			<titleStmt><title level='a'>Tensile acoustic rheometry for rapid and contactless characterization of soft viscoelastic biomaterials</title></titleStmt>
			<publicationStmt>
				<publisher>science direct</publisher>
				<date>04/08/2025</date>
			</publicationStmt>
			<sourceDesc>
				<bibl> 
					<idno type="par_id">10647907</idno>
					<idno type="doi">10.1016/j.biomaterials.2025.123325</idno>
					<title level='j'>Biomaterials</title>
<idno>0142-9612</idno>
<biblScope unit="volume"></biblScope>
<biblScope unit="issue"></biblScope>					

					<author>Weiping Li</author><author>Eric C Hobson</author><author>Kiera Downey</author><author>Timothy L Hall</author><author>Jan P Stegemann</author><author>Cheri X Deng</author>
				</bibl>
			</sourceDesc>
		</fileDesc>
		<profileDesc>
			<abstract><ab><![CDATA[Accurately measuring the viscoelastic properties of biomaterials is critical for understanding their functions inbiological systems and optimizing their development for specific applications. Conventional methods oftenrequire direct physical contact, which hinders longitudinal studies of sterile samples and impose strict requirementsin sample preparation. Here, we introduce tensile acoustic rheometry (TAR), a technique for rapid,contactless characterization of soft viscoelastic biomaterials. TAR uses a dual-mode ultrasound approach to applyan upward force impulse, generating oscillatory tensile and compressive motion in a small, free-standing sample(~30 mm3) with its bottom immobilized on a pre-wetted flat surface by capillary stiction. High frequency ultrasoundpulse echo detection is employed to track this motion via the movement of the top surface of the samplein real time. In this study, we developed a theoretical framework of the tensile-compression motion of the samplefrom which Young’s modulus and viscosity of the sample are determined based on the TAR measurements. TARwas validated across a variety of samples, including engineered hydrogels and commercially available naturalfood products. Results from TAR measurements aligned closely with theoretical predictions, reported values, andshear wave elastography measurements. These findings underscore the versatility and flexibility of TAR as arobust, versatile rheological method for biomaterial characterization with minimal sample preparationrequirements.]]></ab></abstract>
		</profileDesc>
	</teiHeader>
	<text><body xmlns="http://www.tei-c.org/ns/1.0" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:xlink="http://www.w3.org/1999/xlink">
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="1.">Introduction</head><p>Biological tissues respond to mechanical stimuli in ways dependent on both elasticity (recoverable deformation) and viscosity (timedependent deformation) <ref type="bibr">[1]</ref><ref type="bibr">[2]</ref><ref type="bibr">[3]</ref><ref type="bibr">[4]</ref>. Engineered soft viscoelastic materials are used extensively in many elds including medicine, food science, and soft robotics <ref type="bibr">[5]</ref><ref type="bibr">[6]</ref><ref type="bibr">[7]</ref><ref type="bibr">[8]</ref><ref type="bibr">[9]</ref><ref type="bibr">[10]</ref><ref type="bibr">[11]</ref><ref type="bibr">[12]</ref>. For example, hydrogels composed of water-infused networks of hydrophilic polymer chains <ref type="bibr">[5,</ref><ref type="bibr">8]</ref> are an important class of soft biomaterials that can model tissue ECM <ref type="bibr">[7,</ref><ref type="bibr">[13]</ref><ref type="bibr">[14]</ref><ref type="bibr">[15]</ref>, since the interactions between the liquid and solid phases in hydrogels <ref type="bibr">[16]</ref> can recapitulate both the elastic and viscous components of the ECM as important effectors in controlling cell function <ref type="bibr">[13,</ref><ref type="bibr">14]</ref>. Detailed understanding and quantitative characterization of soft materials are critical to provide insights for their applications <ref type="bibr">[17,</ref><ref type="bibr">18]</ref>.</p><p>However, accurate measurement of the viscoelastic properties of biological tissue samples in vitro and soft biomaterials remains challenging, often involving time-consuming and tedious processes.</p><p>Conventional tensile, compressive, indentation, and shear-based rheological techniques are generally low throughput and often requires direct physical contact with the samples. The risk of sample contamination and damage <ref type="bibr">[19]</ref> makes these approaches not feasible for longitudinal characterization or monitoring of samples in sterile condition. Recent techniques such as microrheology have provided new capability of assessing local rheological properties at high spatial resolution by optically tracking the motion of dispersed probe particles within samples <ref type="bibr">[20]</ref> including cell-laden hydrogels <ref type="bibr">[21,</ref><ref type="bibr">22]</ref>. However, embedding exogenous tracers in samples may not always be feasible or desirable <ref type="bibr">[20]</ref>.</p><p>Here we report the development and validation of tensile acoustic rheometry (TAR) for mechanical characterization of soft viscoelastic materials using a dual-mode ultrasound technique, to generate and detect the tensile-compression movement of a free-standing sample without direct physical contact. A theoretical framework of the tensilecompression motion of the sample was developed to allow for the determination of Young's modulus and viscosity based on TAR measurements. TAR was validated across a variety of samples, including engineered hydrogels and natural food samples. These results demonstrate TAR as a robust approach for viscoelastic characterization requiring minimal sample preparation.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.">Materials and methods</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.1.">Cell culture</head><p>Normal human lung broblast cells (NHLF) (Lonza Inc., Walkersville, MD) were cultured in Dulbecco's-modi ed eagle medium (DMEM) supplemented with 10 % FBS and 1 % penicillin streptomycin. Human umbilical vein endothelial cells (HUVEC) were isolated from umbilical cords from the University of Michigan Mott Children's Hospital <ref type="bibr">[23]</ref>. HUVECs were plated in tissue culture ask and grown in fully supplemented VascuLife endothelial cell culture medium (Lifeline Cell Technology, Frederick, MD). Human bone marrow-derived mesenchymal stromal cells (MSC; RoosterBio, Inc.) were maintained in MSC growth medium (Lonza). Cells were maintained at 37 &#8226; C in 100 % humidi ed air with 5 % CO 2 . Medium was changed every other day.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.2.">Cellfree agarose and gelatin samples</head><p>A 20.0 mg/ml (2 %) agarose stock solution was prepared by dissolving agarose powder (Type 1-A; Sigma) into PBS with continuous heating and stirring. The solution was transferred to a 50 &#8226; C water bath and was diluted with pre-heated PBS to generate additional solutions at concentrations of 5.0 mg/ml (0.5 %) and 10.0 mg/ml (1 %). The solutions were placed in a 3d-printed container lled to three heights (4, 6 and 8 mm), and placed at room temperature for 1 h to solidify. A 100 mg/ml (10 %) gelatin stock solution was prepared by dissolving bovine gelatin powder (200 Bloom, 30 mesh; Gelatin Innovations) into PBS with continuous heating and stirring. The solution was diluted with preheated PBS to generate solutions of 50 mg/ml (5 %) and 75 mg/ml (7.5 %) gelatin concentrations. The solutions were placed in a 3Dprinted container with lled to three heights (4, 6 and 8 mm), and placed at room temperature for 1 h to solidify. Round and square punches (diameter 4, 6 and 8 mm) were used to create cylindrically and rectangular-shaped samples with different diameters from the solidi ed samples for TAR tests.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.3.">Fibrin gel with HUVECs and NHLFs</head><p>Fibrinogen (Sigma-Aldrich, St. Louis, MO) was dissolved in phosphate-buffered saline (PBS) at 37 &#8226; C (12.8 mg/mL active clottable protein), and then sterile ltered through a 0.22-&#956;m syringe lter (Millipore Sigma) and kept on ice until ready for use. One ml cell-laden gels were made by mixing brinogen 625 &#956;l brinogen, 20 &#956;l thrombin (Sigma, 50 Unit/ml), 255 &#956;l DMEM, 100 &#956;l FBS and 250k or 500k cells, respectively. The well mixed cell-brinogen suspension was quickly placed in a 24-well plate. After dispensing the hydrogel solution, the 24well plates were incubated at 37 &#8226; C for 30 min for gelation. A punch (diameter 8 mm) was used to create a disk-shaped sample. Fibrin gel with cells were incubated in fully supplemented VascuLife medium at 37 &#8226; C with 100 % humidi ed air containing 5 % CO 2 , and the medium was changed every other day.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.4.">Collagen gels with and without MSCs</head><p>Collagen type I (MP Biomedicals, Solon, OH) stock solution was prepared at 4.0 mg/ml in 0.02 N acetic acid. 500 &#956;l Collagen hydrogels ( nal concentration as 2.0 mg/ml) were made by mixing 150 &#956;l DMEM, 50 &#956;l FBS, 50 &#956;l NaOH (Sigma Aldrich) and 250 &#956;l of 4 mg/ml collagen type I stock solution. One ml cell-laden gels were made by mixing the collagen solution with 250 k or 500k MSCs, respectively. The mixture was pipetted into 48-wells plate and incubated at 37 &#8226; C for 30 min to solidify the mixture. After gelation, hydrogels were moved to 24-well plates and a punch (diameter 8 mm) was used to prepare a disk sample for experiments. The samples were then incubated in MSC growth medium at 37 &#8226; C and 5 % CO 2 . After 24 h, the media was changed to osteogenic media (Lonza) and the media was then changed every other day. Cell-free collagen gels were made in a similar way without addition of MSCs.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.5.">Food samples</head><p>Natural food samples were purchased from local supermarkets including Jello (Mango Flavor Pudding), soft and rm Tofu (House Food), chicken breast (Kirkland), and cooked pork blood (Asian Market). The food samples were cut into cylinders using a punch (diameter 6 mm).</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.6.">TAR setup</head><p>For TAR testing, a sample, typically in the shape of a cylinder was placed on a at smooth surface, e,g., a microscope glass slide, a standard cell culture dish, or a microplate. The sample was free standing except its bottom surface, which was immobilized on the supporting surface by capillary stiction provided by a thin liquid layer between the sample bottom and the supporting surface.</p><p>TAR utilizes a composite ultrasound transducer assembly consisting of two concentric, co-linearly aligned focused ultrasound transducers. The outer annular transducer (center frequency 1.5 MHz) was used as the excitation transducer and the inner transducer (center frequency 7 MHz) worked as the detection transducer. The ultrasound transducer assembly was placed beneath the sample and submerged in a water bath, which provided acoustic coupling for the transducers and the bottom of the supporting plate for the sample. The transducers were directed upward and aimed at the center or another designated location of the top surface of a sample. A 3D motion control platform (Velmex) was used for alignment and positioning of transducers. The excitation transducer was driven by a power ampli er (75A250; Ampli er Research) and a waveform generator (33220A; Agilent) to induce motion in the sample. The detection transducer was controlled by a pulser/receiver (5900 PR; Olympus) to operate in a pulse-echo mode to detect re ection signals from the sample surface, which was digitized with a model 5443 Picoscope (Pico Technology, St Neots, UK). The excitation and detection transducer were synchronized using a pulse generator (Model 565; BNC) for excitation and tracking of sample surface movement.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.7.">TAR measurement of displacement of top surface of a free standing sample</head><p>In TAR, the excitation transducer generates a short excitation pulse (duration 33 &#956;s, acoustic pressure 2.5 MPa) that travels through the sample from bottom to top and produces an upward impulse of force, in the form of the acoustic radiation force, on the top surface of the sample. This force impulse induces a tensile extension of the sample with its bottom xed on the supporting surface, followed by a tensile/ compression oscillatory motion, which is detected by the detection transducer via the movement of the top sample surface using pulse-echo detection at a high pulse repetition frequency (PRF) of 5 KHz, based on the traveling time of the detection pulse to and from the sample surface. Speci cally, to determine the surface movement from the echo signals, a short time window was placed around the surface re ection signal and the phase of the Fourier transform calculated over this time window at the frequency of maximum energy was used to calculate a relative displacement compared to pre-excitation. The MATLAB Phase Unwrap command was used across sequential phase measurements to allow for accurate displacement measurements beyond a single acoustic cycle (approximately 214 &#956;m) as long as the max velocity was less than one half cycle per imaging pulse period (approximately 0.5 m/s). A second Fourier transform of the resulting surface displacement data was performed to obtain the power spectrum of the surface movement, from which the peak frequency and half-width-at-half maximum (HWHM) were determined and used to calculate the shear modulus and viscosity of the sample using a lumped model of the tensile/compressional movement of the sample as described in supplemental information (SI).</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.8.">Shear wave elastography</head><p>Shear wave elastography (SWE) was performed to obtain an independent measure of elastic modulus E of the samples in our TAR experiments using a Logiq E10 imager (GE Healthcare) in elastography mode with an L8-18i probe at 14 MHz. For shear wave imaging, the sample was immersed in phosphate buffered saline (PBS) in a tissue culture dish or a microplate. The probe, secured using a custom-made holder, was placed on the bottom of the dish/microplate coupled with ultrasound gel. A region of interest (ROI) was de ned in the middle of the sample to ensure elastography measurements to minimize boundary effects. SWE measurements were then taken from a 0.5 mm 2 rectangular ROI at the center of the gel sample. The depth of the ROI from the bottom of the plate was kept the same for all measurements. For each sample, ve replicate measurements were taken and averaged to obtain the elastic modulus.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.9.">Finite element analysis (FEA) modeling of TAR using COMSOL</head><p>A COMSOL model was constructed to simulate the behavior of samples during TAR. A 2D axisymmetric model was created for the cylindrical geometry of a sample used in TAR. The Structural Mechanics module was used to model the sample as a linear viscoelastic material with xed boundaries on the bottom and a free boundary condition on other surfaces including the top surface. The parameters used in the simulation include sample dimension, Young's modulus, and viscosity, were based on the values from TAR experiments. The model was used to simulate the response of the sample to ultrasound excitation and determine the surface displacement at the center of the sample, which was measured in TAR.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="2.10.">Statistical analysis</head><p>All experiments were repeated for at least three different samples. For data complying with the normal distribution, two-sample t-test is applied. Nonparametric Mann-Whitney test is performed for the data which doesn't follow a normal distribution. A p-value &lt;0.05 was considered statistically signi cant.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.">Results</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.1.">Theoretical framework of the oscillatory tensile-compression motion of a free-standing sample immobilized by capillary stiction</head><p>TAR is optimized for testing small-volume samples, e.g., cylindrical in shape (e.g., ~3 mm radius and ~5 mm height), placed on a at surface that is pre-wetted with a small drop of liquid. For engineered samples typically stored in a cell culture dish or microplate with medium or PBS, the liquid was brie y removed for a few seconds during the TAR measurement which lasted a few seconds. The small amount of liquid formed a thin layer, or a disk-like plug (with radius R and thickness h), between the bottom of the sample and the supporting surface (Fig. <ref type="figure">1a</ref>). Due to the effects of surface tension, a capillary stiction force <ref type="bibr">[24,</ref><ref type="bibr">25]</ref> is</p><p>, that holds the sample in place on the supporting surface. Here &#963; represents the surface tension, &#952; 1 and &#952; 2 are the contact angle of the liquid plug with the support surface and bottom surface of the sample, respectively. The capillary stiction force immobilizes the bottom of the sample (mass m) on the supporting surface. Thus to lift the sample off the surface, an external upward force is required to overcome both the capillary stiction and the gravitational force (mg). The minimal separation force is given by Refs. <ref type="bibr">[24,</ref><ref type="bibr">25]</ref> F min = F s + mg. As an example, assuming R &#8811; h and &#952; 1 = &#952; 2 = 0 (perfectly wetted surfaces), &#963; = 0.035 Nm -1 (for cell culture medium), and h = 0.1 mm, the lower bound of F min for our experiments can be estimated to be 3 7 N, which is several orders of magnitude greater than the gravitational force of a typical sample in our study.</p><p>If the free-standing sample is subjected to an upward impulse force less than F min , the bottom of the sample remains rmly grounded on the supporting surface. However, if a short upward force impulse is applied to the sample with a xed bottom, the sample sustains an initial upward tensile stretch, resulting in a tensile strain and stress distribution within the continuum of the sample. Notably, the deformed sample will then undergo elastic recovery from the initial stretch, followed by a damped oscillatory tensile-compressive motion driven by the interplay of inertia, elastic restoring forces, and the sample's viscosity.</p><p>For a homogeneous, isotropic, free-standing sample with a xed bottom (cross section A, thickness or height L 0 , mass m, density &#961;, Young's modulus E, and dynamic viscosity &#951;), its vertical oscillatory tensile-compressive motion can be regarded as one-dimensional (1D) to rst order approximation. Assuming linear elasticity, the total restoring force F in the deformed sample is related to the total deformation &#916;l as</p><p>where k eq = EA L0 is the equivalent spring constant. This relationship reveals a Hookean behavior of the tensile-compression motion of the sample, which can be regarded as a mass-spring system. As detailed in the Supplemental Information (SI), the equivalent mass of this mass-spring system is derived as m eq = 1 3 m, assuming a linearly varying velocity pro le along the vertical axis of the sample from the xed bottom to the free top surface. The equivalent damping constant is derived as &#947; eq = A&#951; L0 , considering the viscous energy dissipation caused by the velocity gradient or relative motion within the sample. Thus the tensilecompressive motion of the sample can be modeled as a damped massspring system (Figs. 1b and SI) with a damping coef cient &#915; = &#947; eq 2meq = 3&#951; 2&#961;L 2 0 and intrinsic resonant frequency &#969; 0 = &#773;&#773;&#773;&#773;&#773; &#773; keq meq &#8730; = 1 L0 &#773;&#773;&#773;&#773; 3E &#961; &#8730; , respectively.</p><p>Since &#915; corresponds to the half frequency width at half maximum (HWHM) of the resonant frequency &#969; T , i.e., &#915; = &#916;&#969; 2 , the viscosity and Young's modulus can be obtained respectively as</p><p>and</p><p>) .</p><p>(2)</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.2.">TAR generates and detects tensile-compressional motion in a freestanding sample</head><p>In TAR, a dual-mode ultrasound technique is employed to induce and detect the oscillatory tensile-compressional motion of a free-standing sample. As shown by the example in Fig. <ref type="figure">2a</ref>, a short excitation ultrasound pulse (duration ~ 33 &#956;s) was applied from below, targeting the top surface of a cylindrical 5 % gelatin sample (radius a = 2.8 mm, height L 0 ). The excitation ultrasound pulse generated an upward impulse force in the form of the acoustic radiation force (ARF), or F ARP , acting predominantly on the top surface of the sample. Assuming a plane wavefront and perfect re ection at the air-sample interface, F ARP depends on the acoustic intensity of the excitation pulse I and is estimated as</p><p>c , where c is the speed of sound in the sample material <ref type="bibr">[26,</ref><ref type="bibr">27]</ref>, and r d is the half-width of the ultrasound beam at the focal point on the sample surface. For the excitation pulse used in this study (I = 1.33 &#215; 10 6 W&#8901;m -2 and r d &#8776; 0.1 &#215; 10 -3 m), F ARF &#8776; 6.0 &#215; 10 -3 N, which is signi cantly less than the lower bound of the capillary stiction force estimated in this study. Consequently, the bottom of the sample remained rmly grounded on the supporting surface during TAR, while F ARP induced an upward tensile stretch of the sample as described in the previous section of theoretical framework of TAR. The subsequent tensile-compressional motion, driven by the interlay of inertia, elastic restoring force, and viscosity, essentially represents the impulse response of the sample with a xed bottom surface that is modeled as a damped mass-spring system (SI).</p><p>In TAR, high-sensitivity pulse-echo ultrasound detection at a high pulse repetition frequency (PRF, e.g., 5 KHz), is used to detect the oscillatory tensile-compressive motion of the sample by tracking the sample's top surface displacement in real time based on the temporal shifts in the echo signal from the air-sample interface (Fig. <ref type="figure">2b</ref>). Initially, the height of the sample was determined using pulse-echo detection as L 0 = 5.43 mm, representing the equilibrium position before the application of the excitation pulse. After excitation, TAR detected the damped harmonic motion of the sample's surface (Fig. <ref type="figure">2c</ref>). The power spectrum of this motion exhibited a frequency peak at f 1 = 72.1 Hz with a full width at half maximum (FWHM) &#916;f 1 &#8776; 10.2 Hz (Fig. <ref type="figure">2d</ref>). From this, the damping coef cient was determined as &#915; = &#916;&#969; 2 = &#960;&#916;f 1 &#8776; 16.0 rad&#8901; s -1 , and the Young's modulus as E = (2&#960;f1L0) 2 &#961; 3 = 2.12 kPa and the dynamic viscosity &#951; = 2&#961;L 2 0 &#915; 3 = 0.33 Pa&#8901;s, using Eq. ( <ref type="formula">1</ref>) and Eq. ( <ref type="formula">2</ref>) and assuming &#961; = 1050 kgm -3 . These values are consistent with reported measurements for 5 % gelatin <ref type="bibr">[28,</ref><ref type="bibr">29]</ref>. Notably, real time measurements of the damped surface displacement, which lasted about 0.1 s-0.2 s in a typical samples in TAR, enabled the determination of Young's modulus and viscosity within a fraction of a second. COMSOL simulations were performed to model the dynamic behavior of the sample subjected to the excitation pulse, using the Young's modulus and viscosity values measured in TAR (SI). The simulated surface displacement and corresponding power spectrum agree closely with experimental TAR results (Fig. <ref type="figure">2c</ref> and <ref type="figure">d</ref>). The simulation also revealed an initial non-uniform surface displacement caused by the focused ultrasound beam, which transitioned to uniform displacement within 0.015 s (top, Fig. <ref type="figure">2e</ref>; Supplemental Video 1), after which 1D tensile-compressional oscillatory motion dominated (bottom, Fig. <ref type="figure">2e</ref>-Supplemental Video 2). These results suggest that the initial force impulse generated a surface wave (Rayleigh wave, RW), which rapidly decayed while the 1D tensile-compressional motion persisted longer. The power spectrum con rmed this interpretation, with a secondary peak at f 2 = 178 Hz potentially corresponding to a resonant mode of RW <ref type="bibr">[28,</ref><ref type="bibr">30,</ref><ref type="bibr">31]</ref>, consistent with previous ndings in resonant acoustic rheometry (RAR) <ref type="bibr">[28,</ref><ref type="bibr">30]</ref>. The angular frequency of RWs is related to the shear modulus G of the material and the wavenumber of the surface wave k as &#969; RW = &#773;&#773;&#773;&#773;&#773;&#773; Gk 2 &#961; &#8730; . For TAR, assuming a free boundary condition at r = a, the (0, 1) mode satis es J &#697; 0 (ka) = 0, resulting in ka = 3.832. From this, the shear modulus was determined to be G =</p><p>= 0. 70 kPa for f RW = f 2 = 178 Hz. Assuming the material was incompressible (Poisson ratio &#957; &#8776; 0.5), the Young's modulus was calculated as E = 3G = 2.1 kPa, in agreement with the value derived from f 1 = 72.05 Hz.</p><p>The viscosity of the sample was determined from the HWHM at f 2 <ref type="bibr">[28,</ref><ref type="bibr">30]</ref> using the damping coef cient of RW, &#915; RW &#8776; 0.45&#951;k 2 &#961; = &#960;&#916;f 2 .</p><p>Although &#916;f 2 was challenging to measure due to the low amplitude at f 2 (Fig. <ref type="figure">2d</ref>), the large ratio of the RW damping coef cients to the tensile-compression damping coef cient ( &#915;RW &#915; = 0.3(ka) 2 L 2 0 a 2 &#8776; 16.7) is consistent with the broader peak width at f 2 compared to f 1 , indicating signi cantly higher damping for RW than for tensile-compression motion.</p><p>While TAR is applicable for characterization of solid samples and RAR can characterize both viscoelastic liquid and viscoelastic solids, the consistent results shown in this example provided valuable validation of TAR against RAR measurements of the same sample.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.3.">TAR measurements of viscoelasticity are independent of sample diameter</head><p>We conducted a systematic comparison of the lumped mass-spring model and COMSOL nite element analysis (FEA) simulations to evaluate the effects of various parameters on the frequency spectrum, and thereby the determination of the viscosity and Young's modulus (Eqs. 1 and 2). Overall, the COMSOL simulations and the lumped mass-spring model produced consistent results. Speci cally, the peak frequency of the motion at the center of the sample surface was independent of the sample diameter (Fig. <ref type="figure">3a</ref>) and inversely proportional to the sample thickness (Fig. <ref type="figure">3b</ref>), as described by the relationship</p><p>derived from the mass-spring model. Since TAR directly measures the sample thickness, these results highlight the versatility and simplicity of determining Young's modulus using TAR for samples with varied dimensions.</p><p>While similar trends for the HWFM vs. sample thickness were observed, COMSOL simulations predicted slightly higher values than the damped mass-spring model (theory, Fig. <ref type="figure">3c</ref>). Importantly, consistent results were obtained from both COMSOL simulations and the damped mass-spring model across samples with varying Young's modulus (Fig. <ref type="figure">3d</ref>) and viscosity (Fig. <ref type="figure">3e</ref>). Higher HWHM values were observed in samples with smaller thickness, which corresponded to higher frequencies of tensile-compression motion (Fig. <ref type="figure">3f</ref>). Additionally, larger discrepancies between COMSOL simulations and mass-spring model predictions were noted in samples with lower viscosity (Fig. <ref type="figure">3f</ref>). These differences suggest that energy dissipation mechanisms modeled in COMSOL for a continuum, such as those associated with 3D movements within the sample beyond vertical motion, are not fully captured by the mass-spring model. Similarly, approximation in the mass-spring system, such as the assumption of a linear velocity pro le, may contribute to these discrepancies.</p><p>Taken together, these results validate and highlight the strength of TAR and the lumped mass-spring model for the rapid viscoelastic characterization of samples based on measurements of sample surface movement.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.4.">TAR provides rapid and robust measurements of viscoelastic properties</head><p>We conducted a series of experiments to evaluate TAR in samples of varying dimensions and properties. For cylindrical gelatin samples, TAR measurements demonstrated excellent agreement with predictions from the mass-spring model for the frequency of the surface motion across different sample diameters (Fig. <ref type="figure">4a</ref>) and thicknesses (Fig. <ref type="figure">4b</ref>). The Young's modulus (Fig. <ref type="figure">4c</ref>) and viscosity (Fig. <ref type="figure">4d</ref>) determined from TAR measurements showed that samples with higher gelatin concentrations were stiffer and more viscous, consistent with reported measurements <ref type="bibr">[28]</ref>. Similarly, TAR measurements for cylindrical agarose samples aligned well with mass mass-spring model predictions for frequency, Young's modulus and viscosity as functions of sample dimensions and agarose concentrations (Fig. <ref type="figure">4e-h</ref>). While the viscosity of agarose samples was comparable to that of gelatin samples, their Young's modulus was signi cantly higher, re ecting their greater stiffness. These results not only align with previously reported values <ref type="bibr">[28]</ref> but also highlight TAR's capability to measure a wide range of elasticity values, demonstrating its versatility for viscoelastic characterization.</p><p>We conducted additional TAR experiments with the ultrasound excitation and pulse-echo detection positioned at off-center locations on the sample surface. The results indicated that TAR measurements of surface displacement (Figs. <ref type="figure">S1a</ref> and <ref type="figure">S1b</ref>), as well as the derived Young's modulus and viscosity of the samples (Figs. <ref type="figure">S1c</ref> and <ref type="figure">S1d</ref>) were unaffected by the off-center excitation and detection. These ndings were consistent with the lumped mass-spring model of the tensilecompressional motion during TAR, demonstrating that the measurements are independent of the location of the applied forces on the sample surface. This underscores the validity of the 1D approximation and robustness of TAR.</p><p>TAR measurements using samples with a square surface also exhibited a consistent dominant frequency of surface displacement. While additional components were observed (Fig. <ref type="figure">S2a</ref>), they were present at much lower intensity levels (Fig. <ref type="figure">S2b</ref>). The Young's modulus measurements were unaffected by the shape of the sample surface, but viscosity measurements showed greater variability, particularly for lager square samples. This suggests that different modes of movement in uence the energy distribution at the primary frequency peak, which was used to estimate viscosity.  </p><p>mm) with 5 %, 7.5 %, and 10 % gelatin respectively. n = 18 for all groups. Comparison of TAR measurements (symbols) and mass-spring model prediction (solid lines) 0.5 %, 1.0 %, and 2.0 % agarose samples respectively in terms of frequency of surface movement vs. sample diameter (e) and thickness (f). Young's modulus (g) and viscosity (h) determined from TAR experiments in samples (Diameter 6 mm, thickness 6 mm) with 0.5 %, 1.0 %, and 2.0 % agarose respectively. n = 15 for all groups.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.5.">TAR measurements of Young's modulus compared with shear wave elastography</head><p>Shear wave elastography (SWE) is an ultrasound imaging technique used to assess the elasticity of soft tissues in vivo by measuring the shear wave velocity</p><p>&#8730; , where G is the shear modulus and &#961; is the density of the tissue <ref type="bibr">[32]</ref>. SWE has proven to be a valuable tool for diagnosing and monitoring diseases such as malignant breast lesions, liver brosis, and thyroid nodules <ref type="bibr">[32]</ref>. In this study, we conducted paired experiments on the same samples using SWE and TAR to compare their measurements of Young's modulus across a wide range of materials, including engineered hydrogels (Fig. <ref type="figure">5a</ref>) and commercially purchased food samples of various sizes (Fig. <ref type="figure">5b</ref>). For the food samples tested using both TAR and SWE, we also tested SWE in larger samples (SWE_L, Fig. <ref type="figure">5b</ref>) to examine whether the small size affected measurements based on propagating shear wave in SWE. The values of Young's modulus measured using TAR are slightly higher than those obtained with SWE in both small and larger sized samples, with differences within 10-15 %, which may be attributed to the irregularities in manual sample preparation for TAR. Despite these minor discrepancies, the strong general agreement between TAR and SWE validates the reliability of TAR and provides a useful reference for SWE measurements in complex in vivo environments.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="3.6.">Longitudinal measurements of Young's modulus of cell-laden hydrogels using TAR</head><p>We evaluated the utility of TAR for longitudinal assessment of cellladen hydrogels under sterile condition to monitor changes in their mechanical properties over time. These samples were kept in incubators in sterile condition and were taken out for less than 1 min for TAR measurements when medium was removed for several second for noncontact TAR measurements. After TAR, fresh medium was added and samples were placed into incubator until the next TAR measurement. Experiments were conducted on brin gels embedded with normal human lung broblasts (NHLFs) to detect changes in stiffness of the same samples over time. Compared to cell-free brin gel samples, which maintained constant thickness, cell-laden samples showed cell proliferation (Fig. <ref type="figure">S3a</ref>) and signi cant reduction in thickness dependent on cell seeding density, as indicated by the travel time of the echo signals from the sample surface (Fig. <ref type="figure">S3b</ref>). Samples with a cell seeding density of 1 million cells/ml underwent severe compaction, with thickness decreasing to less than 300 &#956;m within two days of culture, at which point TAR measurements were no longer feasible (Fig. <ref type="figure">S3b</ref>). Additionally, TAR measurements showed that the Young's modulus of cell-laden samples decreased over time, with greater reductions observed at higher cell seeding densities, while the cell-free gels remained mechanically unchanged (Fig. <ref type="figure">S3c</ref>).</p><p>In comparison, brin gels embedded with both NHLFs and HUVECs at a 1:1 ratio showed much less thickness reduction (less than 25 %) over 14 days. Concurrently, vessel-like structures formed within the constructs (top, Fig. <ref type="figure">6a</ref>), and the Young's modulus measured in the same samples decreased gradually over time. This softening effect was more pronounced in samples with higher cell seeding densities (bottom, Fig. <ref type="figure">6a</ref>). We further tested TAR on collagen samples embedded with mesenchymal stem cells (MSCs) cultured in osteogenic differentiation medium (top, Fig. <ref type="figure">6b</ref>). Cell proliferation was observed visually in these samples (top, Fig. <ref type="figure">6b</ref>). The Young's modulus of the collagen constructs began to increase around day 3, with larger increases detected in samples with higher cell densities (bottom, Fig. <ref type="figure">6b</ref>). These results reveal the impact of different cells on the mechanical properties of the hydrogelbased constructs.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.">Discussion</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.1.">Different mechanisms and applicability of TAR and RAR for viscoelastic characterization</head><p>We have previously reported the use of RAR, which also employs a dual-mode ultrasound technique, for non-contact characterization of both viscoelastic liquid and soft materials housed in standard 96-or 48 well microplates <ref type="bibr">[28,</ref><ref type="bibr">30]</ref>. While both TAR and RAR provide rapid ultrasound-based characterization of soft biomaterials, the techniques differ in their underlying physical principles and suitability for speci c applications.</p><p>In RAR, the samples are housed in a sample chamber, e.g. a well of a 96-well microplate, thus the samples are restricted in all boundaries expect the top surface. For these samples, application of an excitation ultrasound generates resonant modes of surface waves in the samples, which can be either liquid or solid, and can monitor phase transitions and the changes in viscoelastic properties in real time. Speci cally, RAR determines viscoelastic properties using the speci c dispersion relation of surface waves, which depends on parameters such as surface tension (for liquids), shear modulus (for solids), and viscosity. Because RAR relies on the formation of resonant surface wave, it is most effective for testing con ned samples with xed boundary conditions and a free surface. This makes RAR particularly suited for studying dynamically evolving biomaterials, such as during coagulation or polymerization/ gelation processes. However, RAR is less applicable for pre-formed samples with irregular shapes that lack con ned boundaries.</p><p>In contrast, TAR is suited for characterizing free-standing samples without stringent restrictions on sample shape. It determines the viscoelastic properties of solid materials based on tensile-compressional motion, which is governed by different equation of motion from RAR, besides the different boundary conditions. Even though such motion is within the continuum, it can be modeled as a lumped mass-spring system and can be assessed by measuring the overall strain or displacement of the sample in the vertical direction. Furthermore, the present of a thin liquid layer/plug between the supporting surface and the bottom of the sample, greatly alleviating the requirement of a highly smooth surface of the sample, as the gravity will pull down the soft sample to be stably positioned on top of the thin liquid layer which provides a capillary stiction to the sample. The tensile-compressive motion of the sample is largely 1D, thus TAR measurements are independent of the sample shape. As such, TAR is versatile, robust, and capable of analyzing a wide variety of samples, independent of the sample shape, as shown in our experimental measurements and theoretical framework. Interestingly, despite the differences in the physical principles underlining RAR and TAR, our results demonstrate consistent measurements of viscoelastic properties between the two techniques, underscoring the reliability and accuracy of both approaches.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.2.">Study limitations and validation of TAR measurements</head><p>The main goal of this study is to demonstrate the capability and operation principle of TAR as a new approach for viscoelastic characterization. Thus, our experiments with hydrogel samples and natural food samples focused on the measurements of their viscoelastic properties, with a limited scope in terms of investigating the mechanisms responsible for the changing viscoelastic properties of the cell-laden hydrogel constructs. We envision future studies will pursue these important objectives once TAR is established.</p><p>In addition, while it is worthwhile to validate TAR by comparing results with those using conventional techniques such as shear rheometry measurement of the same sample. However, since the underlying mechanisms of these techniques are distinctively different from TAR, direct comparison of results thus is challenging and may not be easily obtained as measurement conditions and parameters are not translatable. It is possible to compare TAR and RAR results of the same samples, as shown in the example in Fig. <ref type="figure">2</ref>, but RAR is not optimized for testing of free-standing samples as the tensile-compression movement of the samples dominates the measured signals. Thus, RAR has a much lower signal to noise ratio in the free-standing samples. Furthermore, RAR requires a circular shape to form clean resonant modes of surface waves.</p><p>Furthermore, although SWE is not a conventional method used for typical rheology measurements, SWE has been established as a noncontact measurement technique for tissue elasticity, thus providing a convenient and meaningful validation for our TAR measurements of modulus of the same samples in this study.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="4.3.">Limitations of TAR and considerations for dimensional scaling effects</head><p>In this study, we performed TAR experiments in samples of small volumes with a range of heights, which determined the frequency of the oscillatory tensile-compressive motion of the sample. Thus, the viscosity and modulus measured of the sample was related to this speci c frequency. Even though it is theoretically possible to measure samples of the same materials but different heights to investigate the impact of frequency on the viscoelasticity, the range of frequency is limited by the heights that are attainable in practice. This limits the use of TAR for cases when extremely high strain rate measurements are desired.</p><p>Experiments in this study show that TAR was unable to reliably measure samples with thickness less than 0.3 mm. This limitation is partially related to the frequency of the detection ultrasound transducer used in our prototype system (7 MHz), which corresponds to a wavelength of 0.21 mm. Implementing a higher frequency detection system could overcome this limitation for thinner samples. For instance, a detection frequency of 10 MHz would yield a wavelength of 0.15 mm, making TAR feasible for measurement of thinner samples. However, the application of TAR for such thinner samples is also limited as the thickness approaches the thickness of the liquid plug between the sample and the supporting surface as well as the dif culty in generating tensile-compression motion of a thin sample.</p><p>The accuracy of TAR measurements can be compromised in samples with large cross section area but low stiffness, since 1D assumption is made for the tensile-compression motion in the lumped mass-spring model. Samples with small diameters, or high aspect ratios (thickness &#8811; diameter) will behave more closely as 1D objects, with unidirectional deformation primarily along the vertical axis. However, larger samples have greater freedom to deform in multiple dimensions, leading to multi-axial stress and strain distributions. This behavior reduces the accuracy of the mass-spring model for larger samples. Such dimensional scaling effects are dependent on sample stiffness and represent an important area for future investigation to re ne TAR's applicability and accuracy for various sample Fig. <ref type="figure">6</ref>. Longitudinal TAR measurements of Young's modulus of the same samples at different time points. a. Top: Bright eld images of a brin gel (8 mg/ml) embedded with normal human lung broblasts (NHLFs) and HUVECs at 1:1 ratio with total cell density of 500 k cells/ml. Samples were stained with DAPI (blue) and UEA-I (red) on day 14. Bottom: Measured Young's modulus of cell-free brin gels, brin gels with NHLFs and HUVECs at 250 k cells/ml and 500 k cells/ml respectively. n = 8 for all groups. b. Top: Collagen gels embedded with mesenchymal stem cells (MSCs) at 500 k cells/ml in osteogenic medium. Bottom: Measured Young's modulus of cell-free collagen gels, collagen gels with MSCs at 250 k cells/ml and 500 k cells/ml in osteogenic medium respectively. (For interpretation of the references to colour in this gure legend, the reader is referred to the Web version of this article.) geometry and mechanical properties.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head n="5.">Conclusions</head><p>This study demonstrates the capability of TAR for the characterization of free-standing soft biomaterial samples in small volumes. By utilizing ultrasound-based excitation and detection approach, TAR offers a non-contact approach to quantify the viscoelastic properties of samples of various dimensions and shapes, providing a robust tool for longitudinal monitoring of samples under sterile conditions over time. These capabilities enable TAR for diverse laboratory and industrial applications. </p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>CRediT authorship contribution statement</head></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Declaration of generative AI and AI-assisted technologies in the writing process</head><p>No generative AI and AI-assisted technologies were used in the writing process.</p></div>
<div xmlns="http://www.tei-c.org/ns/1.0"><head>Declaration of competing interest</head><p>The authors declare that they have no known competing nancial interests or personal relationships that could have appeared to in uence the work reported in this paper.</p></div></body>