skip to main content


Title: Evaluation of Robotic-Assisted Carotid Artery Stenting in a Virtual Model Using Motion-Based Performance Metrics
Purpose: Robotic-assisted carotid artery stenting (CAS) cases have been demonstrated with promising results. However, no quantitative measurements have been made to compare manual with robotic-assisted CAS. This study aims to quantify surgical performance using tool tip kinematic data and metrics of precision during CAS with manual and robotic control in an ex vivo model. Materials and Methods: Transfemoral CAS cases were performed in a high-fidelity endovascular simulator. Participants completed cases with manual and robotic techniques in 2 different carotid anatomies in random order. C-arm angulations, table position, and endovascular devices were standardized. Endovascular tool tip kinematic data were extracted. We calculated the spectral arc length (SPARC), average velocity, and idle time during navigation in the common carotid artery and lesion crossing. Procedural time, fluoroscopy time, movements of the deployed filter wire, precision of stent, and balloon positioning were recorded. Data were analyzed and compared between the 2 modalities. Results: Ten participants performed 40 CAS cases with a procedural success of 100% and 0% residual stenosis. The median procedural time was significantly higher during the robotic-assisted cases (seconds, median [interquartile range, IQR]: 128 [49.5] and 161.5 [62.5], p=0.02). Fluoroscopy time differed significantly between manual and robotic-assisted procedures (seconds, median [IQR]: 81.5 [32] and 98.5 [39.5], p=0.1). Movement of the deployed filter wire did not show significant difference between manual and robotic interventions (mm, median [IQR]: 13 [10.5] and 12.5 [11], p=0.5). The postdilation balloon exceeded the margin of the stent with a median of 2 [1] mm in both groups. Navigation with robotic assistance showed significantly lower SPARC values (–5.78±3.14 and –8.63±3.98, p=0.04) and higher idle time values (8.92±8.71 and 3.47±3.9, p=0.02) than those performed manually. Conclusions: Robotic-assisted and manual CAS cases are comparable in the precision of stent and balloon positioning. Navigation in the carotid artery is associated with smoother motion and higher idle time values. These findings highlight the accuracy and the motion stabilizing capability of the endovascular robotic system. Clinical Impact Robotic assistance in the treatment of peripheral vascular disease is an emerging field and may be a tool for radiation protection and the geographic distribution of endovascular interventions in the future. This preclinical study compares the characteristics of manual and robotic-assisted carotid stenting (CAS). Our results highlight, that robotic-assisted CAS is associated with precise navigation and device positioning, and smoother navigation compared to manual CAS.  more » « less
Award ID(s):
2049063
NSF-PAR ID:
10414659
Author(s) / Creator(s):
; ; ; ; ; ; ; ;
Date Published:
Journal Name:
Journal of Endovascular Therapy
ISSN:
1526-6028
Page Range / eLocation ID:
152660282211255
Format(s):
Medium: X
Sponsoring Org:
National Science Foundation
More Like this
  1. Abstract Background

    Percutaneous coronary interventions (PCIs) within left main coronary arteries are high-risk procedures that require optimization of interactions between stent(s) and diseased vessels. Optical Coherence Tomography (OCT) is a widely accepted tool that enhances physicians’ ability to assess proper stent appositions during clinical procedures. The primary aim of this study was to develop complementary post-procedure imaging methodologies to better assess and interpret outcomes of left main PCI procedures, utilizing both reanimated and perfusion-fixed human hearts.

    Methods

    PCIs were performed while obtaining OCT scans within the left main anatomies of six human hearts. Subsequently, each heart was scanned with a micro-CT scanner with optimized parameters to achieve resolutions up to 20 µm. Scans were reconstructed and imported into a DICOM segmentation software to generate computational models of implanted stents and associated coronary vessels. 2D images from OCT that were obtained during PCIs were compared to the 3D models generated from micro-CT reconstructions. In addition, the 3D models were utilized to create virtual reality scenes and enlarged 3D prints for development of “mixed reality” tools relative to bifurcation stenting within human left main coronary arteries.

    Results

    We developed reproducible methodologies for post-implant analyses of coronary artery stenting procedures. In addition, we generated high-resolution 3D computational models, with ~ 20-micron resolutions, of PCIs performed within reanimated and perfusion-fixed heart specimens.

    Conclusions

    Generated computational models of left main PCIs performed in isolated human hearts can be used to obtain detailed measurements that provide further clinical insights on procedural outcomes. The 3D models from these procedures are useful for generating virtual reality scenes and 3D prints for physician training and education.

     
    more » « less
  2. null (Ed.)
    Abstract Endovascular navigation proficiency requires a significant amount of manual dexterity from surgeons. Objective performance measures derived from endovascular tool tip kinematics have been shown to correlate with expertise; however, such metrics have not yet been used during training as a basis for real-time performance feedback. This paper evaluates a set of velocity-based performance measures derived from guidewire motion to determine their suitability for online performance evaluation and feedback. We evaluated the endovascular navigation skill of 75 participants using three metrics (spectral arc length, average velocity, and idle time) as they steered tools to anatomical targets using a virtual reality simulator. First, we examined the effect of navigation task and experience level on performance and found that novice performance was significantly different from intermediate and expert performance. Then we computed correlations between measures calculated online and spectral arc length, our "gold standard" metric, calculated offline (at the end of the trial, using data from the entire trial). Our results suggest that average velocity and idle time calculated online are strongly and consistently correlated with spectral arc length computed offline, which was not the case when comparing spectral arc length computed online and offline. Average velocity and idle time, both time-domain based performance measures, are therefore more suitable measures than spectral arc length, a frequency-domain based metric, to use as the basis of online performance feedback. Future work is needed to determine how to best provide real-time performance feedback to endovascular surgery trainees based on these metrics. 
    more » « less
  3. Background Despite favorable outcomes of surgical pulmonary artery (PA) reconstruction, isolated proximal stenting of the central PAs is common clinical practice for patients with peripheral PA stenosis in association with Williams and Alagille syndromes. Given the technical challenges of PA reconstruction and the morbidities associated with transcatheter interventions, the hemodynamic consequences of all treatment strategies must be rigorously assessed. Our study aims to model, assess, and predict hemodynamic outcomes of transcatheter interventions in these patients. Methods and Results Isolated proximal and “extensive” interventions (stenting and/or balloon angioplasty of proximal and lobar vessels) were performed in silico on 6 patient‐specific PA models. Autoregulatory adaptation of the cardiac output and downstream arterial resistance was modeled in response to intervention‐induced hemodynamic perturbations. Postintervention computational fluid dynamics predictions were validated in 2 stented patients and quantitatively assessed in 4 surgical patients. Our computational methods accurately predicted postinterventional PA pressures, the primary indicators of success for treatment of peripheral PA stenosis. Proximal and extensive treatment achieved median reductions of 14% and 40% in main PA systolic pressure, 27% and 56% in pulmonary vascular resistance, and 10% and 45% in right ventricular stroke work, respectively. Conclusions In patients with Williams and Alagille syndromes, extensive transcatheter intervention is required to sufficiently reduce PA pressures and right ventricular stroke work. Transcatheter therapy was shown to be ineffective for long‐segment stenosis and pales hemodynamically in comparison with published outcomes of surgical reconstruction. Regardless of the chosen strategy, a virtual treatment planning platform could identify lesions most critical for optimizing right ventricular afterload. 
    more » « less
  4. Abstract Objectives

    To investigate the burden of peripheral intravenous catheters (PIVCs) in older hospitalised patients.

    Methods

    A cross‐sectional prospective observational study (2014/2015) to describe the characteristics, indications and outcomes of PIVCs among patients aged ≥65 from 65 Australian hospitals.

    Results

    Amongst 2179 individual PIVCs (in 2041 patients, mean age 77.6 years, 45% female, 58% in NSW), 43% were inserted by doctors and 74% used that day, meaning 25% were ‘idle’. Overall, 18% (393/2179) exhibited signs of PIVC‐related complications. Most commonly exhibited PIVC‐related complications were tenderness (4.1%) and local redness (1.8%). Nearly one in three (29.1%) dressings was soiled, loosened or had come off, and only 36.8% had the time and date documented on the dressing. Both infusing IV medications (aOR 1.74, 95% CI 1.28–2.38,p < 0.001) and inserting the PIVC in a non‐upper limb vein (aOR 3.40 compared to forearm [reference site], 95% CI 1.62–7.17,p < 0.001) were independently associated with PIVC failure. Phlebitis was exhibited in 7% (154) of the patients. Only infusing intravenous medications increased the likelihood of developing symptoms of phlebitis (aOR 1.61, 95% CI 1.01–2.57,= 0.05). Increasing age was inversely associated with symptoms of phlebitis. Among the 1575 patients (79%) who rated their PIVC experience using the Likert scale 0–10 (where 10 = ‘best possible’), the median score was 8 (IQR 6–10). Age in highest quartile (>84 years) was independently associated with lower likelihood of a high score (aOR 0.71, 95% CI 0.54–0.94,= 0.02).

    Conclusions

    Given 1 in 5 PIVCs were identified with having complications, further research should focus on optimising PIVC use in older patients.

     
    more » « less
  5. Patients with percutaneous coronary intervention generally receive either bare metal stents or drug‐eluting stents to restore the normal blood flow. However, due to the lack of stent production with an individual patient in mind, the same level of effectiveness may not be possible in treating two different clinical scenarios. This study introduces for the first time the feasibility of a patient‐specific stenting process constructed from direct 3D segmentation of medical images using direct 3D printing of biodegradable polymer–graphene composite with dual drug incorporation. A biodegradable polymer–carbon composite is prepared doped with graphene nanoplatelets to achieve controlled release of combinatorics as anticoagulation and antirestenosis agents. This study develops a technology prototyped for personalized stenting. An in silico analysis is performed to optimize the stent design for printing and its prediction of sustainability under force exerted by coronary artery or blood flow. A holistic approach covering in silico to in situ–in vivo establishes the structural integrity of the polymer composite, its mechanical properties, drug loading and release control, prototyping, functional activity, safety, and feasibility of placement in coronary artery of swine.

     
    more » « less