Abstract This work describes the apparatus for NIST-F4, an updated cesium atomic fountain at the National Institute of Standards and Technology (NIST), and presents an accuracy evaluation of the fountain as a primary frequency standard. The fountain uses optical molasses to laser cool a cloud of cesium atoms and launch it vertically in a fountain geometry. In high-density mode, the fractional frequency stability of NIST-F4 is , whereτis the measurement time in seconds. The short-term stability is limited by quantum projection noise and by phase noise from the local oscillator, an oven-controlled crystal oscillator operating at 5 MHz. Systematic frequency shifts and their uncertainties have been evaluated, resulting in a systematic (type B) fractional frequency uncertainty .
more »
« less
Evaluation of a caesium fountain frequency standard for antihydrogen spectroscopy
Abstract The performance of a caesium fountain frequency reference for use in precision measurements of trapped antihydrogen in the ALPHA experiment at CERN is evaluated. A description of the fountain is provided together with a characterisation of systematic effects. The impact of the magnetic environment in the Antimatter Factory, where the fountain is installed, on the performance of the fountain is considered and shown to be insignificant. The systematic fractional frequency uncertainty of the fountain is . The short-term frequency stability of the measured frequency from the ALPHA-HM1 maser is , whereas the fountain itself shows a stability limit of . We find a fractional frequency difference of (1.0 ± 2.2 (stat.) ± 6.5 (syst.)) in a comparison with Terrestrial Time via a GNSS Common View satellite link between January 2023 and June 2024. The fountain enables a significant increase in frequency precision in antihydrogen spectroscopic measurements, and paves the way for improved limits on matter–antimatter comparisons.
more »
« less
- Award ID(s):
- 2012117
- PAR ID:
- 10629805
- Publisher / Repository:
- IOP Publishing
- Date Published:
- Journal Name:
- Metrologia
- Volume:
- 62
- Issue:
- 4
- ISSN:
- 0026-1394
- Page Range / eLocation ID:
- 045008
- Format(s):
- Medium: X
- Sponsoring Org:
- National Science Foundation
More Like this
-
-
Abstract We have made improvements in the stability, accuracy, and performance of the NRC-FCs2 fountain clock. The dominant systematic effects have been re-evaluated. Optically-stabilized microwaves are used to improve the short-term stability, now reaching . We have evaluated the distributed cavity phase shift using absorption imaging. This technique dramatically reduces the evaluation time and final uncertainty. We have re-evaluated biases due to microwave leakage and synchronous phase transients with higher accuracy. The total systematic uncertainty of NRC-FCs2 is now in fractional frequency, a factor of 2 improvement over its most recent evaluation.more » « less
-
Abstract A test of lepton flavor universality in and decays, as well as a measurement of differential and integrated branching fractions of a nonresonant decay are presented. The analysis is made possible by a dedicated data set of proton-proton collisions at recorded in 2018, by the CMS experiment at the LHC, using a special high-rate data stream designed for collecting about 10 billion unbiased b hadron decays. The ratio of the branching fractions to is determined from the measured double ratio of these decays to the respective branching fractions of the with and decays, which allow for significant cancellation of systematic uncertainties. The ratio is measured in the range , whereqis the invariant mass of the lepton pair, and is found to be , in agreement with the standard model expectation . This measurement is limited by the statistical precision of the electron channel. The integrated branching fraction in the sameq2range, , is consistent with the present world-average value and has a comparable precision.more » « less
-
Abstract The production of a pair of τ leptons via photon–photon fusion, , is observed for the first time in proton–proton collisions, with a significance of 5.3 standard deviations. This observation is based on a data set recorded with the CMS detector at the LHC at a center-of-mass energy of 13 TeV and corresponding to an integrated luminosity of 138 fb−1. Events with a pair of τ leptons produced via photon–photon fusion are selected by requiring them to be back-to-back in the azimuthal direction and to have a minimum number of charged hadrons associated with their production vertex. The τ leptons are reconstructed in their leptonic and hadronic decay modes. The measured fiducial cross section of is . Constraints are set on the contributions to the anomalous magnetic moment ( ) and electric dipole moments ( ) of the τ lepton originating from potential effects of new physics on the vertex: and (95% confidence level), consistent with the standard model.more » « less
-
Abstract Introduction: We present an extensive theoretical investigation of the electron impact excitation of doubly-ionized titanium (Ti III) to meet the needs of spectral analysis and plasma modeling. OBJECTIVES: The main objective of this work is to extend the currently scarce database of both structure and collision data for Ti III. METHODS: The calculation was performed in the close-coupling approximation using theB-splineR-matrix method. The multi-configuration Hartree–Fock method in combination withB-spline configuration interaction expansions and the non-orthogonal orbitals technique is employed for accurate descriptions of the target wave functions and adequate accounts of the various interactions between the target states. Relativistic effects are treated at the semi-relativistic Breit-Pauli approximation level. RESULTS: The present close-coupling expansion includes 138 fine-structure levels of Ti III belonging to the , , , ( ), ( ), , and configurations. Comprehensive sets of radiative and electron collisional data are reported for all of the possible transitions between the 138 fine-structure levels. Thermally averaged collision strengths are determined using a Maxwellian distribution for a wide range of temperatures from K to K. The accuracy of the calculated radiative parameters is validated by comparing with available values from the NIST database and previous literature. CONCLUSION: Given the lack of sufficient currently available experimental and theoretical data, the electron impact excitation cross sections of the Ti III fine-structure levels presented here are systematic, extensive, and internally consistent, thus making them suitable for many modeling applications.more » « less
An official website of the United States government