skip to main content


Title: Monitoring of irrigated lesion formation with single fiber based multispectral system using machine learning
Abstract

In radiofrequency ablation (RFA) treatment of cardiac arrhythmias, intraprocedural assessment of treatment efficacy relies on indirect measures of adequate tissue destruction. Direct sensing of diffuse reflectance spectral changes at the ablation site using optically integrated RFA catheters has been shown to enable accurate prediction of lesion dimensions, ex vivo. Challenges of optical guidance can be due to obtaining reliable measurements under various catheter‐tissue contact orientations. In this work, addressed this limitation by assessing the feasibility of monitoring lesion progression using single‐fiber reflectance spectroscopy (SFRS). A total of 110 endocardial lesions of various sizes were generated in freshly excised swine right ventricular tissue using a custom‐built, irrigated SFRS‐RFA catheter. Models were developed for assessing catheter‐tissue contact, the presence of nontransmural or transmural lesions and lesion depth percentage. These results support the use of SFRS‐based catheters for irrigated lesion assessment and motivate further exploration of using multi‐SFRS catheters for omnidirectionality.

 
more » « less
NSF-PAR ID:
10370806
Author(s) / Creator(s):
 ;  ;  ;  
Publisher / Repository:
Wiley Blackwell (John Wiley & Sons)
Date Published:
Journal Name:
Journal of Biophotonics
Volume:
15
Issue:
9
ISSN:
1864-063X
Format(s):
Medium: X
Sponsoring Org:
National Science Foundation
More Like this
  1. Atrial fibrillation (AF) is a rapid irregular electrical activity in the upper chamber and the most common sustained cardiac arrhythmia. Many patients require radiofrequency ablation (RFA) therapy to restore sinus rhythm. Pulmonary vein isolation requires distinguishing normal atrial wall from the pulmonary vein tissue, and atrial substrate ablation requires differentiating scar tissue, fibrosis, and adipose tissue. However, current anatomical mapping methods for strategically locating ablation sites by identifying structural substrates in real-time are limited. An intraoperative tool that accurately provides detailed structural information and classifies endocardial substrates could help improve RF guidance during RF ablation therapy. In this work, we propose a 7F NIRS integrated ablation catheter and demonstrate endocardial mapping onex vivoswine (n = 12) and human (n = 5) left atrium (LA). First, pulmonary vein (PV) sleeve, fibrosis and ablation lesions were identified with NIRS-derived contrast indices. Based on these key spectral features, classification algorithms identified endocardial substrates with high accuracy (<11% error). Then, a predictive model for lesion depth was evaluated on classified lesions. Model predictions correlated well with histological measurements of lesion dimensions (R = 0.984). Classified endocardial substrates and lesion depth were represented in 2D spatial maps. These results suggest NIRS integrated mapping catheters can serve as a complementary tool to the current electroanatomical mapping system to improve treatment efficacy.

     
    more » « less
  2. Despite considerable advances in guidance of radiofrequency ablation (RFA) therapy for the treatment of cardiac arrhythmias, success rates have been hampered by a lack of tools for precise intraoperative evaluation of lesion extent. Near‐infrared spectroscopic (NIRS) techniques are sensitive to tissue structural and biomolecular properties, characteristics that are directly altered by radiofrequency (RF) treatment. In this work, a combined NIRS‐RFA catheter is developed for real‐time monitoring of tissue reflectance during RF energy delivery. An algorithm is proposed for processing NIR spectra to approximate nonirrigated lesion depth in both atrial and ventricular tissues. The probe optical geometry was designed to bias measurement influence toward absorption enabling enhanced sensitivity to changes in tissue composition. A set of parameters termed “lesion optical indices” are defined encapsulating spectral differences between ablated and unablated tissue. Utilizing these features, a model for real‐time tissue spectra classification and lesion size estimation is presented. Experimental validation conducted within freshly excised porcine cardiac specimens showed strong concordance between algorithm estimates and post‐hoc tissue assessment.

     
    more » « less
  3. Abstract The disturbances caused by the blood flow and tissue surface motions are major concerns during the motion planning of an intracardiac robotic catheter. Maintaining a stable and safe contact on the desired ablation point is essential for achieving effective lesions during the ablation procedure. In this paper, a probabilistic formulation of the contact stability and the contact safety for intravascular cardiac catheters under the blood flow and surface motion disturbances is presented. Probabilistic contact stability and contact safety metrics, employing a sample-based representation of the blood flow velocity distribution and the heart motion trajectory, are introduced. Finally, the contact stability and safety for an magnetic resonance imaging-actuated robotic catheter under main pulmonary artery blood flow disturbances and left ventricle surface motion disturbances are analyzed in simulation as example scenarios. 
    more » « less
  4. Abstract Background

    Catheter ablation is associated with limited success rates in patients with persistent atrial fibrillation (AF). Currently, existing mapping systems fail to identify critical target sites for ablation. Recently, we proposed and validated several techniques (multiscale frequency [MSF], Shannon entropy [SE], kurtosis [Kt], and multiscale entropy [MSE]) to identify pivot point of rotors using ex‐vivo optical mapping animal experiments. However, the performance of these techniques is unclear for the clinically recorded intracardiac electrograms (EGMs), due to the different nature of the signals.

    Objective

    This study aims to evaluate the performance of MSF, MSE, SE, and Kt techniques to identify the pivot point of the rotor using unipolar and bipolar EGMs obtained from numerical simulations.

    Methods

    Stationary and meandering rotors were simulated in a 2D human atria. The performances of new approaches were quantified by comparing the “true” core of the rotor with the core identified by the techniques. Also, the performances of all techniques were evaluated in the presence of noise, scar, and for the case of the multielectrode multispline and grid catheters.

    Results

    Our results demonstrate that all the approaches are able to accurately identify the pivot point of both stationary and meandering rotors from both unipolar and bipolar EGMs. The presence of noise and scar tissue did not significantly affect the performance of the techniques. Finally, the core of the rotors was correctly identified for the case of multielectrode multispline and grid catheter simulations.

    Conclusion

    The core of rotors can be successfully identified from EGMs using novel techniques; thus, providing motivation for future clinical implementations.

     
    more » « less
  5. Abstract Contact force quality is one of the most critical factors for safe and effective lesion formation during catheter based atrial fibrillation ablation procedures. In this paper, the contact stability and contact safety of a novel magnetic resonance imaging (MRI)-actuated robotic cardiac ablation catheter subject to surface motion disturbances are studied. First, a quasi-static contact force optimization algorithm, which calculates the actuation needed to achieve a desired contact force at an instantaneous tissue surface configuration is introduced. This algorithm is then generalized using a least-squares formulation to optimize the contact stability and safety over a prediction horizon for a given estimated heart motion trajectory. Four contact force control schemes are proposed based on these algorithms. The first proposed force control scheme employs instantaneous heart position feedback. The second control scheme applies a constant actuation level using a quasi-periodic heart motion prediction. The third and the last contact force control schemes employ a generalized adaptive filter-based heart motion prediction, where the former uses the predicted instantaneous position feedback, and the latter is a receding horizon controller. The performance of the proposed control schemes is compared and evaluated in a simulation environment. 
    more » « less