skip to main content


Title: Validation of an enzyme‐linked immunoassay assay for osteocalcin, a marker of bone formation, in dried blood spots
Abstract Objectives

Investigating factors that contribute to bone loss and accretion across populations in remote settings is challenging, particularly where diagnostic tools are scarce. To mitigate this challenge, we describe validation of a commercial ELISA assay to measure osteocalcin, a biomarker of bone formation, from dried blood spots (DBS).

Methods

We validated the Osteocalcin Human SimpleStep ELISA kit from Abcam (ab1951214) using 158 matched plasma and DBS samples. Passing‐Bablok regression analysis assessed the relationships between plasma and DBS osteocalcin concentrations. Dilutional linearity and spike and recovery experiments determined if the DBS matrix interfered with osteocalcin measurement, and intra‐ and inter‐assay coefficients of variation (CVs) were calculated. Limit of detection, analyte stability, and specific forms of osteocalcin measured by the kit were also investigated.

Results

Mean plasma osteocalcin value was 218.2 ng/mL (range 64.6‐618.1 ng/mL). Linear relationships existed between plasma and DBS concentrations of osteocalcin, with no apparent bias in plasma vs DBS concentrations. There was no apparent interference of the DBS matrix with measurement of osteocalcin in DBS. Intra‐assay CV for DBS was ~8%, while average inter‐assay CV was 14.8%. Limit of detection was 0.34 ng/mL. Osteocalcin concentrations were stable in DBS stored at −28°C and room temperature, but not those stored at 37°C. This ELISA kit detects total osteocalcin.

Conclusions

Osteocalcin, a bone formation biomarker, can be measured from DBS. Combined with a previously validated DBS assay for TRACP‐5b, a bone resorption biomarker, these assays have the potential to help researchers disentangle the many factors contributing to bone strength.

 
more » « less
Award ID(s):
1638786
NSF-PAR ID:
10456840
Author(s) / Creator(s):
 ;  ;  ;  ;  ;  ;  
Publisher / Repository:
Wiley Blackwell (John Wiley & Sons)
Date Published:
Journal Name:
American Journal of Human Biology
Volume:
32
Issue:
5
ISSN:
1042-0533
Format(s):
Medium: X
Sponsoring Org:
National Science Foundation
More Like this
  1. Abstract Objectives

    A number of basic questions about bone biology have not been answered, including population differences in bone turnover. In part, this stems from the lack of validated minimally invasive biomarker techniques to measure bone formation and resorption in field‐based population‐level research. The present study addresses this gap by validating a fingerprick dried blood spot (fDBS) assay for tartrate‐resistant acid phosphatase 5b (TRACP‐5b), a well‐defined biomarker of bone resorption and osteoclast number.

    Methods

    We adapted a commercially available enzyme‐linked immunosorbent assay (ELISA) kit from MyBiosource for the quantitative determination of TRACP‐5b levels in serum and plasma for use with DBS. We used a rigorous process of assay modification and validation, including the use of a matched set of 189 adult plasma, fDBS, and venous DBS (vDBS) samples; parameters evaluated included precision, reliability, and analyte stability.

    Results

    Plasma and DBS TRACP‐5b concentrations showed a linear relationship. There were no systematic differences in TRACP‐5b levels in fDBS and vDBS, indicating no significant differences in TRACP‐5b distribution between capillary and venous blood. Parallelism and spike‐and‐recovery results indicated that matrix factors in DBS do not interfere with measurement of TRACP‐5b levels from DBS using the validated kit. Intra‐ and interassay CVs were 5.0% and 12.1%, respectively. DBS samples should preferably be stored frozen but controlled room temperature storage for up to a month may be acceptable.

    Conclusions

    This DBS‐based ELISA assay adds to the methodological toolkit available to human biologists and will facilitate research on bone turnover in population studies.

     
    more » « less
  2. Abstract The spike protein of SARS-CoV-2 engages the human angiotensin-converting enzyme 2 (ACE2) receptor to enter host cells, and neutralizing antibodies are effective at blocking this interaction to prevent infection. Widespread application of this important marker of protective immunity is limited by logistical and technical challenges associated with live virus methods and venous blood collection. To address this gap, we validated an immunoassay-based method for quantifying neutralization of the spike-ACE2 interaction in a single drop of capillary whole blood, collected on filter paper as a dried blood spot (DBS) sample. Samples are eluted overnight and incubated in the presence of spike antigen and ACE2 in a 96-well solid phase plate. Competitive immunoassay with electrochemiluminescent label is used to quantify neutralizing activity. The following measures of assay performance were evaluated: dilution series of confirmed positive and negative samples, agreement with results from matched DBS-serum samples, analysis of results from DBS samples with known COVID-19 status, and precision (intra-assay percent coefficient of variation; %CV) and reliability (inter-assay; %CV). Dilution series produced the expected pattern of dose–response. Agreement between results from serum and DBS samples was high, with concordance correlation = 0.991. Analysis of three control samples across the measurement range indicated acceptable levels of precision and reliability. Median % surrogate neutralization was 46.9 for PCR confirmed convalescent COVID-19 samples and 0.1 for negative samples. Large-scale testing is important for quantifying neutralizing antibodies that can provide protection against COVID-19 in order to estimate the level of immunity in the general population. DBS provides a minimally-invasive, low cost alternative to venous blood collection, and this scalable immunoassay-based method for quantifying inhibition of the spike-ACE2 interaction can be used as a surrogate for virus-based assays to expand testing across a wide range of settings and populations. 
    more » « less
  3. Abstract

    Nalbuphine was a semisynthetic opioid analgesic widely used in the treatment of both acute and chronic pain. We developed and validated a rapid, simple and sensitive method by ultra-performance liquid chromatography–tandem mass spectrometry (MS/MS) for the simultaneous quantitation of nalbuphine in human plasma, and we reported the pharmacokinetic features of patients during general anesthesia for abdominal surgery. Sample separation was achieved on a Kinetex Phenyl-Hexyl column (50 × 2.1 mm, 1.7 μm) after simple protein precipitation with acetonitrile. The mobile phase was composed of acetonitrile and 3 mM of ammonium acetate aqueous solution with 0.1% formic acid. Gradient elution was used in 4.5 min with a flow rate of 0.5 mL/min at 40°C. MS detection using AB Sciex QTRAP 5500 mass spectrometer was characterized by electrospray ionization for positive ions in multiple reaction monitoring mode. Quantitative ion pairs were m/z 358.4 → 340.1 for nalbuphine and m/z 340.0 → 268.3 for nalmefene, which were used as the internal standard (IS). The calibration curves showed good linearity (r2>0.99) over concentration range of 0.1–500 ng/mL. The intra-and inter-batch precisions were within 10.67%, and accuracy ranged from 94.07 to 105.34%. The IS–normalized matrix factors were 1.02–1.03 with RSD% (≤5.82%). The recoveries ranged from 101.09 to 106.30%. In conclusion, a rapid, simple, sensitive and economical analytical method was developed and validated to detect the concentration in plasma samples obtained from patients receiving nalbuphine intravenous injection and was successfully applicated to human pharmacokinetic studies of nalbuphine.

     
    more » « less
  4. Enzyme linked immunosorbent assay (ELISA) is one of the most utilized serological methods to diagnose and identify etiologic agents of many infectious diseases and other physiologically important analytes. ELISA can be used either alone or adjunct to other diagnostic methods such as molecular arrays, and other serological techniques. Most ELISA assays utilize reagents that are proteinaceous in nature, which are not very stable and require cold-chain transport systems. Development of a desirable immunoassay requires stability of reagents used and its ability to be stored at room temperature without sacrificing the activity of the reagents or the protein of interest. Metal organic frameworks (MOFs) are a rapidly emerging and evolving class of porous polymeric materials used in a variety of biosensor applications. In this study, we introduce the use of MOFs to stabilize a universal reporter fusion protein, specifically, avidin-like protein (Tam-avidin2) and the small bioluminescent protein Gaussia luciferase (Gluc) forming the fusion reporter, tamavidin2-Gluc (TA2-Gluc). This fusion protein serves as a universal reporter for any assays that utilize biotin–avidin binding strategy. Using SARS-CoV2 S1 spike antigen as the model target antigen, we demonstrated that encapsulation of TA2-Gluc fusion protein using a nano-porous material, zeolitic imidazolate framework-8 (ZIF-8), allows us to store and preserve this reporter protein at room temperature for over 6 months and use it as a reporter for an ELISA assay. Our optimized assay was validated demonstrating a 0.26 μg mL −1 limit of detection, high reproducibility of assay over days, detection of spiked non-virulent SARS-COV2 pseudovirus in real sample matrix, and detection in real COVID-19 infected individuals. This result can lead to the utilization of our TA2-Gluc fusion protein reporter with other assays and potentially in diagnostic technologies in a point-of-care setting. 
    more » « less
  5. Abstract

    Multiplexed computational sensing with a point‐of‐care serodiagnosis assay to simultaneously quantify three biomarkers of acute cardiac injury is demonstrated. This point‐of‐care sensor includes a paper‐based fluorescence vertical flow assay (fxVFA) processed by a low‐cost mobile reader, which quantifies the target biomarkers through trained neural networks, all within <15 min of test time using 50 µL of serum sample per patient. This fxVFA platform is validated using human serum samples to quantify three cardiac biomarkers, i.e., myoglobin, creatine kinase‐MB, and heart‐type fatty acid binding protein, achieving less than 0.52 ng mL−1limit‐of‐detection for all three biomarkers with minimal cross‐reactivity. Biomarker concentration quantification using the fxVFA that is coupled to neural network‐based inference is blindly tested using 46 individually activated cartridges, which shows a high correlation with the ground truth concentrations for all three biomarkers achieving >0.9 linearity and <15% coefficient of variation. The competitive performance of this multiplexed computational fxVFA along with its inexpensive paper‐based design and handheld footprint makes it a promising point‐of‐care sensor platform that can expand access to diagnostics in resource‐limited settings.

     