Liquid-infused polymers are recognized for their ability to repel foulants, making them promising for biomedical applications including catheter-associated urinary tract infections (CAUTIs). However, the impact of the quantity of free liquid layer covering the surface on protein and bacterial adhesion is not well understood. Here, we explore how the amount of free silicone liquid layer in infused silicone catheter materials influences the adhesion of bacteria and proteins relevant to CAUTIs. To alter the quantity of the free liquid layer, we either physically removed excess liquid from fully infused catheter materials or partially infused them. We then evaluated the impact on bacterial and host protein adhesion. Physical removal of the free liquid layer from the fully infused samples reduced the height of the liquid layer from 60 μm to below detection limits and silicone liquid loss into the environment by approximately 64% compared to controls, without significantly increasing the deposition of protein fibrinogen or the adhesion of the common uropathogen Enterococcus faecalis. Partially infused samples showed even greater reductions in liquid loss: samples infused to 70%–80% of their maximum capacity exhibited about an 85% decrease in liquid loss compared to fully infused controls. Notably, samples with more than 70% infusion did not show significant increases in fibrinogen or E. faecalis adhesion. These findings suggest that adjusting the levels of the free liquid layer in infused polymers can influence protein and bacterial adhesion on their surfaces. Moreover, removing the free liquid layer can effectively reduce liquid loss from these polymers while maintaining their functionality.
more »
« less
Inhibiting host-protein deposition on urinary catheters reduces associated urinary tract infections
Microbial adhesion to medical devices is common for hospital-acquired infections, particularly for urinary catheters. If not properly treated these infections cause complications and exacerbate antimicrobial resistance. Catheter use elicits bladder inflammation, releasing host serum proteins, including fibrinogen (Fg), into the bladder, which deposit on the urinary catheter.Enterococcus faecalisuses Fg as a scaffold to bind and persist in the bladder despite antibiotic treatments. Inhibition of Fg–pathogen interaction significantly reduces infection. Here, we show deposited Fg is advantageous for uropathogensE. faecalis,Escherichia coli,Pseudomonas aeruginosa,K. pneumoniae,A. baumannii, andC. albicans, suggesting that targeting catheter protein deposition may reduce colonization creating an effective intervention for catheter-associated urinary tract infections (CAUTIs). In a mouse model of CAUTI, host-protein deposition was reduced, using liquid-infused silicone catheters, resulting in decreased colonization on catheters, in bladders, and dissemination in vivo. Furthermore, proteomics revealed a significant decrease in deposition of host-secreted proteins on liquid-infused catheter surfaces. Our findings suggest targeting microbial-binding scaffolds may be an effective antibiotic-sparing intervention for use against CAUTIs and other medical device infections.
more »
« less
- Award ID(s):
- 2029378
- PAR ID:
- 10487499
- Publisher / Repository:
- eLife
- Date Published:
- Journal Name:
- eLife
- Volume:
- 11
- ISSN:
- 2050-084X
- Format(s):
- Medium: X
- Sponsoring Org:
- National Science Foundation
More Like this
-
-
null (Ed.)Abstract Goal: This work introduces an integrated system incorporated seamlessly with a commercial Foley urinary catheter for bacterial growth sensing and biofilm treatment. Methods: The system is comprised of flexible, interdigitated electrodes incorporated with a urinary catheter via a 3D-printed insert for impedance sensing and bioelectric effect-based treatment. Each of the functions were wirelessly controlled using a custom application that provides a user-friendly interface for communicating with a custom PCB via Bluetooth to facilitate implementation in practice. Results: The integrated catheter system maintains the primary functions of indwelling catheters - urine drainage, balloon inflation - while being capable of detecting the growth of Escherichia coli, with an average decrease in impedance of 13.0% after 24 hours, tested in a newly-developed simulated bladder environment. Furthermore, the system enables bioelectric effect-based biofilm reduction, which is performed by applying a low-intensity electric field that increases the susceptibility of biofilm bacteria to antimicrobials, ultimately reducing the required antibiotic dosage. Conclusion: Overall, this modified catheter system represents a significant step forward for catheter-associated urinary tract infection (CAUTI) management using device-based approaches, integrating flexible electrodes with an actual Foley catheter along with the control electronics and mobile application. Significance: CAUTIs, exacerbated by the emergence of antibiotic-resistant pathogens, represent a significant challenge as one of the most prevalent healthcare-acquired infections. These infections are driven by the colonization of indwelling catheters by bacterial biofilms.more » « less
-
Bacteria can swim upstream in a narrow tube and pose a clinical threat of urinary tract infection to patients implanted with catheters. Coatings and structured surfaces have been proposed to repel bacteria, but no such approach thoroughly addresses the contamination problem in catheters. Here, on the basis of the physical mechanism of upstream swimming, we propose a novel geometric design, optimized by an artificial intelligence model. UsingEscherichia coli, we demonstrate the anti-infection mechanism in microfluidic experiments and evaluate the effectiveness of the design in three-dimensionally printed prototype catheters under clinical flow rates. Our catheter design shows that one to two orders of magnitude improved suppression of bacterial contamination at the upstream end, potentially prolonging the in-dwelling time for catheter use and reducing the overall risk of catheter-associated urinary tract infection.more » « less
-
null (Ed.)Underactive bladder or detrusor underactivity (DUA), that is, not being able to micturate, has received less attention with little research and remains unknown or limited on pathological causes and treatments as opposed to overactive bladder, although the syndrome may pose a risk of urinary infections or life-threatening kidney damage. Here, we present an integrated expandable electronic and optoelectronic complex that behaves as a single body with the elastic, time-dynamic urinary bladder with substantial volume changes up to ~300%. The system configuration of the electronics validated by the theoretical model allows conformal, seamless integration onto the urinary bladder without a glue or suture, enabling precise monitoring with various electrical components for real-time status and efficient optogenetic manipulation for urination at the desired time. In vivo experiments using diabetic DUA models demonstrate the possibility for practical uses of high-fidelity electronics in clinical trials associated with the bladder and other elastic organs.more » « less
-
Tissue failure at suture lines contributes to complications and readmissions following complex surgeries in distensible organs such as those performed in the lower urinary tract. Excess tension at points of tissue approximation can contribute to abnormal wound healing, urine leaks, infections, and fistula development. A flexible biodegradable adhesive patch that adheres to dynamic tissue and prevents non-targeted adhesion to adjacent tissue is needed to provide support at suture sites throughout the wound healing process. Herein, we have developed a ready-to-use bilayer adhesive patch (BLAP) to reinforce suture lines for application to expandable and dynamic fluid-filled tissues such as the bladder. The external non-adhesive layer of BLAP comprises a bioabsorbable poly(glycerol sebacate) (PGS) elastomer, preventing undesired adhesion to the adjunct tissues. The internal tissue binding layer is composed of PGS modified with L-dopamine (L-DOPA) to allow immediate adhesion to the wet surface of the target tissue. Physical and mechanical properties of the patches were tuned by varying glycerol to sebacate ratios, L-DOPA contents, and curing time to achieve compliance that approximates that of bladder tissue. The candidate PG2S and PG2SD0.018 biomaterials of the designed BLAP demonstrated Young's moduli of 49.4 kPa and 61.5 kPa and stretchability between 174.7% and 223.7%, respectively. BLAP adhered tightly to a porcine bladder repaired cystotomy ex vivo, reinforcing the sutured line and increasing bladder burst pressure more than stand-alone surgical sutures or a commercial bioadhesive glue, Tisseel®. These features, combined with >90% cytocompatibility and biodegradability, render BLAP a promising elastic bioadhesive patch to reinforce suture lines in the bladder. Beyond the urinary tract, BLAP has the potential to be mechanically tuned for a variety of other non-planar, dynamic tissues.more » « less