The novel coronavirus SARS-CoV-2 was first isolated in late 2019; it has spread to all continents, infected over 700 million people, and caused over 7 million deaths worldwide to date. The high transmissibility of the virus and the emergence of novel strains with altered pathogenicity and potential resistance to therapeutics and vaccines are major challenges in the study and treatment of the virus. Ongoing screening efforts aim to identify new cases to monitor the spread of the virus and help determine the danger connected to the emergence of new variants. Given its sensitivity and specificity, nucleic acid amplification tests (NAATs) such as RT-qPCR are the gold standard for SARS-CoV-2 detection. However, due to high costs, complexity, and unavailability in low-resource and point-of-care (POC) settings, the available RT-qPCR assays cannot match global testing demands. An alternative NAAT, RT-LAMP-based SARS-CoV-2 detection offers scalable, low-cost, and rapid testing capabilities. We have developed an automated RT-LAMP-based microfluidic chip that combines the RNA isolation, purification, and amplification steps on the same device and enables the visual detection of SARS-CoV-2 within 40 min from saliva and nasopharyngeal samples. The entire assay is executed inside a uniquely designed, inexpensive disposable microfluidic chip, where assay components and reagents have been optimized to provide precise and qualitative results and can be effectively deployed in POC settings. Furthermore, this technology could be easily adapted for other novel emerging viruses.
more »
« less
RT-LAMP-Based Molecular Diagnostic Set-Up for Rapid Hepatitis C Virus Testing
Hepatitis C virus (HCV) infections occur in approximately 3% of the world population. The development of an enhanced and extensive-scale screening is required to accomplish the World Health Organization’s (WHO) goal of eliminating HCV as a public health problem by 2030. However, standard testing methods are time-consuming, expensive, and challenging to deploy in remote and underdeveloped areas. Therefore, a cost-effective, rapid, and accurate point-of-care (POC) diagnostic test is needed to properly manage the disease and reduce the economic burden caused by high case numbers. Herein, we present a fully automated reverse-transcription loop-mediated isothermal amplification (RT-LAMP)-based molecular diagnostic set-up for rapid HCV detection. The set-up consists of an automated disposable microfluidic chip, a small surface heater, and a reusable magnetic actuation platform. The microfluidic chip contains multiple chambers in which the plasma sample is processed. The system utilizes SYBR green dye to detect the amplification product with the naked eye. The efficiency of the microfluidic chip was tested with human plasma samples spiked with HCV virions, and the limit of detection observed was 500 virions/mL within 45 min. The entire virus detection process was executed inside a uniquely designed, inexpensive, disposable, and self-driven microfluidic chip with high sensitivity and specificity.
more »
« less
- Award ID(s):
- 1942487
- PAR ID:
- 10448706
- Date Published:
- Journal Name:
- Biosensors
- Volume:
- 12
- Issue:
- 5
- ISSN:
- 2079-6374
- Page Range / eLocation ID:
- 298
- Format(s):
- Medium: X
- Sponsoring Org:
- National Science Foundation
More Like this
-
-
Hepatitis C virus (HCV), a member of the Flaviviridae family, is an RNA virus enclosed in an envelope that infects approximately 50 million people worldwide. Despite its significant burden on public health, no vaccine is currently available, and many individuals remain unaware of their infection due to the often asymptomatic nature of the disease. Early detection of HCV is critical for initiating curative treatments, which can prevent long-term complications such as cirrhosis, liver cancer, and decompensated liver disease. However, conventional diagnostic approaches available, such as enzyme immunoassays (EIAs) and polymerase chain reaction (PCR)-based methods, are often costly, time-intensive, and challenging to be implemented in resource-limited settings. This review provides an overview of HCV disease and the structural components of the virus, illustrating how different diagnostic methods target various parts of the viral structure. It examines current diagnostic tests and assays, highlighting their mechanisms, applications, and limitations, which necessitates the development of improved detection methods. Additionally, the paper explores emerging technologies in HCV detection that could offer affordable, accessible, and easy-to-use diagnostic solutions, particularly for deployment in low-resource and point-of-care settings. These advancements have the potential to contribute significantly to achieving the World Health Organization’s (WHO) target of eliminating HCV as a public threat by 2030.more » « less
-
NA (Ed.)Rapid detection of plant diseases before they escalate can improve disease control. Our team has developed rapid nucleic acid extraction methods with microneedles (MN) and combined these with LAMP assays for pathogen detection in the field. In this work, we developed LAMP assays for early blight (Alternaria linariae, A. alternata, and A. solani) and bacterial spot of tomato (Xanthomonas perforans) and validated these LAMP assays and two previously developed LAMP assays for tomato spotted wilt virus and late blight. Tomato plants were inoculated and disease severity was measured. Extractions were performed using MN and LAMP assays were run in tubes (with hydroxynaphthol blue) on a heat block or on a newly designed microfluidic slide chip on a heat block or a slide heater. Fluorescence on the microfluidic chip slides was visualized using EvaGreen and photographed on a smartphone. Plants inoculated with X. perforans or tomato spotted wilt virus tested positive prior to visible disease symptoms, while P. infestans and A. linariae were detected at the time of visual disease symptoms. LAMP assays were more sensitive than PCR and the limit of detection was 1 pg of DNA for both A. linariae and X. perforans. The LAMP assay designed for early blight detected all three species of Alternaria that infect tomato and is thus an Alternaria spp. assay. This study demonstrates the utility of rapid MN extraction followed by LAMP on a microfluidic chip for rapid diagnosis of four important tomato pathogens.more » « less
-
The COVID-19 pandemic demonstrated the public health benefits of reliable and accessible point-of-care (POC) diagnostic tests for viral infections. Despite the rapid development of gold-standard reverse transcription polymerase chain reaction (RT-PCR) assays for SARS-CoV-2 only weeks into the pandemic, global demand created logistical challenges that delayed access to testing for months and helped fuel the spread of COVID-19. Additionally, the extreme sensitivity of RT-PCR had a costly downside as the tests could not differentiate between patients with active infection and those who were no longer infectious but still shedding viral genomes. To address these issues for the future, we propose a novel membrane-based sensor that only detects intact virions. The sensor combines affinity and size based detection on a membrane-based sensor and does not require external power to operate or read. Specifically, the presence of intact virions, but not viral debris, fouls the membrane and triggers a macroscopically visible hydraulic switch after injection of a 40 μL sample with a pipette. The device, which we call the μSiM-DX (microfluidic device featuring a silicon membrane for diagnostics), features a biotin-coated microslit membrane with pores ∼2–3× larger than the intact virus. Streptavidin-conjugated antibody recognizing viral surface proteins are incubated with the sample for ∼1 hour prior to injection into the device, and positive/negative results are obtained within ten seconds of sample injection. Proof-of-principle tests have been performed using preparations of vaccinia virus. After optimizing slit pore sizes and porous membrane area, the fouling-based sensor exhibits 100% specificity and 97% sensitivity for vaccinia virus ( n = 62). Moreover, the dynamic range of the sensor extends at least from 10 5.9 virions per mL to 10 10.4 virions per mL covering the range of mean viral loads in symptomatic COVID-19 patients (10 5.6 –10 7 RNA copies per mL). Forthcoming work will test the ability of our sensor to perform similarly in biological fluids and with SARS-CoV-2, to fully test the potential of a membrane fouling-based sensor to serve as a PCR-free alternative for POC containment efforts in the spread of infectious disease.more » « less
-
Rapid, sensitive and specific detection and reporting of infectious pathogens is important for patient management and epidemic surveillance. We demonstrated a point-of-care system integrated with a smartphone for detecting live virus from nasal swab media, using a panel of equine respiratory infectious diseases as a model system for corresponding human diseases such as COVID-19. Specific nucleic acid sequences of five pathogens were amplified by loop-mediated isothermal amplification on a microfluidic chip and detected at the end of reactions by the smartphone. Pathogen-spiked horse nasal swab samples were correctly diagnosed using our system, with a limit of detection comparable to that of the traditional lab-based test, polymerase chain reaction, with results achieved in ∼30 minutes.more » « less
An official website of the United States government