Title: Characterization of a 220Rn source for low-energy electronic recoil calibration of the XENONnT detector
Low-background liquid xenon detectors are utilized in the investigation of rare events, including dark matter and neutrinoless double beta decay. For their calibration, gaseous 220Rn can be used. After being introduced into the xenon, its progeny isotope 212Pb induces homogeneously distributed, low-energy (<30 keV) electronic recoil interactions. We report on the characterization of such a source for use in the XENONnT experiment. It consists of four commercially available 228Th sources with an activity of 55 kBq. These sources provide a high 220Rn emanation rate of about 8 kBq. We find no indication for the release of the long-lived 228Th above 1.7 mBq. Though an unexpected 222Rn emanation rate of about 3.6 mBq is observed, this source is still in line with the requirements for the XENONnT experiment.  more » « less
Award ID(s):
2112803
PAR ID:
10526544
Author(s) / Creator(s):
; ; ; ;
Publisher / Repository:
SISSA
Date Published:
Journal Name:
Journal of Instrumentation
Volume:
18
Issue:
11
ISSN:
1748-0221
Page Range / eLocation ID:
P11009
Format(s):
Medium: X
Sponsoring Org:
National Science Foundation
More Like this
  1. Abstract The selection of low-radioactive construction materials is of the utmost importance for rare-event searches and thus critical to the XENONnT experiment. Results of an extensive radioassay program are reported, in which material samples have been screened with gamma-ray spectroscopy, mass spectrometry, and $$^{222}$$ 222 Rn emanation measurements. Furthermore, the cleanliness procedures applied to remove or mitigate surface contamination of detector materials are described. Screening results, used as inputs for a XENONnT Monte Carlo simulation, predict a reduction of materials background ( $$\sim $$ ∼ 17%) with respect to its predecessor XENON1T. Through radon emanation measurements, the expected $$^{222}$$ 222 Rn activity concentration in XENONnT is determined to be 4.2 ( $$^{+0.5}_{-0.7}$$ - 0.7 + 0.5 )  $$\upmu $$ μ Bq/kg, a factor three lower with respect to XENON1T. This radon concentration will be further suppressed by means of the novel radon distillation system. 
    more » « less
  2. Abstract The multi-staged XENON program at INFN Laboratori Nazionali del Gran Sasso aims to detect dark matter with two-phase liquid xenon time projection chambers of increasing size and sensitivity. The XENONnT experiment is the latest detector in the program, planned to be an upgrade of its predecessor XENON1T. It features an active target of 5.9 tonnes of cryogenic liquid xenon (8.5 tonnes total mass in cryostat). The experiment is expected to extend the sensitivity to WIMP dark matter by more than an order of magnitude compared to XENON1T, thanks to the larger active mass and the significantly reduced background, improved by novel systems such as a radon removal plant and a neutron veto. This article describes the XENONnT experiment and its sub-systems in detail and reports on the detector performance during the first science run. 
    more » « less
  3. Abstract The selection of low-radioactive construction materials is of utmost importance for the success of low-energy rare event search experiments. Besides radioactive contaminants in the bulk, the emanation of radioactive radon atoms from material surfaces attains increasing relevance in the effort to further reduce the background of such experiments. In this work, we present the$$^{222}$$ 222 Rn emanation measurements performed for the XENON1T dark matter experiment. Together with the bulk impurity screening campaign, the results enabled us to select the radio-purest construction materials, targeting a$$^{222}$$ 222 Rn activity concentration of$$10\,\mathrm{\,}\upmu \mathrm{Bq}/\mathrm{kg}$$ 10 μ Bq / kg in$$3.2\,\mathrm{t}$$ 3.2 t of xenon. The knowledge of the distribution of the$$^{222}$$ 222 Rn sources allowed us to selectively eliminate problematic components in the course of the experiment. The predictions from the emanation measurements were compared to data of the$$^{222}$$ 222 Rn activity concentration in XENON1T. The final$$^{222}$$ 222 Rn activity concentration of$$(4.5\pm 0.1)\,\mathrm{\,}\upmu \mathrm{Bq}/\mathrm{kg}$$ ( 4.5 ± 0.1 ) μ Bq / kg in the target of XENON1T is the lowest ever achieved in a xenon dark matter experiment. 
    more » « less
  4. Abstract The XENONnT detector uses the latest and largest liquid xenon-based time projection chamber (TPC) operated by the XENON Collaboration, aimed at detecting Weakly Interacting Massive Particles and conducting other rare event searches.The XENONnT data acquisition (DAQ) system constitutes an upgraded and expanded version of the XENON1T DAQ system.For its operation, it relies predominantly on commercially available hardware accompanied by open-source and custom-developed software.The three constituent subsystems of the XENONnT detector, the TPC (main detector), muon veto, and the newly introduced neutron veto, are integrated into a single DAQ, and can be operated both independently and as a unified system.In total, the DAQ digitizes the signals of 698 photomultiplier tubes (PMTs), of which 253 from the top PMT array of the TPC are digitized twice, at ×10 and ×0.5 gain.The DAQ for the most part is a triggerless system, reading out and storing every signal that exceeds the digitization thresholds.Custom-developed software is used to process the acquired data, making it available within ∼30 s for live data quality monitoring and online analyses.The entire system with all the three subsystems was successfully commissioned and has been operating continuously, comfortably withstanding readout rates that exceed ∼500 MB/s during calibration.Livetime during normal operation exceeds 99% and is ∼90% during most high-rate calibrations.The combined DAQ system has collected more than 2 PB of both calibration and science data during the commissioning of XENONnT and the first science run. 
    more » « less
  5. Internet-of-things (IoT) devices (e.g., micro camera and microphone) are usually small form factor, low-cost, and low-power, which makes them easy to conceal and deploy in the indoor environment to spy on people for human private information such as location and indoor activities. As a result, these IoT devices introduce a great privacy and ethical threat. Therefore, it is important to reveal these concealed IoT devices in the indoor environment for human privacy protection. This paper presents RFScan, a system that can passively detect, fingerprint, and localize diverse concealed IoT devices in the indoor environment by sensing their unintentional electromagnetic emanations. However, sensing these emanations is challenging due to the weak emanation strength and the interference from the ambient wireless communication signals. To this end, we boost the emanation strength through the non-coherent averaging based on the emanation signal's characteristics and design a novel suppression algorithm to mitigate interference from the wireless communication signals. We further profile emanations across frequency and time that act as the emanation source's unique signature and customize a deep neural network architecture to fingerprint the emanation sources. Furthermore, we can localize the emanation source with an angle-of-arrival (AoA) based triangulation approach. Our experimental results demonstrate the efficiency of the IoT devices' detection, fingerprinting, and localization across different indoor environments. 