succinimide intermediate were not done due to the lack of standards. The simultaneous identification and absolute quantification of mAb deamidation degradation can be performed by CMS, as illustrated in the workflow of Scheme 5c. As a result, the simultaneous quantitation of native and deamidated peptides as well as the succinimide intermediate from deamidation of Asn 318 residue of NIST 8671 mAb can be completed. As shown in Figure 10, the residue N318 site from the heavy chain (HC, Figure 10a) was chosen for examination, and Figure  <ref type="figure">10e</ref>) when the cell was turned on at +1.05 V, while after injection of the NIST mAb digest sample, four oxidation current peaks corresponding to four peptides were recorded shown in Figure <ref type="figure">10f</ref>.</p><p>At the same time, the oxidation products were monitored with online MS detection, which further confirmed the occurrence of electrochemical oxidation of those tryptophan-containing peptides.</p><p>For example, as shown in Figure <ref type="figure">10g</ref>, before electrolysis, a +3 ion of</p><p>VVSVLTVLHQDWLSuc318GK was observed at m/z 597.67; after electrolysis, the oxidation product ions at m/z 602.33 (M + 14 ion), m/z 603.00 (M + 16 ion), and m/z 608.33 (M + 32 ion) were observed in Figure <ref type="figure">10h</ref>. The amount of oxidized VVSVLTVLHQDWLSuc318GK was calculated as 0.050 pmol based on the charge (Q = 1.51 &#215; 10-8C, Figure <ref type="figure">10f</ref>) and the intensity ratios of the oxidation products observed. Based on the EIC peak change for m/z 597.67 upon oxidation (Figures <ref type="figure">10g</ref> and <ref type="figure">h</ref>), the oxidation yield for VVSVLTVLHQDWLSuc318GK was measured as 20.1% . Therefore, the total amount of VVSVLTVLHQDWLSuc318GK was measured to be 0.25 pmol. Simultaneously, the deamidated peptides as well as the unmodified peptide were also quantified by CMS. These results indicate that CMS can be used for monitoring mAb deamidation degradation and also for the quantitation of peptides with modifications when standards are hard to access. Furthermore, since deamidation is a nonenzymatic and spontaneous PTM which is also strongly associated with aging proteome instability and degenerative diseases, novel quantitation strategies such as CMS may facilitate the investigation of the disease mechanisms and the monitoring of the aging process, which would be also useful for forensic analysis. In addition, CMS was also used to quantify amyloid beta (A&#946;)-related peptides.</p><p>Alzheimer's disease (AD) is becoming a major public health problem worldwide. A&#946; peptide and its peptide fragments, derived from the proteolytic processing of amyloid precursor protein (APP), have been considered as the diagnostic biomarker and therapeutic target of AD. <ref type="bibr">[88]</ref><ref type="bibr">[89]</ref><ref type="bibr">[90]</ref> Therefore, quantification of those different A&#946; isoforms and their fragments is significant for early diagnosis of AD. Quantification also facilitates the investigation of disease mechanism and thus benefits drug discovery. MS is widely used and represents an important tool in the field of AD due to its capability of providing both qualitative and quantitative information on the A&#946; involved in AD.</p><p>However, for traditional MS-based absolute quantitation of A&#946; peptides, isotopically-labeled A&#946; peptides are often needed as internal standards, which are time-consuming and expensive. In one experiment, <ref type="bibr">[59]</ref> we reported successful quantifications of two long A&#946; peptide fragments, A&#946;1-16 and A&#946;1-28, using CMS based on the electrochemical oxidation of tyrosine residue without adding isotope labeling standards, for the first time. This result also demonstrated that CMS could also be used for very long peptides.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.3.">Protein quantitation</head><p>MS-based absolute quantitation of protein concentration in biological samples plays a significant role in clinical and pharmaceutical applications such as the evaluation of clinical biomarker candidates <ref type="bibr">[91]</ref> and biotherapeutic bioanalysis/DMPK. <ref type="bibr">[92]</ref> Popular methods for absolute protein quantitation by MS involve the digestion of target proteins and employ isotopelabeled peptide internal standards to quantify chosen surrogate peptides, which are mature and widely used in regulated bioanalysis. <ref type="bibr">[93]</ref> Although these methods have gained success, the synthesis of isotope-labeled peptides are time-consuming and costly. Therefore, it is only feasible to quantify a limited number of proteins at a time due to the high cost and complexity of sample preparation. Thus, the development of methods for accurate protein quantitation without the use of isotope-labeled standards is largely in need. To circumvent this limitation, CMS was also applied for protein quantitation by selecting a surrogate peptide containing electrochemically active residue (e.g., tyrosine, tryptophan and cysteine) to present the corresponding proteins. To evaluate the feasibility of CMS for protein quantitation, several model proteins like &#946;-casein, apomyoglobin, and cytochrome c were first quantified based on tyrosine or tryptophan-containing surrogate peptides and validated with traditional IDMS method with good accuracy. The basic workflow of CMS quantitation is very similar to the commonly used target proteomics quantitation  and <ref type="figure">h</ref>) with oxidation (applied potential: + 1.05 V); EICs of VLVLDTDYK were acquired i) without oxidation and j) with oxidation (applied potential: + 1.05 V). Reprinted with permission from Ref <ref type="bibr">[83]</ref>.</p><p>(e.g., SRM/PRM) including protein extraction, selective enrichment, tryptic digestion, chromatographic separation, and MS detection. To further expand the application of CMS for absolute quantitation of multiple proteins from a mixture sample in one run, three proteins, &#946;-lactoglobulin B (178 amino acids, MW 18,276 Da), &#945;-lactalbumin (142 amino acids, MW 14,175 Da), and carbonic anhydrase (260 amino acids, MW 29,100 Da), in a mixture sample were first digested, and three resulting oxidizable surrogate peptides containing either tyrosine or tryptophan, VLVLDTDYK, LDQWLCEK, and DGPLTGTYR, were selected for CMS analysis (Figure <ref type="figure">11</ref>). <ref type="bibr">[83]</ref> The CMS quantitation result was validated using a traditional isotope dilution method with good accuracy, indicating the capability of CMS for multiplex protein quantitation as long as multiple surrogate peptides representing different proteins can be LC-separated.</p><p>To further explore the CMS utility for real biological sample quantitation, two of the essential circadian clock proteins isolated from Escherichia coli, KaiB and KaiC, were also tested with CMS. <ref type="bibr">[57,</ref><ref type="bibr">83]</ref> The circadian clock is an endogenous timekeeping mechanism that provides many advantages for life in a rhythmically changing environment. Disruption of the circadian clock in humans is correlated with many health issues, such as cancer, heart attacks, obesity, diabetes, fatigue, mood disorders, and, most notably, jet lag. Till now, there is no commercially available protein standard for absolute quantitation. Because proteins are the major determinant of cell physiology, the timely measurement of the protein amount is extremely helpful in understanding the circadian regulation of gene expression and the physiology of cyanobacteria.</p><p>Therefore, the successful quantitation of KaiB <ref type="bibr">[57]</ref> and KaiC <ref type="bibr">[83]</ref> after purification suggests the potential of CMS for real sample quantitation.</p><p>Additionally, to further test the capability of CMS for large molecule quantitation, Vectibix (panitumumab, approximate molecular weight 147 kDa), a recombinant human IgG2 kappa monoclonal antibody that binds specifically to the human epidermal growth factor receptor (EGFR), was also tested and quantified with CMS. LLIYDASNLETGVPSR, a tyrosinecontaining peptide from antibody light chain was identified and separated by LC/MS analysis and used for quantitation by CMS. Without oxidation (Figure <ref type="figure">12a</ref>), peptide LLIYDASNLETGVPSR, chosen as the surrogate peptide, was observed at m/z 874.5 (+2 ion). As shown in Figure <ref type="figure">12b</ref>, after electrolysis, the oxidized product of LLIYDASNLETGVPSR (+2 ion) was detected at m/z 873.5, due to two hydrogen losses from tyrosine oxidation. Figure <ref type="figure">12c</ref> and <ref type="figure">d</ref> show the EIC the +2 ion of LLIYDASNLETGVPSR (m/z 874.5) from 3 &#956;L of 3.4 &#956;M IgG2 antibody digest (10.2 pmol) without and with oxidation, respectively. Note that one IgG2 molecule contains two identical light chains in which the surrogate peptide fragment is located. Therefore, the theoretical amount of surrogate peptide was 20.4 pmol per injection by calculation. The oxidation yield for LLIYDASNLETGVPSR was suggested to be 8.2%, by comparing the peak area of m/z 874.5 before and after oxidation (Figure <ref type="figure">12c</ref> and <ref type="figure">d</ref>). Meanwhile, the amount of the oxidized LLIYDASNLETGVPSR was calculated to be 1.28 pmol, based on the area integration of electric current peak observed from peptide oxidation (Figure <ref type="figure">12f</ref>). Therefore, the measured amount of LLIYDASNLETGVPSR was 15.7 pmol (triplicate average: 15.0 pmol). Compared to the theoretical amount (20.4 pmol) of this surrogate peptide, the measurement error was -26.4%. The successful quantification of mAbs without standards displays the immense potential of CMS for the quantification of large molecule biotherapeutics, which would be extremely useful in the pharmaceutical field like quantitative analysis in bioanalytical and clinical applications. Notably, a major disadvantage of surrogate peptide-based quantitation is the peptide recovery during preparation steps prior to MS, particularly for tryptic digestion. In the case of antibody quantitation by CMS after digestion, the relatively large error of -26.4% could be due to relatively low digestion efficiency for a large antibody protein peptide. <ref type="bibr">[94]</ref> Combining CMS with novel digestion strategies like using surfactants to improve digestion efficiency would improve quantitation accuracy. </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.4.">Quantifying drug impurities of antibody HCP proteins and nitrosamines</head><p>The fact that CMS eliminates the need for standards or a calibration curve makes it ideal as a rapid quantitation method for drug development and the process of monitoring and clearing high drug impurities in a timely manner. Host cell proteins (HCPs) are a major class of impurities derived from biotherapeutics production processes. HCPs that remain in the final drug product, especially for degrading or immunogenic proteins such as cathepsin D or phospholipase B like 2 (PLBL2), have been shown to affect product quality and patient safety due to the immune response. <ref type="bibr">[95,</ref><ref type="bibr">96]</ref> The current standard approach to quantify HCPs is based on enzyme-linked immunosorbent assay (ELISA) <ref type="bibr">[97,</ref><ref type="bibr">98]</ref> which has drawback in its detection selectivity. MS is emerging as an alternative way to characterize HCP contents, but the existing MS methods might be hard to implement since either peptide or protein internal standards are needed for HCP quantitation. These standards take a significant amount time to synthesize. It is extremely important to find a rapid quantitation method-without the use of standards-for timely process optimization and risk assessment so that high-risk HCPs can be tracked and cleared within days of discovery, especially during disease outbreaks such as the COVID-19 pandemic. CMS is an attractive method for this task because it does not require standards. As a demonstration, we applied CMS for the absolute quantitation of PLBL2 spiked in NIST mAb 8671 at two spiking levels (i.e., mAb/PLBL2 = 200:1 and 2000:1). <ref type="bibr">[83]</ref> The workflow was shown in Scheme 5b. NPALWK, a surrogate peptide identified from PLBL2 (sequence shown in Figure <ref type="figure">13a</ref>), was separated and quantified with CMS. When the cell was off (Figure <ref type="figure">13b</ref>), the +2 ion of NPALWK was observed at m/z 364.71. After electrolysis (Figure <ref type="figure">13c</ref>), the oxidation product peaks corresponding to M + 14, M + 16, and M + 32 were observed at m/z 371.70, 372.71, and 380.70, respectively. The oxidation yield for NPALWK was measured to be 16.3% by comparing the EIC peak area of m/z 364.71 before (Figure <ref type="figure">13d</ref>) and after oxidation (Figure <ref type="figure">13e</ref>). Figure <ref type="figure">13g</ref> shows the electric current peak from oxidation of NPALWK, from which the amount of oxidized NPALWK was calculated to be 0.64 pmol. Therefore, the measured amount of NPALWK was 3.9 pmol. For confirmation, an isotope-labeled peptide NPALWK &#770;(labeled at lysine, 13 C6, and 15 N4) was purchased and used as an internal standard to quantify NPALWK in the same digest sample. By this traditional isotope dilution method, NPALWK was measured to be 4.1 pmol. Again, the discrepancy of the quantitation results between CMS and IDMS was small (-4.9%). These results indicate that CMS could be used for absolute quantitation of very low abundant proteins (i.e., PLBL2) in the presence of a highly abundant predominant protein (NIST mAb 8671). CMS thus might be used to track the clearance of a specific known HCP under various process conditions after shotgun proteomics identification. No need for standards would make this CMS strategy more suitable to meet the need for rapid quantitation of HCPs after their identification. when the cell was off and e) when the cell was turned on (applied potential: +1.05 V). Electric current diagrams were collected from oxidation of f) blank solvent and g) NPALWK. Reprinted with permission from Ref <ref type="bibr">[83]</ref>.</p><p>CMS can also be used for quantifying small molecular drug impurities such as Nnitrosamines. Carcinogenic N-nitrosamines were recently found in the sartan family of drugs and caused many drug recalls. <ref type="bibr">[99]</ref> The detection and quantification of these chemicals is therefore especially important. Methods reported for N-nitrosamine quantitation rely on the use of standards and are just applicable to simple N-nitrosamines; <ref type="bibr">[100]</ref><ref type="bibr">[101]</ref><ref type="bibr">[102]</ref><ref type="bibr">[103]</ref> there is an urgent need to quantify Nnitrosamines derived from drugs with complicated structures that lack standards. In our study, <ref type="bibr">[60]</ref> CMS was applied for the absolute quantitation of six simple N-nitrosamines: Nnitrosodiethylamine (NDEA), N-nitroso-4-phenylpiperidine (NPhPIP), N-nitrosodiphenylamine (NDPhA), N-nitrosodibutylamine (NDBA), N-nitrosodipropylamine (NDPA), and Nnitrosopiperidine (NPIP). This quantification resulted in an excellent measurement accuracy (quantitation error &#8804;1.1%). Briefly, N-nitrosamine is first converted into electrochemically active hydrazine via zinc reduction under acidic condition and the resulting hydrazine can then be easily quantified using CMS (Scheme 6a). In addition, as a demonstration of the method utility, a drug-like N-nitrosamine, (R)-N-(2-(6-chloro-5-methyl-1&#8242;-nitroso-2,3-dihydrospiro[indene-1,4&#8242;-piperidin]-3-yl)propan-2yl)acetamide (N-nitrosamine VII), was also synthesized (Scheme 6b) and successfully quantified using CMS at 15 ppb level in a complex formulation matrix, following solvent extraction, Nnitrosamine isolation, and reductive conversion to hydrazine 7. As shown in Figures <ref type="figure">14 a</ref> and <ref type="figure">b</ref>, the oxidation product ion at m/z 348 had increased intensity when +0.3 V potential (vs Ag/AgCl) was applied to WE. The protonated hydrazine 7 ion of m/z 350 shown in Figure <ref type="figure">14d</ref> was smaller by 16.9% than that of the peak shown in Figure <ref type="figure">14c</ref>, suggesting the oxidation yield to be 16.9%.</p><p>On the other hand, the oxidation current peak was detected, as shown in Figure <ref type="figure">14f</ref> (Figure <ref type="figure">14e</ref> shows the background current diagram for the blank solvent sample under the same +0.3 V potential as a contrast). On the basis of the integration of the current peak area, the amount of the oxidized 7 was calculated to be 10.0 pmol (about 3.5 ng). Considering the extraction and isolation yield (43.9%), the side product ratio (4.4%), the reduction yield (58.1%), and the dilution factor 10, the CMS measured amount of N-nitrosamine VII was 410 pmol on the average from a triplicate measurement. The measurement error was -1.1% compared with the theoretical amount of 412 pmol. The standard-free feature of CMS makes it a simple and powerful approach for Nnitrosamine absolute quantitation. It has great potential for analysis of other drug impurities or metabolites, which could significantly impact pharmaceutical quality control, food safety analysis, and water treatment applications in the future. Although zinc-based chemical reduction shows some limitations, especially for controlling the N-nitrosamine and zinc reaction ratio, other reductants with better selectivity could overcome this issue. Besides, modified electrodes could be used for directly reducing or oxidizing N-nitrosamines, potentially simplifying the quantitation process. </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.">Challenges and future perspectives</head><p>To summarize, the method of CMS is surveyed in this review by addressing applications ranging from small molecule metabolites to large molecule biotherapeutics in a detailed manner.</p><p>Standard-free CMS is well suitable for the absolute quantitation of target molecules for which the generation of standards is often difficult or not available, such as small drug molecules with complicated structures and peptides with PTMs like phosphopeptides and deamidated peptides.</p><p>No need for standards and calibration curves would significantly shorten the process of method development compared to traditional quantitation strategies. It enables the fast-screening analysis during drug discovery as well as the rapid quantification and monitoring of high-risk drug impurities for timely process optimization and quality control. Moreover, coupling CMS with versatile direct ambient sampling methods like CPSI-MS makes it feasible for on-site quantitation and forensic analysis. The selective redox reaction could be achieved by specific modification (e.g., enzymes, antibodies, aptamers, lectins, or molecularly imprinted polymer cavities) on WE or simply by varying the electrochemical redox potential. However, although the above CMS publications showed promising results, we have to admit that there are some bottlenecks in terms of the current CMS format which need more development to be widely used. One major challenge is that real biological samples can sometimes be very complex. Although combining CMS with LC separation has been used in our previous work, resolving all electroactive target analytes as well as the interferences would still be challenging due to the fact that unlike MS, the overlapped current response cannot be differentiated. Thereby, for extremely complicated clinical samples such as biomarker quantitation in human body fluids and large molecule bioanalysis/DMPK, pre-treatment steps prior to CMS including immune-affinity enrichment, protein/peptide fractionation, or multidimensional online separation would be helpful and make CMS powerful tool to reduce time and cost, especially for large-scale quantitation. Another drawback of CMS is that only the analytes with electroactive functional groups could be quantified, thus hindering the expansion of CMS. To address this, we are developing electrochemically active tags which can covalently link to non-active target analytes such as peptides that lack Y, W, or C amino acid residues. For example, this tag can be designed with an amine-reactive group that specifically labels lysine residue, allowing the absolute quantitation of any tryptic peptides. We believe that this can make CMS an excellent quantitation tool for a very broad range of analytes regardless of the electrochemically reactive groups.</p><p>In terms of quantitation sensitivity, peptides such as WGG can be well quantified at an amount as low as 75 fmol. Further possible improvement in CMS quantitation sensitivity could be achieved using microelectrode instead of currently used millimeter sized electrodes. We believe this standard-free quantitation approach has the potential for more interesting applications in the near future.</p></div></body>