Title: Assessment of a pH optode for oceanographic moored and profiling applications
Abstract As global ocean monitoring programs and marine carbon dioxide removal methods expand, so does the need for scalable biogeochemical sensors. Currently, pH sensors are widely used to measure the ocean carbonate system on a variety of autonomous platforms. This paper assesses a commercially available optical pH sensor (optode) distributed by PyroScience GmbH for oceanographic applications. Results from this study show that the small, solid‐state pH optode demonstrates a precision of 0.001 pH and relative accuracy of 0.01 pH using an improved calibration routine outlined in the manuscript. A consistent pressure coefficient of 0.029 pH/1000 dbar is observed across multiple pH optodes tested in this study. The response time is investigated for standard and fast‐response versions over a range of temperatures and flow rates. Field deployments include direct comparison to ISFET‐based pH sensor packages for both moored and profiling platforms where the pH optodes experience sensor‐specific drift rates up to 0.006 pH d−1. In its current state, the pH optode potentially offers a viable and scalable option for short‐term field deployments and laboratory mesocosm studies, but not for long term deployments with no possibility for recalibration like on profiling floats.  more » « less
Award ID(s):
2300400 2110258 2300399
PAR ID:
10569742
Author(s) / Creator(s):
; ; ; ; ; ; ; ; ; ;
Publisher / Repository:
ASLO
Date Published:
Journal Name:
Limnology and Oceanography: Methods
Volume:
22
Issue:
11
ISSN:
1541-5856
Page Range / eLocation ID:
805 to 822
Format(s):
Medium: X
Sponsoring Org:
National Science Foundation
More Like this
  1. Increasing interest in the deployment of optical oxygen sensors, or optodes, on oceanographic moorings reflects the value of dissolved oxygen (DO) measurements in studies of physical and biogeochemical processes. Optodes are well-suited for moored applications but require careful, multi-step calibrations in the field to ensure data accuracy. Without a standardized set of protocols, this can be an obstacle for science teams lacking expertise in optode data processing and calibration. Here, we provide a set of recommendations for the deployment andin situcalibration of data from moored optodes, developed from our experience working with a set of 60 optodes deployed as part of the Gases in the Overturning and Horizontal circulation of the Subpolar North Atlantic Program (GOHSNAP). In particular, we detail the correction of drift in moored optodes, which occurs in two forms: (i) an irreversible, time-dependent drift that occurs during both optode storage and deployment and (ii) a reversible and pressure-and-time-dependent drift that is detectable in some optodes deployed at depths greater than 1,000 m. The latter is virtually unidentified in the literature yet appears to cause a low-bias in measured DO on the order of 1 to 3µmol kg−1per 1,000 m of depth, appearing as an exponential decay over the first days to months of deployment. Comparisons of our calibrated DO time series against serendipitous mid-deployment conductivity-temperature-depth (CTD)-DO profiles, as well as biogeochemical (BGC)-ARGO float profiles, suggest the protocols described here yield an accuracy in optode-DO of ∼1%, or approximately 2.5 to 3µmol kg−1. We intend this paper to serve as both documentation of the current best practices in the deployment of moored optodes as well as a guide for science teams seeking to collect high-quality moored oxygen data, regardless of expertise. 
    more » « less
  2. Abstract Ocean acidification alters the oceanic carbonate system, increasing potential for ecological, economic, and cultural losses. Historically, productive coastal oceans lack vertically resolved high‐resolution carbonate system measurements on time scales relevant to organism ecology and life history. The recent development of a deep ion‐sensitive field‐effect transistor (ISFET)‐based pH sensor system integrated into a Slocum glider has provided a platform for achieving high‐resolution carbonate system profiles. From May 2018 to November 2019, seasonal deployments of the pH glider were conducted in the central Mid‐Atlantic Bight. Simultaneous measurements from the glider's pH and salinity sensors enabled the derivation of total alkalinity and calculation of other carbonate system parameters including aragonite saturation state. Carbonate system parameters were then mapped against other variables, such as temperature, dissolved oxygen, and chlorophyll, over space and time. The seasonal dynamics of carbonate chemistry presented here provide a baseline to begin identifying drivers of acidification in this vital economic zone. 
    more » « less
  3. Abstract Although increasing efforts have been devoted to the development of non‐invasive wearable electrochemical sweat sensors for monitoring physiological and metabolic information, most of them still suffer from poor stability and specificity over time and fluctuating temperatures. This study reports the design and fabrication of a long‐term stable and highly sensitive flexible electrochemical sensor based on nanocomposite‐modified porous graphene by facile laser treatment for detecting biomarkers such as glucose in sweat. The laser‐reduced and patterned stable conductive nanocomposite on the porous graphene electrode provides the resulting glucose sensor with an excellent sensitivity of 1317.69 µA mm−1cm−2and an ultra‐low limit of detection of 0.079 µm. The sensor can also detect pH and exhibit extraordinary stability to maintain more than 91% sensitivity over 21 days in ambient conditions. Taken together with a temperature sensor based on the same material system, the dual glucose and pH sensor integrated with a flexible microfluidic sweat sampling network further results in accurate continuous on‐body glucose detection calibrated by the simultaneously measured pH and temperature. The low‐cost, highly sensitive, and long‐term stable platform could facilitate the early identification and continuous monitoring of different biomarkers for non‐invasive disease diagnosis and treatment evaluation. 
    more » « less
  4. Abstract Marine animals equipped with sensors provide vital information for understanding their ecophysiology and collect oceanographic data on climate change and for resource management. Existing methods for attaching sensors to marine animals mostly rely on invasive physical anchors, suction cups, and rigid glues. These methods can suffer from limitations, particularly for adhering to soft fragile marine species such as squid and jellyfish, including slow complex operations, unreliable fixation, tissue trauma, and behavior changes of the animals. However, soft fragile marine species constitute a significant portion of ocean biomass (>38.3 teragrams of carbon) and global commercial fisheries. Here we introduce a soft hydrogel-based bioadhesive interface for marine sensors that can provide rapid (time <22 s), robust (interfacial toughness >160 J m−2), and non-invasive adhesion on various marine animals. Reliable and rapid adhesion enables large-scale, multi-animal sensor deployments to study biomechanics, collective behaviors, interspecific interactions, and concurrent multi-species activity. These findings provide a promising method to expand a burgeoning research field of marine bio-sensing from large marine mammals and fishes to small, soft, and fragile marine animals. 
    more » « less
  5. Abstract Multiplexed solid‐contact ion‐selective electrodes (SCISEs) are fabricated using printed circuit board (PCB) and mesoporous carbon black (MCB) as ion‐to‐electron transducer (solid contact). Four sensor configurations were examined and showed that in addition to MCB, the sensor configuration plays crucial role in the stability of the potential response. The enhanced sensor stability was also linked with suppression of transmembrane flux of water. The sensors exhibited near‐Nernstian sensitivity (58.1 mV/dec for K+ISEs and −55.1 mV/dec for NO3ISEs), low detection limits (1.5–2.2 μM), and good short‐term stability (∼0.1 mV/min). Sensors can be stored dry and used without preconditioning. This work demonstrates a promising approach to combining PCB technology and carbon black for large‐scale production of low cost ISEs for point‐of‐care testing, wearables, orin situfield measurements. 
    more » « less