  1. Abstract

    Ion‐sensitive field effect transistor‐based pH sensors have been shown to perform well in high frequency and long‐term ocean sampling regimes. The Honeywell Durafet is widely used due to its stability, fast response, and characterization over a large range of oceanic conditions. However, potentiometric pH monitoring is inherently complicated by the fact that the sensors require careful calibration. Offsets in calibration coefficients have been observed when comparing laboratory to field‐based calibrations and prior work has led to the recommendation that an in situ calibration be performed based on comparison to discrete samples. Here, we describe our work toward a self‐calibration apparatus integrated into a SeapHOx pH, dissolved oxygen, and CTD sensor package. This Self‐Calibrating SeapHOx is capable of autonomously recording calibration values from a high quality, traceable, primary reference standard: equimolar tris buffer. The Self‐Calibrating SeapHOx's functionality was demonstrated in a 6‐d test in a seawater tank at Scripps Institution of Oceanography (La Jolla, California, U.S.A.) and was successfully deployed for 2 weeks on a shallow, coral reef flat (Lizard Island, Australia). During the latter deployment, the tris‐based self‐calibration using 15 on‐board samples exhibited superior reproducibility to the standard spectrophotometric pH‐based calibration using > 100 discrete samples. Standard deviations of calibration pH using tris ranged from 0.002 to 0.005 whereas they ranged from 0.006 to 0.009 for the standard spectrophotometric pH‐based method; the two independent calibration methods resulted in a mean pH difference of 0.008. We anticipate that the Self‐Calibrating SeapHOx will be capable of autonomously providing climate quality pH data, directly linked to a primary seawater pH standard, and with improvements over standard calibration techniques.

     
    more » « less
  2. null (Ed.)
    Abstract. Equimolal tris (2-amino-2-hydroxymethyl-propane-1,3-diol) buffer in artificialseawater is a well characterized and commonly used standard for oceanographic pH measurements. We evaluated the stability of tris pH when stored in purportedly gas-impermeable bags across a variety of experimental conditions, including bag type and storage in air vs. seawater over300 d. Bench-top spectrophotometric pH analysis revealed that the pH of tris stored in bags decreased at a rate of 0.0058±0.0011 yr−1 (mean slope ±95 % confidence interval of slope). The upper and lower bounds of expected pH change att=365 d, calculated using the averages and confidence intervals of slope and intercept of measured pH change vs. time data, were −0.0042 and −0.0076 from initial pH. Analyses of total dissolved inorganic carbonconfirmed that a combination of CO2 infiltration and/or microbialrespiration led to the observed decrease in pH. Eliminating the change in pH of bagged tris remains a goal, yet the rate of pH change is lower than many processes of interest and demonstrates the potential of bagged tris for sensor calibration and validation of autonomous in situ pH measurements. 
    more » « less