Title: A Validated UHPLC–MS/MS Method for Determination of Nalbuphine in Human Plasma and Application for Pharmacokinetic Study of Patients Undergoing General Anesthesia
Abstract

Nalbuphine was a semisynthetic opioid analgesic widely used in the treatment of both acute and chronic pain. We developed and validated a rapid, simple and sensitive method by ultra-performance liquid chromatography–tandem mass spectrometry (MS/MS) for the simultaneous quantitation of nalbuphine in human plasma, and we reported the pharmacokinetic features of patients during general anesthesia for abdominal surgery. Sample separation was achieved on a Kinetex Phenyl-Hexyl column (50 × 2.1 mm, 1.7 μm) after simple protein precipitation with acetonitrile. The mobile phase was composed of acetonitrile and 3 mM of ammonium acetate aqueous solution with 0.1% formic acid. Gradient elution was used in 4.5 min with a flow rate of 0.5 mL/min at 40°C. MS detection using AB Sciex QTRAP 5500 mass spectrometer was characterized by electrospray ionization for positive ions in multiple reaction monitoring mode. Quantitative ion pairs were m/z 358.4 → 340.1 for nalbuphine and m/z 340.0 → 268.3 for nalmefene, which were used as the internal standard (IS). The calibration curves showed good linearity (r2>0.99) over concentration range of 0.1–500 ng/mL. The intra-and inter-batch precisions were within 10.67%, and accuracy ranged from 94.07 to 105.34%. The IS–normalized matrix factors were 1.02–1.03 with RSD% (≤5.82%). The recoveries ranged from 101.09 to 106.30%. In conclusion, a rapid, simple, sensitive and economical analytical method was developed and validated to detect the concentration in plasma samples obtained from patients receiving nalbuphine intravenous injection and was successfully applicated to human pharmacokinetic studies of nalbuphine.

 
more » « less
NSF-PAR ID:
10477465
Author(s) / Creator(s):
; ; ; ; ; ; ; ; ;
Publisher / Repository:
Oxford University Press
Date Published:
Journal Name:
Journal of Chromatographic Science
Volume:
61
Issue:
8
ISSN:
0021-9665
Format(s):
Medium: X Size: p. 758-765
Size(s):
["p. 758-765"]
Sponsoring Org:
National Science Foundation
More Like this
  1. Abstract

    Previously compound12showed great anti‐trypanosome activity without toxicity in anin vivostudy. In the current study, a sensitive and rapid high‐performance liquid chromatography–tandem mass spectrometry (HPLC–MS/MS) method was developed and validated to investigate its pharmacokinetics in mouse plasma. A protein precipitation method was applied to extract the compound, and it was then separated using a Kinetex C18column with mobile phase consisting of acetonitrile–0.1% formic acid water (50:50, v/v) at a flow rate of 300 μl/min. The analytes were detected with the multiple reaction monitoring in negative electrospray ionization source for quantitative response of the compounds. Compound12was detected atm/z477.0 → 367.2, while the internal standard compound14was detected atm/z499.2 → 268.2. Inter‐ and intra‐day precision was <5.22 and 2.79% respectively, while the accuracy range was within ±9.65%. The method was successfully applied to evaluate the pharmacokinetics of compound12in mouse plasma with two formulations (20% Cremophor EL or sesame oil) and drug administration routes (oral and intraperitoneal injection). We observed a better drug serum concentration with the Cremophor formulation, and the two different drug administration routes did not show significant differences from the drug distribution.

     
    more » « less
  2. After attending this presentation, attendees will gain knowledge in the strategy to achieve high-throughput and simultaneous analysis of cannabinoids and appreciate a validated LC-UV method for analysis of twelve cannabinoids in hemp oil. This presentation will first impact the forensic science community by introducing three fast LC separations of twelve cannabinoids that can be used with either UV or mass spectrometric (MS) detection. It will further impact the forensic science community by introducing a validated LC-UV method for high-throughput and simultaneous analysis of twelve cannabinoids in hemp oil, which can be routinely used by cannabis testing labs. In recent years, the use of products of Cannabis sativa L. for medicinal purposes has been in a rapid growth, although their preparation procedure has not been clearly standardized and their quality has not been well regulated. To analyze the therapeutic components, i.e. cannabinoids, in products of Cannabis sativa L., LC-UV has been frequently used, because LC-UV is commonly available and usually appropriate for routine analysis by the cannabis growers and commercial suppliers. In the literature, a few validated LC-UV methods have been described. However, so far, all validated LC-UV methods only focused in the quantification of eleven or less cannabinoids. Therefore, a method able to simultaneously analyze more cannabinoids in a shorter run time is still in high demand, because more and more cannabinoids have been isolated and many of them have shown medicinal properties. In this study, the LC separation of twelve cannabinoids, including cannabichromene (CBC), cannabidiolic acid (CBDA), cannabidiol (CBD), cannabidivarinic acid (CBDVA), cannabidivarin (CBDV), cannabigerolic acid (CBGA), cannabigerol (CBG), cannabinol (CBN), delta-8 tetrahydrocannabinol (Δ8-THC), delta-9 tetrahydrocannabinolic acid A (Δ9-THCA A), delta-9 tetrahydrocannabinol (Δ9-THC), and tetrahydrocannabivarin (THCV), has been systematically optimized using a Phenomenex Luna Omega 3 µm Polar C18 150 mm × 4.6 mm column with regard to the effects of the type of organic solvent, i.e. methanol and acetonitrile, the content of the organic solvent, and the pH of the mobile phase. The optimization has resulted in three LC conditions at 1.0 mL/minute able to separate the twelve cannabinoids: 1) a mobile phase consisting of water and methanol, both containing 0.1% formic acid (pH 2.69), with a gradient elution at 75% methanol for the first 3 minutes and then linearly increase to 100% methanol at 12.5 minutes; 2) a mobile phase consisting of water and 90% (v/v) acetonitrile in water, both containing 0.1% formic acid and 20 mM ammonium formate (pH 3.69), with an isocratic elution at 75% acetonitrile for 14 minutes; and 3) a mobile phase consisting of water and 90% (v/v) acetonitrile in water, both containing 0.03% formic acid and 20 mM ammonium formate (pH 4.20), with an isocratic elution at 75% acetonitrile for 14 minutes. In order to demonstrate the effectiveness of the achieved LC separations, a LC-UV method is further validated for the high-throughput and simultaneous analysis of twelve cannabinoids. The method used the mobile phase at pH 3.69, which resulted in significant improvement in throughput compared to other validated LC-UV methods published so far. The method used flurbiprofen as the internal standard. The linear calibration range of all the cannabinoids were between 0.1 to 25 ppm with R2≥0.9993. The LOQ (S/N=10) of the cannabinoids was between 17.8 and 74.2 ppb. The validation used a hemp oil containing 3.2 wt% CBD and no other cannabinoids, which was reported by the vendor with a certificate of analysis, as the matrix to prepare control samples: the hemp oil was first extracted using liquid-liquid extraction (LLE) with methanol; cannabinoids were then spiked into the extract at both 0.5 ppm and 5 ppm level. Afterwards, the recovery, precision (%RSD) and accuracy (%Error) of the control samples were assessed and the results met the requirements by the ISO/IEC 17025 and ASTM E2549-14 guidelines. 
    more » « less
  3. Lipid screening of biological substrates is an important step during biomarker detection and identification. In this work, a fast workflow is described capable of rapid screening for lipid components from biological tissues at ambient pressure based on liquid microjunction extraction in tandem with nano-electrospray ionization (nESI) with ultra-high resolution mass spectrometry, i.e. , liquid extraction surface analysis (LESA) coupled to Fourier-transform ion cyclotron resonance (tandem) mass spectrometry (LESA-FT-ICR-MS/MS). Lipid profiles are presented for thin tissue sections of mouse brain (MB) and liver (ML) samples, analyzed in both positive and negative mode by data-dependent acquisition (DDA) tandem FT-ICR-MS/MS. Candidate assignments were based on fragmentation patterns using mostly SimLipid software and accurate mass using mostly the LipidMaps database (average sub-ppm mass error). A typical, single point surface analysis (<1 mm spatial sampling resolution) lasted less than 15 minutes and resulted in the assignment of (unique and mulitple) lipid identifications of ∼190 (MB) and ∼590 (ML) m / z values. Despite the biological complexity, this led to unique identifications of distinct lipid molecules (sub-ppm mass error) from 38 different lipid classes, corresponding to 10–30% of the lipid m / z identifications. 
    more » « less
  4. Stable potassium (K) isotopes (41K/39K) have shown great promise as novel chemical tracers for a wide range of bio-, geo-, and cosmo-chemical processes, but high precision stable K isotope analysis remains a challenge for plasma source mass spectrometry due to intense argon-related interferences produced directly from argon plasma. Here we provide an assessment on the analytical figures of merit of a new generation collision-cell equipped multi-collector inductively coupled plasma mass spectrometer (MC-ICP-MS), Sapphire from Nu Instruments, for K isotope analysis based on our extensive tests over a duration of ~8 months. Because use of helium and hydrogen as collision/reaction gases can reduce argon-related interferences to negligible levels at optimal flow rates, the collision-cell mode can operate at low mass resolution during K isotope analysis, providing >2 orders of magnitude higher K sensitivity (>1000 V per μg mL-1 K), as compared to the widely used “cold plasma” method, and the capability for direct 40K measurement. One challenge of the collision/reaction cell analysis on Sapphire is its higher susceptibility to matrix effects, requiring effective sample purification prior to analysis. Also, the collision-cell mode on Sapphire shows a pronounced effect associated with concentration (or ion intensity) mismatch between the sample and the bracketing standard during analysis, and this effect may not be fully eliminated through conventional concentration matching practice. Instead, we developed a correction method for this concentration/ion intensity mismatch effect. Our method reduces the burden to the operator and increases sample throughput. This method allows for accurate K isotope analysis with an intermediate precision of ≤0.05 ‰ (2SD) to be routinely achieved using the collision cell on Sapphire, representing a major advance to stable K isotope analysis. 
    more » « less
  5. This study develops and optimizes a new protocol to measure lithium isotope ratios using a single collector quadrupole inductively coupled plasma mass spectrometer (Q-ICP-MS) operated under hot plasma (1550 W) conditions with a sample–standard bracketing method. Our Q-ICP-MS method reduces sample consumption to 2.5 ng of Li and achieves a high long-term precision of 1.1‰ (2SD). This Q-ICP-MS method exhibits high matrix tolerance (Na/Li < 100), suitable for ng-sized and high-matrix geological samples. We also developed a dual-column system for Li separation, with large loading capacity (29.6 meq), complete recovery (∼100%) and satisfactory purification (Na/Li m m −1 < 1), as well as a fixed elution range for Li fractions (28–60 mL). This new chromatography method has been applied to chemically diverse materials, producing consistent results. In addition, we report the Li isotope compositions of 13 geostandards, and our measurements agree well with reported data within analytical uncertainties. This study documents that Li element concentration and Li isotope composition can be routinely measured using a single collector ICP-MS, which is convenient and commercially affordable for future Li isotope research across the fields of Earth and Environmental Sciences. 
    more » « less