Title: Genetic algorithm search for the worst‐case MRI RF exposure for a multiconfiguration implantable fixation system modeled using artificial neural networks
PurposeThis paper presents a method to search for the worst‐case configuration leading to the highest RF exposure for a multiconfiguration implantable fixation system under MRI. MethodsA two‐step method combining an artificial neural network and a genetic algorithm is developed to achieve this purpose. In the first step, the level of RF exposure in terms of peak 1‐g and/or 10‐g averaged specific absorption rate (SAR1g/10g), related to the multiconfiguration system, is predicted using an artificial neural network. A genetic algorithm is then used to search for the worst‐case configuration of this multidimensional nonlinear problem within both the enumerated discrete sample space and generalized continuous sample space. As an example, a generic plate system with a total of 576 configurations is used for both 1.5T and 3T MRI systems. ResultsThe presented method can effectively identify the worst‐case configuration and accurately predict the SAR1g/10gwith no more than 20% of the samples in the studied discrete sample space, and can even predict the worst case in the generalized continuous sample space. The worst‐case prediction error in the generalized continuous sample space is less than 1.6% for SAR1gand less than 1.3% for SAR10gcompared with the simulation results. ConclusionThe combination of an artificial neural network with genetic algorithm is a robust technique to determine the worst‐case RF exposure level for a multiconfiguration system, and only needs a small amount of training data from the entire system.  more » « less
Award ID(s):
1440107
PAR ID:
10454358
Author(s) / Creator(s):
 ;  ;  ;  ;  
Publisher / Repository:
Wiley Blackwell (John Wiley & Sons)
Date Published:
Journal Name:
Magnetic Resonance in Medicine
Volume:
84
Issue:
5
ISSN:
0740-3194
Page Range / eLocation ID:
p. 2754-2764
Format(s):
Medium: X
Sponsoring Org:
National Science Foundation
More Like this
  1. Abstract PurposeRecent numerical and empirical results proved that high permittivity materials (HPM) used in pads placed near the subject or directly integrated with coils can increase the SNR and reduce the specific absorption rate (SAR) in MRI. In this paper, we propose an analytical investigation of the effect on the magnetic field distribution of a layer of HPM surrounding an anatomy‐mimicking cylindrical sample. MethodsThe study is based on a reformulation of the Mie scattering for cylindrical geometry, following an approach recently introduced for spherical samples. The total field in each medium is decomposed in terms of inward and outward electromagnetic waves, and the fields are expressed as series of cylindrical harmonics, whose coefficients can be interpreted as classical reflection and transmission coefficients. ResultsOur new formulation allows a quantitative evaluation of the effect of the HPM layer for varying permittivity and thickness, and it provides an intuitive understanding of such effect in terms of propagation and scattering of the RF field. ConclusionWe show how HPM can filter out the modes that only contribute to the noise or RF power deposition, resulting in higher SNR or lower SAR, respectively. Our proposed framework provides physical insight on how to properly design HPM for MRI applications. 
    more » « less
  2. PurposeThe paper presents a novel method to reduce the RF‐induced heating of active implantable medical devices during MRI. MethodsWith the addition of an energy decoying and dissipating structure, RF energy can be redirected toward the dissipating rings through the decoying conductor. Three lead groups (45 cm‐50 cm) and 4 (50 cm‐100 cm) were studied in 1.5 Tesla MR systems by simulation and measurement, respectively. In vivo modeling was performed using human models to estimate the RF‐induced heating of an active implantable medical device for spinal cord treatment. ResultIn the simulation study, it was shown that the peak 1g‐averaged specific absorption rate near the lead‐tips can be reduced by 70% to 80% compared to those from the control leads. In the experimental measurements during a 2‐min exposure test in a 1.5 Telsa MR system, the temperature rises dropped from the original 18.3℃, 25.8℃, 8.1℃, and 16.1℃ (control leads 1‐4) to 5.4℃, 6.9℃, 1.6℃, and 3.3℃ (leads 1‐4 with the energy decoying and dissipation structure). The in vivo calculation results show that the maximum induced temperature rise among all cases can be substantially reduced (up to 80%) when the energy decoying and dissipating structures were used. ConclusionOur studies confirm the effectiveness of the novel technique for a variety of scanning scenarios. The results also indicate that the decoying conductor length, number of rings, and ring area must be carefully chosen and validated. 
    more » « less
  3. PurposeThe Radiofrequency (RF)‐induced heating for an active implantable medical device (AIMD) with dual parallel leads is evaluated in this paper. The coupling effects between dual parallel leads are studied via simulations and experiments methods. The global transfer function technique is used to assess the RF‐induced heating for dual‐lead AIMDs inside four human body models. MethodsRF‐induced heating for spinal cord stimulator systems with 60 and 90 cm length leads are studied at three parallel dual‐lead configurations (closely spaced, 8 mm spaced, and 40 mm spaced) and a single‐lead configuration. The global transfer function method is used to develop the AIMD models of different configurations and is used for lead‐tip heating assessments inside human body models. ResultsIn simulation studies, the peak 1g specific absorption rate/temperatrue rises of dual parallel leads systems is lower than those from the single‐lead system. In experimental American Society for Testing and Materials phantom studies, the temperature rises for the single‐lead AIMD system can be 2.4 times higher than that from dual‐lead AIMD systems. For the spinal cord stimulator systems used in the study, the statistical analysis shows the RF‐induced heating of dual‐lead configurations are also lower than those from the single‐lead configuration inside all four human body models. ConclusionFor the AIMD system in this study, it shows that the coupling effects between the dual parallel leads of AIMD systems can reduce RF‐induced heating. The global transfer function for different spatial distance dual‐lead configurations can potentially provide a method for the RF‐induced heating evaluation for dual‐lead AIMD systems. 
    more » « less
  4. PurposeTo develop an improved k‐space reconstruction method using scan‐specific deep learning that is trained on autocalibration signal (ACS) data. TheoryRobust artificial‐neural‐networks for k‐space interpolation (RAKI) reconstruction trains convolutional neural networks on ACS data. This enables nonlinear estimation of missing k‐space lines from acquired k‐space data with improved noise resilience, as opposed to conventional linear k‐space interpolation‐based methods, such as GRAPPA, which are based on linear convolutional kernels. MethodsThe training algorithm is implemented using a mean square error loss function over the target points in the ACS region, using a gradient descent algorithm. The neural network contains 3 layers of convolutional operators, with 2 of these including nonlinear activation functions. The noise performance and reconstruction quality of the RAKI method was compared with GRAPPA in phantom, as well as in neurological and cardiac in vivo data sets. ResultsPhantom imaging shows that the proposed RAKI method outperforms GRAPPA at high (≥4) acceleration rates, both visually and quantitatively. Quantitative cardiac imaging shows improved noise resilience at high acceleration rates (rate 4:23% and rate 5:48%) over GRAPPA. The same trend of improved noise resilience is also observed in high‐resolution brain imaging at high acceleration rates. ConclusionThe RAKI method offers a training database‐free deep learning approach for MRI reconstruction, with the potential to improve many existing reconstruction approaches, and is compatible with conventional data acquisition protocols. 
    more » « less
  5. Abstract PurposeTo introduce a method for the estimation of the ideal current patterns (ICP) that yield optimal signal‐to‐noise ratio (SNR) for realistic heterogeneous tissue models in MRI. Theory and MethodsThe ICP were calculated for different surfaces that resembled typical radiofrequency (RF) coil formers. We constructed numerical electromagnetic (EM) bases to accurately represent EM fields generated by RF current sources located on the current‐bearing surfaces. Using these fields as excitations, we solved the volume integral equation and computed the EM fields in the sample. The fields were appropriately weighted to calculate the optimal SNR and the corresponding ICP. We demonstrated how to qualitatively use ICP to guide the design of a coil array to maximize SNR inside a head model. ResultsIn agreement with previous analytic work, ICP formed large distributed loops for voxels in the middle of the sample and alternated between a single loop and a figure‐eight shape for a voxel 3‐cm deep in the sample's cortex. For the latter voxel, a surface quadrature loop array inspired by the shape of the ICP reached of the optimal SNR at 3T, whereas a single loop placed above the voxel reached only of the optimal SNR. At 7T, the performance of the two designs decreased to and , respectively, suggesting that loops could be suboptimal at ultra‐high field MRI. ConclusionICP can be calculated for human tissue models, potentially guiding the design of application‐specific RF coil arrays. 
    more » « less