<?xml-model href='http://www.tei-c.org/release/xml/tei/custom/schema/relaxng/tei_all.rng' schematypens='http://relaxng.org/ns/structure/1.0'?><TEI xmlns="http://www.tei-c.org/ns/1.0">
	<teiHeader>
		<fileDesc>
			<titleStmt><title level='a'>Control of Powered Ankle–Foot Prostheses on Compliant Terrain: A Quantitative Approach to Stability Enhancement</title></titleStmt>
			<publicationStmt>
				<publisher>MDPI</publisher>
				<date>02/01/2026</date>
			</publicationStmt>
			<sourceDesc>
				<bibl> 
					<idno type="par_id">10698763</idno>
					<idno type="doi">10.3390/act15020107</idno>
					<title level='j'>Actuators</title>
<idno>2076-0825</idno>
<biblScope unit="volume">15</biblScope>
<biblScope unit="issue">2</biblScope>					

					<author>Chrysostomos Karakasis</author><author>Camryn Scully</author><author>Robert Salati</author><author>Panagiotis Artemiadis</author>
				</bibl>
			</sourceDesc>
		</fileDesc>
		<profileDesc>
			<abstract><ab><![CDATA[<p>Walking on compliant terrain presents a substantial challenge for individuals with lower-limb amputation, further elevating their already high risk of falling. While powered ankle–foot prostheses have demonstrated adaptability across speeds and rigid terrains, control strategies optimized for soft or compliant surfaces remain underexplored. This work experimentally validates an admittance-based control strategy that dynamically adjusts the quasi-stiffness of powered prostheses to enhance gait stability on compliant ground. Human subject experiments were conducted with three healthy individuals walking on two bilaterally compliant surfaces with ground stiffness values of 63 and25kNm, representative of real-world soft environments. Controller performance was quantified using phase portraits and two walking stability metrics, offering a direct assessment of fall risk. Compared to a standard phase-variable controller developed for rigid terrain, the proposed admittance controller reduced short-term maximum Lyapunov exponents by an average of 7%, indicating improved local dynamic stability. These results support the potential of adaptive prostheses control to enhance gait stability on compliant surfaces, contributing to the development of more robust human–prosthesis interaction.</p>]]></ab></abstract>
		</profileDesc>
	</teiHeader>
	<text><body><div/></body></text>
</TEI>
