Abstract Parahydrogen-induced polarization of13C nuclei by side-arm hydrogenation (PHIP-SAH) for [1-13C]acetate and [1-13C]pyruvate esters with application of PH-INEPT-type pulse sequences for1H to13C polarization transfer is reported, and its efficiency is compared with that of polarization transfer based on magnetic field cycling (MFC). The pulse-sequence transfer approach may have its merits in some applications because the entire hyperpolarization procedure is implemented directly in an NMR or MRI instrument, whereas MFC requires a controlled field variation at low magnetic fields. Optimization of the PH-INEPT-type transfer sequences resulted in13C polarization values of 0.66 ± 0.04% and 0.19 ± 0.02% for allyl [1-13C]pyruvate and ethyl [1-13C]acetate, respectively, which is lower than the corresponding polarization levels obtained with MFC for1H to13C polarization transfer (3.95 ± 0.05% and 0.65 ± 0.05% for allyl [1-13C]pyruvate and ethyl [1-13C]acetate, respectively). Nevertheless, a significant13C NMR signal enhancement with respect to thermal polarization allowed us to perform13C MR imaging of both biologically relevant hyperpolarized molecules which can be used to produce useful contrast agents for the in vivo imaging applications.
more »
« less
New clinical opportunities of low-field MRI: heart, lung, body, and musculoskeletal
Abstract Contemporary whole-body low-field MRI scanners (< 1 T) present new and exciting opportunities for improved body imaging. The fundamental reason is that the reduced off-resonance and reduced SAR provide substantially increased flexibility in the design of MRI pulse sequences. Promising body applications include lung parenchyma imaging, imaging adjacent to metallic implants, cardiac imaging, and dynamic imaging in general. The lower cost of such systems may make MRI favorable for screening high-risk populations and population health research, and the more open configurations allowed may prove favorable for obese subjects and for pregnant women. This article summarizes promising body applications for contemporary whole-body low-field MRI systems, with a focus on new platforms developed within the past 5 years. This is an active area of research, and one can expect many improvements as MRI physicists fully explore the landscape of pulse sequences that are feasible, and as clinicians apply these to patient populations.
more »
« less
- Award ID(s):
- 1828736
- PAR ID:
- 10471730
- Publisher / Repository:
- Springer Science + Business Media
- Date Published:
- Journal Name:
- Magnetic Resonance Materials in Physics, Biology and Medicine
- Volume:
- 37
- Issue:
- 1
- ISSN:
- 1352-8661
- Format(s):
- Medium: X Size: p. 1-14
- Size(s):
- p. 1-14
- Sponsoring Org:
- National Science Foundation
More Like this
-
-
Abstract Magnetic resonance imaging (MRI) is a powerful and widely used in vivo imaging technique that enables whole body imaging without ionizing radiation. In clinical practice,1H MRI is employed for imaging anatomical and physiological states via monitoring of protons in water and lipids. In order to monitor biochemical processes at the molecular level, several research groups are exploring responsive MRI agents that alter their signal upon interaction with an analyte or biological environment of interest. Fluorine (19F) MRI agents are promising due to the19F nucleus having similar magnetic resonance (MR) properties to proton and the absence of endogenous19F in living systems, resulting in no background signal. In order to make responsive19F MR agents for molecular imaging and analysis, fluorinated platforms must be developed in which their19F MR signal changes after interacting with a target analyte. A promising strategy is to use paramagnetic metals to modulate the19F MR signal by altering the relaxation rates and/or chemical shift of an appended19F imaging tag. In this concept, we provide an overview of the theoretical principles and molecular design strategies that have been exploited in the design of responsive19F MR agents, with a specific focus on agents based on small molecule paramagnetic metal ion chelates.more » « less
-
Abstract PurposeTo demonstrate speech‐production real‐time MRI (RT‐MRI) using a contemporary 0.55T system, and to identify opportunities for improved performance compared with conventional field strengths. MethodsExperiments were performed on healthy adult volunteers using a 0.55T MRI system with high‐performance gradients and a custom 8‐channel upper airway coil. Imaging was performed using spiral‐based balancedSSFPand gradient‐recalled echo (GRE) pulse sequences using a temporal finite‐difference constrained reconstruction. Speech‐production RT‐MRI was performed with three spiral readout durations (8.90, 5.58, and 3.48 ms) to determine trade‐offs with respect to articulator contrast, blurring, banding artifacts, and overall image quality. ResultsBoth spiral GRE and bSSFP captured tongue boundary dynamics during rapid consonant‐vowel syllables. Although bSSFP provided substantially higher SNR in all vocal tract articulators than GRE, it suffered from banding artifacts at TR > 10.9 ms. Spiral bSSFP with the shortest readout duration (3.48 ms, TR = 5.30 ms) had the best image quality, with a 1.54‐times boost in SNR compared with an equivalent GRE sequence. Longer readout durations led to increased SNR efficiency and blurring in both bSSFP and GRE. ConclusionHigh‐performance 0.55T MRI systems can be used for speech‐production RT‐MRI. Spiral bSSFP can be used without suffering from banding artifacts in vocal tract articulators, provide better SNR efficiency, and have better image quality than what is typically achieved at 1.5 T or 3 T.more » « less
-
Abstract PurposeTo develop a small‐tip multidimensional RF pulse design procedure that incorporates linear time‐invariant gradient imperfections and concomitant field effects. This could be particularly important for contemporary low‐field MRI systems with high‐performance gradients. Theory and MethodsWe developed an extension of the small‐tip excitation k‐space formalism, where concomitant fields were approximated as a Bloch‐Siegert shift in the rotating frame. This was evaluated using realistic simulations of 2D selective excitation at various field strengths (0.2T, 0.55T, 1.5T, 3T, and 7T) with single and parallel transmit. Simulated excitation profiles from the original and extended k‐space formalisms were compared. Experimental validations were performed at 0.55T with a single‐channel transmit. ResultsThe extended formalism provides improved 2D excitation profiles in all scenarios simulated, compared against the original formalism. The proposed method corrects the concomitant field effects on 2D selective excitations forB0 > 0.2T when the magnitude of theB0is far larger than that of nonrotating concomitant fields. Simulation and phantom experiments at 0.55T match well for both original and proposed methods, with the proposed method providing sharper and more accurate excitation profiles at off‐isocenter distances up to 15 cm. The impact of the proposed method is greatest in scenarios where concomitant fields are substantial, such as low field strengths and off‐isocenter. ConclusionConcomitant fields can be modeled as a Bloch‐Siegert shift in the rotating frame during multidimensional RF pulse design, resulting in improved excitation profiles with sharp edges. This is important to consider for off‐isocenter excitations and imaging at low field strengths with strong gradients.more » « less
-
Abstract Fetal magnetic resonance imaging (MRI) is an important adjunct modality for the evaluation of fetal abnormalities. Recently, low-field MRI systems at 0.55 Tesla have become available which can produce images on par with 1.5 Tesla systems but with lower power deposition, acoustic noise, and artifact. In this article, we describe a technical innovation using low-field MRI to perform diagnostic quality fetal MRI.more » « less