Abstract Studying and quantifying the mechanics of blood clots is essential to better diagnosis and prognosis of, as well as therapy for, thromboembolic pathologies such as strokes, heart attacks, and pulmonary embolisms. Unfortunately, mechanically testing blood clots is complicated by their softness and fragility, thus making the use of classic mounting techniques, such as clamping, challenging. This is particularly true for mechanical testing under large deformation. Here, we describe protocols for creating in vitro blood clots and securely mounting these samples on mechanical test equipment. To this end, we line 3D‐printed molds with a hook‐and‐loop fabric that, after coagulation, provides a secure interface between the sample and device mount. In summary, our molding and mounting protocols are ideal for performing large‐deformation mechanical testing, with samples that can withstand substantial deformation without delaminating from the apparatus. © 2021 Wiley Periodicals LLC. Basic Protocol 1: Cube‐shaped blood clot preparation Basic Protocol 2: Sheet‐shaped blood clot preparation
more »
« less
Mechanical properties of clot made from human and bovine whole blood differ significantly
Thromboembolism – that is, clot formation and the subsequent fragmentation of clot – is a leading cause of death worldwide. Clots’ mechanical properties are critical determinants of both the embolization process and the pathophysiological consequences thereof. Thus, understanding and quantifying the mechanical properties of clots is important to our ability to treat and prevent thromboembolic disease. However, assessing these properties from in vivo clots is experimentally challenging. Therefore, we and others have turned to studying in vitro clot mimics instead. Unfortunately, there are significant discrepancies in the reported properties of these clot mimics, which have been hypothesized to arise from differences in experimental techniques and blood sources. The goal of our current work is therefore to compare the mechanical behavior of clots made from the two most common sources, human and bovine blood, using the same experimental techniques. To this end, we tested clots under pure shear with and without initial cracks, under cyclic loading, and under stress relaxation. Based on these data, we computed and compared stiffness, strength, work-to-rupture, fracture toughness, relaxation time constants, and prestrain. While clots from both sources behaved qualitatively similarly, they differed quantitatively in almost every metric. We also correlated each mechanical metric to measures of blood composition. Thereby, we traced this inter-species variability in clot mechanics back to significant differences in hematocrit, but not platelet count. Thus, our work suggests that the results of past studies that have used bovine blood to make in vitro mimics – without adjusting blood composition – should be interpreted carefully. Future studies about the mechanical properties of blood clots should focus on human blood alone.
more »
« less
- Award ID(s):
- 2105175
- PAR ID:
- 10517632
- Publisher / Repository:
- Elsevier
- Date Published:
- Journal Name:
- Journal of the Mechanical Behavior of Biomedical Materials
- Volume:
- 154
- Issue:
- C
- ISSN:
- 1751-6161
- Page Range / eLocation ID:
- 106508
- Format(s):
- Medium: X
- Sponsoring Org:
- National Science Foundation
More Like this
-
-
Abstract Thromboembolic diseases are a significant cause of mortality and are clinically treated enzymatically with tissue plasminogen activator (tPA). Interestingly, prior studies in fibrin fibers and fibrin gels have demonstrated that thrombolysis may be mechanically sensitive. This study aims to expand mechano‐lytic studies to whole blood clots. Furthermore, this study investigates not only how mechanics impacts lysis but also how lysis impacts mechanics. Therefore, clots made from whole human blood are exposed to tPA while the clots are either stretched or unstretched. After, the resulting degree of clot lysis is measured by weighing the clots and by measuring the concentration of D‐dimer in the surrounding bath. Additionally, each clot's mechanical properties are measured. This study finds that mechanical stretch accelerates loss in clot weight but does not impact the lysis rate as measured by D‐dimer. Moreover, lysis not only removes clot volume but also reduces the remaining clot's stiffness and toughness. In summary, tPA‐induced lysis of whole clot appears mechanically insensitive, but stretch reduces clot weight. Furthermore, results show that thrombolysis weakens clot. This suggests that thrombolysis may increase the risk of secondary embolizations but may also ease clot removal during thrombectomy, for example.more » « less
-
Abstract Tissue mimicking materials are designed to represent real tissue in applications such as medical device testing and surgical training. Thanks to progress in 3D‐printing technology, tissue mimics can now be easily cast into arbitrary geometries and manufactured with adjustable material properties to mimic a wide variety of tissue types. However, it is unclear how well 3D‐printable mimics represent real tissues and their mechanics. The objective of this work is to fill this knowledge gap using the Stratasys Digital Anatomy 3D‐Printer as an example. To this end, we created mimics of biological tissues we previously tested in our laboratory: blood clots, myocardium, and tricuspid valve leaflets. We printed each tissue mimic to have the identical geometry to its biological counterpart and tested the samples using identical protocols. In our evaluation, we focused on the stiffness of the tissues and their fracture toughness in the case of blood clots. We found that the mechanical behavior of the tissue mimics often differed substantially from the biological tissues they aim to represent. Qualitatively, tissue mimics failed to replicate the traditional strain‐stiffening behavior of soft tissues. Quantitatively, tissue mimics were stiffer than their biological counterparts, especially at small strains, in some cases by orders of magnitude. In those materials in which we tested toughness, we found that tissue mimicking materials were also much tougher than their biological counterparts. Thus, our work highlights limitations of at least one 3D‐printing technology in its ability to mimic the mechanical properties of biological tissues. Therefore, care should be taken when using this technology, especially where tissue mimicking materials are expected to represent soft tissue properties quantitatively. Whether other technologies fare better remains to be seen.more » « less
-
When a thrombus breaks off and embolizes it can occlude vital vessels such as those of the heart, lung, or brain. These thromboembolic conditions are responsible for 1 in 4 deaths worldwide. Thrombus resistance to embolization is driven by its intrinsic fracture toughness as well as other, non-surface-creating dissipative mechanisms. In our current work, we identify and quantify these latter mechanisms toward future studies that aim to delineate fracture from other forms of dissipation. To this end, we use an in vitro thrombus mimic system to produce whole blood clots and explore their dissipative mechanics under simple uniaxial extension, cyclic loading, and stress-relaxation. We found that whole blood clots exhibit Mullins-like effect, hysteresis, permanent set, strain-rate dependence, and nonlinear stress-relaxation. Interestingly, we found that performing these tests under dry or submerged conditions did not change our results. However, performing these tests under room temperature or body temperature conditions yielded differences. Importantly, because we use venous blood our work is most closely related to venous in vivo blood clots. Overall, we have demonstrated that whole blood clots show several dissipative phenomena – similarly to hydrogels – that will be critical to our understanding of thrombus embolization.more » « less
-
Thrombus formation in blood-contacting medical devices is a major concern in the medical device industry, limiting the clinical efficacy of these devices. Further, a locally formed clot within the device has the potential to detach from the surface, posing a risk of embolization. Clot embolization from blood-contacting cardiovascular devices can result in serious complications like acute ischemic stroke and myocardial infarction. Therefore, clot embolization associated with device-induced thrombosis can be life-threatening and requires an enhanced fundamental understanding of embolization characteristics to come up with advanced intervention strategies. Therefore, this work aims to investigate the adhesive characteristics of blood clots on common biocompatible materials used in various cardiovascular devices. This study focuses on characterizing the adhesion strength of blood clots on materials such as polytetrafluoroethylene (PTFE), polyurethane (PU), polyether ether ketone (PEEK), nitinol, and titanium, frequently used in medical devices. In addition, the effect of incubation time on clot adhesion is explored. Results from this work demonstrated strongest clot adhesion to titanium with 3 h of incubation resulting in 1.06 ± 0.20 kPa detachment stresses. The clot adhesion strength on titanium was 51.5% higher than PEEK, 35.9% higher than PTFE, 63.1% higher than PU, and 35.4% higher than nitinol. Further, adhesion strength increases with incubation time for all materials. The percentage increase in detachment stress over incubation time (ranging from 30 min to 3 h) for polymers ranged from at least 108.75% (PEEK), 140.74% (PU), to 151.61% (PTFE). Whereas, for metallic surfaces, the percentage rise ranged from 70.21% (nitinol) to 89.28% (titanium). Confocal fluorescence imaging of clot remnants on the material surfaces revealed a well-bounded platelet-fibrin network at the residual region, representing a comparatively higher adhesive region than the non-residual zone of the surface.more » « less
An official website of the United States government