Title: Measurements of volatile organic compounds in ambient air by gas-chromatography and real-time Vocus PTR-TOF-MS: calibrations, instrument background corrections, and introducing a PTR Data Toolkit
Volatile organic compound (VOC) emissions and subsequent oxidation contribute to the formation of secondary pollutants and poor air quality in general. As more VOCs at lower mixing ratios have become the target of air quality investigations, their quantification has been aided by technological advancements in proton-transfer-reaction time-of-flight mass spectrometry (PTR-TOF-MS). However, such quantification requires appropriate instrument background measurements and calibrations, particularly for VOCs without calibration standards. This study utilized a Vocus PTR-TOF-MS coupled with a gas chromatograph for real-time and speciated measurements of ambient VOCs in Boulder, Colorado, during spring 2021. The aim of these measurements was to understand and characterize instrument response and temporal variability as to inform the quantification of a broader range of detected VOCs. Fast, frequent calibrations were made every 2 h in addition to daily multipoint calibrations. Sensitivities derived from the fast calibrations were 5 ± 6 % (average and 1 standard deviation) lower than those derived from the multipoint calibrations due to an offset between the calibrations and instrument background measurement. This offset was caused, in part, by incomplete mixing of the standard with diluent. These fast calibrations were used in place of a normalization correction to account for variability in instrument response and accounted for non-constant reactor conditions caused by a gradual obstruction of the sample inlet. One symptom of these non-constant conditions was a trend in fragmentation, although the greatest observed variability was 6 % (1 relative standard deviation) for isoprene. A PTR Data Toolkit (PTR-DT) was developed to assess instrument performance and rapidly estimate the sensitivities of VOCs which could not be directly calibrated on the timescale of the fast calibrations using the measured sensitivities of standards, molecular properties, and simple reaction kinetics. Through this toolkit, the standards' sensitivities were recreated within 1 ± 8 % of the measured values. Three clean-air sources were compared: a hydrocarbon trap, zero-grade air and ultra-high purity nitrogen, and a catalytic zero-air generator. The catalytic zero-air generator yielded the lowest instrument background signals for the majority of ions, followed by the hydrocarbon trap. Depending on the ionization efficiency, product ion fragmentation, ion transmission, and instrument background, standards' limits of detection (5 s measurement integration) derived from the catalytic zero-air generator and the fast calibration sensitivities ranged from 2 ppbv (methanol) to 1 pptv (decamethylcyclopentasiloxane; D5 siloxane) with most standards having detection limits below 20 pptv. Finally, applications of measurements with low detection limits are considered for a few low-signal species including sub-parts-per-trillion by volume (pptv) enhancements of icosanal (and isomers; 1 min average) in a plume of cooking emissions, and sub-parts-per-trillion by volume enhancements in dimethyl disulfide in plumes containing other organosulfur compounds. Additionally, chromatograms of hexamethylcyclotrisiloxane, octamethylcyclotetrasiloxane, and decamethylcyclopentasiloxane (D3, D4, and D5 siloxanes, respectively), combined with high sensitivity, suggest that online measurements can reasonably be associated with the individual isomers.  more » « less
Award ID(s):
2206655
PAR ID:
10588855
Author(s) / Creator(s):
; ; ;
Publisher / Repository:
Copernicus
Date Published:
Journal Name:
Atmospheric Measurement Techniques
Volume:
16
Issue:
21
ISSN:
1867-8548
Page Range / eLocation ID:
5261 to 5285
Subject(s) / Keyword(s):
Sponsoring Org:
National Science Foundation
More Like this
  1. Abstract. Furanoids are significant contributors to volatile organic compound hydroxyl radical reactivity in biomass burning emissions, yet their accurate measurement using proton-transfer-reaction time-of-flight mass spectrometry (PTR-ToF-MS) remains challenging due to potential interferences and measurement uncertainties. In this study, we conduct detailed laboratory characterizations of furan (C4H4O, protonated m/z 69.033), 2(3H)-furanone (C4H4O2, m/z 85.028), 2-furaldehyde (C5H4O2, m/z 97.028), 2,5-dimethylfuran (C6H8O, m/z 97.065), and maleic anhydride (C4H2O3, m/z 99.008). Sensitivities for these compounds were found to have minimal dependence (less than 15 %) on both sample humidity and drift tube electric field strength (E/N). At high E/N (150 Td), fragmentation was observed for 2-furaldehyde (∼ 8 %) at m/z 69.033, potentially interfering furan measurements, while hydrolysis products corresponding to m/z+18 ions were detected for 2(3H)-furanone, 2-furaldehyde, and maleic anhydride. The hydrolysis of maleic anhydride to maleic acid is most prominent, with the hydrate accounting for 7 %–31 % of the parent ion signal across E/N conditions. Gas standard recertification confirmed the long-term stability of furanoids, and 21 other VOCs, in compressed gas mixtures, with changes in mixing ratios of less than 5 % over seven years, although PTR-ToF-MS instrument sensitivities decreased by ∼ 30 % during this time, likely due to aging of the microchannel plate (MCP). While the stability of gas standards and the minimal humidity and fragmentation effects support the accurate measurement of furanoids by PTR-ToF-MS, discrepancies with co-deployed gas chromatography-mass spectrometry highlight the need to further investigate potential isomeric and fragment interferences, particularly in aged BB smoke. 
    more » « less
  2. Scented wax products, such as candles and wax warmers/melts, are popular fragranced consumer products that are commonly used in residential buildings. As scented wax products are intentionally fragranced to produce pleasant smellscapes for occupants, they may represent an important source of volatile organic compounds (VOCs) to indoor atmospheres. The aim of this study is to evaluate terpene emission factors (EFs) and inhalation intake fractions (iFs) for scented wax products to better understand their impact on indoor chemistry and chemical exposures. Full-scale emission experiments were conducted in the Purdue zEDGE Test House using a variety of scented candles (n = 5) and wax warmers/melts (n = 14) under different outdoor air exchange rates (AERs). Terpene concentrations were measured in real-time using a proton transfer reaction time-of-flight mass spectrometer (PTR-TOF-MS). PTR-TOF-MS measurements revealed that scented candle and wax warmer/melt products emit a variety of monoterpenes (C10H16) and oxygen-containing monoterpenoids (C10H14O, C10H16O, C10H18O, C10H20O), with peak concentrations in the range of 10^−1 to 10^2 ppb. Monoterpene EFs were much greater for scented wax warmers/melts (C10H16 EFs ~ 10^2 mg per g wax consumed) compared to scented candles (C10H16 EFs ~ 10^−1 to 100 mg per g wax consumed). Significant emissions of reactive terpenes from both products, along with nitrogen oxides (NO, NO2) from candles, depleted indoor ozone (O3) concentrations. Terpene iFs were similar between the two products (iFs ~ 10^3 ppm) and increased with decreasing outdoor AER. Terpene iFs during concentration decay periods were similar to, or greater than, iFs during active emission periods for outdoor AERs ≤ 3.0 h^−1. Overall, scented wax warmers/melts were found to release greater quantities of monoterpenes compared to other fragranced consumer products used in the home, including botanical disinfectants, hair care products, air fresheners, and scented sprays. 
    more » « less
  3. Abstract. Obtaining quantitative information for molecular species present in aerosols from real-time mass spectrometers such as an extractive electrospray time-of-flight mass spectrometer (EESI) and an aerosol mass spectrometer (AMS) can be challenging. Typically, molecular species are calibrated directly through the use of pure standards. However, in some cases (e.g., secondary organic aerosol (SOA) formed from volatile organic compounds (VOCs)), direct calibrations are impossible, as many SOA species can either not be purchased as pure standards or have ambiguous molecular identities. In some cases, bulk OA sensitivities are used to estimate molecular sensitivities. This approach is not sufficient for EESI, which measures molecular components of OA, because different species can have sensitivities that vary by a factor of more than 30. Here, we introduce a method to obtain EESI calibration factors when standards are not available, and we provide a thorough analysis of the feasibility, performance, and limitations of this new technique. In this method, complex aerosol mixtures were separated with high-performance liquid chromatography (HPLC) followed by aerosol formation via atomization. The separated aerosols were then measured by an EESI and an AMS, which allowed us to obtain sensitivities for some species present in standard and SOA mixtures. Pure compounds were used to test the method and characterize its uncertainties, and obtained sensitivities were consistent within ±20 % when comparing direct calibrations vs. HPLC calibrations for a pure standard and within a factor of 2 for a standard mixture. In some cases, species were not completely resolved by chromatography, and positive matrix factorization (PMF) of AMS data enabled further separation. This method should be applicable to other real-time MS techniques. Improvements in chromatography are possible that would allow better separation in complex mixtures. 
    more » « less
  4. Abstract. A new ion source (IS) utilizing vacuum ultraviolet (VUV) light is developed and characterized for use with iodide–chemical ionization massspectrometers (I−-CIMS). The VUV-IS utilizes a compact krypton lamp that emits light at two wavelengths corresponding to energies of∼10.030 and 10.641 eV. The VUV light photoionizes either methyl iodide (ionization potential, IP = 9.54 ± 0.02 eV)or benzene (IP = 9.24378 ± 0.00007 eV) to form cations and photoelectrons. The electrons react with methyl iodide to formI−, which serves as the reagent ion for the CIMS. The VUV-IS is characterized by measuring the sensitivity of a quadrupole CIMS (Q-CIMS) toformic acid, molecular chlorine, and nitryl chloride under a variety of flow and pressure conditions. The sensitivity of the Q-CIMS, with theVUV-IS, reached up to ∼700 Hz pptv−1, with detection limits of less than 1 pptv for a 1 min integration period. Thereliability of the Q-CIMS with a VUV-IS is demonstrated with data from a month-long ground-based field campaign. The VUV-IS is further tested byoperation on a high-resolution time-of-flight CIMS (TOF-CIMS). Sensitivities greater than 25 Hz pptv−1 were obtained for formic acid andmolecular chlorine, which were similar to that obtained with a radioactive source. In addition, the mass spectra from sampling ambient air wascleaner with the VUV-IS on the TOF-CIMS compared to measurements using a radioactive source. These results demonstrate that the VUV lamp is a viablesubstitute for radioactive ion sources on I−-CIMS systems for most applications. In addition, initial tests demonstrate that the VUV-IS canbe extended to other reagent ions by the use of VUV absorbers with low IPs to serve as a source of photoelectrons for high IP electron attachers,such as SF6-. 
    more » « less
  5. Abstract We present measurements of volatile organic compounds (VOCs) and other trace gases taken in Salt Lake City, Utah in August and September 2022. As part of the Salt Lake regional Smoke, Ozone and Aerosol Study (SAMOZA), 35 VOCs were measured with two methods: a proton‐transfer‐reaction time‐of‐flight mass spectrometer (PTR‐ToF‐MS) and 2,4‐dinitrophenylhydrazine (DNPH) cartridges analyzed by high‐performance liquid chromatography (HPLC). Over two months, the total measured VOCs averaged 32 ± 24 ppb (mean ± standard deviation) with the hourly maximum at 141 ppb, and the total calculated OH reactivity averaged 3.7 ± 3.0 s−1(maximum at 20.7 s−1). Among them, methanol and ethanol were the most abundant VOCs, making up 42% of the ambient mixing ratio. Isoprene and monoterpenes contributed 25% of the OH reactivity from VOCs, while formaldehyde and acetaldehyde made up another 30%. The positive matrix factorization analysis showed 5 major sources of VOCs, with 32% of abundance being attributed to secondary production/biogenic sources, 44% from the combination of traffic and personal care products, 15% from industrial solvent use, and the rest from biomass burning (10%). Moderate smoke‐impacted days elevated various hazardous air pollutants (HAPs) on average by 45%–217% compared to smoke‐free days. The ratio of OH reactivity from NOxto that from VOCs showed that ozone production was mostly VOC‐limited throughout the campaign, consistent with our modeling study. VOCs and NOxboth showed increased OH reactivity due to smoke influence. NOxfeatured increased reactivity on weekdays compared to weekends, an effect not shown for VOC reactivity during SAMOZA. 
    more » « less