Optical phase and birefringence signals occur in cells and thin, semi-transparent biomaterials. A dual-modality quantitative phase and polarization microscope was designed to study the interaction of cells with extracellular matrix networks and to relate optical pathlength and birefringence signals within structurally anisotropic biomaterial constructs. The design was based on an existing, custom-built digital holographic microscope, to which was added a polarization microscope utilizing liquid crystal variable retarders. Phase and birefringence channels were calibrated, and data was acquired sequentially from cell-seeded collagen hydrogels and electrofabricated chitosan membranes. Computed phase height and retardance from standard targets were accurate within 99.7% and 99.8%, respectively. Phase height and retardance channel background standard deviations were 35 nm and 0.6 nm, respectively. Human fibroblasts, visible in the phase channel, aligned with collagen network microstructure, with retardance and azimuth visible in the polarization channel. Electrofabricated chitosan membranes formed in 40 µm tall microfluidic channels possessed optical retardance ranging from 7 to 11 nm, and phase height from 37 to 39 µm. These results demonstrate co-registered dual-channel acquisition of phase and birefringence parameter maps from microstructurally-complex biospecimens using a novel imaging system combining digital holographic microscopy with voltage-controlled polarization microscopy.
A liquid crystal variable retarder (LCVR) enables fast, automated control of retardance that can be used as a variable waveplate in polarimetric instruments. However, precise control of the polarization state requires calibration of the LCVR. A manufacturer calibration curve is typically supplied for a single specific wavelength and temperature, but for applications under different conditions, additional calibration is needed. Calibration is typically performed with crossed polarizers to generate an intensity curve that is converted to retardance, but this method is prone to noise when retardance is close to zero. Here, we demonstrate a simple common-path Sagnac interferometer to measure retardance and provide open source software for automated generation of calibration curves for retardance as a function of wavelength and voltage. We also provide a curve fitting method and closed-form functional representation that outputs the voltage needed to achieve a desired retardance given a specified wavelength.
- Award ID(s):
- 1845801
- Publication Date:
- NSF-PAR ID:
- 10372117
- Journal Name:
- Applied Optics
- Volume:
- 59
- Issue:
- 34
- Page Range or eLocation-ID:
- Article No. 10673
- ISSN:
- 1559-128X; APOPAI
- Publisher:
- Optical Society of America
- Sponsoring Org:
- National Science Foundation
More Like this
-
-
Polarization aberrations are found in most optical components due to a materials differing response to s- and p-polarizations. This differing response can manifest either as diattenuation, retardance, or both. Correction of polarization aberrations, such as these, are critical in many applications such as interferometry, polarimetry, display, and high contrast imaging, including astronomy. In this work, compensators based on liquid crystal polymer and anti-reflection thin-films are presented to correct polarization aberrations in both transmission and reflection configurations. Our method is versatile, allowing for good correction in transmission and reflection due to optical components possessing differing diattenuation and retardance dispersions. Through simulation and experimental validation we show two designs, one correcting the polarization aberrations of a dichroic spectral filter over a 170nm wavelength band, and the other correcting the polarization aberration of an aluminum-coated mirror over a 400nm wavelength band and a 55-degree cone of angles. The measured performance of the polarization aberration compensators shows good agreement with theory.
-
Abstract: 100 words Jurors are increasingly exposed to scientific information in the courtroom. To determine whether providing jurors with gist information would assist in their ability to make well-informed decisions, the present experiment utilized a Fuzzy Trace Theory-inspired intervention and tested it against traditional legal safeguards (i.e., judge instructions) by varying the scientific quality of the evidence. The results indicate that jurors who viewed high quality evidence rated the scientific evidence significantly higher than those who viewed low quality evidence, but were unable to moderate the credibility of the expert witness and apply damages appropriately resulting in poor calibration. Summary: <1000 words Jurors and juries are increasingly exposed to scientific information in the courtroom and it remains unclear when they will base their decisions on a reasonable understanding of the relevant scientific information. Without such knowledge, the ability of jurors and juries to make well-informed decisions may be at risk, increasing chances of unjust outcomes (e.g., false convictions in criminal cases). Therefore, there is a critical need to understand conditions that affect jurors’ and juries’ sensitivity to the qualities of scientific information and to identify safeguards that can assist with scientific calibration in the courtroom. The current project addresses thesemore »
-
Many correlations exist between spectral reflectance or transmission with various phenotypic responses from plants. Of interest to us are metabolic characteristics, namely, how the various polarimetric components of plants may correlate to underlying environmental, metabolic, and genotypic differences among different varieties within a given species, as conducted during large field experimental trials. In this paper, we overview a portable Mueller matrix imaging spectropolarimeter, optimized for field use, by combining a temporal and spatial modulation scheme. Key aspects of the design include minimizing the measurement time while maximizing the signal-to-noise ratio by mitigating systematic error. This was achieved while maintaining an imaging capability across multiple measurement wavelengths, spanning the blue to near-infrared spectral region (405–730 nm). To this end, we present our optimization procedure, simulations, and calibration methods. Validation results, which were taken in redundant and non-redundant measurement configurations, indicated that the polarimeter provides average absolute errors of (5.3±2.2)×10−3and (7.1±3.1)×10−3, respectively. Finally, we provide preliminary field data (depolarization, retardance, and diattenuation) to establish baselines of barren and non-barren
Zea maize hybrids (G90 variety), as captured from various leaf and canopy positions during our summer 2022 field experiments. Results indicate that subtle variations in retardance and diattenuation versus leaf canopy position may be presentmore » -
Abstract The recent availability of Schottky-type field emission electron microprobes provides incentive to consider analyzing micrometer-sized features. Yet, to quantify sub-micrometer-sized features, the electron interaction volume must be reduced by decreasing accelerating voltage. However, the K lines of the transition elements (e.g., Fe) then cannot be excited, and the L lines must be used. The Fe L α 1,2 line is the most intense of the L series but bonding effects change its atomic parameters because it involves a valence band electron transition. For successful traditional electron probe microanalysis, the mass absorption coefficient (MAC) must be accurately known, but the MAC of Fe L α 1,2 radiation by Fe atoms varies from one Fe-compound to another and is not well known. We show that the conventional method of measuring the MAC by an electron probe cannot be used in close proximity to absorption edges, making its accurate determination impossible. Fortunately, we demonstrate, using a set of Fe–silicide compounds, that it is possible to derive an accurate calibration curve, for a given accelerating voltage and takeoff angle, which can be used to quantify Fe in Fe–silicide compounds. The calibration curve can be applied to any spectrometer without calibration and gives accuratemore »