skip to main content


Title: Ultracompliant Carbon Nanotube Direct Bladder Device
Abstract

The bladder, stomach, intestines, heart, and lungs all move dynamically to achieve their purpose. A long‐term implantable device that can attach onto an organ, sense its movement, and deliver current to modify the organ function would be useful in many therapeutic applications. The bladder, for example, can suffer from incomplete contractions that result in urinary retention with patients requiring catheterization. Those affected may benefit from a combination of a strain sensor and electrical stimulator to better control bladder emptying. The materials and design of such a device made from thin layer carbon nanotube (CNT) and Ecoflex 00–50 are described and demonstrate its function with in vivo feline bladders. During bench‐top characterization, the resistive and capacitive sensors exhibit stability throughout 5000 stretching cycles under physiology conditions. In vivo measurements with piezoresistive devices show a high correlation between sensor resistance and volume. Stimulation driven from platinum‐silicone composite electrodes successfully induce bladder contraction. A method for reliable connection and packaging of medical grade wire to the CNT device is also presented. This work is an important step toward the translation of low‐durometer elastomers, stretchable CNT percolation, and platinum‐silicone composite, which are ideal for large‐strain bioelectric applications to sense or modulate dynamic organ states.

 
more » « less
NSF-PAR ID:
10459215
Author(s) / Creator(s):
 ;  ;  ;  ;  ;  ;  ;  
Publisher / Repository:
Wiley Blackwell (John Wiley & Sons)
Date Published:
Journal Name:
Advanced Healthcare Materials
Volume:
8
Issue:
20
ISSN:
2192-2640
Format(s):
Medium: X
Sponsoring Org:
National Science Foundation
More Like this
  1. The past few years have witnessed a growing interest in wireless and batteryless implants, due to their potential in long-term biomedical monitoring of in-body conditions such as internal organ movements, bladder pressure, and gastrointestinal health. Early proposals for batteryless implants relied on inductive near-field coupling and ultrasound harvesting, which require direct contact between the external power source and the human body. To overcome this near-field challenge, recent research has investigated the use of RF backscatter in wireless micro-implants because of its ability to communicate with wireless receivers that are placed at a distance outside the body (∼0.5 m), allowing a more seamless user experience. Unfortunately, existing far-field backscatter designs remain limited in their functionality: they cannot perform biometric sensing or secure data transmission; they also suffer from degraded harvesting efficiency and backscatter range due to the impact of variations in the surrounding tissues. In this paper, we present the design of a batteryless, wireless and secure system-on-chip (SoC) implant for in-body strain sensing. The SoC relies on four features: 1) employing a reconfigurable in-body rectenna which can operate across tissues adapting its backscatter bandwidth and center frequency; 2) designing an energy efficient 1.37 mmHg strain sensing front-end with an efficiency of 5.9 mmHg·nJ/conversion; 3) incorporating an AES-GCM security engine to ensure the authenticity and confidentiality of sensed data while sharing the ADC with the sensor interface for an area efficient random number generation; 4) implementing an over-the-air closed-loop wireless programming scheme to reprogram the RF front-end to adapt for surrounding tissues and the sensor front-end to achieve faster settling times below 2 s. 
    more » « less
  2. Abstract

    Inkjet printing is rapidly emerging as a means to fabricate low‐cost electronic devices; however, its widespread adoption is hindered by the complexity of the inks and the relatively high processing temperatures, limiting it to only a few metals and substrates. A new approach for inkjet printing is described, based on commercially available, particle‐free inks formulated from inorganic metal salts and their subsequent low‐temperature conversion to metallic structures by a non‐equilibrium, inert gas plasma. This single, general method is demonstrated for a library of metals including gold (Au), silver (Ag), copper (Cu), palladium (Pd), platinum (Pt), lead (Pb), bismuth (Bi), and tin (Sn). As one figure of merit, the resistivities of the printed metals are measured to be between 2× and 10× of the respective bulk metals. Uniquely, it is found that the printed metal films exhibit a very large surface area because of the plasma‐initiated nucleation and growth process, making this technique attractive for sensing device applications. A Bi‐based trace Pb sensor, an Au‐based amyloid‐β42sensor, and an Au‐based strain gauge are fabricated as representative chemical, biological, and mechanical sensors, and are found to exhibit enhanced sensitivity compared to analogues made with conventional methods.

     
    more » « less
  3. Abstract

    Thermoregulatory garments composed of liquid‐cooled plastic tubes have users ranging from astronauts to multiple sclerosis patients and are emerging as a flexible cooling solution for wearable electronics and high‐power robotics. Despite the plethora of applications, the current cooling systems are cumbersome to use due to their excessive size. In this work this issue is resolved by developing soft, thermally conductive silicone–aluminum composite tubes. To achieve optimal device performance, the material must be designed to balance the decrease in bulk thermal resistance and the increase in interfacial tube‐substrate resistance due to composite stiffening. Thus, to enable the rational design of such tubes, a closed form thermomechanical model that predicts cooling performance as a function of tube geometry and filler fraction is developed and experimentally validated. Predictions via this model and experiments are used to reveal how the tube's geometrical and material design can be adjusted to minimize the required length of tubing and maximize the heat extracted from a metallic surface and skin. Lastly, through a holistic analysis, this work demonstrates that besides significantly increasing overall cooling capability, the use of low‐resistance tubing can provide a multifold reduction in the cooling system size and enable novel operating modes.

     
    more » « less
  4. Many soft robotic components require highly stretchable, electrically conductive materials for proper operation. Often these conductive materials are used as sensors or as heaters for thermally responsive materials. However, there is a scarcity of stretchable materials that can withstand the high strains typically experienced by soft robots, while maintaining the electrical properties necessary for Joule heating ( e.g. , uniform conductivity). In this work, we present a silicone composite containing both liquid and solid inclusions that can maintain a uniform conductivity while experiencing 200% linear strains. This composite can be cast in thin sheets enabling it to be wrapped around thermally responsive soft materials that increase their volume or stretchability when heated. We show how this material opens up possibilities for electrically controllable shape changing soft robotic actuators, as well as all-silicone actuation systems powered only by electrical stimulus. Additionally, we show that this stretchable composite can be used as an electrode material in other applications, including a strain sensor with a linear response up to 200% strain and near-zero signal noise. 
    more » « less
  5. Microbial adhesion to medical devices is common for hospital-acquired infections, particularly for urinary catheters. If not properly treated these infections cause complications and exacerbate antimicrobial resistance. Catheter use elicits bladder inflammation, releasing host serum proteins, including fibrinogen (Fg), into the bladder, which deposit on the urinary catheter.Enterococcus faecalisuses Fg as a scaffold to bind and persist in the bladder despite antibiotic treatments. Inhibition of Fg–pathogen interaction significantly reduces infection. Here, we show deposited Fg is advantageous for uropathogensE. faecalis,Escherichia coli,Pseudomonas aeruginosa,K. pneumoniae,A. baumannii, andC. albicans, suggesting that targeting catheter protein deposition may reduce colonization creating an effective intervention for catheter-associated urinary tract infections (CAUTIs). In a mouse model of CAUTI, host-protein deposition was reduced, using liquid-infused silicone catheters, resulting in decreased colonization on catheters, in bladders, and dissemination in vivo. Furthermore, proteomics revealed a significant decrease in deposition of host-secreted proteins on liquid-infused catheter surfaces. Our findings suggest targeting microbial-binding scaffolds may be an effective antibiotic-sparing intervention for use against CAUTIs and other medical device infections.

     
    more » « less