Title: Optimizing the sensitivity of high repetition rate broadband transient optical spectroscopy with modified shot-to-shot detection
A major limitation of transient optical spectroscopy is that relatively high laser fluences are required to enable broadband, multichannel detection with acceptable signal-to-noise levels. Under typical experimental conditions, many condensed phase and nanoscale materials exhibit fluence-dependent dynamics, including higher order effects such as carrier–carrier annihilation. With the proliferation of commercial laser systems, offering both high repetition rates and high pulse energies, have come new opportunities for high sensitivity pump-probe measurements at low pump fluences. However, experimental considerations needed to fully leverage the statistical advantage of these laser systems have not been fully described. Here, we demonstrate a high repetition rate, broadband transient spectrometer capable of multichannel shot-to-shot detection at 90 kHz. Importantly, we find that several high-speed cameras exhibit a time-domain fixed pattern noise resulting from interleaved analog-to-digital converters, which is particularly detrimental to the conventional “ON/OFF” modulation scheme used in pump-probe spectroscopy. Using a modified modulation and data processing scheme, we achieve a noise level of 10−5 in 4 s for differential transmission, an order of magnitude lower than for commercial 1 kHz transient spectrometers for the same acquisition time. We leverage the high sensitivity of this system to measure the differential transmission of monolayer graphene at low pump fluence. We show that signals on the order of 10−6 OD can be measured, enabling a new data acquisition regime for low-dimensional materials.  more » « less
Award ID(s):
2112550 2004683
PAR ID:
10422401
Author(s) / Creator(s):
; ; ;
Date Published:
Journal Name:
Review of Scientific Instruments
Volume:
94
Issue:
4
ISSN:
0034-6748
Format(s):
Medium: X
Sponsoring Org:
National Science Foundation
More Like this
  1. Over the past decade, the proliferation of pulsed laser sources with high repetition rates has facilitated a merger of ultrafast time-resolved spectroscopy with imaging microscopy. In transient absorption microscopy (TAM), the excited-state dynamics of a system are tracked by measuring changes in the transmission of a focused probe pulse following photoexcitation of a sample. Typically, these experiments are done using a photodiode detector and lock-in amplifier synchronized with the laser and images highlighting spatial heterogeneity in the TAM signal are constructed by scanning the probe across a sample. Performing TAM by instead imaging a spatially defocused widefield probe with a multipixel camera could dramatically accelerate the acquisition of spatially resolved dynamics, yet approaches for such widefield imaging generally suffer from reduced signal-to-noise due to an incompatibility of multipixel cameras with high-frequency lock-in detection. Herein, we describe implementation of a camera capable of high-frequency lock-in detection, thereby enabling widefield TAM imaging at rates matching those of high repetition rate lasers. Transient images using a widefield probe and two separate pump pulse configurations are highlighted. In the first, a widefield probe was used to image changes in the spatial distribution of photoexcited molecules prepared by a tightly focused pump pulse, while in the second, a widefield probe detected spatial variations in photoexcited dynamics within a heterogeneous organic crystal excited by a defocused pump pulse. These results highlight the ability of high-sensitivity lock-in detection to enable widefield TAM imaging, which can be leveraged to further our understanding of excited-state dynamics and excitation transport within spatially heterogeneous systems. 
    more » « less
  2. ABSTRACT We report single-shot transient absorption (SSTA) measurements of an organic film of 3,3’-Diethyloxatricarbocyanine iodide (DOTCI). In SSTA, the pump-probe time delay is spatially encoded by using a tilted pump pulse. Translation of the sample during SSTA measurements averages over any spatial heterogeneity in the film. We demonstrate that exciton dynamics measured with the single-shot technique agrees with traditional transient absorption measurements of the same film. A signal-to-noise ratio of ∼40 is achieved in 10 s. The ability to measure exciton dynamics in organic films will enable future SSTA measurements of exciton dynamics during the molecular aggregation events that result in film formation. 
    more » « less
  3. This manuscript describes the first application of ultrafast-laser-absorption spectroscopy (ULAS) to characterizing high-pressure (up to 40 bar), multi-phase combustion gases. Single-shot measurements of temperature and CO were acquired at 5 kHz in AP-HTPB propellant flames with and without aluminum. An ultrafast light source was used to produce broadband pulses of light near 4.96 𝜇m at a repetition rate of 5 kHz and a high-speed mid-infrared imaging spectrometer was used to image the pulses across an 86 nm bandwidth with a spectral resolution of 0.7 nm. Measurements of temperature and CO concentration were obtained by least-squares fitting simulated absorbance spectra of CO to measured spectra. A system of corrective optics was used to diminish the e˙ect of beam steering during high-pressure experiments, greatly increasing the pressure capabilities of the diagnostic. The diagnostic was used to characterize AP-HTPB propellant flames in an argon bath gas at pressures of 1, 10, 20, and 40 bar. An aluminized AP-HTPB propellant was also characterized at 10 and 20 bar to demonstrate that ULAS can provide high-fidelity measurements in particulate-laden flames. The results demonstrate that ULAS is capable of providing single-shot temperature and species measurements at high pressures with 1-𝜎 precisions less than 1.1% and 3% for temperature and species respectively, despite non-absorbing transmission losses in excess of 90%. 
    more » « less
  4. Mid-infrared microscopy is an important tool for biological analyses, allowing a direct probe of molecular bonds in their low energy landscape. In addition to the label-free extraction of spectroscopic information, the application of broadband sources can provide a third dimension of chemical specificity. However, to enable widespread deployment, mid-infrared microscopy platforms need to be compact and robust while offering high speed, broad bandwidth, and high signal-to-noise ratio. In this study, we experimentally showcase the integration of a broadband, high-repetition-rate dual-comb spectrometer (DCS) in the mid-infrared range with a scanning microscope. We employ a set of 1-GHz mid-infrared frequency combs, demonstrating their capability for high-speed and broadband hyperspectral imaging of polymers and ovarian tissue. The system covers 1000 cm−1 at νc = 2941 cm−1 with 12.86 kHz spectra acquisition rate and 5 µm spatial resolution. Taken together, our experiments and analysis elucidate the trade-off between bandwidth and speed in DCS as it relates to microscopy. This provides a roadmap for the future advancement and application of high-repetition-rate DCS hyperspectral imaging. 
    more » « less
  5. BoxCARS and pump-probe geometries are common implementations of two-dimensional infrared (2D IR) spectroscopy. BoxCARS is background-free, generally offering greater signal-to-noise ratio, which enables measuring weak vibrational echo signals. Pulse shapers have been implemented in the pump-probe geometry to accelerate data collection and suppress scatter and other unwanted signals by precise control of the pump-pulse delay and carrier phase. Here, we introduce a 2D-IR optical setup in the BoxCARS geometry that implements a pulse shaper for rapid acquisition of background-free 2D IR spectra. We show a signal-to-noise improvement using this new fast-scan BoxCARS setup versus the pump-probe geometry within the same configuration. 
    more » « less